K150050 · Siemens Medical Solutions USA, Inc. · IYN · Feb 9, 2015 · Radiology
Device Facts
Record ID
K150050
Device Name
ACUSON P500 Diagnostic Ultrasound System
Applicant
Siemens Medical Solutions USA, Inc.
Product Code
IYN · Radiology
Decision Date
Feb 9, 2015
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1550
Device Class
Class 2
Attributes
3rd-Party Reviewed
Indications for Use
The ACUSON P500 ultrasound imaging system is intended for the following applications: Fetal, Abdominal. Small Parts. OB/GYN. Pelvic. Cardiovascular. Musculoskeletal, Superficial Musculoskeletal, and Peripheral Vascular applications. The system also provides the ability to measure anatomical structures fetal, abdominal, small orqan, transrectal, transvaginal, cardiovascular, peripheral vessel, musculoskeletal (conventional), and musculoskeletal (superficial) and calculation packages that provide information to the clinician that may be used adjunctively with other medical data obtained by a physician for clinical diagnosis purposes.
Device Story
ACUSON P500 is a mobile, software-controlled diagnostic ultrasound system. It acquires harmonic ultrasound echo data via transducers and displays images in B-Mode, M-Mode, Pulsed/Continuous Wave Doppler, Color Doppler, Color M-mode, Tissue Doppler, and Power Doppler. Used in clinical settings by physicians to visualize anatomical structures and fluid flow. The system performs post-processing of received echoes to generate on-screen displays and provides measurement/calculation packages. Clinicians use these outputs adjunctively with other medical data to support clinical diagnosis. Benefits include non-invasive visualization of internal structures and blood flow, aiding in diagnostic decision-making.
Clinical Evidence
No clinical data. The device relies on bench testing, including acoustic output measurements, biocompatibility testing, and verification of electrical, thermal, and mechanical safety standards.
Technological Characteristics
Mobile diagnostic ultrasound system; 64-channel digital architecture. Supports B, M, PW/CW Doppler, Color, Power, and Tissue Harmonic Imaging modes. Patient contact materials tested to ISO 10993-1. Complies with UL 60601-1, IEC 60601-2-37, AIUM/NEMA UD-2/UD-3. Connectivity includes DICOM. Software developed per IEC 62304. Sterilization/disinfection per manufacturer guidelines.
Indications for Use
Indicated for diagnostic ultrasound imaging and fluid flow analysis in fetal, abdominal, small parts, OB/GYN, pelvic, cardiovascular, musculoskeletal, superficial musculoskeletal, and peripheral vascular applications. Intended for use by physicians for clinical diagnosis.
Regulatory Classification
Identification
An ultrasonic pulsed doppler imaging system is a device that combines the features of continuous wave doppler-effect technology with pulsed-echo effect technology and is intended to determine stationary body tissue characteristics, such as depth or location of tissue interfaces or dynamic tissue characteristics such as velocity of blood or tissue motion. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo is a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an image of three stylized human profiles facing to the right, with a design that resembles an abstract bird or wing shape above them.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
February 9, 2015
Siemens Medical Solutions USA, Inc. % Mr. Mark Job Responsible Third Party Official Regulatory Technology Services LLC 1394 25th Street, NW BUFFALO MN 55313
Re: K150050
Trade/Device Name: Acuson P500 Ultrasound System Regulation Number: 21 CFR 892.1550 Regulation Name: Ultrasonic pulsed doppler imaging system Regulatory Class: II Product Code: IYN, IYO, ITX Dated: January 5, 2015 Received: January 12, 2015
Dear Mr. Job:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Robert A Ochs
Robert Ochs. Ph.D. Acting Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K150050
Device Name Acuson P500 Ultrasound System
#### Indications for Use (Describe)
The ACUSON P500 ultrasound system is intended for diagnostic ultrasound imaging and fluid flow analysis for the following applications: Fetal, Abdominal, Small Parts, OB/GYN, Pelvic, Cardiovascular, Musculoskeletal, Superficial Musculoskeletal, and Peripheral Vascular applications.
The system also provides the ability to measure anatomical structures fetal, abdominal, small organ, transrectal, transvaginal, cardiovascular, peripheral vessel, musculoskeletal (conventional), and musculoskeletal (superficial) and calculation packages that provide information to the clinician that may be used adjunctively with other medical data obtained by a physician for clinical diagnosis purposes.
