The sponsor utilized four years of post-market clinical performance data and a clinical study manuscript to demonstrate that the device reduces complications (specifically POAF and retained blood) compared to other products in routine clinical settings.
Reduction in Interventions for Post Operative Effusions and Atrial Fibrillation with Active Clearance of Chest Drainage Catheters; Clinical observational/outcomes study; Follow-up/Duration: 4 years (post-market data)
Patients undergoing cardiothoracic surgery; Number of Sites: 2 (Heart Center, Paracelsus Medical University, Germany; St. Charles Medical Center, USA)
Other chest drainage products
Reduction in POAF (Post-Operative Atrial Fibrillation), reduction in retained blood, and reduction in interventions for post-operative effusions
Indications for Use
The PleuraFlow System is indicated for use during cardiothoracic surgical procedures and chest trauma. Its active clearance technology proactively removes clots formed inside the chest tube to prevent or minimize chest tube occlusion with clot. A patent chest tube enables evacuation of blood and fluid from the operative site after closure of the surgical wound and reduces retained blood. The product is indicated for adult and pediatric patients including infant, preadolescent and adolescent patients under clinical settings.
Device Story
PleuraFlow System is a wound drain catheter system designed to maintain chest tube patency. It consists of a silicone chest tube (20FR-32FR) and a clearance apparatus. The apparatus includes a guide tube and a PTFE-coated clearance wire with a 105-degree angled distal loop. A magnetic shuttle advances the clearance wire/loop within the chest tube to mechanically break up or prevent clot formation and tube occlusion. Used in clinical settings by healthcare providers to ensure continuous drainage of blood and fluid post-surgery. By preventing chest tube clogging, the device reduces retained blood, which is associated with complications like post-operative atrial fibrillation (POAF).
Clinical Evidence
No new clinical trials were required for this submission. Evidence relies on previously cleared design validation and verification (K093565) and four years of post-market performance data. Literature cited indicates that active clearance of chest drainage catheters reduces retained blood and post-operative atrial fibrillation (POAF) interventions.
Technological Characteristics
System comprises a silicone chest tube (20FR, 24FR, 28FR, 32FR) with barium stripe and a clearance apparatus (PTFE-coated wire with distal loop). Operates via manual mechanical advancement of the clearance wire using a magnetic shuttle. No electronic energy source or software algorithms involved.
Indications for Use
Indicated for adult and pediatric patients (infant, preadolescent, adolescent) undergoing cardiothoracic surgery or experiencing chest trauma requiring evacuation of blood and fluid from the operative site.
Regulatory Classification
Identification
A powered suction pump is a portable, AC-powered or compressed air-powered device intended to be used to remove infectious materials from wounds or fluids from a patient's airway or respiratory support system. The device may be used during surgery in the operating room or at the patient's bedside. The device may include a microbial filter.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features the department's name in a circular arrangement around a symbol. The symbol consists of three stylized human profiles facing right, with a stylized eagle or bird-like element above them.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
June 2, 2015
ClearFlow Incorporated Dr. Dov Gal, DVM Vice President Regulatory Affairs, Quality Assurance and Clinical Affairs 1630 South Sunkist Street, Suite E Anaheim, California 92806
Re: K150042
Trade/Device Name: PleuraFlow® System Regulation Number: 21 CFR 878.4780 Regulation Name: Powered suction pump Regulatory Class: Class II Product Code: OTK, GBX Dated: April 14, 2015 Received: April 20, 2015
Dear Dr. Gal:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{1}------------------------------------------------
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
## Jennifer R. Stevenson -S
For Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
### Indications for Use
510(k) Number (if known) 150042
Device Name PleuraFlow® System
Indications for Use (Describe)
The PleuraFlow System is indicated for use during cardical procedures and chest trauma. Its active clearance technology proactively removes clots formed inside to prevent or minimize chest tube occlusion with clot. A patent chest tube enables evacuation of blood and fluid from the operative site after closure of the surgical wound and reduces retained blood. The product is indicated for adult and patients including infant, preadolescent and adolescent patients under clinical settings.
