K150024 · Guenter Bissinger Medizintechnik GmbH · GEI · Feb 20, 2015 · General, Plastic Surgery
Device Facts
Record ID
K150024
Device Name
Bissinger Monopolar Cables
Applicant
Guenter Bissinger Medizintechnik GmbH
Product Code
GEI · General, Plastic Surgery
Decision Date
Feb 20, 2015
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4400
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Cables for electrosurgery are designed to conduct electrical power from the output of a high-frequency electrosurgical generator to the electrosurgical instrument. Do not exceed a maximum output of 6250 Vp of your generator.
Device Story
Bissinger Monopolar Cables are reusable, non-sterile cables designed to conduct high-frequency electrical power from an electrosurgical generator to an electrosurgical instrument. The device acts as a passive conduit for electrical energy; it does not process signals or perform automated analysis. Used in clinical settings by surgeons or trained medical staff, the cables facilitate cutting and coagulation during electrosurgical procedures. The device is connected between the generator and the instrument; it does not affect clinical decision-making directly but enables the delivery of energy required for the surgical task. Benefits include reliable power transmission for electrosurgical instruments. The device is reprocessed via steam sterilization (autoclave) between uses.
Clinical Evidence
Bench testing only. No clinical data provided. Testing included electrical safety per IEC 60601 and IEC 60601-2-2:2009, validation of steam sterilization parameters (132°C, 4-minute exposure), and validation of manual cleaning and reprocessing protocols. Biocompatibility is supported by the use of materials with an established history of safe use per ISO 10993-1.
Technological Characteristics
Reusable monopolar cable; silicone construction. Length: 3.080 m; Socket: 4 mm; Plug: 8 mm. Connectivity: Bovie electrosurgical units. Energy source: High-frequency electrosurgical generator (max 6250 Vp). Sterilization: Moist heat (autoclave) at 132°C for 4 minutes. Standards: IEC 60601-2-2:2009, ISO 10993-1.
Indications for Use
Indicated for use in electrosurgical procedures to conduct electrical power from a high-frequency generator to an electrosurgical instrument. No specific patient population, age, or gender restrictions are stated.
Regulatory Classification
Identification
An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.
Predicate Devices
Hans Hermann Laparoscopes and Accessories (K051610)
Unipolar Endoscopic Coagulator-cutter and Accessories (K932456)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image is the logo for the Department of Health and Human Services (HHS). The logo features the words "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" arranged in a circular fashion. Inside the circle is a stylized image of three human profiles facing right, with flowing lines extending from the first profile.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
February 20, 2015
Güenter Bissinger Medizintechnik GmbH Matthias Bissinger Managing Director Hans-Theisen-Straße 1 79331 Teningen Baden-Wuerttemberg Germany
Re: K150024
Trade/Device Name: Monopolar Cables Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical cutting and coagulation device and accessories Regulatory Class: Class II Product Code: GEI Dated: December 1, 2014 Received: January 7, 2015
Dear Mr. Bissinger:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations. Title 21. Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical
{1}------------------------------------------------
device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# Jennifer R. Stevenson -S
For Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K150024 I
Device Name Monopolar Cables
Indications for Use (Describe)
Cables for electrosurgery are designed to conduct electrical power from the output of a high-frequency electrosurgical generator to the electrosurgical instrument.
Do not exceed a maximum output of 6250 Vp of your generator.
Type of Use (Select one or both, as applicable)
2 Prescription Use (Part 21 CFR 801 Subpart D)
- Over-The-Counter Use (21 CFR 801 Subpart C)
## PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
### FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW!*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for Bissinger. The logo features a stylized letter 'b' in red, with the center of the 'b' forming the pupil of an eye. Below the eye is the word 'bissinger' in lowercase, with a line underneath the 'er' at the end of the word.
K
# VOLUME 006
# 510(k) Summary
DATE OF APPLICATION: 2014-12-01
| APPLICANT: | Guenter Bissinger Medizintechnik GmbH<br>Hans-Theisen-Str. 1,<br>D-79331 Teningen ,<br>Germany<br>Tel.: +1.603.570.2039<br>Fax: +1.603.570.2055<br>E-Mail: info@bissinger.com<br>Internet: www.bissinger.com |
|-----------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| CONTACT PERSON: | Matthias Bissinger<br>Tel.: +49.7461 914330<br>E-Mail: bissinger@bissinger.com |
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows a logo for "bissinger". The logo features a stylized letter "b" in red, with a white outline resembling an eye around it. Below the "b", the word "bissinger" is written in a smaller, black, sans-serif font.
