The AD-2129A TENS device is intended for temporary relief of pain associated with sore and aching muscles due to strain from exercise or normal household and work activities. It is also intended for symptomatic relief and management of chronic, intractable pain and relief of pain associated with arthritis.
Device Story
AD-2129A is a microprocessor-controlled Transcutaneous Electrical Nerve Stimulator (TENS) for OTC home use. Device inputs include user-selected operation modes and pulse output strength via buttons. The device transforms these inputs into programmed electrical pulses delivered through electrode adhesive pads to the skin, causing minimal muscle stimulation. Output is displayed on an LCD (mode, strength, time). The device provides symptomatic pain relief by stimulating nerves/muscles. It features automatic no-load trip and automatic shut-off. It is intended for patients seeking relief from muscular pain, chronic intractable pain, or arthritis-related pain.
Clinical Evidence
Bench testing only. No clinical data provided. Testing included biocompatibility (ISO 10993-5, ISO 10993-10), electromagnetic compatibility (IEC 60601-1-2), and electrical safety (IEC 60601-1, IEC 60601-2-10).
Technological Characteristics
Microprocessor-controlled TENS; ABS housing; 2 output channels; 8 modes; monophasic rectangular waveform; frequency 2-125Hz; pulse duration 60-125us. Powered by 4x1.5V AAA batteries. Self-adhesive electrodes (substrate and wire). Standards: IEC 60601-1, IEC 60601-2-10, IEC 60601-1-2, ISO 10993-1, -5, -10.
Indications for Use
Indicated for temporary relief of pain associated with sore/aching muscles due to exercise or normal household/work activities; symptomatic relief and management of chronic, intractable pain; and relief of pain associated with arthritis. For OTC use.
Regulatory Classification
Identification
A transcutaneous electrical nerve stimulator for pain relief is a device used to apply an electrical current to electrodes on a patient's skin to treat pain.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the seal of the U.S. Department of Health and Human Services. The seal features a stylized representation of three human profiles facing right, stacked on top of each other. The profiles are black and are set against a white background. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the left side of the seal.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
August 14, 2015
Andon Health Co., Ltd. Liu Yi President No. 3, Jin Ping Street, Ya An Road, Nankai District Tianjin, CN 300190
Re: K150043
Trade/Device Name: AD-2129A Transcutaneous Electrical Nerve Stimulators (TENS) Regulation Number: 21 CFR 882.5890 Regulation Name: Transcutaneous Electrical Nerve Stimulator for Pain Relief Regulatory Class: Class II Product Code: NUH, NYN Dated: July 14, 2015 Received: July 16, 2015
Dear Liu Yi:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in
{1}------------------------------------------------
the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# Michael J. Hoffmann -S
for Carlos L. Peña, PhD, MS Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
#### Indications for Use
510(k) Number (if known) K150043
Device Name
AD-2129A Transcutaneous Electrical Nerve Stimulators (TENS)
Indications for Use (Describe)
The AD-2129A TENS device is intended for temporary relief of pain associated with sore and aching muscles due to strain from exercise or normal household and work activities. It is also intended for symptomatic relief and management of chronic, intractable pain and relief of pain associated with arthritis.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
|-------------------------------------------------|--|
Prescription Use (Part 21 CFR 801 Subpart D)
|X | Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirement of SMDA and 21 CFR 807.92.
