K143728 · Suzhou Minhua Medical Apparatus · DXQ · May 15, 2015 · Cardiovascular
Device Facts
Record ID
K143728
Device Name
Minhua Disposable Blood Pressure Cuff
Applicant
Suzhou Minhua Medical Apparatus
Product Code
DXQ · Cardiovascular
Decision Date
May 15, 2015
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.1120
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The Minhua Disposable Blood Pressure Cuff is an accessory used in conjunction with non-invasive blood pressure monitoring systems for determination of a person's blood pressure. The cuff is non-sterile and for single-patient use. It is available in neonatal, pediatric and adult sizes. The cuff is not designed, sold, or intended for use except as indicated.
Device Story
Device is an inelastic sleeve with an integrated inflatable bladder and tubing; wraps around patient limb; secured via hook and loop fastener. Connects to non-invasive blood pressure monitoring systems to facilitate indirect blood pressure determination. Single-patient use; non-sterile. Used in clinical settings to assist healthcare providers in patient monitoring. Benefits include accurate blood pressure measurement across diverse patient sizes, including neonates, pediatrics, adults, and overweight subjects.
Clinical Evidence
Bench testing only. Testing performed per ANSI/AAMI SP 10 (2002+A1:2003+A2:2006+(R)2008) for sphygmomanometers and ISO 10993 series (Part 1, 5, and 10) for biocompatibility, cytotoxicity, irritation, and skin sensitization.
Technological Characteristics
Materials: Non-woven fabric, PVC tubing, nylon brushed fabric. Sensing principle: Inflatable bladder for indirect pressure measurement. Pressure limit: 0-300 mmHg. Connectivity: Tubing connection to external monitoring system. Biocompatibility: ISO 10993 compliant. Form factor: Various sizes (neonatal to thigh).
Indications for Use
Indicated for use in neonatal, pediatric, and adult patient populations for the indirect measurement of blood pressure when used with non-invasive blood pressure monitoring systems.
Regulatory Classification
Identification
A blood pressure cuff is a device that has an inflatable bladder in an inelastic sleeve (cuff) with a mechanism for inflating and deflating the bladder. The cuff is used in conjunction with another device to determine a subject's blood pressure.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features the department's name in a circular arrangement around a symbol. The symbol is a stylized representation of three human profiles facing right, stacked on top of each other. The profiles are rendered in a simple, line-art style.
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-002
May 15, 2015
Suzhou Minhua Medical Apparatus Supplies Co., Ltd. c/o Rhonda Alexander, Senior Regulatory Specialist Registrar Corp. Medical Device Division 144 Research Drive Hampton, Virginia 23666
Re: K143728
> Trade/Device Name: Minhua Disposable Blood Pressure Cuff Regulation Number: 21 CFR 870.1120 Regulation Name: Blood Pressure Cuff Regulatory Class: Class II Product Code: DXQ Dated: March 24, 2015 Received: March 25, 2015
Dear Ms. Alexander:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Page 2 - Rhonda Alexander
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
Kenneth J. Cavanaugh -S
for
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
# Indications for Use
510(k) Number (if known)
Device Name Minhua Disposable Blood Pressure Cuff
## Indications for Use (Describe)
The Minhua Disposable Blood Pressure Cuff is an accessory used in conjunction with non-invasive blood pressure monitoring systems for determination of a person's blood pressure. The cuff is non-sterile and for single-patient use. It is available in neonatal, pediatric and adult sizes. The cuff is not designed, sold, or intended for use except as indicated.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
## PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
#### FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows a logo for MINHUA. The logo is in blue and features a stylized font. There is a red dot above the letter 'i' in MINHUA. The logo has a modern and clean design.
Suzhou Minhua Medical Apparatus Supplies Co .,LTD Tel: +0086-13601885073 fax:+0086-021-59841086 Email:QD@minhuayiliao.com
## 510(k) SUMMARY (21 CFR 807.92)
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirement of 21 CFR 807.92
The assigned 510(k) number is:
Premarket Notification [510(k)] Summary
## A. General Information
| Submitter's Name: | Suzhou Minhua Medical Apparatus Supplies Co., LTD |
|-------------------|---------------------------------------------------|
| Address: | You Yi Industrial park, Songlin Town, |
| | WuJiang ,CHINA 215222 |
| Telephone: | +86 21 59840221-8003 |
| Fax Number: | +86 21 59841086 |
| Contact Person: | Ji Dong Yang, Vice General Manager |
| E-mail: | QD@minhuayiliao.com |
| Date Prepared: | September. 03, 2014 |
## B. Device
| Trade Name: | Minhua Disposable Blood Pressure Cuff |
|--------------------|---------------------------------------|
| Common Name: | Blood Pressure Cuff |
| Product Code: | DXQ |
| Class: | Class 2 |
| Regulation Number: | 21 CFR 870.1120 |
| Review Panel: | Cardiovascular |
## C. Identification of Legally Marketed Predicate Device
| Name: | Medline Disposable Blood pressure cuffs |
|---------------|-----------------------------------------|
| Manufacture: | Medline Industries, Inc. |
| K Number: | K071244 |
| Date Cleared: | March 5, 2008 |
## D. Description of the Device
The device comprises tubing attached to an inelastic sleeve with an integrated inflatable bladder that is wrapped around the patient's limb, secured by hook and loop closure, and used to measure the patient's blood pressure. The device tubing is connected to a non-invasive blood pressure measurement system.
