← Product Code [OMP](/productcode/OMP) · K150006

# Prevena Incision Management System with Peel & Place Dressing, Prevena Incision Management System with Customizable Dressing (K150006)

_Kci USA, Inc. · OMP · Nov 6, 2015 · General, Plastic Surgery · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K150006

## Device Facts

- **Applicant:** Kci USA, Inc.
- **Product Code:** [OMP](/productcode/OMP.md)
- **Decision Date:** Nov 6, 2015
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 878.4780
- **Device Class:** Class 2
- **Review Panel:** General, Plastic Surgery
- **Attributes:** Therapeutic

## Indications for Use

The Prevena Incision Management System is intended to manage the environment of surgical incisions that continue to drain following sutured or stapled closed by maintaining a closed environment and removing exudate via the application of negative pressure wound therapy.

## Device Story

The Prevena Incision Management System is a negative pressure wound therapy (NPWT) device used to manage surgical incisions. It consists of a dressing applied over the incision site, a disposable canister, and a powered suction pump. The pump delivers continuous -125 mmHg negative pressure to the dressing for up to 7 days. The system maintains a closed environment, protects the incision from external contamination, and removes exudate into the canister. It is used in clinical settings, such as the operating room, and is operated by healthcare professionals. The device includes visual and audible alarms for system status, such as air leaks, full canisters, or low battery. The system helps draw incision edges together and manage fluid, potentially benefiting the patient by supporting the healing environment of the surgical site.

## Clinical Evidence

No clinical tests were conducted. Bench testing included performance testing of negative pressure delivery, software verification and validation, and battery life testing. Usability testing was performed with patient users to assess the impact of labeling changes resulting from software modifications; all usability goals were met.

## Technological Characteristics

The system utilizes NPWT via a powered suction pump delivering -125 mmHg. Components include an occlusive dressing (Peel & Place or Customizable), disposable canister, and therapy unit. Connectivity is limited to internal alarm systems. Software is used for therapy control and alarm management. The device is designed for continuous operation for up to 7 days.

## Regulatory Identification

A powered suction pump is a portable, AC-powered or compressed air-powered device intended to be used to remove infectious materials from wounds or fluids from a patient's airway or respiratory support system. The device may be used during surgery in the operating room or at the patient's bedside. The device may include a microbial filter.

## Predicate Devices

- Prevena Incision Management System ([K121883](/device/K121883.md))

## Reference Devices

- Prevena Peel & Place Dressing ([K100821](/device/K100821.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image contains the logo for the U.S. Department of Health & Human Services. The logo features the department's name in a circular arrangement around a stylized symbol. The symbol consists of three human profiles facing right, stacked on top of each other, creating a sense of depth and unity. The overall design is simple, clean, and conveys a sense of government authority and public service.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

November 6, 2015

KCI USA Incoprorated Ms. Melanie Avila Senior Manager, Regulatory Affairs 6203 Farinon Drive San Antonio, Texas 78249

Re: K150006

Trade/Device Name: Prevena Incision Management System Regulation Number: 21 CFR 878.4780 Regulation Name: Powered suction pump Regulatory Class: Class II Product Code: OMP Dated: October 12, 2015 Received: October 13, 2015

Dear Ms. Avila:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical

{1}------------------------------------------------

device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

# David Krause -S

- Binita S. Ashar, M.D., M.B.A., F.A.C.S. for Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure

{2}------------------------------------------------

### Indications for Use

510(k) Number (if known) K150006

Device Name Prevena Incision Management System

#### Indications for Use (Describe)

The Prevena Incision Management System is intended to manage the environment of surgical incisions that continue to drain following sutured or stapled closed by maintaining a closed environment and removing exudate via the application of negative pressure wound therapy.

| Type of Use (Select one or both, as applicable) |                                             |
|-------------------------------------------------|---------------------------------------------|
| X Prescription Use (Part 21 CFR 801 Subpart D)  | Over-The-Counter Use (21 CFR 801 Subpart C) |

