← Product Code [PIO](/productcode/PIO) · K150034

# AMT Enteral Transition Adapters (K150034)

_Applied Medical Technology, Inc. · PIO · Jun 9, 2015 · Gastroenterology, Urology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K150034

## Device Facts

- **Applicant:** Applied Medical Technology, Inc.
- **Product Code:** [PIO](/productcode/PIO.md)
- **Decision Date:** Jun 9, 2015
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 876.5980
- **Device Class:** Class 2
- **Review Panel:** Gastroenterology, Urology

## Indications for Use

The AMT Enteral Transition Adapters are intended to facilitate enteral specific connections between AAMI/CN3(PS) compliant connectors and non ISO 80369-1 compliant legacy enteral connectors.

## Device Story

AMT Enteral Transition Adapters facilitate connectivity between new ISO 80369-1 compliant (Enfit) enteral devices and legacy enteral feeding equipment (luer slip, catheter tip, bolus). Device consists of five adapter configurations: Male Enfit to Male Luer, Female Enfit to Female Luer, Female Enfit to Y-Port, Female Enfit to Bolus, and Male Enfit to Christmas-tree. Adapters are used by clinicians or patients in clinical or home settings to ensure compatibility during the industry-wide transition to Enfit connectors. By enabling use of existing legacy inventory with new feeding sets, the adapters prevent supply chain disruption and ensure continuity of enteral nutrition delivery. Adapters are provided non-sterile for single-user use.

## Clinical Evidence

Bench testing only. Performance testing included tensile strength, fluid leakage, stress cracking, resistance to separation (axial load/unscrewing), resistance to overriding, disconnection from unscrewing, and flow testing. Biocompatibility data was leveraged from previous 510(k) K123716 for permanent (>30 days) and prolonged (<29 days) indirect mucosal contact.

## Technological Characteristics

Non-sterile, single-user adapters. Materials are DEHP and Latex free. Five configurations: Male Enfit to Male Luer, Female Enfit to Female Luer, Female Enfit to Y-Port, Female Enfit to Bolus, and Male Enfit to Christmas-tree. Mechanical connection interface designed to bridge ISO 80369-1 compliant connectors and legacy luer/bolus connectors.

## Regulatory Identification

A gastrointestinal tube and accessories is a device that consists of flexible or semi-rigid tubing used for instilling fluids into, withdrawing fluids from, splinting, or suppressing bleeding of the alimentary tract. This device may incorporate an integral inflatable balloon for retention or hemostasis. This generic type of device includes the hemostatic bag, irrigation and aspiration catheter (gastric, colonic, etc.), rectal catheter, sterile infant gavage set, gastrointestinal string and tubes to locate internal bleeding, double lumen tube for intestinal decompression or intubation, feeding tube, gastroenterostomy tube, Levine tube, nasogastric tube, single lumen tube with mercury weight balloon for intestinal intubation or decompression, and gastro-urological irrigation tray (for gastrological use).

## Special Controls

*Classification.* (1) Class II (special controls). The barium enema retention catheter and tip with or without a bag that is a gastrointestinal tube and accessory or a gastronomy tube holder accessory is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I (general controls) for the dissolvable nasogastric feed tube guide for the nasogastric tube. The class I device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 876.9.

## Predicate Devices

- Enteral Specific Transition Connectors (Cedic S.R.L.) ([K140581](/device/K140581.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol that resembles a stylized human figure or a bird in flight, composed of three overlapping profiles.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

June 9, 2015

Applied Medical Technology, Inc. Joshua D. Meinke OA/Regulatory Supervisor 8006 Katherine Boulevard Brecksville, OH 44141

Re: K150034

> Trade/Device Name: AMT Enteral Transition Adapters Regulation Number: 21 CFR§ 876.5980 Regulation Name: Gastrointestinal tube and accessories Regulatory Class: II Product Code: PIO Dated: March 16. 2015 Received: March 18, 2015

Dear Joshua D. Meinke,

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.

{1}------------------------------------------------

You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely vours.

Image /page/1/Picture/8 description: The image shows the text "Herbert P. Lerner -S" in a simple, sans-serif font. The text is black and appears to be the main focus of the image. The background is a light gray, and there is a faint watermark-like design behind the text.

Benjamin R. Fisher, Ph.D. for Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

## Indications for Use

510(k) Number (if known) K150034

#### Device Name AMT Enteral Transition Adapters

#### Indications for Use (Describe)

The AMT Enteral Transition Adapters are intended to facilitate enteral specific connections between AAMI/CN3(PS) compliant connectors and non ISO 80369-1 compliant legacy enteral connectors.