### 1.0 submitter's information
| Name: | Andon Health Co., Ltd. |
|----------------------|--------------------------------------------------------------------------|
| Address: | No 3, Jinping Street Ya An Road, Nankai District, Tianjin,<br>P.R. China |
| Phone number: | 86-22-6052 6161 |
| Fax number: | 86-22-6052 6162 |
| Contact: | Liu Yi |
| Date of Application: | 1/7/2015 |
#### 2.0 Device name
Classification Name: Stimulator, Nerve, Transcutaneous, Over-The-Counter Device name: Transcutaneous Electrical Nerve Stimulators (TENS) Model: AD-2129A
## 3.0 Classification
Production code: NUH, NYN Regulation number: 21 CFR 882.5890 Classification: II Panel: Neurology
#### 4.0 Predicate device information
| Information | Predicate 1 | Predicate 2 |
|----------------|---------------|---------------------|
| 510(k) number: | K131159 | K140168 |
| Manufacturer | Chattem, Inc. | EasyMed Instruments |
{4}------------------------------------------------
| | | Co., Ltd. |
|------------------------|-------------------------------------|-------------------------------------|
| Trade/Proprietary Name | Smart Relief | EasyStim TN28_OTC |
| Classification Name | Transcutaneous Nerve<br>Stimulator. | Transcutaneous Nerve<br>Stimulator. |
| Regulation number | 882.5890 | 21 CFR 882.5890 |
| Classification | Class II | Class II |
| Product Code | NUH, NYN | NUH |
#### 5.0 Intended use
The AD-2129A TENS device is intended for temporary relief of pain associated with sore and aching muscles due to strain from exercise or normal household and work activities, It is also intended for symptomatic relief and management of chronic, intractable pain and relief of pain associated with arthritis.
#### 6.0 Device description
The AD-2129A Transcutaneous Electrical Nerve Stimulator(TENS) is transcutaneous electrical nerve stimulator for relief of muscular pain and sold without prescription.
The device consists of a microprocessor, buttons, electrical pads, and display. Keys can control the device to choose the operation modes, adjust pulse output strength, then the channel that effectively transfers your desired choice of programmed electrical pulses directly through electrode adhesive pads to the suggested area of the body where the electrodes are placed, causing minimal muscle stimulation. The LCD display can show user the mode and strength chosen and other information like date and time
Self-adhesive electrodes are used in this devices, and they are intended for use as the disposable, conductive adhesive interface between the patient's skin and the Electrical Stimulator. The electrodes is for OTC use, and mainly consists of substrate and wire.
#### 7.0 Performance summary
TENS conforms to the following standards:
{5}------------------------------------------------
- · IEC 60601-1, Medical Electrical Equipment Part 1: General Requirements for Safety, 2005.
- · IEC 60601-2-10, Medical electrical equipment Part 2: Particular requirements for the basic safety and essential performance of nerve and muscle stimulators
- · IEC 60601-1-2, Medical Electrical Equipment Part 1-2: General Requirements for Safety - Collateral standard: Electromagnetic Compatibility - Requirements and Tests, 2007.
- · ISO 10993: Biological evaluation of medical devices Part 1: Evaluation and testing within a risk management process.
- · ISO 10993: Biological evaluation of medical devices Part 5: Test for in vitro cytotoxicity.
- · ISO 10993: Biological evaluation of medical devices Part 10: Test for irritation and skin sensitization.
### 8.0 Non-clinical Testing Summary
The following testing was performed on the TENS devices in accordance with the requirements of the design control regulations and established quality assurance procedures.
#### (1) Biocompatibility of materials
When use the TENS device, one side of the electrode will contact the user, and the materials used on this side is tested according to ISO 10993-5 and ISO 10993-10, and the result shows it meet the applicable requirements.
#### (2) Electromagnetic Compatibility
Electromagnetic Compatibility test has been performed on the TENS devices according to the identical standard of IEC 60601-1-2, and the test result show that, the device meet all the applicable requirements.
#### (3) Electrical Safety Testing
Electrical Safety Test has been performed according to IEC 60601-1, and the test result shows that, the device meet all the applicable requirements. Also,
{6}------------------------------------------------
particular safety test has been performed according to IEC 60601-2-10.
### 9.0 Comparison to the predicate device and the conclusion
The applicant device TENS is substantially equivalent to Smart Relief (K131159) and EasyStim TN28_OTC(K140168).
