K150059 · Becton, Dickinson Company · FPA · Apr 28, 2015 · General Hospital
Device Facts
Record ID
K150059
Device Name
BD FlowSmart Set
Applicant
Becton, Dickinson Company
Product Code
FPA · General Hospital
Decision Date
Apr 28, 2015
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5440
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The BD FlowSmart™ set is intended for the subcutaneous infusion of medication, including insulin, from an external infusion pump.
Device Story
BD FlowSmart™ is a single-use, sterile subcutaneous infusion set; interfaces with external infusion pumps to deliver medication (e.g., insulin). Device consists of an adhesive patch, a flexible 28G perforated plastic cannula, and detachable tubing. Operates via 90-degree manual or automatic insertion into subcutaneous tissue. Used in home or clinical settings by patients or caregivers. Tubing connects to pump reservoir; adhesive base secures cannula to skin. Perforated cannula design facilitates medication delivery. Healthcare providers/patients use the device to manage chronic conditions requiring continuous subcutaneous infusion; benefits include reliable medication delivery and pump compatibility via Paradigm® or Luer Lock connections.
Clinical Evidence
No clinical trials were conducted. Evidence consists of non-clinical bench testing, including biocompatibility (ISO 10993), functional performance (ISO 10555-1, ISO 1135-1, ISO ID26), animal testing per USDA standards/Animal Welfare Act, and human factor evaluations (IEC 62366:2007). Results demonstrated the device meets requirements for its intended use.
Technological Characteristics
Single-use sterile subcutaneous infusion set. Features: 30G insertion needle, 28G plastic cannula (6mm length), adhesive base, and detachable tubing. Connectivity: Paradigm® and Luer Lock compatible. Materials evaluated per ISO 10993. Sterilization: Sterile. Form factor: Perpendicular (90-degree) insertion. No software or electronic components.
Indications for Use
Indicated for subcutaneous infusion of medication, including insulin, from an external infusion pump in patients requiring such therapy.
Regulatory Classification
Identification
An intravascular administration set is a device used to administer fluids from a container to a patient's vascular system through a needle or catheter inserted into a vein. The device may include the needle or catheter, tubing, a flow regulator, a drip chamber, an infusion line filter, an I.V. set stopcock, fluid delivery tubing, connectors between parts of the set, a side tube with a cap to serve as an injection site, and a hollow spike to penetrate and connect the tubing to an I.V. bag or other infusion fluid container.
Special Controls
*Classification.* Class II (special controls). The special control for pharmacy compounding systems within this classification is the FDA guidance document entitled “Class II Special Controls Guidance Document: Pharmacy Compounding Systems; Final Guidance for Industry and FDA Reviewers.” Pharmacy compounding systems classified within the intravascular administration set are exempt from the premarket notification procedures in subpart E of this part and subject to the limitations in § 880.9.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, stacked on top of each other, representing the department's focus on people and health.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
April 28, 2015
Becton, Dickinson Company Ms. Avital Merl Regulatory Affairs Manager 1 Becton Drive Franklin Lakes, NJ 07666
Re: K150059
Trade/Device Name: BD FlowSmart™ Set Regulation Number: 21 CFR 880.5440 Regulation Name: Intravascular Administration Set Regulatory Class: II Product Code: FPA, FPK Dated: March 24, 2015 Received: March 25, 2015
Dear Ms. Merl:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Page 2 - Ms. Avital Merl
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers. International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Tina
Kiang -S
for Erin I. Keith, M.S. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use Statement
K150059 510(k) Number (if known):
Device Name: BD FlowSmart™ Set
Indications for Use:
The BD FlowSmart™ set is intended for the subcutaneous infusion of medication, including insulin, from an external infusion pump.
