J2A Dental Implant System, J2C Dental Implant System
Applicant
Kj Meditech Co., Ltd.
Product Code
DZE · Dental
Decision Date
Apr 2, 2015
Decision
SESE
Submission Type
Special
Regulation
21 CFR 872.3640
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The J2A Dental Implant System and J2C Dental Implant System are in partially or fully edentulous mandibles and maxillae, in support of single or multiple-unit restorations including; cemented retained, or overdenture restorations, and terminal or intermediate abutment support for fixed bridgework. The 12A Dental Implant System and J2C Dental Implant System are for single and two stage surgical procedures. These systems are intended for delayed loading.
Device Story
J2A and J2C Dental Implant Systems are internal hexagon dental implants; surgically placed in maxilla or mandible bone to replace missing teeth. Implants consist of titanium alloy fixtures and abutments; surface treated with Resorbable Blast Media (RBM) or TiN coating. Used by dental surgeons in clinical settings for single or two-stage procedures; intended for delayed loading. Output is a stable foundation for dental restorations; aids in restoring masticatory function and aesthetics for edentulous patients.
Clinical Evidence
Bench testing only. No clinical data provided. Substantial equivalence supported by risk analysis, biocompatibility, and performance testing of the predicate device (K121047).
Technological Characteristics
Materials: Ti 6Al 4V ELI (ASTM F136). Surface: RBM (fixtures), TiN coating (abutments). Design: Internal hexagon connection, bone-level, tapered body. Sterilization: Gamma. Connectivity: None (mechanical device).
Indications for Use
Indicated for partially or fully edentulous patients (mandibles and maxillae) requiring single or multiple-unit restorations (cemented, screw-retained, or overdenture) or terminal/intermediate abutment support for fixed bridgework. Suitable for single or two-stage surgical procedures with delayed loading.
Regulatory Classification
Identification
An endosseous dental implant is a prescription device made of a material such as titanium or titanium alloy that is intended to be surgically placed in the bone of the upper or lower jaw arches to provide support for prosthetic devices, such as artificial teeth, in order to restore a patient's chewing function.
Special Controls
*Classification.* (1) Class II (special controls). The device is classified as class II if it is a root-form endosseous dental implant. The root-form endosseous dental implant is characterized by four geometrically distinct types: Basket, screw, solid cylinder, and hollow cylinder. The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)(2)
*Classification.* Class II (special controls). The device is classified as class II if it is a blade-form endosseous dental implant. The special controls for this device are:(i) The design characteristics of the device must ensure that the geometry and material composition are consistent with the intended use;
(ii) Mechanical performance (fatigue) testing under simulated physiological conditions to demonstrate maximum load (endurance limit) when the device is subjected to compressive and shear loads;
(iii) Corrosion testing under simulated physiological conditions to demonstrate corrosion potential of each metal or alloy, couple potential for an assembled dissimilar metal implant system, and corrosion rate for an assembled dissimilar metal implant system;
(iv) The device must be demonstrated to be biocompatible;
(v) Sterility testing must demonstrate the sterility of the device;
(vi) Performance testing to evaluate the compatibility of the device in a magnetic resonance (MR) environment;
(vii) Labeling must include a clear description of the technological features, how the device should be used in patients, detailed surgical protocol and restoration procedures, relevant precautions and warnings based on the clinical use of the device, and qualifications and training requirements for device users including technicians and clinicians;
(viii) Patient labeling must contain a description of how the device works, how the device is placed, how the patient needs to care for the implant, possible adverse events and how to report any complications; and
(ix) Documented clinical experience must demonstrate safe and effective use and capture any adverse events observed during clinical use.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the seal of the U.S. Department of Health and Human Services. The seal features the department's name in a circular arrangement around the perimeter. At the center of the seal is a stylized representation of a human figure, composed of three overlapping profiles facing to the right. The seal is rendered in black and white.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
April 2, 2015
KJ Meditech Co., Ltd. C/O Ms. Priscilla Chung Regulatory Affairs Consultant LK Consulting Group USA, Inc. 2651 East Chapman Avenue, Suite 110 Fullerton, California 92831
Re: K150060
Trade/Device Name: J2A Dental Implant System, J2C Dental Implant System Regulation Number: 21 CFR 872.3640 Regulation Name: Endosseous Dental Implant Regulatory Class: II Product Code: DZE, NHA Dated: March 02, 2015 Received: March 06, 2015
Dear Ms. Chung:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Tina
Kiang -S
for Erin I. Keith, M.S. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
#### Indications for Use
# 510(k) Number (if known)
K150060
#### Device Name J2A Dental Implant System, J2C Dental Implant System
#### Indications for Use (Describe)
The I2A Dental Implant System and I2C Dental Implant System are in partially or fully edentulous mandibles and maxillae, in support of single or multiple-unit restorations including; cemented retained, or overdenture restorations, and terminal or intermediate abutment support for fixed bridgework. The 12A Dental Implant System and J2C Dental Implant System are for single and two stage surgical procedures. These systems are intended for delayed loading.
