# VS3-IR-MMS System (K150018)

_Visionsense, Ltd. · OWN · Jun 17, 2015 · Gastroenterology, Urology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K150018

## Device Facts

- **Applicant:** Visionsense, Ltd.
- **Product Code:** [OWN](/productcode/OWN.md)
- **Decision Date:** Jun 17, 2015
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 876.1500
- **Device Class:** Class 2
- **Review Panel:** Gastroenterology, Urology
- **Attributes:** Real-World Evidence

## Real-World Evidence

| Submission | Device | Sponsor | RWD Sources | RWE Use Summary | Key Tags |
| --- | --- | --- | --- | --- | --- |
| K150018 · Jun 17, 2015 | VS3-IR-MMS System | Visionsense, Ltd. | Clinical evaluation in human subjects (outside the US) | The device was evaluated in human subjects during free flap surgical procedures to demonstrate that it functions as intended for the visual assessment of blood flow and tissue perfusion. | Free flaps; Human subjects; Tissue perfusion; Blood flow assessment |

### Clinical Evidence

| Study Design | Population | Comparator | Key Endpoints |
| --- | --- | --- | --- |
| Clinical evaluation (observational/case series) | Human subjects undergoing free flap procedures; Sample Size: Several | Not applicable for this study | Device functionality and safety (adverse events) |

## Indications for Use

The Endoscope Module of the VS3-IR-MMS System is intended for viewing internal surgical sites during general surgical procedures, for use in visualization of ventricles and structures within the brain during neurological surgical procedures, viewing internal surgical sites during anterior spinal procedures, such as nucleotomy, discectomy, and foraminotomy, and shoulder and knee arthroscopic procedures. The Iridium Module of the VS3-IR-MMS System is intended for capturing and viewing fluorescent images for the visual assessment of blood flow, as an adjunctive method for the evaluation of tissue perfusion, and related tissue-transfer circulation in tissue and free flaps used in plastic, micro- and reconstructive surgical procedures.

## Device Story

VS3-IR-MMS System is a surgical visualization platform comprising a base endoscope system and an Iridium Module. The system captures high-definition 3D video of surgical sites; the Iridium Module adds fluorescence imaging capabilities. During surgery, the physician administers Indocyanine Green (ICG) dye intravenously. The system's Laser Light Source (LLS) provides 805nm excitation light to the surgical field; an IR camera captures the resulting 825-850nm fluorescence emission. The system processes these signals to display real-time fluorescent images on a monitor, allowing the surgeon to visually assess blood flow and tissue perfusion in real-time. Used in OR settings by surgeons; output assists in evaluating tissue viability during reconstructive procedures. Benefits include improved intraoperative assessment of tissue-transfer circulation, potentially reducing complications in plastic and reconstructive surgeries.

## Clinical Evidence

Bench testing verified imaging quality using ICG-diluted samples and chicken skin models. Simulated use testing performed on a human hand outside the US. Clinical evaluation performed on human subjects in free flap procedures outside the US demonstrated intended function with no reported adverse events.

## Technological Characteristics

System includes HD 3D endoscope, 3D camera, CCU, display monitor, and Iridium Module (MMS scope, Laser Light Source). Excitation wavelength 805nm; emission 825-850nm. Complies with IEC 60601-1, IEC 60601-1-2, IEC 60601-1-4, and IEC 60825-1. Connectivity includes removable storage for recording/playback.