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows a logo with the word "MINHUA" in blue, stylized with a unique font. The letters are connected and have a geometric appearance. There is a red dot above the "i" in "MINHUA". The logo has a modern and corporate feel.
Suzhou Minhua Medical Apparatus Supplies Co .,LTD Tel: +0086-13601885073 fax:+0086-021-59841086 Email:QD@minhuayiliao.com
## E. Intended Use
Indirect measurement of blood pressure.
#### F .Indications for use
The Minhua Disposable Blood Pressure Cuff is an accessory used in conjunction with non-invasive blood pressure monitoring systems for determination of a person's blood pressure. The cuff is non-sterile and for single-patient use. It is available in neonatal, pediatric and adult sizes. The cuff is not designed, sold, or intended for use except as indicated.
#### G. Various size
| Model | Cuff Size | Arm Range | Bladder Range |
|-------------|-----------|-----------|---------------|
| 002DC01NJT1 | 100x26mm | 30-60mm | 68x26mm |
| 002DC01NJT2 | 100x26mm | 30-60mm | |
| 002DC02NJT1 | 119x31mm | 40-80mm | 76 x 31mm |
| 002DC02NJT2 | 119x31mm | 40-80mm | |
| 002DC03NJT1 | 146x41mm | 60-110mm | 96 x 41mm |
| 002DC03NJT2 | 146x41mm | 60-110mm | |
| 002DC04NJT1 | 181x48mm | 70-130mm | 123 x 48mm |
| 002DC04NJT2 | 181x48mm | 70-130mm | |
| 002DC05NJT1 | 191x56mm | 80-150mm | 132 x 56mm |
| 002DC05NJT2 | 191x56mm | 80-150mm | |
| 002DC06NJT1 | 195x55mm | 80-140mm | 85 x 55mm |
| 002DC06NJT2 | 195x55mm | 80-140mm | |
| 002DC07NJT1 | 274x83mm | 130-200mm | 135 x 83mm |
| 002DC07NJT2 | 274x83mm | 130-200mm | |
| 002DC09NJT1 | 335x107mm | 180-260mm | 185 x 107mm |
| 002DC09NJT2 | 335x107mm | 180-260mm | |
| 002DC11NJT1 | 524x143mm | 260-350mm | 265 x 143mm |
| 002DC11NJT2 | 524x143mm | 260-350mm | |
| 002DC13NJT1 | 605x174mm | 320-420mm | 325 x 174mm |
| 002DC13NJT2 | 605x174mm | 320-420mm | |
| 002DC15NJT1 | 745x198mm | 420-500mm | 425 x 198mm |
| 002DC15NJT2 | 745x198mm | 420-500mm | |
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image contains a logo with the word "MINHUA" in blue, stylized with a unique font. The "i" in "MINHUA" is dotted with a red circle. The logo has a modern and corporate feel, with the letters connected by a blue line at the bottom.