Type of Use (Select one or both, as applicable)
|× Prescription Use (Part 21 CFR 801 Subpart D)
■ Over-The-Counter Use (21 CFR 801 Subpart C)
#### PLEASE DO NOT WRITE BELOW THIS LINE – CONTINUE ON A SEPARATE PAGE IF NEEDED.
#### FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
{3}------------------------------------------------
#### 510(k) Summary (k150060)
This summary of 510(K) – substantial equivalence information is being submitted in accordance with requirements of 21 CFR Part 807.92.
Date: 04/02/2015
## 1. Submitter
| | Submitter |
|---------|---------------------------------------------------------------|
| Name | KJ Meditech Co., Ltd. |
| Address | 959-21 Daechon-dong, Buk-gu,Gwang-ju, 500-470,<br>South Korea |
| Phone | +82-62-972-5476 |
| Fax | +82-62-973-2809 |
| Contact | Huykki Moon, CEO |
# 2. U.S Agent/Contact Person
Priscilla Chung LK Consulting Group USA, Inc. 2651 E Chapman Ave Ste 110, Fullerton, CA 92831 Phone: 714.202.5789 Fax: 714-409-3357 Email: juhee.c@lkconsultinggroup.com
## 3. Device
- . Trade Name: J2A Dental Implant System
- J2C Dental Implant System
- Common Name: Dental Implant System
- Classification Name: Endosseous Dental Implant
- Product Code: DZE, NHA
- . Classification regulation: 21CFR872.3640
## 4. Predicate Device:
- Primary Predicate Device: HERO II Dental Implant System by KJ Meditech Co., Ltd. (K121047)
- . Reference Predicate Devices: TS Fixture System by OSSTEM IMPLANT CO.,LTD (K121995) GS III SYSTEM by OSSTEM IMPLANT CO.,LTD (K091208)
{4}------------------------------------------------
## 5. Description:
The J2A Dental Implant System and J2C Dental Implant System are internal hexagon type dental implant systems made of Titanium 6AL 4V ELI alloy intended to be surgically placed in the bone of the upper or lower jaw arches for loading after a conventional healing period. The implants may be used to replace one or more missing teeth. The systems are similar to other commercially available products based on the intended use, the technology used, the claims, the material composition employed and performance characteristics. The surface of these systems have been treated with R.B.M (Resorbable Blast Media). The size information is as below.