Type of Use (Select one or both, as applicable)
[X] Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# Section 5: 510(k) Summary
The following information is provided as required by 21 CFR § 807.87 for PleuraFlow® System 510(k) premarket notification. In response to the Safe Medical Devices Act of 1990, the following is a summary of the safety and effectiveness information upon which the substantial equivalence determination is based.
| Date of Submission: | April 14, 2015 |
|--------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Applicant: | Clear Flow, Inc.<br>1630 S. Sunkist St., Suite E<br>Anaheim CA<br>92806 |
| Primary Contact Person: | Dov Gal, DVM<br>ClearFlow, Inc.<br>1630 S. Sunkist St., Suite E<br>Anaheim CA<br>92806<br>Phone: 714-916-5014<br>Fax: 714-916-5019<br>Email: dgal@clearflow.com |
| CEO/President: | Paul Molloy<br>ClearFlow, Inc.<br>1630 S. Sunkist St., Suite E<br>Anaheim CA<br>92806<br>Phone: 714-905-5271<br>Fax: 714-916-5019<br>Email: pmolloy@clearflow.com |
| Device Proprietary Name: | PleuraFlow® System |
| Device Common Name: | Introduction/drainage; wound drain catheter system. |
{4}------------------------------------------------
### Regulatory Class and Name: Class II, Powered Suction Pump
#### Product Codes: OTK and GBX
- Indication For Use: The PleuraFlow System is indicated for use during cardiothoracic surgical procedures and chest trauma. Its active clearance technology proactively removes clots formed inside the chest tube to prevent or minimize chest tube occlusion with clot. A patent chest tube enables evacuation of blood and fluid from the operative site after closure of the surgical wound and reduces retained blood. The product is indicated for adult and pediatric patients including infant, preadolescent and adolescent patients under clinical settings.
- Predicate Device: Predicate device is the PleuraFlow Catheter System (K093565) by ClearFlow, Inc., AKA Clear Systems, Inc.
- The PleuraFlow System is comprised of a silicone Chest Tube Device Description: and a Clearance Apparatus. The PleuraFlow Chest Tube is available in four (4) standard sizes (20FR, 24FR, 28FR and 32FR). Each Chest Tube has a Cut Length of 19 inches (48.3 cm) with graduated measurements in centimeters from the distal eyelet. Each Chest Tube has a barium stripe to facilitate visualization. The Chest Tube is connected to a Clearance Apparatus, which is connected to the tubing from the drainage canister. The Clearance Apparatus consists of a Guide Tube and a PTFE-coated Clearance Wire with a Loop set on its distal end, bent at a 105-degree angle. The Clearance Apparatus is advanced into the PleuraFlow Chest Tube using a magnetic Shuttle. When indicated, the Clearance Wire and Loop is advanced and retracted within the PleuraFlow Chest Tube to proactively prevent or break up and clear any tube obstructions or clogging to keep the tube patent.
{5}------------------------------------------------
- The PleuraFlow System is indicated for use during Indications for Use: cardiothoracic surgical procedures and chest trauma. Its active clearance technology proactively removes clots formed inside the chest tube to prevent or minimize chest tube occlusion with clot. A patent chest tube enables evacuation of blood and fluid from the operative site after closure of the surgical wound and reduces retained blood. The product is indicated for adult and pediatric patients including infant, preadolescent and adolescent patients under clinical settings.
- The safety and effectiveness of the PleuraFlow System has Performance Data: been previously demonstrated through design validation and verification that were cleared under 510(k) (K093565). Performance has been further demonstrated through post-market data.
Use of the device over the last four (4) years has shown that the product has significantly reduced the complications for patients who have received treatment with PleuraFlow versus other products during and after surgery.
Studies have linked both chest tube clogging and retained pericardial blood with POAF.2, 10 There is a substantial body of literature illustrating that shunting blood through pericardial windows to divert the blood to the pleural spaces can reduce POAF.1% demonstrates by adhering to a protocol developed to maximize chest tube patency, POAF can be reduced. Consistent with prior studies, this link appears to be related to reducing RBS. A manuscript titled "Reduction in Interventions for Post Operative Effusions and Atrial Fibrillation with Active Clearance of Chest Drainage Catheters" was submitted to the Journal of Thoracic and Cardiovascular Surgery. Authors are Sirch J.1, Ledwon M.1, Püski T.1, Grossmann I.1, Boyle EM.2, Pfeiffer S.1,Fischlein T.1 From the following institutions:
{6}------------------------------------------------
| 1. Cardiac Surgery, Heart Center, Paracelsus Medical | |
|-------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| University, Nurenberg, Germany, | |
| 2. St. Charles Medical Center, Bend, Oregon, United States. | |
| Conclusion: | The evaluation of the PleuraFlow System does not raise any<br>additional concerns regarding safety and effectiveness and may<br>therefore be considered substantially equivalent to the<br>predicate device. |
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.