K150002 · Saset (Chengdu), Inc. · IYN · Jun 23, 2015 · Radiology
Device Facts
Record ID
K150002
Device Name
iNSIGHT Color Doppler Ultrasound System
Applicant
Saset (Chengdu), Inc.
Product Code
IYN · Radiology
Decision Date
Jun 23, 2015
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1550
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The iNSIGHT Color Doppler Ultrasound System is intended for visualization of ultrasound of internal organs, for medical diagnostic purposes only. The main applications are: General radiology, Abdominal (excluding fetal Doppler), Vascular, OB/GYN, Urology, Breast, Superficial Organ (Small Parts), Musculoskeletal, and Cardiology. Each application includes a set of exams, including the specific measurements, reports, pictograms, annotations and system presets.
Device Story
iNSIGHT Color Doppler Ultrasound System is a preprogrammed diagnostic imaging system. It acquires ultrasound signals via various transducers (convex, phased, linear, volume, cavity) to produce images for clinical diagnosis. Imaging modes include B, M, PW, CW, CDI, PDI, DPDI, TDI, MC, THI, Spatial Compounding, Panoramic, Freehand 3D, 4D, Strain, and ARFI Elasticity. Used in clinical settings by healthcare professionals to visualize internal organs and perform measurements/reports. Output is displayed on the system for clinician interpretation to aid in medical diagnosis. Benefits include non-invasive visualization of anatomical structures and blood flow.
Clinical Evidence
No clinical data. Bench testing only. Compliance with IEC 60601-1, IEC 60601-1-2, IEC 60601-2-37, NEMA UD 2-2004, and ISO 10993-5/10 standards verified. Accuracy of strain and elasticity quantitative measurements was validated via bench testing.
Technological Characteristics
Color Doppler ultrasound system with various transducers (convex, phased, linear, volume, cavity). Materials comply with ISO 10993-5 (cytotoxicity) and ISO 10993-10 (irritation/sensitization). Operates under Track 1 acoustic output limits. Electrical safety per IEC 60601-1; EMC per IEC 60601-1-2; performance per IEC 60601-2-37. Supports multiple imaging modes including B, M, PW, CW, CDI, PDI, DPDI, TDI, MC, THI, Spatial Compounding, Panoramic, 3D/4D, and ARFI.
Indications for Use
Indicated for diagnostic ultrasound imaging or fluid flow analysis of internal organs in general radiology, abdominal, vascular, OB/GYN, urology, breast, superficial organ, musculoskeletal, and cardiology applications. Patient population includes adults and pediatrics.
Regulatory Classification
Identification
An ultrasonic pulsed doppler imaging system is a device that combines the features of continuous wave doppler-effect technology with pulsed-echo effect technology and is intended to determine stationary body tissue characteristics, such as depth or location of tissue interfaces or dynamic tissue characteristics such as velocity of blood or tissue motion. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
Acuson S2000 and S3000 Diagnostic Ultrasound Systems (K130881)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the seal of the U.S. Department of Health and Human Services. The seal is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the seal is an abstract symbol that resembles three human profiles facing to the right, stacked on top of each other.
June 23, 2015
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
Saset (Chengdu) Inc. % Ms. Diana Hong General Manager Mid-Link Consulting Co., Ltd. P.O. Box 120-119 Shanghai 200120 CHINA
Re: K150002
Trade/Device Name: iNSIGHT Color Doppler Ultrasound System Regulation Number: 21 CFR 892.1550 Regulation Name: Ultrasonic pulsed doppler imaging system Regulatory Class: II Product Code: IYN, IYO, ITX Dated: June 15, 2015 Received: June 17, 2015
Dear Ms. Hong:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
Michael D'Hara
For
Robert Ochs, Ph.D. Acting Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
### Indications for Use
510(k) Number (if known) K150002
#### Device Name
Trade Name: iNSIGHT Color Doppler Ultrasound System
Models of iNSIGHT Color Doppler Ultrasound System: iNSIGHT 25C, iNSIGHT 35R, iNSIGHT 23R, iNSIGHT 11R, iNSIGHT 10B and iMago c21; Models of Probes: SH3C50, SH3P30, SH4DC54, SH6E47 and SH7L38
#### Indications for Use (Describe)
The iNSIGHT Color DopplerUltrasound System is intended for visualization of ultrasound of internal organs, for medical diagnostic purposes only. The main applications are: General radiology, Abdominal (excluding fetal Doppler), Vascular, OB/GYN, Urology, Breast, Superficial Organ (Small Parts), Musculoskeletal, and Cardiology. Each application includes a set of exams, including the specific measurements, reports, pictograms, annotations and system presets.