## Regulatory Identification

An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

## Special Controls

*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

In combination with the general controls of the FD&C Act, the integrated operating table-electromechanical surgical system is subject to the following special controls:

1. (1) Premarket clinical performance testing, or a combination of premarket clinical performance testing and postmarket surveillance (in accordance with special control (2)), must include the following:
   1. (i) Objective performance measures (e.g., rate and number of conversions to other surgical modalities, rate of device related adverse events (including tissue injury, hematoma, and increased blood loss), and their severity, cause, and outcomes) must be reported with relevant descriptive comparator performance measures.
   2. (ii) The data must demonstrate the performance of the device for providing accurate and precise control of attached surgical instruments in range of clinical conditions relevant to the device's intended use.
   3. (iii) The test dataset must include data collected from a patient population representative of the intended patient population under anticipated conditions of use.
2. (2) Data obtained from postmarket surveillance must demonstrate, in consideration of the premarket data obtained in accordance with special control (1), that the device performs in accordance with special control (1), unless FDA determines, based on the totality of the premarket data, that data from postmarket surveillance is not required to demonstrate that the device performs as intended. Such postmarket surveillance must be conducted per a protocol determined appropriate by FDA to demonstrate that the device performs as intended (in consideration of the premarket data obtained in accordance with special control (1)), and must include initiation, enrollment, and reporting requirements to ensure timely periodic updates to FDA on post-market surveillance progress and outcomes.
3. (3) Animal performance testing must evaluate the extent of port site trauma due to repositioning of table during surgical procedures when utilizing robotic minimally invasive and laparoscopic approaches
4. (4) The device manufacturer must develop, and update as necessary, a device-specific use training program that ensures proper device setup/use/shutdown, accurate control of instruments to perform the intended surgical procedures, troubleshooting and handling during unexpected events or emergencies, and safe practices to mitigate use error.
5. (5) The device manufacturer may only distribute the device to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program.
(6) Human factors assessment must demonstrate that the user can correctly use the device system across all intended use environments with the provided instructions and training materials, including patient access during normal operating conditions and emergency situations, and effects arising from the integrated nature of the operating table and robotic surgical arms.
(7) Labeling must include:
(i) A detailed summary of clinical performance testing conducted with the device, including study population, results, adverse events, and comparisons to any comparator groups identified;
(ii) A statement in the labeling that the safety and effectiveness for the representative specific procedures was based on evaluation of the device as a surgical tool and did not include evaluation of outcomes related to the treatment of the patient's underlying disease or condition, unless FDA determines that it can be removed or modified based on clinical performance data submitted to FDA;
(iii) Identification of compatible devices;
(iv) The list of surgical procedures for which the device has been determined to be safe with clinical justification;
(v) Reprocessing instructions for reusable components;
(vi) A shelf life for any sterile components;
(vii) A description of the device-specific use training program;
(viii) A statement that the device is only for distribution to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program; and
(ix) A summary of any completed postmarket surveillance data collected as required by special control (2), including updated labeling to accurately reflect outcomes observed in postmarket surveillance.
(8) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use and must include:
(i) Device motion accuracy and repeatability;
(ii) System testing;
(iii) Instrument reliability;
(iv) Crosstalk;
(v) Table motion control;
(vi) Thermal effects on tissue;
(vii) User-device interface performance;
(viii) Workspace access testing; and
(ix) Performance testing with compatible devices.
(9) Software verification, validation, and hazard analysis must be performed.
(10) Electromagnetic compatibility and electrical, thermal, and mechanical safety testing must be performed.
(11) Performance data must demonstrate the sterility of all patient-contacting device components.
(12) Performance data must support the shelf life of the device components provided sterile by demonstrating continued sterility and package integrity over the labeled shelf life.
(13) Performance data must validate the reprocessing instructions for the reusable components of the device.
(14) Performance data must demonstrate that all patient-contacting components of the device are biocompatible.
(15) Performance data must demonstrate that all patient-contacting components of the device are non-pyrogenic.
(16) The device manufacturer must submit a report to the FDA annually on the anniversary of initial marketing authorization for the device, until such time as FDA may terminate such reporting, which comprises the following information:
(i) Cumulative summary, by year, of complaints and adverse events since date of initial marketing authorization; and
(ii) Identification and rationale for changes made to the device, labeling, or device specific use training program, which did not require submission of a premarket notification during the reporting period.