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary Prepared December 15, 2014
| Sponsor: | Siemens Medical Solutions, Inc.,<br>Ultrasound Division<br>685 East Middlefield Road<br>Mountain View, California 94043 | | |
|----------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------|--|--|
| Manufacturing Facility | SIEMENS LTD SEOUL<br>2nd -30 floor, 143, Sunhwan-ro,<br>Jungwon-gu, Seongnam-si, Gyeonggi-do,<br>Republic of Korea | | |
| Contact Person: | Shelly Pearce<br>Telephone:<br>(650) 279-0134<br>(650) 694-5580<br>Fax: | | |
| Device Name: | ACUSON P500 Ultrasound System | | |
| Common Name: | Diagnostic Ultrasound System with Accessories | | |
| Classification:<br>a market of the comments of the comments of | | | |
Regulatory Class: = Review Category: Tier II Classification Panel: Radiology
| Ultrasonic Pulsed Doppler Imaging System | FR # 892.1550 | Product Code 90-IYN |
|------------------------------------------|---------------|---------------------|
| Ultrasonic Pulsed Echo Imaging System | FR # 892.1560 | Product Code 90-IYO |
| Diagnostic Ultrasound Transducer | FR # 892.1570 | Product Code 90-ITX |
# A. Legally Marketed Predicate Devices
The ACUSON P500 Ultrasound System is a multi-purpose diagnostic ultrasound system with accessories and proprietary software, and is substantially equivalent to our current product, the ACUSON X300 ultrasound system (K121699) and ACUSON S2000 ultrasound system (K140959).
# B. Device Description:
The ACUSON P500 is a multi-purpose mobile, software controlled, diagnostic ultrasound system with an on-screen display for thermal and mechanical indices related to potential bioeffect mechanisms. Its function is to acquire harmonic ultrasound echo data and display it in B-Mode, M-Mode, Pulsed (PW) Doppler Mode, Continuous (CW) Doppler Mode, Color Doppler Mode, Color M mode, Tissue Doppler Image, Power (Amplitude) Doppler Mode and a combination of modes on a Flat Panel Display.
{4}------------------------------------------------
# C. Intended Use
The ACUSON P500 ultrasound imaging system is intended for the following applications: Fetal, Abdominal. Small Parts. OB/GYN. Pelvic. Cardiovascular. Musculoskeletal, Superficial Musculoskeletal, and Peripheral Vascular applications.
The system also provides the ability to measure anatomical structures fetal, abdominal, small orqan, transrectal, transvaginal, cardiovascular, peripheral vessel, musculoskeletal (conventional), and musculoskeletal (superficial) and calculation packages that provide information to the clinician that may be used adjunctively with other medical data obtained by a physician for clinical diagnosis purposes.
# D. Substantial Equivalence
The ACUSON P500 is substantially equivalent to the ACUSON X300, cleared via K121699, and ACUSON S2000. All systems transmit ultrasonic energy into patients, then perform post processing of received echoes to generate onscreen display of anatomic structures and fluid flow within the body. All systems allow for specialized measurements of structures and flow, and calculations.
| | Feature / Characteristic | Predicate Device<br>ACUSON X300<br>(K121699) | Predicate Device<br>ACUSON S2000<br>(K140959) | Submission Device<br>ACUSON P500 |
|--------------------------------------------------------------------------------------------|----------------------------------------------|-----------------------------------------------|-----------------------------------------------|----------------------------------|
| | Indications for Use: | | | |
| ■ | Fetal | √ | √ | √ |
| ■ | Abdominal | √ | √ | √ |
| ■ | Small Organ | √ | √ | √ |
| ■ | Cardiac | √ | √ | √ |
| ■ | Transrectal | √ | √ | √ |
| ■ | Transvaginal | √ | √ | √ |
| ■ | Peripheral vessel | √ | √ | √ |
| ■ | Musculo-skeletal (conventional) | √ | √ | √ |
| ■ | Musculo-skeletal (superficial) | √ | √ | √ |
| ■ | Emergency Medicine | √ | √ | √ |
| | Center Frequencies Supported: | | | |
| ■ | 2.0 MHz | √ | √ | √ |
| ■ | 2.5 MHz | √ | √ | √ |
| ■ | 3.0 MHz | √ | √ | √ |
| ■ | 3.5 MHz | √ | √ | √ |
| ■ | 4.0 MHz | √ | √ | √ |
| ■ | 5.0 MHz | √ | √ | √ |
| ■ | 5.5 MHz | √ | √ | √ |
| ■ | 6.0 MHz | √ | √ | √ |
| ■ | 6.5 MHz | √ | √ | √ |
| ■ | 7.5 MHz | √ | √ | √ |
| ■ | 8.0 MHz | √ | √ | √ |
| ■ | 9.0 MHz | √ | √ | √ |
| ■ | 10.0 MHz | √ | √ | √ |
| Feature / Characteristic | Predicate Device<br>ACUSON X300<br>(K121699) | Predicate Device<br>ACUSON S2000<br>(K140959) | Submission Device<br>ACUSON P500 | |
| 11.0 MHz | √ | √ | √ | |
