The DPR Cable is intended: - Spinal applications including sublaminar and intraspinous process wiring for trauma applications. - Trochanteric reattachment after trochanteric osteotomy following total hip arthroplasty. - Sternotomy indications including the "rewiring" of osteomized sternums. - Trauma surgery indications including olecranon, ankle, patella and some shoulder fracture rewiring. The device is intended for single patient use only.
Device Story
DPR Cable is a flexible, multi-strand ultra-high molecular weight polyethylene (UHMWPE) cerclage cable used for stabilization of bony segments. Fibers incorporate bismuth trioxide for radiographic visualization. Device is supplied sterile for single-patient use. During surgery, clinician wraps cable around bone segments and secures via knot. Provides high-strength fixation for spinal, orthopedic, and sternal trauma applications. MR safe. Benefits include high tensile and fatigue strength compared to metallic cable systems, providing stable bone fixation.
Clinical Evidence
Bench testing only. Biomechanical testing included fatigue strength, tensile force, creep, wear-debris, and knot strength. Biocompatibility testing included cytotoxicity and leaching assessments. Results indicate performance equivalent to predicate and reference devices.
Technological Characteristics
Flexible, multi-strand UHMWPE (Dyneema Purity) fibers with bismuth trioxide (Bi2O3) for radiopacity. 4 mm diameter. Fixation via knot. Sterilized via EtO. Supplied in double-layer packaging. MR safe.
Indications for Use
Indicated for patients requiring bone fixation via cerclage, including spinal sublaminar/intraspinous wiring for trauma, trochanteric reattachment post-osteotomy, sternotomy rewiring, and fracture rewiring (olecranon, ankle, patella, shoulder). Single patient use only.
Regulatory Classification
Identification
A bone fixation cerclage is a device intended to be implanted that is made of alloys, such as cobalt-chromium-molybdenum, and that consists of a metallic ribbon or flat sheet or a wire. The device is wrapped around the shaft of a long bone, anchored to the bone with wire or screws, and used in the fixation of fractures.
Predicate Devices
Titanium Alloy Songer Cable System/ Atlas Cable System (K920201)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing right, stacked on top of each other, representing the department's focus on people and health.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
October 29, 2015
DSM Biomedical Mr. Tom Maguire Senior Director Global Clinical and Regulatory 735 Pennsylvania Drive Exton, Pennsylvania 19341
Re: K143716
Trade/Device Name: DSM Biomedical DPR Cable Regulation Number: 21 CFR 888.3010 Regulation Name: Bone fixation cerclage Regulatory Class: Class II Product Code: JDQ Dated: September 16, 2015 Received: September 17, 2015
Dear Mr. Maguire:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
Page 2 – Mr. Tom Maguire
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# Mark N. Melkerson -S
Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Original 510(k) Premarket Notification DSM Biomedical DPR Cable
Image /page/2/Picture/2 description: The image shows the logo for DSM, a company focused on health, nutrition, and materials. The logo features a colorful, abstract circular design on the left, composed of overlapping shapes in shades of blue, purple, green, and orange. To the right of the circular design is the company name, "DSM," in bold, blue letters. Below the company name is the tagline, "BRIGHT SCIENCE. BRIGHTER LIVING."
#### Indications for Use Statement
The DPR Cable is intended for:
- Spinal applications include sublaminar and intraspinous process wiring for trauma ● applications.
- o Trochanteric reattachment after trochanteric osteotomy following total hip arthroplasty.
- Sternotomy indications including the "rewiring" of osteomized sternums.
- Trauma surgery indications including olecranon, ankle, patella and some shoulder fracture rewiring.
The device is intended for single patient use only.
Prescription Use
Over-The-Counter Use__
(Part 21 CFR 801 Subpart D)
(21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of Center for Devices and Radiological Health (CDRH)
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the DSM logo. The logo consists of a colorful abstract design on the left and the letters "DSM" in bold blue on the right. Below the letters, there is a tagline that reads "BRIGHT SCIENCE. BRIGHTER LIVING." in a smaller blue font.
# 510(k) Summary
| K143716 | |
|----------------------------|-----------------------------------------------------------------------------------------------------------------------------------|
| Submitted By: | DSM Biomedical<br>735 Pennsylvania Drive<br>Exton, PA 19341 |
| Contact Person: | Tom Maguire<br>Sr. Director, Global Clinical & Regulatory Affairs<br>E: Tom.Maguire@dsm.com<br>P: 484-713-2146<br>F: 484-713-2903 |
| Date Prepared: | October 26, 2015 |
| Device: | |
| Trade Name: | DPR Cable |
| Common/Usual Name: | Orthopedic Wire |
| Classification Name: | Bone Fixation Cerclage |
| Classification Regulation: | 21 CFR 888.3010 |
| Device Class: | Class II |
| Device Code: | JDQ |
| Advisory Panel: | General and Plastic Surgery |
| Predicate: | Titanium Alloy Songer Cable System/ Atlas Cable System<br>(Medtronic Sofamor Danek , K920201) |
| Reference Devices: | Iso-Elastic Cable (Kinamed Inc., K102834)<br>Orthopedic Polyethylene Cable (Smith&Nephew,<br>K924141) |
# Device Description:
The DPR Cable is a flexible, multi-strand ultra-high molecular weight polyethylene
{4}------------------------------------------------
(UHMWPE) cerclage cable. The device is intended to provide stabilization of bony segments.
