K150055 · Bio-Detek, Inc. · MKJ · May 2, 2015 · Cardiovascular
Device Facts
Record ID
K150055
Device Name
OneStep Pediatric CPR Multi-Function Electrode
Applicant
Bio-Detek, Inc.
Product Code
MKJ · Cardiovascular
Decision Date
May 2, 2015
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.5310
Device Class
Class 3
Attributes
Therapeutic, Pediatric
Indications for Use
Defibrillation Cardioversion Noninvasive Pacing ECG Monitoring CPR Feedback For use with ZOLL® Defibrillators, such as: R Series X Series Trained Personnel only, Including: Physicians Nurses Paramedics Emergency Medical Technicians Cardiovascular Laboratory Technicians The OneStep Pediatric CPR electrodes are indicated for use on a patient less than 8 years of age or weighing less than 55 lbs (25kg).
Device Story
Multi-function electrode (MFE) for pediatric patients; connects to ZOLL R Series and X Series defibrillators. Inputs: electrical signals for ECG/defibrillation/pacing; motion data from dual accelerometers (anterior and posterior pads) for CPR depth feedback. Operation: dual-sensor technology compensates for soft/compressible surfaces during CPR, ensuring accurate depth measurement. Output: therapy delivery (defibrillation/pacing) and real-time CPR quality metrics displayed on host defibrillator. Used in hospital environments by trained clinicians. Benefits: enables accurate CPR feedback on non-firm surfaces, improving resuscitation quality for pediatric patients.
Clinical Evidence
No clinical data. Substantial equivalence supported by bench testing, including EMC (IEC 60601-1-2), electrical/mechanical safety (IEC 60601-1, IEC 60601-2-4), biocompatibility (ISO 10993-1, -5, -10), and usability testing.
Technological Characteristics
Solid hydrogel pads with pure tin conductive elements; PET backing; adhesive perimeter. Dual-sensor motion tracking (accelerometers) on anterior and posterior pads. Proprietary connector for ZOLL defibrillators. Supports self-test and expiration date identification. Standards: IEC 60601-1, IEC 60601-2-4, ISO 10993-1/5/10.
Indications for Use
Indicated for patients < 8 years of age or < 55 lbs (25kg) requiring defibrillation, cardioversion, noninvasive pacing, ECG monitoring, or CPR feedback. For use by trained personnel including physicians, nurses, paramedics, EMTs, and cardiovascular lab technicians.
Regulatory Classification
Identification
An automated external defibrillator (AED) system consists of an AED and those accessories necessary for the AED to detect and interpret an electrocardiogram and deliver an electrical shock (e.g., battery, pad electrode, adapter, and hardware key for pediatric use). An AED system analyzes the patient's electrocardiogram, interprets the cardiac rhythm, and automatically delivers an electrical shock (fully automated AED), or advises the user to deliver the shock (semi-automated or shock advisory AED) to treat ventricular fibrillation or pulseless ventricular tachycardia.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the seal of the Department of Health & Human Services - USA. The seal is circular, with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. In the center of the seal is a stylized image of three human profiles facing to the right, with a triple-line design above them.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
May 2, 2015
Bio-Detek, Inc. % Tanmay Shukla Sr. Regulatory Affairs Specialist Zoll Medical Corporation 269 Mill Road Chelmsford, Massachusetts 01824-4105
Re: K150055
Trade/Device Name: OneStep Pediatric CPR Multi-function Electrode Regulation Number: 21 CFR 870.5310 Regulation Name: Automated External Defibrillator Regulatory Class: Class III Product Code: MKJ, LIX Dated: April 3, 2015 Received: April 6, 2015
Dear Tanmay Shukla:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply
{1}------------------------------------------------
with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
Arthrell Stein
for Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Section 4 - Indications for Use
510(k) number (if known):
Device Name: OneStep Pediatric CPR Multi-Function Electrode
#### Intended Use:
- Defibrillation ●
- Cardioversion ●
- Noninvasive Pacing o
- ECG Monitoring ●
- CPR Feedback .
For use with ZOLL® Defibrillators, such as:
- R Series ●
- X Series .
Trained Personnel only, Including:
- Physicians ●
- Nurses ●
- Paramedics ●
- Emergency Medical Technicians ●
- Cardiovascular Laboratory Technicians ●
The OneStep Pediatric CPR electrodes are indicated for use on a patient less than 8 years of age or weighing less than 55 lbs (25kg).