#### 1. Device Name
| Product code: | GEI |
|----------------------|----------------------------------------------------------------|
| Trade Names: | Bissinger Monopolar Cables |
| Common Name: | electrosurgical, cutting & coagulation & accessories |
| Classification Name: | Electrosurgical cutting and coagulation device and accessories |
#### Classification Product Code / Subsequent Code 2.
## 2.1. Product Code
| Device | Medical<br>Specialty | Review Panel | Product Code | Device Class | Regulation Number |
|-------------------------------------------------------------------------|----------------------|--------------------------------|--------------|--------------|-------------------|
| Electrosurgical cutting<br>and coagulation de-<br>vice and accessories. | Part 878 | General & Plastic Sur-<br>gery | GEI | 2 | 878.4400 |
#### Predicate Device 3.
Bissinger's Monopolar Cables are substantially equivalent to the following predicate devices, already cleared by the FDA:
| Predicate Device | 510(k) Number | 510(k) Holder |
|----------------------------------------------|---------------|----------------------------|
| Hans Hermann Laparoscopes and<br>Accessories | K051610 | HANS HERMANN GMBH |
| Suffer Electrosurgical Cables | K073450 | Suffer Medizintechnik GmbH |
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows a logo for Bissinger. The logo features a stylized letter 'b' that is shaped like an eye, with a dark red iris and a black pupil. The word 'bissinger' is written in lowercase letters below the eye-shaped 'b'. The font is simple and modern, and the overall design is clean and eye-catching.
#### Description of the Device 4.
Bissinger monopolar cables are non-sterile, reusable Monopolar Cables fitting Bovie Electrosurgical units. They are designed to conduct electrical power from the output of a high frequency generator to the instrument.
#### Indications for Use 5.
Cables for electrosurgery are designed to conduct electrical power from the output of a high-frequency electrosurgical generator to the electrosurgical instrument.
Do not exceed a maximum output of 6250 Vp of your generator.
#### Technological Characteristics 6.
| | NEW DEVICE | Predicate Device 1 | Predicate Device 2 | Reference DEVICE | RE-<br>SULT |
|--------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------|
| 510(k)<br>Submitter/<br>Holder | Bissinger Medizintechnik<br>GmbH | HANS HERMANN<br>GMBH | Sutter Medizintechnik<br>GmbH | Jarit Surgical Instru-<br>ments Inc. | NA |
| Trade<br>Name | Bissinger Monopolar<br>Cables | Hans Hermann Lapa-<br>roscopes and acces-<br>sories | Sutter Electrosurgical<br>Cables | Unipolar Endoscopic<br>Coagulator-cutter and<br>Accessories | NA |
| Device | Bissinger Monopolar<br>Cables | Laparoscopes | Sutter Electrosurgical<br>Cables | JARIT SURGICAL<br>INSTRUMENTS | NA |
| 510(k)<br>Number | | K051610 | K073450 | K932456 | NA |
| Intended<br>use | Cables for electrosur-<br>gery are designed to<br>conduct electrical power<br>from the output of a<br>high-frequency electro-<br>surgical generator to the<br>electrosurgical<br>instru-<br>ment.<br>Do not exceed a maxi-<br>mum output of 6250 Vp<br>of your generator. | The Laparosopes and<br>accessories are<br>intended for use in<br>providing access to<br>and visualization of<br>body cavities, organs,<br>and canals to perform<br>various diagnostic<br>and therapeutic<br>surgical procedures.<br>The arthroscope is<br>indicated for illumina-<br>tion during joint ex-<br>aminations, arthros-<br>copies, biopsies an (<br>diagnosis of joint<br>disease in minimally<br>invasive procedures<br>of the knee, shoulder,<br>wrist (carpal tunnel<br>syndrome), temporal<br>mandibular joint,<br>ankle and elbow. The<br>bipolar electrodes are<br>use to coagulate and<br>to remove or destroy<br>tissue by the use of<br>bipolar HF current. | To electrically con-<br>nect monopolar<br>/bipolar electrosurgi-<br>cal instruments to an<br>electrosurgical gen-<br>erator. | For use by, or as di-<br>rected by, a surgeon in<br>endoscopic surgery.For<br>use in<br>endoscopic<br>surgery. For use when a<br>rigid endoscopic instru-<br>ment for grasping<br>and/or dissecting if soft<br>tissue is determined to<br>be appropriate by the<br>surgeon.<br>Monopolar<br>electrosurgical current<br>can be used for coagu-<br>lation and/or cutting as<br>determined necessary<br>and appropriate by the<br>surgeon. | Sub-<br>stan-<br>tial<br>Equiv-<br>alent |
| Type | Monopolar Cable | Monopolar silicone<br>Cables | Monopolar silicone<br>Cables | Jarit 600-290 | Sub-<br>stan-<br>tialEqu<br>ivalent |
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows a logo with the word "bissinger" in lowercase letters. Above the word is a stylized graphic that combines the letter "b" with an eye. The "b" is colored in a dark red, and the eye is outlined in black with a black pupil. The word "bissinger" is in a simple, sans-serif font.