J2A Dental Implant System
- · 3.75mm Dia. x 7mm(L) / 8.5mm (L) / 10.0mm(L) / 13.0mm(L) / 13.0mm(L) / 15mm(L)
- · 4.00mm Dia. x 7mm(L) / 8.5mm (L) / 10.0mm(L) / 13.0mm(L) / 13.0mm(L) / 15mm(L)
- · 4.30mm Dia. x 7mm(L) / 8.5mm (L) / 10.0mm(L) / 13.0mm(L) / 13.0mm(L) / 15mm(L)
- · 4.50mm Dia. x 7mm(L) / 8.5mm (L) / 10.0mm(L) / 13.0mm(L) / 13.0mm(L) / 15mm(L)
- · 5.00mm Dia. x 7mm(L) / 8.5mm (L) / 10.0mm(L) / 13.0mm(L) / 13.0mm(L) / 15mm(L) • 5.50mm Dia. x 7mm(L) / 8.5mm (L) / 10.0mm(L) / 13.0mm(L) / 13.0mm(L) / 15mm(L)
- · 6.00mm Dia. x 7mm(L) / 8.5mm (L) / 10.0mm(L) / 13.0mm(L) / 13.0mm(L) / 15mm(L)
## J2C Dental Implant System
- 3.75mm Dia. x 7mm(L) / 8.5mm (L) / 10.0mm(L) / 13.0mm(L) / 13.0mm(L) / 15mm(L)
- 4.00mm Dia. x 7mm(L) / 8.5mm (L) / 10.0mm(L) / 13.0mm(L) / 13.0mm(L) / 15mm(L)
- · 4.30mm Dia. x 7mm(L) / 8.5mm (L) / 10.0mm(L) / 13.0mm(L) / 13.0mm(L) / 15mm(L)
- 4.50mm Dia. x 7mm(L) / 8.5mm (L) / 10.0mm(L) / 13.0mm(L) / 13.0mm(L) / 15mm(L)
- · 5.00mm Dia. x 7mm(L) / 8.5mm (L) / 10.0mm(L) / 13.0mm(L) / 13.0mm(L) / 15mm(L)
- 5.50mm Dia. x 7mm(L) / 8.5mm (L) / 10.0mm(L) / 11.5mm(L) / 13.0mm(L) / 15mm(L)
- 6.00mm Dia. x 7mm(L) / 8.5mm (L) / 10.0mm(L) / 13.0mm(L) / 13.0mm(L) / 15mm(L)
There are one-piece abutments which are compatible with both J2A and J2C fixtures with the following features. Only straight abutments are provided.
| Material / Surface Treatment | Ti 6Al 4V ELI, ASTM F136 / TiN Coating |
|------------------------------|--------------------------------------------------------------------------------------------------------------|
| Dimensions | Diameter: Ø4mm ~ Ø7mm Cuff: 1mm~5mm Height: 4.0mm~7.0mm Connection platform: Narrow platform & Wide platform |
There are also two-piece abutments which are compatible with both J2A and J2C fixtures with the following features. Only straight abutments are provided.
| Material / Surface Treatment | Ti 6Al 4V ELI, ASTM F136 / TiN Coating |
|---------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------|
| Dimensions | • Diameter: Ø4mm ~ Ø7mm<br>• Cuff: 1mm~5mm<br>• Height: 4.0mm~7.0mm<br>• Connection platform: Narrow platform & Wide platform |
| Cover screws made of Titanium 6Al 4V ELI alloy are also provided both for I2A and I2C | |
Cover screws made of Titanium 6AL 4V ELI alloy are also provided both for J2A and J2C fixtures.
{5}------------------------------------------------
#### 6. Indication for use:
The J2A Dental Implant System and J2C Dental Implant System are indicated for use in partially or fully edentulous mandibles and maxillae, in support of single or multiple-unit restorations including; cemented retained, screw retained, or overdenture restorations, and terminal or intermediate abutment support for fixed bridgework. The J2A Dental Implant System and J2C Dental Implant System are for single and two stage surgical procedures. These systems are intended for delayed loading.
{6}------------------------------------------------
### 7. Basis for Substantial Equivalence
### 7.1. Fixtures and Cover Screws
The I2A Dental Implant and J2C Dental Implant have the identified predicate devices. They are similar in fundamental scientific technology in that they are all threaded root form implants constructed of titanium with RBM roughened surfaces. They all share same internal hexagon abutment connection system with interface. The subject and predicate devices are both bone-level implants that share similar body shape design walled neck and tapered body design.
