← Product Code [NKB](/productcode/NKB) · K143684

# NuVasive Reline System (K143684)

_Nu Vasive, Incorporated · NKB · Apr 23, 2015 · Orthopedic · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K143684

## Device Facts

- **Applicant:** Nu Vasive, Incorporated
- **Product Code:** [NKB](/productcode/NKB.md)
- **Decision Date:** Apr 23, 2015
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 888.3070
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic, Pediatric

## Indications for Use

When used as a pedicle screw fixation system, the NuVasive Reline System is intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the posterior thoracic, lumbar, and sacral spine: 1. Degenerative disc disease (as defined by back pain of discogenic origin with degeneration of the disc confirmed by patient history and radiographic studies) 2. Degenerative spondylolisthesis with objective evidence of neurologic impairment 3.Fracture 4. Dislocation 5.Scoliosis 6. Kyphosis 7. Spinal tumor and/or 8. Failed previous fusion (pseudoarthrosis) When used for posterior non-cervical screw fixation in pediatric patients. NuVasive Reline System implants are indicated as an adjunct to treat adolescent idiopathic scoliosis. Additionally the NuVasive Reline System is intended to treat pediatric patients diagnosed with the following conditions: spondylolisthesis / spondylolysis, and fracture caused by tumor and/or trauma. Pediatric pedicle screw fixation is limited to a posterior approach and is intended to be used with autograft and/or allograft. The NuVasive Reline System is also indicated for the treatment of severe spondylolisthesis (Grades 3 and 4) of the L5-S1 vertebral joint in skeletally mature patients receiving fusion by autogenous bone graft, having the device fixed or attached to the lumbar and sacral spine (L3 to sacrum), with removal of the implants after attainment of a solid fusion. When used as an anterolateral non-pedicle screw system in the thoracic and lumbar spine, the NuVasive Reline System is also intended for the following indications: 1. Degenerative disc disease (as defined by back pain of discogenic origin with degeneration of the disc confirmed by patient history and radiographic studies) 2. Spinal stenosis 3. Spondylolisthesis 4. Spinal deformities 5.Fracture 6. Pseudoarthosis 7. Tumor resection and/or 8. Failed previous fusion In order to achieve additional levels of fixation, the Reline System rods may be connected to the Armada System.

## Device Story

Pedicle screw system comprising screws, hooks, rods, lock screws, transverse connectors, rod-to-rod connectors, and iliac connectors; used for spinal immobilization and stabilization. Components configured to match patient anatomy/pathology; rods connectable to Armada System for additional fixation. Used in surgical settings by surgeons; provides rigid mechanical support to spinal segments to facilitate fusion. Benefits include stabilization of unstable or deformed spinal segments.

## Clinical Evidence

Bench testing only. Testing included Dynamic Compression Bending (ASTM F1717), Static Axial Slip (ASTM F1798), Static and Dynamic Flexion-Extension moment (ASTM F1798), and Static Tulip Pull-off.

## Technological Characteristics

Pedicle screw spinal system; components include screws, hooks, rods, and connectors. Mechanical fixation device. Testing performed per ASTM F1717 and ASTM F1798 standards.

## Regulatory Identification

(1) Rigid pedicle screw systems are comprised of multiple components, made from a variety of materials that allow the surgeon to build an implant system to fit the patient's anatomical and physiological requirements. Such a spinal implant assembly consists of a combination of screws, longitudinal members (e.g., plates, rods including dual diameter rods, plate/rod combinations), transverse or cross connectors, and interconnection mechanisms (e.g., rod-to-rod connectors, offset connectors).(2) Semi-rigid systems are defined as systems that contain one or more of the following features (including but not limited to): Non-uniform longitudinal elements, or features that allow more motion or flexibility compared to rigid systems.