The DPR Cable is made of woven ultra-high molecular weight polyethylene fibers which incorporate bismuth trioxide (Bi₂O₃). The addition of bismuth trioxide allows for radiographic visualization both during and after surgical procedures. The device is supplied sterile in double-layer packages. The DPR Cable is intended for single patient use only. DPR Cable is MR safe.
# Intended Use:
The DPR Cable is intended:
- Spinal applications including sublaminar and intraspinous process wiring for . trauma applications.
- . Trochanteric reattachment after trochanteric osteotomy following total hip arthroplasty.
- . Sternotomy indications including the "rewiring" of osteomized sternums.
- Trauma surgery indications including olecranon, ankle, patella and some shoulder . fracture rewiring.
The device is intended for single patient use only.
{5}------------------------------------------------
# Technological Characteristics:
| Characteristic | DSM Biomedical<br>DPR Cable<br>K143716 | Medtronic Sofamor Danek<br>Titanium Alloy Songer Cable<br>System/Atlas™ Cable System<br>K920201 |
|-------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications for<br>Use | The DPR Cable is intended:<br>Spinal applications including<br>sublaminar and intraspinous<br>process wiring for trauma<br>applications. Trochanteric reattachment after<br>trochanteric osteotomy following<br>total hip arthroplasty. Sternotomy indications including<br>the “rewiring” of osteomized<br>sternums. Trauma surgery indications<br>including the olecranon, ankle,<br>patella and some shoulder<br>fracture rewiring. The device is intended for single<br>patient use only. | The Atlas™ Cable System can be utilized<br>anywhere monofilament wires have been<br>previously found to be indicated. The<br>indications are:<br>Spinal applications would include<br>sublaminar and intraspinous<br>process wiring for trauma<br>applications. Another<br>application would be the use of<br>Atlas™ Cable System for<br>instrumentation involved in the<br>correction of scoliotic,kyphotic<br>and lordotic deformities. The<br>stainless steel system may also<br>be used with other stainless steel<br>spinal implants such as the Unit<br>Rod or Luque Rod or wherever<br>"wiring" may help secure the<br>attachment of other implants.<br>The titanium system may also be<br>used with other titanium<br>implants. Trochanteric reattachment after<br>trochanteric osteotomy following |
| Characteristic | DSM Biomedical<br>DPR Cable<br>K143716 | Medtronic Sofamor Danek<br>Titanium Alloy Songer Cable<br>System/Atlas™ Cable System<br>K920201 |
| | | total hip arthroplasty.<br>Sternotomy indications would include the “re-wiring” of osteomizing sternums. Trauma surgery indications would include olecranon, ankle, patella and some shoulder rewiring. |
| Materials of<br>Composition | Dyneema Purity® UHMWPE and Bi2O3 | Titanium Steel and Titanium Alloy |
| Device<br>Characteristics | Flexible, high strength | High strength |
| Radiopaque | Yes | Yes |
| Fixation | Knot | Integral Crimp |
| Dimensions | 4 mm | 1 mm diameter |
| Tensile Strength<br>of fixated loop | 2289+/-35 | 1005+/-49 |
| Fatigue Strength | 1559N | <44.5N |
| Biocompatible | Yes | Yes |
| Reusable | Single Use Only | Single Use Only |
| Sterilization<br>Method | EtO | Irradiation |
| Packaging | Double layer | Not available |
{6}------------------------------------------------
{7}------------------------------------------------
### Biocompatibility and Performance Data:
# Biocompatibility testing
The biocompatibility of the materials used in the manufacturer of the DPR Cable, Dyneema Purity® fibers and bismuth trioxide, are established by their long historical clinical use in medical devices and the combination of these materials does not raise any new questions of safety and effectiveness. Cytotoxicity and leaching was performed on the Dyneema Purity® RP material.
Biomechanical testing included fatigue strength, tensile force, creep and wear-debris and knot strength. Testing results indicate that the device is equivalent to the predicate device.
#### Substantial Equivalence:
The DPR Cable is substantially equivalent in intended use, tensile strength, and fixation strength to the predicate device (Titanium Alloy Songer Cable System/Atlas™ Cable System, Medtronic Sofamor Danek, K920201). The DPR Cable is substantially equivalent to the material characteristics and creep and wear performance of the reference devices (Iso-Elastic Cable, Kinamed Inc., K102834 and Orthopedic Polyethylene Cable, Smith & Nephew, K924141).
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.