Over-The-Counter Use Prescription Use X AND/OR (Part 21 CFR 801 Subpart D) (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE – CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Page __ of ___
{3}------------------------------------------------
## 510(k) Summary
| Submitted By: | Avital Merl<br>Regulatory Affairs Manager, BD Medical<br>1 Becton Drive<br>Franklin Lakes, NJ 07417<br>Tel: 201 847 4739<br>Fax: 201 847 5307 | | |
|----------------|-----------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------|--|
| Date Prepared: | April 28, 2015 | | |
| Device Name: | Trade Name: | BD FlowSmart™ Set | |
| | Common Name: | Subcutaneous Infusion Set | |
| | Classification: | Class II device; 21 CFR 880.5440,<br>Intravascular Administration set) | |
| | Product Code: | FPA (intravascular administration set)<br>FPK (tubing, intravascular administration set) | |
#### Legally marketed predicate devices to which substantial equivalence is being claimed: K011071- Medtronic's MiniMed Quickset K130468- Ypsomed's Orbit Infusion Set
#### Device Description:
The BD FlowSmart™ set is a subcutaneous administration set intended to interface with commercially available infusion pumps with suitable connections. The infusion set features a flexible perforated catheter perpendicular to an adhesive patch and detachable tubing. The tubing is connected on one end to the medication reservoir of an infusion pump and on the other end to the patient, attached to the skin by an adhesive base. It is a single-use sterile device.
### Intended Use:
The BD FlowSmart™ set is intended for the subcutaneous infusion of medication, including insulin, from an external infusion pump.
### Comparison with Predicate Devices:
The BD FlowSmart™ set has a similar intended use as its predicate devices for the subcutaneous delivery of fluids and medication, including insulin, from an external infusion pump. It also shares several similarities in technology compared to its predicate devices. These technological characteristics include an adhesive base, catheter, and tubing connection components. The needle insertion method, 90 degree angle of insertion, and tubing connection types are also similar features to its predicate devices. The technological differences include feature of a single 6mm cannula length, a 30G insertion needle and 28G plastic cannula. Pre-clinical testing and published literature support substantial equivalence of the subject device despite these technological differences. The table below provides a side by side comparison of the subject device compared to its predicates.
{4}------------------------------------------------
| Feature | Subject Device: BD<br>FlowSmart™ Set | Primary Predicate Device:<br>MinMed Quick-Set | Secondary Predicate<br>Device: Orbit Infusion Set |
|----------------------------|-------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------|
| 510(k) Number | NA | K011071 | K130468 |
| Manufacturer | BD | Medtronic | Ypsomed |
| Intended Use | For the subcutaneous<br>infusion of medication,<br>including insulin, from<br>an external infusion<br>pump. | For the subcutaneous infusion<br>of medication, including<br>insulin, from an external<br>infusion pump. | For the subcutaneous<br>delivery of fluids and<br>medication, such as insulin,<br>from an external infusion<br>pump. |
| System<br>Components | Infusion set + tubing | Infusion set+ tubing | Infusion set + tubing |
| Needle insertion<br>method | Manual or Automatic | Manual or Automatic | Manual |
| Insertion Needle<br>Gauge | 30G | 27G | 28G |
| Angle of Insertion | Straight 90 degrees | Straight 90 degrees | Straight 90 degrees |
| Plastic Cannula<br>Length | 6mm | 6mm/9mm | 6mm/9mm |
| Plastic Cannula<br>Gauge | 28G | 25G | 26G |
| Tubing<br>Connection Type | Paradigm® and Luer<br>Lock | Paradigm® | Luer Lock |
| Tubing Length<br>(cm) | 61 and 107 | 45, 58, 81, and 109 | 45, 60, 75, and 105 |
| Connection<br>Positions | Multiple | Fixed | Multiple |
| Replacement<br>frequency | Disposable, replaced<br>every 72 hours | Disposable, replaced every 72<br>hours | Disposable, replaced every 72<br>hours |
| Provided Sterile | YES | YES | YES |
### Testing:
BD has performed non-clinical testing to demonstrate substantial equivalence. This testing includes device material biocompatibility (ISO 109993: Biological evaluation of medical devices), device functional performance (ISO 10555-1: Sterile single use intravascular catheters, ISO ID26: Medical Electrical Equipment. Part 2: Particular requirements for the safety of infusion pumps, ISO 1135-1: Sterilization of healthcare products), animal testing (United States Department of Agriculture (USDA) standards, US Animal Welfare Act), and human factor evaluations (IEC 62366:2007). Results of these tests were found to be acceptable and demonstrated that the BD FlowSmart™ set met requirements for its intended use and is as safe and as effective as its predicate devices.
### Conclusion:
The analysis and testing performed demonstrate the BD FlowSmart™ set is substantially equivalent to its predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.