Type of Use (Select one or both, as applicable)
| <span style="font-family: Arial;"> <span style="font-size: 10pt;"> <input checked="true" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D) </span> </span> |
|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| <span style="font-family: Arial;"> <span style="font-size: 10pt;"> <input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </span> |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
System: iNSIGHT Color Doppler Ultrasound System
Transducer: SH3C50 Convex Array Transducer
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical Application | Mode of Operation | | | | | | | |
|-----------------------------|------------------------------------|---|---|-----|-----|------------------|-----------------------|---------------------|
| | | B | M | PWD | CWD | Color<br>Doppler | Combined<br>(Specify) | Other*<br>(Specify) |
| General<br>(Track 1 Only) | Specific<br>(Track 1&3) | | | | | | | |
| Ophthalmic | Ophthalmic | | | | | | | |
| Fetal<br>Imaging &<br>Other | Fetal | N | N | N | | N | BMDC | 1,2,3 |
| | Abdominal | N | N | N | | N | BMDC | 1,2,3 |
| | Intra-operative(Specify) | | | | | | | |
| | Intra-operative(Neuro) | | | | | | | |
| | Laparoscopic | | | | | | | |
| | Pediatric | N | N | N | | N | BMDC | 1,2,3 |
| | Small Organ(Specify) | | | | | | | |
| | Neonatal Cephalic | | | | | | | |
| | Adult Cephalic | | | | | | | |
| | Trans-rectal | | | | | | | |
| | Trans-vaginal | | | | | | | |
| | Trans-urethral | | | | | | | |
| | Trans-esoph.(non-Card.) | | | | | | | |
| | Musculo-skeletal<br>(Conventional) | | | | | | | |
| | Musculo-skeletal<br>(Superficial) | | | | | | | |
| | Intravascular | | | | | | | |
| | Other (Specify) | | | | | | | |
| Cardiac | Cardiac Adult | | | | | | | |
| | Cardiac Pediatric | | | | | | | |
| | Intravascular (Cardiac) | | | | | | | |
| | Trans-esoph.(Cardiac) | | | | | | | |
| | Intra-cardiac | | | | | | | |
| | Other (Specify) | | | | | | | |
| Peripheral<br>Vessel | Peripheral vessel | | | | | | | |
| | Other (Specify) | | | | | | | |
N = new indication; P = previously cleared by FDA; E = added under this appendix
- 1. Tissue Harmonic Imaging (THI)
- 3. Freehand 3D imaging
- 5. STIC
- 7. ARFI strain image
- 9.4D Imaging
- 2. Spatial Compounding Imaging
- 4. Panoramic B and Color mode
- 6. Strain image
- 8. Tissue Doppler Imaging (TDI)
{4}------------------------------------------------
System: iNSIGHT Color Doppler Ultrasound System
Transducer: SH3P30 Phased Array Transducer
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical Application | Mode of Operation | | | | | | | |
|-----------------------------|------------------------------------|---|---|-----|-----|------------------|-----------------------|---------------------|
| General<br>(Track 1 Only) | Specific<br>(Track 1&3) | B | M | PWD | CWD | Color<br>Doppler | Combined<br>(Specify) | Other*<br>(Specify) |
| Ophthalmic | Ophthalmic | | | | | | | |
| Fetal<br>Imaging &<br>Other | Fetal | | | | | | | |
| | Abdominal | | | | | | | |
| | Intra-operative(Specify) | | | | | | | |
| | Intra-operative(Neuro) | | | | | | | |
| | Laparoscopic | | | | | | | |
| | Pediatric | | | | | | | |
| | Small Organ(Specify) | | | | | | | |
| | Neonatal Cephalic | | | | | | | |
| | Adult Cephalic | N | N | N | | N | BMDC | 1 |
| | Trans-rectal | | | | | | | |
| | Trans-vaginal | | | | | | | |
| | Trans-urethral | | | | | | | |
| | Trans-esoph.(non-Card.) | | | | | | | |
| | Musculo-skeletal<br>(Conventional) | | | | | | | |