## Predicate Devices

- Novadaq Technologies SPY Imaging System ([K063345](/device/K063345.md))

## Reference Devices

- VS3 System ([K123467](/device/K123467.md))
- VS3 System ([K131434](/device/K131434.md))
- VS3 System ([K141002](/device/K141002.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features the department's emblem, which consists of a stylized human figure with three faces in profile, representing health and well-being. The emblem is surrounded by the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" in a circular arrangement.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

June 17, 2015

Visionsense LTD. % Mr. Gerard J. Prud'homme Hogan Lovells US LLP 555 13th Street, North West Washington, District of Columbia 20004

Re: K150018

Trade/Device Name: VS3-IR-MMS System Regulation Number: 21 CFR 876.1500 Regulation Name: Endoscope and Accessories Regulatory Class: Class II Product Code: OWN Dated: May 19, 2015 Received: May 19, 2015

Dear Mr. Prud'homme:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set

{1}------------------------------------------------

forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours.

# Jennifer R. Stevenson -S

For Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

## Indications for Use

510(k) Number (if known) K150018

Device Name

VS3-IR-MMS System

#### Indications for Use (Describe)

The Endoscope Module of the VS3-IR-MMS System is intended for viewing internal surgical sites during general surgical procedures, for use in visualization of ventricles and structures within the brain during neurological surgical procedures, viewing internal surgical sites during anterior spinal procedures, such as nucleotomy, discectomy, and foraminotomy, and shoulder and knee arthroscopic procedures. The Iridium Module of the VS3-IR-MMS System is intended for capturing and viewing fluorescent images for the visual assessment of blood flow, as an adjunctive method for the evaluation of tissue perfusion, and related tissue-transfer circulation in tissue and free flaps used in plastic, micro- and reconstructive surgical procedures.

| Type of Use (Select one or both, as applicable)                                      |  |
|--------------------------------------------------------------------------------------|--|
| <div> <span>☑</span> Prescription Use (Part 21 CFR 801 Subpart D)             </div> |  |
| <div> <span>☐</span> Over-The-Counter Use (21 CFR 801 Subpart C)             </div>  |  |

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

#### 510(k) Summary

#### Visionsense's VS3-IR-MMS System

#### Submitter's Name, Address. Telephone Number, Contact Person and Date Prepared

Visionsense Ltd. 20 Hamagshimim Street Petach Tikva 49348 lsrael Phone: 972-3-9244339 Facsimile: +972-3-9244-335

Contact Person: Alex Chanin, CEO

Date Prepared: May 19, 2015

#### Name of Device and Name/Address of Sponsor

VS3-IR-MMS Svstem

#### Common or Usual Name / Classification Name

Surgical Microscope / Angiographic X-ray System / Endoscopes and Accessories

#### Predicate Device

Novadaq Technologies SPY Imaging System (K063345)

#### Intended Use / Indications for Use

The Endoscope Module of the VS3-IR-MMS System is intended for viewing internal surgical sites during general surgical procedures, for use in visualization of ventricles and structures within the brain during neurological surgical procedures, viewing internal surgical sites during anterior and posterior spinal procedures, such as nucleotomy, discectomy, and foraminotomy, and shoulder and knee arthroscopic procedures. The Iridium Module of the VS3-IR-MMS System is intended for capturing and viewing fluorescent images for the visual assessment of blood flow, as an adjunctive method for the evaluation of tissue perfusion, and related tissue-transfer circulation in tissue and free flaps used in plastic, micro- and reconstructive surgical procedures.