| Similarities and differences comparison | | | |
|----------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Characteristics | Subject device | Predicate device(1) | Predicate device(2) |
| Product name | AD-2129A TENS device | Smart Relief | EasyStim TN28_OTC |
| 510(K)number | To be assigned | K131159 | K140168 |
| Product code | NUH, NYN | NUH, NYN | NUH |
| Regulation No. | 21 CFR 882.5890 | 21 CFR 882.5890 | 21 CFR 882.5890 |
| Intended use | The device is intended<br>for temporary relief of pain<br>associated with sore and<br>aching muscles due to<br>strain from exercise or<br>normal household and<br>work activities, It is also<br>intended for symptomatic<br>relief and management of<br>chronic, intractable pain<br>and relief of pain<br>associated with arthritis. | To be used for temporary<br>relief of pain associated with<br>sore and aching muscles<br>due to strain from exercise<br>or normal household and<br>work activities.<br>To be used for symptomatic<br>relief and management of<br>chronic, intractable pain and<br>relief of pain associated with<br>arthritis. | This device is intended for<br>the relief of pain associated<br>with sore or aching muscles<br>of the lower back, arms, or<br>legs due to strain from<br>exercise or normal<br>household and work<br>activities. |
| Presentation or<br>OTC | OTC | OTC | OTC |
| Environment of<br>use | Home use | Not specified | Clinics, hospital and home<br>environments |
| Number of<br>Outputs mode | 8 | 1 | 8 |
| Number of<br>Outputs | 2 | 1 | 2 |
| Waveform | Monophasic rectangular | Asymmetrical Biphasic<br>rectangular | Biphasic rectangular<br>Monophasic rectangular |
| Maximum<br>output<br>voltage(max)<br>500 Ω, 2 kΩ,<br>and 10 kΩ | 500 Ω: 48V<br>2 kΩ: 91.2V<br>10 kΩ: 46V | Not specified | 500 Ω: 68V<br>2 kΩ: 102V<br>10 kΩ: 110V |
| Maximum | 500 Ω: 96mA | | 500 Ω: 133mA |
| | | | |
| Output<br>Current 500 Ω,<br>2 kΩ, and 10<br>kΩ | 2KΩ: 4.33mA<br>10 kΩ:0.44 mA | Not specified | 2KΩ: 51mA<br>10 kΩ:11 mA |
| Maximum<br>Phase Charge,<br>(μC) @ 500 Ω | 11.92uC | Not specified | 20.02uC |
| Maximum<br>Average<br>Current(500ohm) | 8.64mA | Not specified | 3.0375mA |
| Maximum<br>Current Density,<br>(mA/cm² @ 500 Ω) | 0.38mA/cm² | Not specified | 0.188mA/cm² |
| Maximum<br>Average Power<br>Density, (W/cm²<br>@ 500 Ω | 1.62mW/cm² | Not specified | 7.52mW/cm² |
| Frequency<br>(Hz) | 2-125HZ | 1-100 Hz | 1-150Hz |
| Pulse<br>Duration(us) | 60-125 | Not specified | 10-250us, in steps of 50us |
| Burst Mode | None | None | Yes |
| Timer<br>range(min) | 15 minutes for all<br>programs | Not specified | 20min, 25min, 30min, 40min<br>depending on preset<br>program |
| Indication<br>display: | -On/Off status<br>-Output strength<br>-Output mode<br>-Time to cut-off | Not specified | -On/Off status<br>-Low battery<br>-Voltage/Current level<br>-Output mode<br>-Time to cut-off |
| Power Source | 4×1.5 size AAA | 3 V lithium battery | 2 Alkaline AA 1.5V (LR6)<br>Batteries |
| Dimension | 140mm×63mm×30mm | 64mmX38mmX13mm | 66mm×136mm×30.7mm |
| Weight | 88g (exclude battery) | 20g | 146.5 grams |
| Housing<br>Materials | ABS | ABS | ABS |
| Microprocessor<br>control | Yes | Yes | Yes |
| Automatic<br>Overload trip | No | Not specified | Yes |
| Automatic<br>no-load trip | Yes | Not specified | Yes |
| Automatic<br>shut-off | Yes | Not specified | Yes |
| User override<br>control | No | Not specified | Yes |
| Electrode<br>compliance<br>with 21 CFR<br>898 | Yes | Yes | Yes |
| Electrode cable | Yes | No | Yes |
{7}------------------------------------------------
{8}------------------------------------------------
The new device AD-2129A TENS has the same intended use, design, technological characteristics as the predicate device K131159 and K140168. Only pulse frequency, pulse strength and the appearance are different, but the Electrical Safety test and the EMC test in this submission provides demonstrated that these small differences do not raise any new questions of safety and effectiveness to the new devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.