The manufacturing facility, raw materials for the surface treatments (RBA treatment), surface treatment method, manufacturing process, and cleaning process of the J2A and I2C Dental Implant are identical to the predicate device (Hero I and IS Dental Implant System, K121047).
| Item | Subject Device<br>(Modified Device 1) | Subject Device<br>(Modified Device 2) | Primary Predicate<br>Device<br>(Unmodified Device) | Reference Predicate<br>Device 1 | Reference Predicate Device 2 |
|--------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(K) Number | K150060 | K150060 | K121047 | K121995 | K091208 |
| Device Name | J2A Dental Implant<br>System | J2C Dental Implant<br>System | HERO II Dental<br>Implant System | TS Fixture System | GS III SYSTEM |
| Manufacturer | KJ Meditech Co., Ltd. | KJ Meditech Co., Ltd. | KJ Meditech Co., Ltd. | OSSTEM IMPLANT<br>CO.,LTD | OSSTEM IMPLANT<br>CO.,LTD |
| Indications for<br>Use | The J2A Dental<br>Implant System is<br>indicated for use in<br>partially or fully<br>edentulous mandibles<br>and maxillae, in<br>support of single or<br>multiple-unit<br>restorations including;<br>cemented retained, | The J2C Dental Implant<br>System is indicated for<br>use in partially or fully<br>edentulous mandibles<br>and maxillae, in<br>support of single or<br>multiple-unit<br>restorations including;<br>cemented retained,<br>screw retained, or | The HERO II Dental<br>Implant System is<br>indicated for use in<br>partially or fully<br>edentulous mandibles<br>and maxillae, in<br>support of single or<br>multiple-unit<br>restorations including;<br>cemented retained, | The TS Fixture System<br>is indicated for use in<br>partially or fully<br>edentulous mandibles<br>and maxillae, in support<br>of single or multiple-unit<br>restorations including;<br>cemented retained,<br>screw retained, or<br>overdenture restorations, | The GS III System is indicated<br>for use in partially or fuhlly<br>edentulous mandibles and<br>maxillae, in support of single<br>or multiple-unit restorations<br>including; cemented retained,<br>screw retained, or overdent-<br>ure restorations, and terminal<br>or intermediate abutment<br>support for fixed bridgework. |
| | screw retained, or<br>overdenture<br>restorations, and<br>terminal or<br>intermediate abutment<br>support for fixed<br>bridgework. The J2A<br>Dental Implant System<br>is for single and two<br>stage surgical<br>procedures. The system<br>is intended for delayed<br>loading. | overdenture<br>restorations, and<br>terminal or<br>intermediate abutment<br>support for fixed<br>bridgework. The J2C<br>Dental Implant System<br>is for single and two<br>stage surgical<br>procedures. The system<br>is intended for delayed<br>loading. | screw retained, or<br>overdenture<br>restorations, and<br>terminal or<br>intermediate abutment<br>support for fixed<br>bridgework. The HERO<br>II Dental Implant<br>System is for single and<br>two stage surgical<br>procedures. The system<br>is intended for delayed<br>loading. | and final or temporary<br>abutment support for<br>fixed bridgework. It is<br>intended for delayed<br>loading.<br>TS Fixture System is<br>compatible with<br>abutment in the ET/SS<br>Implant<br>System. | The GS II System is for single<br>and two stage surgical<br>procedures.<br>It is not for immediate load. |