| | Musculo-skeletal<br>(Superficial) | | | | | | | |
| | Intravascular | | | | | | | |
| | Other (Specify) | | | | | | | |
| Cardiac | Cardiac Adult | N | N | N | N | N | BMDC | 1,5,8 |
| | Cardiac Pediatric | N | N | N | N | N | BMDC | 1,5,8 |
| | Intravascular (Cardiac) | | | | | | | |
| | Trans-esoph.(Cardiac) | | | | | | | |
| | Intra-cardiac | | | | | | | |
| | Other (Specify) | | | | | | | |
| Peripheral<br>Vessel | Peripheral vessel | | | | | | | |
| | Other (Specify) | | | | | | | |
N = new indication; P = previously cleared by FDA; E = added under this appendix
- 1. Tissue Harmonic Imaging (THI)
- 3. Freehand 3D imaging
- 5. STIC
- 7. ARFI strain image
- 9.4D Imaging
- 2. Spatial Compounding Imaging
- 4. Panoramic B and Color mode
- 6. Strain image
- 8. Tissue Doppler Imaging (TDI)
{5}------------------------------------------------
System: iNSIGHT Color Doppler Ultrasound System
Transducer: SH4DC54 Volume Transducer
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical Application | | Mode of Operation | | | | | | |
|-----------------------------|------------------------------------|-------------------|---|-----|-----|------------------|-----------------------|---------------------|
| General<br>(Track 1 Only) | Specific<br>(Track 1&3) | B | M | PWD | CWD | Color<br>Doppler | Combined<br>(Specify) | Other*<br>(Specify) |
| Ophthalmic | Ophthalmic | | | | | | | |
| Fetal<br>Imaging &<br>Other | Fetal | N | N | N | | N | BMDC | 1,2,3,9 |
| | Abdominal | N | N | N | | N | BMDC | 1,2,3,9 |
| | Intra-operative(Specify) | | | | | | | |
| | Intra-operative(Neuro) | | | | | | | |
| | Laparoscopic | | | | | | | |
| | Pediatric | | | | | | | |
| | Small Organ(Specify) | | | | | | | |
| | Neonatal Cephalic | | | | | | | |
| | Adult Cephalic | | | | | | | |
| | Trans-rectal | | | | | | | |
| | Trans-vaginal | | | | | | | |
| | Trans-urethral | | | | | | | |
| | Trans-esoph.(non-Card.) | | | | | | | |
| | Musculo-skeletal<br>(Conventional) | | | | | | | |
| | Musculo-skeletal<br>(Superficial) | | | | | | | |
| | Intravascular | | | | | | | |
| | Other (Specify) | | | | | | | |
| Cardiac | Cardiac Adult | | | | | | | |
| | Cardiac Pediatric | | | | | | | |
| | Intravascular (Cardiac) | | | | | | | |
| | Trans-esoph.(Cardiac) | | | | | | | |
| | Intra-cardiac | | | | | | | |
| Peripheral<br>Vessel | Peripheral vessel | | | | | | | |
| | Other (Specify) | | | | | | | |
N = new indication; P = previously cleared by FDA; E = added under this appendix
- 1. Tissue Harmonic Imaging (THI)
- 3. Freehand 3D imaging
- 5. STIC
- 7. ARFI strain image
- 9.4D Imaging
- 2. Spatial Compounding Imaging
- 4. Panoramic B and Color mode
- 6. Strain image
- 8. Tissue Doppler Imaging (TDI)
{6}------------------------------------------------
System: iNSIGHT Color Doppler Ultrasound System
Transducer: SH6E47 Cavity Transducer
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical Application | | Mode of Operation | | | | | | |
|-----------------------------|------------------------------------|-------------------|---|-----|-----|------------------|---------------------------|---------------------|
| General<br>(Track 1 Only) | Specific<br>(Track 1&3) | B | M | PWD | CWD | Color<br>Doppler | Combined<br>(Specify) | Other*<br>(Specify) |
| Ophthalmic | Ophthalmic | | | | | | | |
| Fetal<br>Imaging &<br>Other | Fetal | | | | | | | |
| | Abdominal | | | | | | | |
| | Intra-operative(Specify) | | | | | | | |
| | Intra-operative(Neuro) | | | | | | | |