## Special Controls

*Classification.* (1) Class II (special controls), when intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine: severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra; degenerative spondylolisthesis with objective evidence of neurologic impairment; fracture; dislocation; scoliosis; kyphosis; spinal tumor; and failed previous fusion (pseudarthrosis). These pedicle screw spinal systems must comply with the following special controls:(i) Compliance with material standards;
(ii) Compliance with mechanical testing standards;
(iii) Compliance with biocompatibility standards; and
(iv) Labeling that contains these two statements in addition to other appropriate labeling information:
 “Warning: The safety and effectiveness of pedicle screw spinal systems have been established only for spinal conditions with significant mechanical instability or deformity requiring fusion with instrumentation. These conditions are significant mechanical instability or deformity of the thoracic, lumbar, and sacral spine secondary to severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra, degenerative spondylolisthesis with objective evidence of neurologic impairment, fracture, dislocation, scoliosis, kyphosis, spinal tumor, and failed previous fusion (pseudarthrosis). The safety and effectiveness of these devices for any other conditions are unknown.”
“Precaution: The implantation of pedicle screw spinal systems should be performed only by experienced spinal surgeons with specific training in the use of this pedicle screw spinal system because this is a technically demanding procedure presenting a risk of serious injury to the patient.”
(2) Class II (special controls), when a rigid pedicle screw system is intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion in the treatment of degenerative disc disease and spondylolisthesis other than either severe spondylolisthesis (grades 3 and 4) at L5-S1 or degenerative spondylolisthesis with objective evidence of neurologic impairment. These pedicle screw systems must comply with the following special controls:
(i) The design characteristics of the device, including engineering schematics, must ensure that the geometry and material composition are consistent with the intended use.
(ii) Non-clinical performance testing must demonstrate the mechanical function and durability of the implant.
(iii) Device components must be demonstrated to be biocompatible.
(iv) Validation testing must demonstrate the cleanliness and sterility of, or the ability to clean and sterilize, the device components and device-specific instruments.
(v) Labeling must include the following:
(A) A clear description of the technological features of the device including identification of device materials and the principles of device operation;
(B) Intended use and indications for use, including levels of fixation;
(C) Identification of magnetic resonance (MR) compatibility status;
(D) Cleaning and sterilization instructions for devices and instruments that are provided non-sterile to the end user; and
(E) Detailed instructions of each surgical step, including device removal.
(3) Class II (special controls), when a semi-rigid system is intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion for any indication. In addition to complying with the special controls in paragraphs (b)(2)(i) through (v) of this section, these pedicle screw systems must comply with the following special controls:
(i) Demonstration that clinical performance characteristics of the device support the intended use of the product, including assessment of fusion compared to a clinically acceptable fusion rate.
(ii) Semi-rigid systems marketed prior to the effective date of this reclassification must submit an amendment to their previously cleared premarket notification (510(k)) demonstrating compliance with the special controls in paragraphs (b)(2)(i) through (v) and paragraph (b)(3)(i) of this section.

## Predicate Devices

- Reline System ([K142499](/device/K142499.md))
- GSB Global Balance System ([K132014](/device/K132014.md))
- SpheRx II Pedicle Screw System ([K091502](/device/K091502.md))
- Synthes USS Universal Spine System ([K111358](/device/K111358.md))
- Medtronic CD HORIZON Spinal System ([K141604](/device/K141604.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is an abstract image of an eagle with three human profiles incorporated into its design.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

April 23, 2015

NuVasive, Incorporated Ms. Olga Lewis Regulatory Affairs Specialist 7475 Lusk Boulevard San Diego, California 92121

Re: K143684

Trade/Device Name: NuVasive® Reline™ System Regulation Number: 21 CFR 888.3070 Regulation Name: Pedicle screw spinal system Regulatory Class: Class III Product Code: NKB, OSH, MNH, MNI, KWP, KWQ Dated: March 24, 2015 Received: March 25, 2015

Dear Ms. Lewis:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set

{1}------------------------------------------------

forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

# Lori A. Wiggins -S

for Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

## Indications for Use

510(k) Number (if known) K143684

Device Name NuVasive® Reline™ System

## Indications for Use (Describe)

When used as a pedicle screw fixation system, the Nu Vasive Reline System is intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the following acute and chronic instabilities or deformities of the posterior thoracic, lumbar, and sacral spine:

1. Degenerative disc disease (as defined by back pain of discogenic origin with degeneration of the disc confirmed by patient history and radiographic studies)

- 2. Degenerative spondylolisthesis with objective evidence of neurologic impairment
- 3. Fracture
- 4. Dislocation
- 5. Scoliosis
- 6. Kyphosis
- 7. Spinal tumor and/or

8. Failed previous fusion (pseudoarthrosis)

When used for posterior non-cervical screw fixation in pediatric patients, Nu Vasive Reline System implants are indicated as an adjunct to fusion to treat adolescent idiopathic scoliosis. Additionally the NuVasive Reline System is intended to treat pediatric patients diagnosed with the following conditions: spondylolysis, and fracture caused by tumor and/or trauma. Pediatric pedicle screw fixation is limited to a posterior approach and is intended to be used with autograft and/or allograft.

The Nu Vasive Reline System is also indicated for the treatment of severe spondylolisthesis (Grades 3 and 4) of the L5-S1 vertebral joint in skeletally mature patients receiving fusion by autogenous bone graft, having the device fixed or attached to the lumbar and sacral spine (L3 to sacrum), with removal of the implants after attainment of a solid fusion.

When used as an anterolateral non-pedicle screw system in the thoracic and lumbar spine, the NuVasive Reline System is also intended for the following indications:

1. Degenerative disc disease (as defined by back pain of discogenic origin with degeneration of the disc confirmed by patient history and radiographic studies)

- 2. Spinal stenosis
- 3. Spondylolisthesis
- 4. Spinal deformities
- 5. Fracture
- 6. Pseudoarthosis
- 7. Tumor resection and/or
- 8. Failed previous fusion

In order to achieve additional levels of fixation, the Reline System rods may be connected to the Armada System.