|×| Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# SECTION - 5

# 510(K) Summary

AMT Enteral Transition Adapters

| Date Prepared:                    | June 8, 2015                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        |
|-----------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Submitter:                        | Joshua D. Meinke<br>QA/Regulatory Supervisor<br>Applied Medical Technology, Inc.<br>8006 Katherine Boulevard<br>Brecksville, OH 44141<br>Phone: 440-717-4252<br>Fax: 440-717-4200<br>Email: Josh.Meinke@appliedmedical.net<br>Contact Person: Joshua Meinke                                                                                                                                                                                                                                                                                                                                                                                         |
| Device<br>Information:            | Trade Name: AMT Enteral Transition Adapters<br>Common Name: Enteral Transition Adapter<br>Classification Name: Gastroenterology and Urology (21 CFR 876.5980)<br>Regulatory Class: II<br>Product Code: PIO<br>FDA Identification: Facilitate enteral specific connections between<br>AAMI/CN3(PS) compliant connectors and non ISO 80369-1<br>compliant legacy enteral connectors.                                                                                                                                                                                                                                                                  |
| Predicate<br>Device:              | Enteral Specific Transition Connectors (Cedic S.R.L.)<br>(cleared under K140581)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |
| Indications for<br>Use:           | The AMT Enteral Transition Adapters are intended to facilitate enteral specific<br>connections between AAMI/CN3(PS) compliant connectors and non ISO 80369-1<br>compliant legacy enteral connectors.                                                                                                                                                                                                                                                                                                                                                                                                                                                |
| Device<br>Description:            | New standard ISO-80369-1 is pushing feeding tube manufacturers to transition<br>current extension set designs to incorporate enteral only connectors. Most enteral<br>feeding devices currently on the market utilize luer slip or catheter tip connectors to<br>connect enteral giving sets and syringes to enteral receiving sets. Standard luer slip<br>and catheter tip connectors raise risk levels for patients as devices outside of the<br>enteral market use similar connections, including IVs. Therefore, the ISO/IEC 80369<br>series was developed to help reduce the possibility of misconnections across different<br>medical fields. |
|                                   | The FDA recently cleared an updated AMT enteral extension set in 510(k) K142989.<br>This extension set remains identical to our previously marketed device, but includes<br>Enfit connectors. However, launching a enteral extension sets with Enfit connectors<br>will not be simultaneous across all manufacturers and distributors. Therefore,<br>transition adapters are necessary for a limited time.                                                                                                                                                                                                                                          |
|                                   | As the new Enfit connectors are not backwards compatible with current luer and<br>bolus connections currently used for enteral products, AMT has designed several<br>transition adapter configurations to help users make the switch to the new<br>connections. AMT estimates that there will be a time of at least a year and possibly<br>up to three years where users and distributors will still have access to enteral devices<br>with luer / bolus connectors. The transition adapters AMT has designed in this<br>510(k) will allow users to use their old giving/receiving sets until the market is<br>flushed of old inventory.            |
|                                   | The AMT Enteral Transition Adapters come in 5 configurations and are made up of 5<br>separate components. The five configurations include:<br>• Male Enfit to Male Luer Adapter<br>• Female Enfit to Female Luer Adapter<br>• Female Enfit to Y-Port (Bolus/Luer) Adapter<br>• Female Enfit to Bolus Adapter<br>• Male Enfit to Christmas-tree Adapter                                                                                                                                                                                                                                                                                              |
|                                   | These 5 transition adapter configurations will allow AMT to offer users with options<br>for all of the common enteral connectors currently on the market. All components<br>and materials have been previously cleared in 510(k)s for similar use. It is planned to<br>include these transition adapters kitted with enteral devices previously cleared<br>through the 510(k) process, as well as selling the adapters as stand-alone items.                                                                                                                                                                                                        |
| Technological<br>Characteristics: | The AMT Enteral Transition Adapters are provided non-sterile for single user use<br>only. They are made from DEHP and Latex free materials. The Enteral Transition<br>Adapters are provided in a number of configurations providing different connection<br>possibilities to enteral devices currently on the market. Description of the five<br>different transition adapters AMT will offer are listed below:                                                                                                                                                                                                                                     |
|                                   | Transition Adapter Types:<br>• Male Enfit to Male Luer Adapter (P/N: TRN101)<br>• Female Enfit to Female Luer Adapter (P/N: TRN201)<br>• Female Enfit to Y-Port (Bolus/Luer) Adapter ) (P/N: TRN203)<br>• Female Enfit to Bolus Adapter (P/N: TRN202)<br>• Male Enfit to Christmas-tree Adapter (P/N: TRN102)                                                                                                                                                                                                                                                                                                                                       |
| Biocompatibility<br>Testing:      | No additional biocompatibility testing was performed for this submission. The<br>device and materials in finished form for the AMT Enteral Transition Adapters were<br>previously tested under 510(k) K123716. All AMT Enteral Transition adapters have<br>been tested for permanent (greater than 30 days) indirect mucosal contact, except for<br>the Male Enfit to Christmas-tree Adapter (P/N: TRN 102) which has been tested to<br>prolonged (less than 29 days) indirect mucosal contact.                                                                                                                                                     |

{4}------------------------------------------------

Applied Medical Technology, Inc. -510(k) Submission

{5}------------------------------------------------

| Performance<br>Testing: | AMT conducted various performance tests on all components contained within AMT<br>Enteral Transition Adapters. The tests carried out included:<br>• Tensile strength<br>• Fluid leakage<br>• Stress cracking<br>• Resistance to separation from axial load<br>• Resistance to separation from unscrewing<br>• Resistance to overriding<br>• Disconnection from unscrewing<br>• Flow testing<br><br>Incompatibility testing was conducted on the Enfit connectors as recommended in the<br>ISO 80369-1 standard. Although the Enfit connectors will not misconnect to other<br>connectors in the 80369 series, misconnections with non ISO/IEC 80369 series<br>connectors may still occur. |
|-------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Conclusions:            | The AMT Enteral Transition Adapters are substantially equivalent to the predicate<br>device cleared under K140581 in intended use, patient population, design,<br>biocompatibility, testing criteria, and method of operation.                                                                                                                                                                                                                                                                                                                                                                                                                                                            |

---

**Source:** [https://fda-staging.innolitics.com/device/K150034](https://fda-staging.innolitics.com/device/K150034)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