| | Laparoscopic | | | | | | | |
| | Pediatric | | | | | | | |
| | Small Organ(Specify) | | | | | | | |
| | Neonatal Cephalic | | | | | | | |
| | Adult Cephalic | | | | | | | |
| | Trans-rectal | | | | | | | |
| | Trans-vaginal | N | N | N | | N | BMDC | 1,2 |
| | Trans-urethral | | | | | | | |
| | Trans-esoph.(non-Card.) | | | | | | | |
| | Musculo-skeletal<br>(Conventional) | | | | | | | |
| | Musculo-skeletal<br>(Superficial) | | | | | | | |
| | Intravascular | | | | | | | |
| | Other (Specify) | | | | | | | |
| Cardiac | Cardiac Adult | | | | | | | |
| | Cardiac Pediatric | N | N | N | | N | BMDC | 1,2 |
| | Intravascular (Cardiac) | | | | | | | |
| | Trans-esoph.(Cardiac) | | | | | | | |
| | Intra-cardiac | | | | | | | |
| | Other (Specify) | | | | | | | |
| Peripheral<br>Vessel | Peripheral vessel | | | | | | | |
| | Other (Specify) | | | | | | | |
| Clinical Application | | Mode of Operation | | | | | | |
| General<br>(Track 1 Only) | Specific<br>(Track 1&3) | B | M | PWD | CWD | Color<br>Doppler | Combine<br>d<br>(Specify) | Other*<br>(Specify) |
| Ophthalmic | Ophthalmic | | | | | | | |
| Fetal<br>Imaging &<br>Other | Fetal | | | | | | | |
| | Abdominal | N | N | N | | N | BMDC | 1,2,3,4,6,7 |
| | Intra-operative(Specify) | | | | | | | |
| | Intra-operative(Neuro) | | | | | | | |
| | Laparoscopic | | | | | | | |
| | Pediatric | N | N | N | | N | BMDC | 1,2,3,4,6,7 |
| | Small Organ(Specify) | N | N | N | | N | BMDC | 1,2,3,4,6,7 |
| | Neonatal Cephalic | | | | | | | |
| | Adult Cephalic | N | N | N | | N | BMDC | 1,2,3,4,6,7 |
| | Trans-rectal | | | | | | | |
| | Trans-vaginal | | | | | | | |
| | Trans-urethral | | | | | | | |
| | Trans-esoph.(non-Card.) | | | | | | | |
| | Musculo-skeletal<br>(Conventional) | N | N | N | | N | BMDC | 1,2,3,4,6,7 |
| | Musculo-skeletal<br>(Superficial) | N | N | N | | N | BMDC | 1,2,3,4,6,7 |
| | Intravascular | | | | | | | |
| | Other (Specify) | | | | | | | |
| Cardiac | Cardiac Adult | | | | | | | |
| | Cardiac Pediatric | | | | | | | |
| | Intravascular (Cardiac) | | | | | | | |
| | Trans-esoph.(Cardiac) | | | | | | | |
| | Intra-cardiac | | | | | | | |
| | Other (Specify) | | | | | | | |
| Peripheral<br>Vessel | Peripheral vessel | N | N | N | | N | BMDC | 1,2,3,4,6,7 |
| | Other (Specify) | | | | | | | |
N = new indication; P = previously cleared by FDA; E = added under this appendix
- 1. Tissue Harmonic Imaging (THI)
- 3. Freehand 3D imaging
- 5. STIC
- 7. ARFI strain image
- 9.4D Imaging
- 2. Spatial Compounding Imaging
- 4. Panoramic B and Color mode
- 6. Strain image
- 8. Tissue Doppler Imaging (TDI)
{7}------------------------------------------------
System: iNSIGHT Color Doppler Ultrasound System
Transducer: SH7L38 Linear Array Transducer
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
N = new indication; P = previously cleared by FDA; E = added under this appendix
* Examples of other modes of operation may include: A-mode, Amplitude Doppler, 3-D Imaging, Harmonic Imaging, Tissue Motion Doppler, and Color Velocity Imaging.
- 1. Tissue Harmonic Imaging (THI)
- 3. Freehand 3D imaging
- 5. STIC
- 7. ARFI strain image
- 9.4D Imaging
- 4. Panoramic B and Color mode 6. Strain image
- 8. Tissue Doppler Imaging (TDI)
2. Spatial Compounding Imaging
{8}------------------------------------------------
### Tab #7 510(k) Summary
This 510(k) Summary of 510(k) safety and effectiveness information is being submitted in accordance with requirements of SMDA 1990 and Title 21, CFR Section 807.92.