| Design | • Implant Type: Bone<br>Level Implant<br>• Connection Type:<br>Internal Hexagon<br>• Neck Design:<br>Straight walled neck<br>• Body Design:<br>Tapered design | • Implant Type: Bone<br>Level Implant<br>• Connection Type:<br>Internal Hexagon<br>• Neck Design:<br>Straight walled neck<br>with micro-thread<br> | • Implant Type: Bone<br>Level Implant<br>• Connection Type:<br>Internal Hexagon<br>• Neck Design:<br>Straight walled neck<br>• Body Design:<br>Tapered design | • Implant Type: Bone<br>Level Implant<br>• Connection Type:<br>Internal Hexagon<br>• Neck Design: Straight<br>walled neck<br>• Body Design:<br>Tapered design | • Implant Type: Bone Level<br>Implant<br>• Connection Type:<br>Internal Hexagon<br>• Neck Design: Straight<br>walled neck with micro-<br>thread<br>• Body Design:<br>Tapered design |
| Appearance | Image: Implant 1 | Image: Implant 2 | Image: Implant 3 | Image: Implant 4 | Image: Implant 5 |
| Endosseous<br>Implant Material | Ti 6Al 4V ELI, ASTM<br>F136 | Ti 6Al 4V ELI, ASTM<br>F136 | Ti 6Al 4V ELI, ASTM<br>F136 | Grade 4 Pure Titanium<br>(ASTM F67) | Grade 4 Pure Titanium<br>(ASTM F67) |
| Surface<br>Treatment | RBM Treatment | RBM Treatment | RBM Treatment | SA Treatment | SA Treatment |
| Sterilization<br>Method | Same - No<br>modification | Same - No<br>modification | Gamma | Gamma | Gamma |
| Implant<br>Diameters | 3.75mm, 4.0mm,<br>4.3mm, 4.5mm, 5.0<br>mm, 5.5mm, 6.0mm | 3.75mm, 4.0mm,<br>4.3mm, 4.5mm, 5.0<br>mm, 5.5mm, 6.0mm | 3.75mm, 4.0mm,<br>4.5mm, 5.0 mm,<br>6.0mm | 3.75mm, 3.77mm,<br>4.2mm, 4.25mm,<br>4.6mm, 4.63mm,<br>4.65mm, 5.05mm,<br>5.08mm, 5.1mm | 3.75mm~5.1mm |
| Implant Lengths | 7.0mm - 15.0 mm | 7.0mm - 15.0 mm | 8.5mm - 15.0 mm | 7.0mm - 15.0 mm | 7.0 - 15.0 mm |
| Cover Screw | Image: Ti 6Al 4V ELI, ASTM F136 | Image: Ti 6Al 4V ELI, ASTM F136 | Image: Ti 6Al 4V ELI, ASTM F136 | Image: Ti 6Al 4V ELI, ASTM F136/Anodizing | - |
510(k) Submission.
510(k) summary , 4 / 9 page
{7}------------------------------------------------
510(k) Submission.
510(k) summary , 5 / 9 page
{8}------------------------------------------------
#### 7.2. Abutments
The I2A and J2C Abutments have the same intentified predicate devices. They are similar in fundamental scientific technology and design. The subject and the predicate devices offer one-piece as well as two-piece type abutments.
The manufacturing facility, raw materials for the surface treatments (TiN coating), surface treatment method, manufacturing process, and cleaning process of the J2A and J2C Abutments are identical to the predicate device (Hero I and IS Dental Implant System, K121047).
| Item | Subject Device<br>(Modified Device) | Primary Predicate Device<br>(Unmodified Device) | Reference Predicate Device |
|---------------|-------------------------------------|-------------------------------------------------|----------------------------|
| 510(K) Number | K150060 | K121047 | K121995 |
| Device Name | J2A and J2C Dental Implant System | HERO II Dental Implant System | TS Fixture System |
510(k) Submission.