# Monoplar Cables
## Page 4 of 5
| | NEW DEVICE | Predicate Device 1 | Predicate Device 2 | Reference DEVICE | RE-<br>SULT |
|---------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------|-------------------------------------------|
| Design | Length: 3,080 m.<br>Socket: Ø 4 mm<br>Plug Ø 8 mm<br>Connection: Bovie | Length: 3 m,<br>Connector: 4 mm<br>Connection: Erbe,<br>Martin, Berchtold, | Length: 3.5 mtr, 4,5<br>mtr. Connector: 4 mm<br>Connection:<br>Val-<br>leylab,<br>Conmed,<br>Bovie, Bowa, Erbe, K.<br>Storz, R. Wolf, Martin,<br>Berchtold, Aesculap | Length: 3,080 m.<br>Socket: Ø 4 mm<br>Plug Ø 8 mm<br>Connection: Bovie | Sub-<br>stan-<br>tialEqu<br>ivalent |
| Materials /<br>Biocom-<br>patibility | Material: Silicon<br>Cables do not have<br>direct contact with the<br>human body or only with<br>uninjured skin, so that<br>biocompatibility must not<br>be evaluated. | Silicon | Silicon | Silicon | Sub-<br>stan-<br>tialequ<br>iva-<br>lenrt |
| Sterility | Reusable<br>Instruments.<br>Delivery non sterile.<br>The instruments must be<br>sterilized in moist heat<br>(Autoclave).<br>Pre-vacuum Cycle<br>- Temperature: 132°C<br>- Exposure time: 4 min.<br>- Drying time: at least 20<br>min. | Reusable<br>Instru-<br>ments. Delivery non<br>sterile.<br>The instruments must<br>be sterilized in moist<br>heat<br>(Autoclave).<br>Pre-vacuum Cycle<br>- Temperature:<br>132°C, max 137°C<br>- Exposure time: 3<br>min.<br>- Drying time: at least<br>10 min. | Sutter Electrosurgical<br>Cables are supplied<br>non-sterile and can<br>be reused after clean-<br>ing and steam sterili-<br>zation.<br>Steam-sterilize in the<br>autoclave in accord-<br>ance with DIN EN<br>13060 / DIN EN 285.<br>Fractioned<br>pre-<br>vacuum sterilization:<br>Temperature 134 °C<br>(273 °F), 3 minutes;<br>max.<br>Temperature<br>138 °C (280 °F),<br>max. duration<br>20<br>minutes. | Similar: Reusable In-<br>struments | Sub-<br>stan-<br>tialEqu<br>ivalent |
| Compati-<br>bility with<br>environ-<br>ment and<br>other<br>devices | Generators: Bovie | Generators: Erbe,<br>Martin, Berchtold, | Generators: Erbe,<br>Martin, Valleylab,<br>Conmed, Berchtold,<br>Bowie, Bowa, Aescu-<br>lap | Generators: Bovie | Sub-<br>stan-<br>tialEqu<br>ivalent |
| Standards<br>met | IEC 60601-2-2: 2009 | IEC 60601-2-2:2009 | IEC 60601-2-2:2007 | IEC 60601-2-2:2009 | Sub-<br>stan-<br>tialEqu<br>ivalent |
| Energy<br>applied | Power maxima 6.250 Vp | Power maxima 6.250<br>Vp | Maximum Monopolar<br>cablesVoltage 10000<br>Vpp | Power maxima 6.250<br>Vp | Sub-<br>stan-<br>tialEqu<br>ivalent |
## 7. Testing
Testing in order to proof safety and effectiveness of Bissinger's Monopolar Cables has been performed according to recognized consensus Standards and results are conforming to the respective requirements.
#### Electrical Safety 7.1.
The devices subject to this submission have been tested according to the requirements of IEC 60601 and IEC 60601-2-2:2009.
{7}------------------------------------------------
#### 7.2. Sterilization
The sterilization process has been validated under consideration of recognized standards. Testing shows that the Products can be steam sterilized with a sufficient sterility assurance level by use of standard sterilization parameters.
#### 7.3. Reprocessing
Reprocessability was tested and validated under consideration of recognized standards.
#### 7.4. Manual cleaning
Manual cleaning validation was performed under consideration of recognized standards.
#### 8. Biocompatibility
All requirements of biocompatibility are met through the composition of the used materials which demonstrate the appropriate levels of biocompatibility for its clinical use. The used materials are also used in many other medical devices and have an established history of safe use and biocompatibility outlined in ISO 10993-1.
#### Substantial Equivalence Summary / Conclusion 9.
Based on available 510(k) information provided herein, Bissinger Monopolar Cables are considered substantial equivalent to the predicate devices in terms of indications for use, material, technology, design and performance specifications.
There are no differences between the devices which would raise new issues of safety or effectiveness.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.