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

## 510(k) SUMMARY Prevena Incision Management Peel & Place Dressing and Customizable Dressing

| Submitter Information [21 CFR 807.929(a)(1)]                                              |                                                                                                                                                                                                                                                                                            |
|-------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Name                                                                                      | KCI USA, Inc. (Kinetic Concepts, Inc.)                                                                                                                                                                                                                                                     |
| Address                                                                                   | 6203 Farinon Drive<br>San Antonio, TX 78249                                                                                                                                                                                                                                                |
| Phone number                                                                              | 210-515-4059                                                                                                                                                                                                                                                                               |
| Fax number                                                                                | 210-255-6727                                                                                                                                                                                                                                                                               |
| Establishment<br>Registration Number                                                      | 3009897021                                                                                                                                                                                                                                                                                 |
| Name of contact person                                                                    | Melanie Avila, Senior Manager, Regulatory Affairs                                                                                                                                                                                                                                          |
| Date prepared                                                                             | October 12, 2015                                                                                                                                                                                                                                                                           |
| Name of the device [21 CFR 807.92(a)(2)]                                                  |                                                                                                                                                                                                                                                                                            |
| Trade or proprietary<br>name                                                              | Prevena™ Incision Management System                                                                                                                                                                                                                                                        |
| Common or usual name                                                                      | Negative Pressure Wound Therapy System                                                                                                                                                                                                                                                     |
| Classification name                                                                       | Negative Pressure Wound Therapy Powered Suction Pump (and components)                                                                                                                                                                                                                      |
| Classification panel                                                                      | General and Plastic Surgery                                                                                                                                                                                                                                                                |
| Regulation                                                                                | 878.4780                                                                                                                                                                                                                                                                                   |
| Product Code(s)                                                                           | OMP                                                                                                                                                                                                                                                                                        |
| Legally marketed<br>device(s) to which<br>equivalence is claimed<br>[21 CFR 807.92(a)(3)] | Prevena™ Incision Management System, cleared under 510(k) K121883                                                                                                                                                                                                                          |
| Device description<br>[21 CFR 807.92(a)(4)]                                               | Negative pressure wound therapy system                                                                                                                                                                                                                                                     |
| Indications for use<br>[21 CFR 807.92(a)(5)]                                              | The Prevena ™ Incision Management System is intended to manage the<br>environment of surgical incisions that continue to drain following sutured or<br>stapled closure by maintaining a closed environment and removing exudate<br>via the application of negative pressure wound therapy. |

{4}------------------------------------------------

#### Comparison of the Technological Characteristics with the Predicate Device [21 CFR 807.92(a)(6)]

Neqative Pressure Wound Therapy is the technological principal for both the subject and predicate devices. Application of negative pressure to an incision site that is closed via staples nelps draw the incision edges together and removes fluid from the incision site. The occlusive drape of the dressing provides a negative pressure environment and protects the incision from external contamination.

#### At a high level, the subject device and predicate device are based on the following same technoloqical elements:

- The dressings that are applied over the incision site in the operating room are identical. One of . the following dressings may be selected by the surgeon, based on incision length and geometry:
	- The Prevena Peel & Place Dressing (cleared under K100821) which can be used for linear o incisions up to 8 inches, or
	- The Prevena Customizable Dressing (cleared under K121883), which can be configured o for non-linear incisions or linear incisions longer than 8 inches
- A negative pressure pump (therapy unit) is required that can provide -125 mmHg of negative ● pressure continuously to the dressing for a maximum of 7 days.
- . The dressings are connected to the therapy unit via a disposable canister.
- Incision fluid is collected into the disposable canister
- The therapy unit provides alarms that indicate when negative pressure wound therapy may be ● compromised (e.g., visual and audible alarms indicating an air leak in the system or when the canister is full or when batteries are low).

The following technological differences exist between the subject and predicate devices:

- The volume and mute time has been chanqed for the Leak Alert and the mute time has been ● changed for the Low Battery Alert to minimize nuisance alerts under the new air leak threshold
- . The air leak threshold has been slightly increased to allow for more air in the system before therapy unit alarms
- The labeling has been updated to reflect the proposed change in mute times

#### Performance Data [21 CFR 807.92(b)]

#### Summary of non-clinical tests conducted for determination of substantial equivalence [21 CFR 807.92(b)(1)]

The Prevena 125 Therapy Unit was evaluated to ensure conformance to design specifications following minor changes to software to reduce nuisance alerts.

The following bench tests were conducted:

- . Performance testing to confirm therapy unit delivers negative pressure wound therapy within specification
- Software verification and validation testing to confirm that therapy unit performs as designed ●
- Battery life testing to demonstrate that changes do not affect battery life. .

#### Summary of clinical tests conducted for determination of substantial equivalence or of clinical information [21 CFR 807.92(b)(2)]

No clinical tests were necessary. However:

- . Usability testing was conducted with patient users to assess the impact of proposed labeling changes made as a result of software changes. The results demonstrated that all usability qoals were met, and recommended changes to the Patient Guide have been addressed.
#### Conclusions drawn [21 CFR 807.92(b)(3)]

The Prevena Incision Management System and its predicate (K121883) are substantially equivalent in terms of safety, function and indications for use.

---

**Source:** [https://fda-staging.innolitics.com/device/K150006](https://fda-staging.innolitics.com/device/K150006)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