{9}------------------------------------------------
| Manufacturer | KJ Meditech Co., Ltd. | KJ Meditech Co., Ltd. | OSSTEM IMPLANT CO.,LTD |
|------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications for<br>Use | The J2A and J2C Dental Implant<br>Systems are indicated for use in<br>partially or fully edentulous<br>mandibles and maxillae, in support<br>of single or multiple-unit<br>restorations including; cemented<br>retained, screw retained, or<br>overdenture restorations, and<br>terminal or intermediate abutment<br>support for fixed bridgework. The<br>J2A and J2C Dental Implant<br>Systems are for single and two stage<br>surgical procedures. The system is<br>intended for delayed loading. | The HERO II Dental Implant System<br>is indicated for use in partially or fully<br>edentulous mandibles and maxillae, in<br>support of single or multiple-unit<br>restorations including; cemented<br>retained, screw retained, or<br>overdenture restorations, and terminal<br>or intermediate abutment support for<br>fixed bridgework. The HERO II Dental<br>Implant System is for single and two<br>stage surgical procedures. The system<br>is intended for delayed loading. | The TS Fixture System is indicated<br>for use in partially or fully<br>edentulous mandibles and maxillae,<br>in support of single or multiple-unit<br>restorations including; cemented<br>retained, screw retained, or<br>overdenture restorations, and final<br>or temporary abutment support for<br>fixed bridgework. It is intended for<br>delayed loading.<br>TS Fixture System is compatible<br>with abutment in the ET/SS Implant<br>System. |
| One Piece<br>Abutment | • Ti 6Al 4V ELI, ASTM F136 /<br>TiN Coating (Entire body)<br>• Diameter: Ø4mm ~ Ø7mm<br>Cuff: 1mm~5mm<br>Height: 4.0mm~7.0mm | • Ti 6Al 4V ELI, ASTM F136 / TiN<br>Coating (Partial)<br>• Diameter: Ø4mm ~ Ø6mm<br>Cuff: 1mm~4mm<br>Height: 4.0mm~7.0mm | • Ti 6Al 4V ELI, ASTM F136 /<br>TiN Coating (Partial)<br>• Diameter: Ø4mm ~ Ø7mm<br>Cuff: 1mm~5mm<br>Height: 4.0mm~7.0mm |
{10}------------------------------------------------
| Two Piece<br>Abutment | Image: Dental abutment and screw<br>Ti 6Al 4V ELI, ASTM F136 /<br>TiN Coating (Entire body) Diameter: Ø4mm ~ Ø7mm Cuff: 1mm~5mm Height: 4.0mm~7.0mm | Image: Dental abutment and screw<br>Ti 6Al 4V ELI, ASTM F136 / TiN<br>Coating (Entire body) Diameter: Ø4mm ~ Ø7mm Cuff: 1mm~4mm Height: 4.0mm~7.0mm | Image: Dental abutment and screw<br>Ti 6Al 4V ELI, ASTM F136 /<br>TiN Coating (Entire body) Diameter: Ø4mm ~ Ø7mm Cuff: 1mm~5mm Height: 4.0mm~7.0mm |
|-----------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------|
|-----------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------|
{11}------------------------------------------------
### 8. Non-Clinical Testing
Risk analysis was conducted according to ISO14971 to demonstrate that the following modifications to the company's own predicate, Hero II Dental Implant System (K121047) is substantially equivalent:
- Fixture thread design change ●
- . Fixture size addition
- Coating area expansion for one-piece type abutments
- Abutment size addition .
Additional predicate devices of similar design and dimensions were identified to determine substantial equivalence. In addition, based on the risk analysis the shelf-life, sterilization, biocompatibility, and performance testing of the predicate, Hero II Dental Implant System (K121047), is applicable to the subject device.
## 9. Conclusion
The subject devices and the predicate device have the same intended use and have the same technological characteristics. The subject and predicate implants are all made of commercially pure titanium and have the same surface treatments.
Overall, the J2A Dental Implant system and the J2C Dental Implant system have the following similarities to the predicate device:
- * have the same intended use,
- * use the same operating principle,
- * incorporate the same basic design,
- * incorporate the same material and the surface treatment.
Based on the similarities, we conclude that the J2A Dental Implant system and the J2C Dental Implant system are substantially equivalent to the predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.