| Modes: | | | | |
| B | √ | √ | √ | |
| M | √ | √ | √ | |
| PWD (Pulsed Wave Doppler) | √ | √ | √ | |
| SCW (Steerable CW) | √ | √ | √ | |
| D (Color Doppler) | √ | √ | √ | |
| Power Doppler<br>(Amplitude Doppler) | √ | √ | √ | |
| Combined (BM, BC, BCM, BCD) | √ | √ | √ | |
| THI (Tissue Harmonic Imaging) | √ | √ | √ | |
| M-THI | √ | √ | √ | |
| Features: | | | | |
| 3D Imaging (3-Scape) | √ | √ | -- | |
| 3D Measurements | √ | √ | -- | |
| 4D Basic Imaging (fourSight 4D) | √ | √ | -- | |
| Panoramic 2D Imaging (SieScape) | √ | √ | -- | |
| Cardiac Imaging<br>physiological signal display<br>multiplane TEE<br>fourSight TEE Imaging | √ | √ | -- | |
| Dual-Beam Processing | √ | √ | √ | |
| Quad-Beam Processing | √ | √ | √ | |
| SynAps | √ | √ | -- | |
| Clip Capture | √ | √ | √ | |
| Tissue Grayscale Optimization (TGO) | √ | √ | -- | |
| Spectral DTI | √ | √ | -- | |
| Axius OB Automated Calipers | √ | √ | -- | |
| Steerable CW | √ | √ | √ | |
| Stress Echo | √ | √ | -- | |
| 2D(color) DTI | √ | √ | -- | |
| Image Arena reporting S/W | √ | √ | -- | |
| DIMAQ (PIMS Workplace) | √ | √ | √ | |
| Vascular Enhancement (Clarify VE) | √ | √ | -- | |
| eSielmage | -- | √ | √ | |
| Advance SieClear | -- | √ | √ | |
| Multiple Frequency Imaging(MultiHertz) | √ | √ | √ | |
| Digital Architecture | √ | √ | √ | |
| Fully integrated DICOM | √ | √ | √ | |
| Dynamic TCE | √ | √ | √ | |
| AHP | √ | √ | -- | |
| Monitor: 15.4"WXGA (FPD) | √ (17" FPD) | √ (19" FPD) | √ | |
| Wireless | √ | √ | √ | |
| Feature / Characteristic | Predicate Device<br>ACUSON X300<br>(K121699) | Predicate Device<br>ACUSON S2000<br>(K140959) | Submission Device<br>ACUSON P500 | |
| # Channels | 128 | 192 | 64 | |
| Output Display Standard<br>(Track 3) | √ | √ | √ | |
| Patient Contact Materials | Tested to ISO 10993-1 | Tested to ISO 10993-1 | Tested to ISO 10993-1 | |
| UL60601-1 Certified | √ | √ | √ | |
The submission device is substantially equivalent to the predicate with regard to both intended use and technological characteristics.
{5}------------------------------------------------
{6}------------------------------------------------
# E. A brief discussion of nonclinical tests submitted, referenced, or relied on in the 510(k) for a determination of substantial equivalence
The device has been evaluated for acoustic output, biocompatibility, cleaning and disinfection effectiveness as well as thermal, electrical, electromagnetic and mechanical safety and has been found to conform with applicable medical device safety standards. The system complies with the following voluntary standards:
- UL 60601-1, Safety Requirements for Medical Equipment 国
- IEC 60601-2-37 Diagnostic Ultrasound Safety Standards 보
- CSA C22.2 No. 601-1. Safety Requirements for Medical Equipment l
- AIUM/NEMA UD-3, Standard for Real Time Display of Thermal and Mechanical 전 Acoustic Output Indices on Diagnostic Ultrasound Equipment
- 피 AIUM/NEMA UD-2, Acoustic Output Measurement Standard for Diagnostic Ultrasound
- 93/42/EEC Medical Devices Directive 별
- 트 Safety and EMC Requirements for Medical Equipment
- EN/IEC 60601-1 L
- I EN/IEC 60601-1-1
- 트 EN/IEC 60601-1-2
- 트 EN/IEC 60601-1-6
- . EN/IEC 62304
- EN/IEC 62366 家
- I EN/IEC 60601-2-18
- 트 EN/IEC 60601-2-25
- ISO 10993-1 Biocompatibility 피
Cleared patient contact materials, electrical and mechanical safety are unchanged.
# F. A summary discussion of the clinical tests submitted, referenced, or relied on for a determination of substantial equivalence.
Since the P500 uses the same technology and principles as existing devices, clinical data is not required.
### G. Summary
Intended uses and other key features are consistent with traditional clinical practice and FDA guidelines. The design and development process of the manufacturer conforms with 21 CFR 820 Quality System Requlation and ISO 13485:2003 quality system standards. The product is designed to conform with applicable medical device safety standards and compliance is verified through independent evaluation with ongoing factory surveillance. Diagnostic ultrasound has accumulated a long history of safe and effective performance. Therefore it is the opinion of Siemens Medical that the P500 is substantially equivalent with respect to safety and effectiveness to devices currently cleared for market.
{7}------------------------------------------------
The ACUSON P500 is verified and validated according to the company's design control process.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.