The Intended Use of the Biomet Proximal Humerus Plating System is for internal fixation of humeral fractures. The indications for the Proximal Humerus Plating System are as follows: The Biomet Proximal Humerus Plating System is indicated for fixation of fractures and fracture dislocations, fusions, osteotomies and non-unions of the humerus, particularly in osteopenic bone.
Device Story
System consists of two plating families (Low and High) with left/right configurations; plates and screws provide internal fixation for humeral fractures. Plates and screws manufactured from Titanium Alloy (ASTM F136); multidirectional screws from Cobalt Chromium (ASTM F1537). System includes instrumentation for fracture alignment and stabilization. Used by orthopedic surgeons in clinical settings to stabilize bone segments; facilitates healing of fractures, non-unions, and osteotomies. Provides subchondral support in the humeral head. Implants provided sterile (gamma irradiation) or non-sterile (steam sterilization by user).
Clinical Evidence
No clinical data provided. Substantial equivalence based on non-clinical performance testing, including axial load construct testing, engineering evaluations, and galvanic corrosion assessment.
Technological Characteristics
Plates: Type II anodized Titanium Alloy (Ti-6Al-4V ELI, ASTM F136). Screws: Titanium Alloy (ASTM F136) or Cobalt Chromium (ASTM F1537). Instrumentation: Stainless steel (ASTM F899). Design: Anatomical proximal humerus plates with subchondral support. Sterilization: Gamma irradiation (sterile) or steam sterilization (non-sterile).
Indications for Use
Indicated for fixation of fractures, fracture dislocations, fusions, osteotomies, and non-unions of the humerus, particularly in patients with osteopenic bone.
Regulatory Classification
Identification
Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.
Anatomic Locked Plating System Long Plate Line Extension (K101421)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image contains the logo for the U.S. Department of Health & Human Services. The logo features a stylized representation of an eagle or bird with three human profiles incorporated into its design. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the bird symbol.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
March 30, 2015
Biomet Manufacturing Ms. Julie Largent Regulatory Affairs Specialist P.O. Box 587 56 East Bell Drive Warsaw, Indiana 46582
Re: K143697
Trade/Device Name: Biomet Proximal Humerus Plating System Regulation Number: 21 CFR 888.3030 Regulation Name: Single/multiple component metallic bone fixation appliances and accessories Regulatory Class: Class II Product Code: HRS Dated: February 18, 2015 Received: February 19, 2015
Dear Ms. Largent:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing
{1}------------------------------------------------
# Page 2 - Ms. Julie Largent
(21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# Mark N. Melkerson -S
Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known)
K143697 (paqe 1/1)
Device Name
Biomet Proximal Humerus Plating System
Indications for Use (Describe)
The Biomet Proximal Humerus Plating System is indicated for fixation of fracture dislocations, fusions, osteotomies and non-unions of the humerus, particularly in osteopenic bone.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
|-------------------------------------------------|--|
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the word "BIOMET" in a bold, sans-serif font. The letters are all capitalized and appear to be a dark gray or black color. A thin, horizontal line runs underneath the word, emphasizing the text.
# 510(k) Summary
In accordance with 21 CFR §807.92 and the Safe Medical Devices Act of 1990, the following information is provided for the Biomet Proximal Humerus Plating System 510(k) premarket notification. The submission was prepared in accordance with the FDA guidance document, 'Format for Traditional and Abbreviated 510(k)s', issued on August 12, 2005.