#### Technological Characteristics / Principles of Operation

The VS3-IR-MMS System consists of the cleared base system and the Iridium Module that is the subject of this submission. The currently marketed VS3 system with endoscope functionality was cleared under the following product codes: GCJ (General Surgery Endoscope), GWG (Neurological Endoscope), and HRX (Orthopedic Arthroscope). The cleared base system consists of the following components and accessories: High Definition 3D Endoscope, High Definition 3D Camera with 3 control buttons and a focus adjustment knob, 2D Coupler, CCU, Light Source, and Display Monitor. The VS3-IR-MMS System that is the subject of this submission contains the previously cleared components of the cleared base system listed above, as well as a High Definition IR Fluorescence Scope (MMS), a Laser Light Source (LLS), associated software and VS, Iridium IR Fluorescence Kit. The new Iridium Module, which uses the newly added High Definition IR Fluorescence Scope (MMS) 

{4}------------------------------------------------

and a Laser Light Source (LLS), as well as the cleared High Definition 3D Camera, Camera Control Unit (CCU), and Display Monitor, is used to generate florescence excitation illumination for external use in capturing and viewing fluorescent images for the visual assessment of blood flow, as an adjunctive method for the evaluation of tissue perfusion, and related tissue-transfer circulation in tissue and free flaps used in plastic, micro- and reconstructive surgical procedures.

In sum, the proposed VS3-IR-MMS System is built on the predicate base VS3 System that has already been cleared for the treatment of viewing internal surgical sites during general surgical procedures and for use in visualization of ventricles and structures within the brain during neurological surgical procedures, viewing internal surgical sites during anterior and posterior spinal procedures, such as nucleotomy, discectomy, and foraminotomy, and shoulder and knee arthroscopic procedures (see K123467. K131434. K141002). The Iridium Module added to the cleared base system results in the following changes to the cleared system:

- . Addition of a new scope ("MMS"), which is intended to provide a view of the surgical field from 20cm to 45cm above the patient.
- . Addition of a Laser Light Source ("LLS") used to generate florescence excitation illumination.
- Software modifications to allow the new MMS with LLS to be used with the cleared system.
- Addition of the VS, Iridium IR Fluorescence Kit which contains the sterile ICG imaging agent used for fluorescence imaging.
- . Additional indications for use with the new Iridium Module, which includes the MMS with LLS and associated software (described above) to include capturing and viewing fluorescent images for the visual assessment of blood flow, as an adjunctive method for the evaluation of tissue perfusion, and related tissue-transfer circulation in tissue and free flaps used in plastic, micro- and reconstructive surqical procedures.

The Iridium Module that is the subject of this submission is designed to work with the Indocyanine Green (ICG) IR fluorescence imaging agent supplied in the VSg Iridium IR Fluorescence Kit. Each VS . Iridium IR Fluorescence Kit contains the following:

- o Six 25mg vials of sterile ICG imaging agent.
- o Six 10ml vials of sterile Water for Injection.

ICG in blood has maximum excitation at 805nm and an emission band between 825nm and 850nm. VS, Iridium Module provides excitation light to the surgical field to excite the dye molecules and captures emission from the dye using an IR camera.

The VS3-IR-MMS System permits recording surgical procedures, storing them on removable storage devices, and plaving the procedures back.

{5}------------------------------------------------

#### Performance Data

The subject device conforms to the following recognized standards:

| Standards No. | Standards<br>Organization | Standards Title                                                                                                                         |
|---------------|---------------------------|-----------------------------------------------------------------------------------------------------------------------------------------|
| 60601-1-2     |                           | Medical electrical equipment - Part 1-2: General requirements for                                                                       |
| (19-2)        | IEC                       | basic safety and essential performance - collateral standard:<br>electromagnetic compatibility - requirements and tests (Edition<br>3). |
| Standards No. | Standards<br>Organization | Standards Title                                                                                                                         |
| 60601-1       |                           | Medical electrical equipment - Part 1: General requirements for                                                                         |
| (19-4)        | IEC                       | basic safety and essential performance.                                                                                                 |
| Standards No. | Standards                 | Standards Title                                                                                                                         |
|               | Organization              |                                                                                                                                         |
|               |                           | Medical electrical equipment - Part 1: General requirements for                                                                         |
| 60601-1-4     | IEC                       | safety, Collateral standard: Programmable electrical medical                                                                            |
|               |                           | systems.                                                                                                                                |
| Standards No. | Standards                 | Standards Title                                                                                                                         |
|               | Organization              |                                                                                                                                         |
| 60825-1       |                           | Safety of laser products - Part 1: Equipment classification, and                                                                        |
| (12-273)      | IEC                       | requirements.                                                                                                                           |

The system software was validated and performs as intended per the pre-specified requirements.