Suzhou Minhua Medical Apparatus Supplies Co .,LTD Tel: +0086-13601885073 fax:+0086-021-59841086 Email:QD@minhuayiliao.com
## H. Comparison to the Predicate Device
| Features | Su Zhou Minhua Medical Apparatus Supplies<br>Co., Ltd | Medline Industries, Inc. |
|------------------------------|---------------------------------------------------------------------------------------|----------------------------------------------------------|
| Intended use | Indirect measurement of blood pressure | Indirect measurement of blood pressure |
| Patient<br>Populations | Adults/Pediatrics | Adults/Pediatrics |
| Tube<br>Configuration | One or two tube | One or two tube |
| Sterility | Not supplied sterile | Not supplied sterile |
| Pressure limits | 0-300mmHg | 0-300mmHg |
| Biocompatibility | Comply with ISO <b>10993</b> biocompatibility evaluation | Comply with ISO <b>10993</b> biocompatibility evaluation |
| cuff | Non-woven fabric, Nontoxic | Non-woven fabric, Nontoxic |
| Tube port<br>connector | PVC:stable;corrosion resistance;heat-resistance | PVC:stable;corrosion resistance;heat-resistance |
| Hook & loop<br>fastener | Non-woven fabric+Nylon brushed fabric :stable, corrosion resistance; heat-resistance; | Non-woven fabric+Nylon brushed fabric |
| Conform to AHA bladder sizes | | |
| | Recommendations | Recommendations |
| | 002DC01NJT1 100x26mm (3-6cm) | MDS 9741 100x26mm (3-6cm) |
| Neonat al 1 | 002DC01NJT2 100x26mm (3-6cm) | MDS 9751 100x26mm (3-6cm) |
| Neonat al 2 | 002DC02NJT1 119x31mm (4-8cm)<br>002DC02NJT2 119x31mm (4-8cm) | MDS 9742 119x31mm (4-8cm)<br>MDS 9752 119x31mm (4-8cm) |
| Neonat al 3 | 002DC03NJT1 146x41mm (6-11cm)<br>002DC03NJT21 146x41mm (6-11cm) | MDS 9743 146x41mm (6-11cm)<br>MDS 9753 146x41mm (6-11cm) |
| Neonat al 4 | 002DC04NJT1 181x48mm (7-13cm)<br>002DC04NJT2 181x48mm (7-13cm) | MDS 9744 181x48mm (7-13cm)<br>MDS 9754 181x48mm (7-13cm) |
| Neonat al 5 | 002DC05NJT1 191x56mm (8-15cm)<br>002DC05NJT2 191x56mm (8-15cm) | MDS 9745 191x56mm (8-15cm)<br>MDS 9755 191x56mm (8-15cm) |
| | 002DC06NJT1 195x55mm (8-14 cm) | MDS 9710 195x55mm (8-14cm) |
{6}------------------------------------------------
Suzhou Minhua Medical Apparatus Supplies Co .,LTD Tel: +0086-13601885073 fax:+0086-021-59841086 Email:QD@minhuayiliao.com
| Size | Child | 002DC07NJT1 274X83mm (13-20cm) | Child | MDS 9711 274X83mm (13-20cm) |
|------|-------------|---------------------------------|-------------|------------------------------|
| | | 002DC07NJT2 274X83mm (13-20cm) | | MDS 9721 274X83mm (13-20cm) |
| | Small adult | 002DC09NJT1 335x107mm (18-26cm) | Small adult | MDS 9712 335x107mm (18-26cm) |
| | | 002DC09NJT2 335x107mm (18-26cm) | | MDS 9722 335x107mm (18-26cm) |
| | Adult | 002DC11NJT1 524X143mm (26-35cm) | Adult | MDS 9713 524X143mm (26-35cm) |
| | | 002DC11NJT2 524X143mm (26-35cm) | | MDS 9723 524X143mm (26-35cm) |
| | Big adult | 002DC13NJT1 605x174mm (32-42cm) | Big adult | MDS 9714 605x174mm (32-42cm) |
| | | 002DC13NJT2 605x174mm (32-42cm) | | MDS 9724 605x174mm (32-42cm) |
| | Thigh | 002DC15NJT1 745x198mm (42-50cm) | Thigh | MDS 9715 745x198mm (42-50cm) |
| | | 002DC15NJT2 745x198mm (42-50cm) | | MDS 9725 745x198mm (42-50cm) |
The Minjua Disposable Blood Pressure Cuff has the same intended use, basic construction, and technology specification as the predicated devices are wrapped the patient's arm or leg and secured by a hook and loop fastener commonly called Velcro. Both devices are available in the same size and range and are intended for the same patient populations. The materials of both devices are all conformed to ISO 10993. As with the predicate, we provide extended sizes of the cuffs in order to accommodate special groups, such as overweight subjects. Based on the performance testing in this submission, the slight difference between these blood pressure cuffs does not raise any safety or effectiveness issue.
## I. Summary of testing
The Minhua Disposable Blood Pressure Cuff has been tested according to the following standards: *ANSI/AAMI SP 10, Manual, electronic or automated sphygmomanometers, 2002+A1:2003+A2:2006+(R)2008
*ISO 10993-1, Biological evaluation of medical devices -- Part 1: Evaluation and testing within a risk management process, 2009
*ISO 10993-5, Biological evaluation of medical devices- Part 5: Tests for In Vitro cytotoxicity, 2009
*ISO 10993-10, Biological evaluation of medical devices -- Part 10: Tests for irritation and skin
{7}------------------------------------------------
Image /page/7/Picture/1 description: The image shows a logo for MINHUA. The logo is blue and features a stylized design with the letters "MINHUA" incorporated into the design. There is a red dot above the letter "i" in MINHUA. The logo has a modern and corporate look.
Suzhou Minhua Medical Apparatus Supplies Co .,LTD Tel: +0086-13601885073 fax:+0086-021-59841086 Email:QD@minhuayiliao.com
sensitization :2010
## J. Conclusion of Substantial Equivalence
Based on the comparison of intended use, design, materials and performance, we conclude that the new device is substantially equivalent to the predicate. The differences between the devices do not raise new questions of safety and effectiveness.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.