| Sponsor: | Biomet Inc.<br>56 East Bell Drive<br>PO Box 587<br>Warsaw, IN 46581<br>Establishment Registration Number: 1825034 |
|-----------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact: | Julie Largent<br>Regulatory Affairs Specialist<br>305-269-6391 |
| Date: | March 12, 2015 |
| Subject Device: | Trade Name: Biomet Proximal Humerus Plating System<br>Common Name: Plate, Fixation, Bone<br><br>Classification Name:<br>• HRS - Single/multiple component metallic bone fixation |
- appliances and accessories (21 CFR 888.3030)
#### Legally marketed devices to which substantial equivalence is claimed:
- ACE Proximal Humerus Plates K955472 ●
- S Shoulder Plating System K060290
- Anatomic Locked Plating System Long Plate Line Extension K101421
#### Device Description
The Biomet Proximal Humeral Plating System offers two plating families (Low and High) with both left and right configurations in various lengths. Both plating families in the Biomet Proximal Humerus Plating System incorporate the same spatial subchondral support as the Biomet S3 proximal humerus plates (K060290) in the head of the plates. The plates and screws will be manufactured from Titanium Alloy (ASTM F136) and are either type II or color anodized. The system will also include multidirectional screws that are manufactured from Cobalt Chromium (ASTM F1537). The plates are anatomically designed to match the anatomy of the proximal humerus. The system will also include instrumentation to aid the user in alignment and stabilization of fractures to the skeletal system.
{4}------------------------------------------------
#### Intended Use and Indications for Use
The Intended Use of the Biomet Proximal Humerus Plating System is for internal fixation of humeral fractures. The indications for the Proximal Humerus Plating System are as follows: The Biomet Proximal Humerus Plating System is indicated for fixation of fractures and fracture dislocations, fusions, osteotomies and non-unions of the humerus, particularly in osteopenic bone.
### Summary of Technological Characteristics
The rationale for substantial equivalence is based on consideration of the following characteristics:
- Intended Use: The Intended Use of the Biomet Proximal Humerus Plating System is the same as to the intended uses cleared in K955472 and K060290 and similar to K101421.
- Indications for Use: The indications for the Biomet Proximal Humerus Plating System is similar to and based on the Indications for Use cleared in K955472, K060290, and K101421.
- Materials: The Biomet Proximal Humerus Plating System plates are manufactured from Type II anodized Titanium Alloy (Ti-6Al-4V ELI). The locking and non-locking screws included in the system are also manufactured from Titanium Alloy (Ti-6Al-4V ELI ASTM F-136) and are either type II anodized or color anodized. The system also contains multi-directional screws that are manufactured from Cobalt Chromium (ASTM F1537). Titanium alloys as well as implant grade cobalt chromium are commonly used materials in orthopedic implants and are used in predicate devices cleared via K955472 and K101421. The patient contact instrumentation is comprised of stainless steel that meets ASTM F-899.
- . Design Features: The design features for the Biomet Proximal Humerus Plating System are similar to currently marketed devices K955472, K060290 and K101421. The design differences have not identified any issues that would impact the safety and effectiveness of the devices.
- . Sterilization: The implants are offered to the user either in the sterile or non-sterile configuration. The non-sterile devices will be required to be steam sterilized by the user prior to use while the sterile implants will be terminally sterilized by gamma irradiation. These sterilization configurations are the same as the predicate devices currently marketed and cleared via K060290 and K101421.
## Summary of Performance Data (Nonclinical and/or Clinical)
- . Non-Clinical Tests
- Non-clinical performance testing included axial load construct testing, o engineering evaluations and galvanic corrosion assessment to determine substantial equivalence of the Biomet Proximal Humerus Plating System. Results indicate that the subject plating system is substantially equivalent to legally marketed devices offering a reasonable assurance of safety and effectiveness.
- Clinical Tests
- o No clinical tests are provided for basis of substantial equivalence.
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image contains the word "BIOMET" in all capital letters. The letters are bold and gray. The font is sans-serif and the letters are evenly spaced.
#### Substantial Equivalence Conclusion
The Biomet Proximal Humerus Plating System has shown to be substantially equivalent to the predicate devices. Results of non-clinical tests/engineering justification and the similarities with legally marketed predicated devices indicate the device will perform within the intended uses and no new issues of safety and effectiveness have been raised.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.