Prescription Use X (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
# (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
{3}------------------------------------------------
# INCORPORATED
# 510(k) Summary:
Submitter's Name and Address: Bio-Detek, Inc. A Subsidiary of ZOLL®Medical Corporation 525 Narragansett Park Drive Pawtucket, RI 02861 Tel. (401) 729-1400
Contact Person: Tanmay Shukla Sr. Regulatory Affairs Specialist ZOLL Medical Corporation 269 Mill Road Chelmsford, MA 01824 Tel. (978) 421-9171 tshukla@zoll.com
Date Summary Prepared: January 9, 2015
Device Name / Proprietary Name: ZOLL OneStep Pediatric CPR Multi-Function Electrode (MFE)
Common Name: Automated External Defibrillator Multi-Function Electrodes
Classification Name: Multi-Function Electrode, with CPR Aid Accessory to an Automated External Defibrillator (21 CFR 870.5310 - Product Code MKJ)
Substantial Equivalence: ZOLL OneStep Pediatric CPR MFE (K120907) OneStep CPR II MFE (K133441)
{4}------------------------------------------------
## Description of Device:
As with the currently marketed OneStep Pediatric CPR Multi-Function Electrode (reviewed and cleared with K120907), the modified OneStep Pediatric CPR Multi-Function Electrode (MFE) is intended for use with ZOLL R Series and ZOLL X Series defibrillators for ECG Monitoring, Defibrillation, External Noninvasive Pacing, Cardioversion and CPR Feedback for use on patient less than 8 years of age or weighing less than 55lbs (25kg) in either the hospital environment. The currently marketed OneStep Pediatric CPR MFE provides users with accurate depth compression feedback when CPR is performed on a firm surface – a known recommended practice. A recent publication by the AHA acknowledges that "accelerometers are insensitive to mattress compression" and stresses the need for "continued development of optimal and widely available CPR monitoring." So. with the current submission. we are proposing to modify the OneStep Pediatric CPR MFE to incorporate a second motion sensor on the posterior electrode thereby allowing rescuers to obtain accurate depth compression feedback when CPR is performed on soft/ compressible surfaces. Dual sensor technology, utilized in the proposed OneStep Pediatric CPR MFE. has been reviewed and cleared by the agency with OneStep CPR II MFE (K133441).
As with the predicate device, the proposed electrode and wire harness with a ZOLL proprietary connector will have a pre-connect feature that enhances the user's ability to deliver immediate therapy. And like the predicate device, the OneStep Pediatric CPR electrode is also designed to support the defibrillator's self test and expiration date identification. Each electrode pad is structurally comprised of a solid hydrogel with a pure tin electrical conductive element having a polyethylene terephtalate (PET) backing with an adhesive perimeter suitable for coupling to patient skin during rescue and/or treatment.
Device Name: OneStep Pediatric CPR Multi-Function Electrode
#### Intended Use:
- Defibrillation
- Cardioversion
- Noninvasive Pacing ●
- ECG Monitoring o
- CPR Feedback ●
For use with ZOLL® Defibrillators, such as:
- R Series ●
- X Series .
By Trained Personnel only. Including:
- Physicians ●
- Nurses ●
- Paramedics
{5}------------------------------------------------
- Emergency Medical Technicians 0
- o Cardiovascular Laboratory Technicians
The OneStep Pediatric CPR electrodes are indicated for use on a patient less than 8 years of age or weighing less than 55 lbs (25kg).
#### Substantial Equivalence - Non-Clinical Evidence:
The proposed ZOLL OneStep Pediatric CPR Multi-Function Electrode (MFE) is identical to the currently marketed ZOLL OneStep Pediatric CPR MFE (K120907) except that the proposed electrode introduces an additional motion sensor on the posterior pad which allows the users to obtain accurate compression depth feedback while performing CPR on soft/ compressible surfaces. The change is similar to one implemented for OneStep CPR II MFE (K133441).The intended use of the OneStep Pediatric CPR Multi-Function Electrodes as described in the Indications for Use and labeling has not changed as a result of this submission.
#### Substantial Equivalence - Clinical Evidence:
N/A - Clinical evidence was not necessary to show substantial equivalence
## Comparison of Technological Characteristics
From a technological standpoint, the only difference between the proposed ZOLL OneStep Pediatric CPR Multi-Function Electrode (MFE) and the currently marketed ZOLL OneStep Pediatric CPR MFE (K120907) is the introduction of the dual sensor technology in the proposed device. Dual sensor technology, utilized in the proposed OneStep Pediatric CPR MFE, has been reviewed and cleared by the agency with OneStep CPR II MFE (K133441).
#### Performance Testing:
The proposed OneStep Pediatric CPR Multi-Function Electrode has been subjected to extensive performance testing to ensure the device meets all of its functional requirements and performance specifications as defined in applicable National/International recognized standards.
- EMC Testing in accordance with IEC 60601-1-2. .
- Electrical, mechanical, bio-compatibility and simulated use testing per the o applicable requirement of international recognized standards IEC 60601-1, IEC 60601-2-4, ISO 10993-1, ISO 10993-5 and ISO 10993-10.
- Testing to qualify the use of the electrode with R Series and X Series devices. ●
- Usability testing ●
{6}------------------------------------------------
# Conclusion:
The information provided in this 510(k) demonstrates that the proposed ZOLL OneStep Pediatric CPR Multi-Function Electrode (MFE) is substantially equivalent to indicated commercially distributed devices with regard to performance, safety and effectiveness.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.