In addition, bench testing was conducted to verify that the subject device can detect and visualize anatomy and blood flow per the proposed indications for use of the Iridium Module. Specifically, 10mg of ICG was dilated in sterile water to a concentration equal to that of ICG in the blood of a representative human (i.e., 2 ug/ml). Next, the ICG in sterile water diluted to representative human concentration was imaged directly to verify good quality imaging. Then, the sample of ICG in sterile water diluted to representative human concentration was covered with chicken skin and imaged to simulate visualization through skin.

On a related note, outside of the United States, a human hand was used in a simulated use testing environment to demonstrate that the Iridium Module has the capability to visualize blood flow corresponding to ICG dye fluorescence of blood vessels in the skin.

Lastly, an evaluation of the device was performed in free flaps in several human subjects outside the US and demonstrated that the device functions as intended with no adverse events reported using the system.

In all instances, the VS3-IR-MMS System using the Iridium Module functioned as intended and met all performance acceptance criteria.

### Substantial Equivalence

The endoscope functionality of the base VS3-IR-MMS System was previously cleared. The proposed Iridium Module of the VS3-R-MMS has the same intended uses and similar indications, technological characteristics, and principles of operation as its predicate device (K063345). The minor technological differences between the VS3-IR-MMS System Iridium Module and its predicate device do not raise different questions of safety or effectiveness. Performance data demonstrate that the VS3-IR-MMS System Iridium Module is as safe and effective as the predicate device. Thus, 

{6}------------------------------------------------

the VS3-IR-MMS Iridium Module is substantially equivalent. See the below comparison table of the Iridium Module of the updated VS3-IR-MMS and the Novadaq Spy Imaging device (K063345).

| Attributes            | VS3-IR-MMS (proposed)                                | Spy (K063345)                                        |
|-----------------------|------------------------------------------------------|------------------------------------------------------|
| Manufacturer          | Visionsense Ltd.                                     | Novadaq Technologies Inc                             |
| Drape Usage           | Yes                                                  | Yes                                                  |
| Imaging Agent         | IR fluorescence dye (e.g.,<br>Indocyanine Green-ICG) | IR fluorescence dye (e.g.,<br>Indocyanine Green-ICG) |
| Imaging Head          | Silicon Image Sensor                                 | Silicon Image Sensor                                 |
| Imaging Device        | No direct or indirect patient contact                | No direct or indirect patient<br>contact             |
| Light Source          | Infrared Laser                                       | Infrared Laser                                       |
| Excitation Wavelength | 805nm                                                | 805nm                                                |
| Field of View         | 19cm x 14cm<br>(40cm nominal distance)               | 19cm x 14cm<br>(30cm nominal distance)               |
| Imaging               | Fluorescent Imaging                                  | Fluorescent Imaging                                  |
| Emission Band         | 825nm to 850nm                                       | 825nm to 850nm                                       |
| Emission Capture      | IR camera                                            | IR camera                                            |

### Conclusions

The VS3-IR-MMS System Iridium Module is substantially equivalent to the predicate device. Performance and clinical testing demonstrates that the newly added components to the proposed device performs substantially equivalent to the predicate device, and any differences in technological characteristics do not raise different questions of safety or efficacy compared to the predicate device.

---

**Source:** [https://fda-staging.innolitics.com/device/K150018](https://fda-staging.innolitics.com/device/K150018)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).