The assigned 510(k) Number: K150002
- 1. Date of Preparation: 12/19/2014
- 2. Sponsor Identification
### Saset (Chengdu) Inc.
A-601, Huoju Building, #16 Chuangye RD B Hi-Tech Avenue, Hi-Tech Zone Chengdu, Sichuan, 610041, China
Establishment Registration Number: 3007751490
Contact Person: Mr. Dong Chyuan Liu Position: General Manager Tel: 0086-28-85130926-808 Fax: 0086-28-85130926-100 Email: dongcliu@163.com
- 3. Designated Submission Correspondent
Ms. Diana Hong (Primary Contact Person) Ms. Lee Fu (Alternative Contact Person)
Mid-Link Consulting Co., Ltd P.O. Box 120-119, Shanghai, 200120, China
Tel: +86-21-22815850, Fax: 240-238-7587 Email: info@mid-link.net
{9}------------------------------------------------
#### 4. Identification of Proposed Device
Trade Name: iNSIGHT Color Doppler Ultrasound System Common Name: Diagnostic Ultrasound System and Transducers
Models of iNSIGHT Color Doppler Ultrasound System: iNSIGHT 37C, iNSIGHT 25C, iNSIGHT 35R, iNSIGHT 23R, iNSIGHT 11R, iNSIGHT 10B and iMago c21
Models of Probes: SH3C50, SH3P30, SH4DC54, SH6E47 and SH7L38
Regulatory Information Classification Name: Ultrasound Pulsed Doppler Imaging System; Classification: II; Product Code: IYN: Regulation Number: 21 CFR 892.1550; Review Panel: Radiology:
Classification Name: Ultrasonic Pulsed Echo Imaging System; Classification: II; Product Code: IYO; Regulation Number: 21 CFR 892.1560; Review Panel: Radiology;
Classification Name: Diagnostic Ultrasound Transducer; Classification: II; Product Code: ITX; Regulation Number: 21 CFR 892.1570; Review Panel: Radiology:
#### Intended Use Statement:
The iNSIGHT Color Doppler Ultrasound System is intended for visualization of ultrasound of internal organs, for medical diagnostic purposes only. The main applications are: General radiology, Abdominal (excluding fetal Doppler), Vascular, OB/GYN, Urology, Breast, Superficial Organ (Small Parts), Musculoskeletal, and Cardiology. Each application includes a set of exams, including the specific measurements, reports, pictograms, annotations and system presets.
#### Device Description
The iNSIGHT Color Doppler Ultrasound Systems are integrated preprogrammed color ultrasound imaging system, capable of producing a resolution intended for clinical diagnostic imaging applications. Their basic function is to acquire ultrasound signal and display the images in the following imaging operations: B, M, PW, CW, CDI, PDI, DPDI, TDI, MC and THI imaging mode, Spatial Compounding Imaging, Panoramic B and Color mode, Freehand 3D, 4D Imaging, Strain imaging, ARFI Elasticity Imaging.
{10}------------------------------------------------
- ર . Identification of Predicate Device
510(k) Number: K090059 Product Name: iMago c21 Diagnostic Ultrasound System Model Name: iMago c21
510(k) Number: K130881 Product Name: Acuson S2000 and S3000 Diagnostic Ultrasound Systems Model Name: Acuson S2000
- 6. Non-Clinical Test Conclusion
Non clinical tests were conducted to verify that the proposed device met all design specifications as was Substantially Equivalent (SE) to the predicate device. The test results demonstrated that the proposed device complies with the following standards:
- > IEC 60601-1:2005 + CORR. 1: 2006 + CORR. 2:2007 + AM 1: 2012 Medical electrical equipment - Part 1: General requirements for basic safety and essential performance
- > IEC 60601-1-2: 2007 General requirements for basic safety and essential performance -Collateral standard: Electromagnetic compatibility - Requirements and tests.
- A IEC 60601-2-37: 2007 Medical electrical equipment - Part2 -37 Particular requirements for the basic safety and essential performance of ultrasonic medical diagnostic and monitoring equipment.
- A NEMA UD 2-2004 (R2009) Acoustic output measurement standard for diagnostic ultrasound equipment
- > ISO 10993-5:2009, Biological Evaluation of Medical Device, Part 5-Tests for Vitro Cytotoxicity:
- > ISO 10993-10:2010, Biological evaluation of medical devices Part 10: Tests for irritation and skin sensitization;
Discussion about the testing of strain and elasticity quantitative measurements:
The bench testing for the features providing strain and elasticity quantitative has been conducted to show that the accuracy of the strain and elasticity imaging of the proposed device; the result has been analyzed that they are accepted for clinical use.
- 7. Clinical Test Conclusion
No clinical study is included in this submission.
{11}------------------------------------------------
#### 8. Substantially Equivalent (SE) Comparison
| Item | Proposed Device | Predicate Device | Predicate Device |
|--------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | | K090059 | K130881 |
| Product Code | IYN, IYO and ITX | IYN, IYO and ITX | IYN/ IYO/ ITX/OBJ |
| Intended Use | The iNSIGHT Color Diagnostic Ultrasound System is intended for visualization of ultrasound of internal organs, for medical diagnostic purposes only. The main applications are: General radiology, Abdominal (excluding fetal Doppler), Vascular, OB/GYN, Urology, Breast, Superficial Organ (Small Parts), Musculoskeletal, and Cardiology. Each application includes a set of exams, including the specific measurements, reports, pictograms, annotations and system presets. | The iMago c21 medical ultrasound system is intended for visualization of ultrasound of internal organs, for medical diagnostic purposes only. The main applications are: General radiology, Abdominal (excluding fetal Doppler), Vascular, OB/GYN, Urology, Breast, Superficial Organ (Small Parts), Musculoskeletal, and Cardiology. Each application includes a set of exams, including the specific measurements, reports, pictograms, annotations and system presets. | The S2000 and S3000 ultrasound imaging systems are intended for the following applications: Fetal, Abdominal, Intraoperative, Pediatric, Small Parts, Transcranial, OB/GYN, Cardiac, Pelvic, Neonatal/Adult Cephalic, Vascular, Musculoskeletal, Superficial Musculoskeletal, and Peripheral vascular applications.<br>The system also provides the ability to measure anatomical structures {fetal, abdominal, intraoperative, intraoperative neurological, pediatric, small organ, neonatal cephalic, adult cephalic, cardiac, trans-esophageal, transrectal, transvaginal, peripheral vessel, musculo-skeletal (conventional), musculo-skeletal (superficial) and neonatal cardiac} and calculation packages that provide information that provide information to the clinician that may be used adjunctively with other clinical diagnosis purposes.<br>The arterial health package (AHP) software provides the physician with the capability to measure Intima Media Thickness and the option to reference normative tables that have been validated and published in peer-reviewed studies. The information is intended to provide the physician with an easily understood tool for communicating with patients regarding state of their cardiovascular system. This feature should be utilized according to the "ASE Consensus Statement; Use of Carotid Ultrasound to Identify Subclinical Vascular Disease and Evaluate Cardiovascular Disease Risk; A |
### Table 1 Comparison of Technology Characteristics
{12}------------------------------------------------
| | | | | consensus Statement from the Amerian<br>Association of Echocardiography; Carotid<br>Intima-Media Thickness Task Force,<br>Endorsed by the Society for Vascular<br>Imaging" |
|------------------------|---------------|------------------------------------|------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Configuration | | Color Doppler Ultrasound<br>System | Color Doppler Ultrasound<br>System | Color Doppler Ultrasound System |
| | | Probes | Probes | Probes |
| | | / | / | Ultrasound Catheter |
| Acoustic Track | | Track 1 | Track 1 | Track 3 |
| Biocom | cytotoxicity | ISO 10993-5 | ISO 10993-5 | ISO 10993-5 |
| patibilit | sensitization | ISO 10993-10 | ISO 10993-10 | ISO 10993-10 |
| y | irritation | ISO 10993-10 | ISO 10993-10 | ISO 10993-10 |
| Electrical Safety | | IEC 60601-1 | IEC 60601-1 | IEC 60601-1 |
| EMC | | IEC 60601-1-2 | IEC 60601-1-2 | IEC 60601-1-2 |
| Particular Performance | | IEC 60601-2-37 | IEC 60601-2-37 | IEC 60601-2-37 |
#### 9. Substantially Equivalent (SE) Conclusion
Based on the comparison and analysis above, the proposed devices are determined to be Substantially Equivalent (SE) to the predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.