Type of Use (Select one or both, as applicable)

> Prescription Use (Part 21 CFR 801 Subpart D)

_ Over-The-Counter Use (21 CFR 801 Subpart C)

{3}------------------------------------------------

PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.

## FOR FDA USE ONLY

Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)

This section applies only to requirements of the Paperwork Reduction Act of 1995.

## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the Nuvasive logo. The logo consists of a purple and gray abstract shape on the left, followed by the word "NUVASIVE" in gray. Below the word "NUVASIVE" is the phrase "Speed of Innovation" in a smaller font, also in gray. The logo appears to be for a company or organization called Nuvasive.

## 510(k) Summary

In accordance with Title 21 of the Code of Federal Regulations, Part 807, and in particular 21 CFR §807.92, the following summary of information is provided:

#### Submitted by: A.

Olga Lewis Regulatory Affairs Specialist NuVasive, Incorporated 7475 Lusk Blvd. San Diego, California 92121 Telephone: (858) 909-1800

Date Prepared: December 23, 2014

#### B. Device Name

| Trade or Proprietary Name: | NuVasive® Reline™ System                               |
|----------------------------|--------------------------------------------------------|
| Common or Usual Name:      | Pedicle Screw System                                   |
| Classification Name:       | Pedicle Screw Spinal System, Spinal Interlaminal       |
|                            | Fixation Orthosis, Spinal Intervertebral Body Fixation |
|                            | Orthosis                                               |
| Device Class:              | Class III                                              |
| Classification:            | 21 CFR & 888.3050, 888.3060, 888.3070                  |
| Product Code:              | NKB, OSH, MNI, MNH, KWQ,KWP                            |

#### C. Predicate Devices

The subject device is substantially equivalent to the primary predicate device Reline System (K142499) and additional predicate devices. GSB Global Balance System (K132014), SpheRx II Pedicle Screw System (K091502), Synthes USS Universal Spine System (K111358), and Medtronic CD HORIZON Spinal System (K141604).

#### D. Device Description

The NuVasive Reline System is a pedicle screw system that consists of a variety screws, hooks, rods, lock screws, transverse connectors, rod-to-rod connectors, iliac connectors and associated general instruments. Implant components are available in a variety sizes and can be rigidly locked into a variety of different configurations to suit the individual pathology and anatomical conditions of the patient. This 510(k) is to add new components to the Reline System.

#### E. Intended Use

When used as a pedicle screw fixation system, the NuVasive Reline System is intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the posterior thoracic, lumbar, and sacral spine:

- 1. Degenerative disc disease (as defined by back pain of discogenic origin with degeneration of the disc confirmed by patient history and radiographic studies)
- 2. Degenerative spondylolisthesis with objective evidence of neurologic impairment
- 3.Fracture

{5}------------------------------------------------

Image /page/5/Picture/1 description: The image shows the logo for Nuvasive. The logo consists of a purple leaf-like shape on the left, followed by the word "NUVASIVE" in gray. Below the word "NUVASIVE" is the phrase "Speed of Innovation" in a smaller font.

- 4. Dislocation
- 5.Scoliosis
- 6. Kyphosis
- 7. Spinal tumor and/or
- 8. Failed previous fusion (pseudoarthrosis)

When used for posterior non-cervical screw fixation in pediatric patients. NuVasive Reline System implants are indicated as an adjunct to treat adolescent idiopathic scoliosis. Additionally the NuVasive Reline System is intended to treat pediatric patients diagnosed with the following conditions: spondylolisthesis / spondylolysis, and fracture caused by tumor and/or trauma. Pediatric pedicle screw fixation is limited to a posterior approach and is intended to be used with autograft and/or allograft.

The NuVasive Reline System is also indicated for the treatment of severe spondylolisthesis (Grades 3 and 4) of the L5-S1 vertebral joint in skeletally mature patients receiving fusion by autogenous bone graft, having the device fixed or attached to the lumbar and sacral spine (L3 to sacrum), with removal of the implants after attainment of a solid fusion.

When used as an anterolateral non-pedicle screw system in the thoracic and lumbar spine, the NuVasive Reline System is also intended for the following indications:

- 1. Degenerative disc disease (as defined by back pain of discogenic origin with degeneration of the disc confirmed by patient history and radiographic studies)
- 2. Spinal stenosis
- 3. Spondylolisthesis
- 4. Spinal deformities
- 5.Fracture
- 6. Pseudoarthosis
- 7. Tumor resection and/or
- 8. Failed previous fusion

In order to achieve additional levels of fixation, the Reline System rods may be connected to the Armada System.

#### F. Technological Characteristics

As was established in this submission, the subject NuVasive Reline System is substantially equivalent to other predicate devices cleared by the FDA for commercial distribution in the United States. The subject device was shown to be substantially equivalent and have equivalent technological characteristics to its predicate device through comparison in areas including design, labeling/intended use, material composition, and function.

#### G. Performance Data

Nonclinical testing was performed to demonstrate that the subject NuVasive Reline System is substantially equivalent to the predicate device. The following testing was performed:

- o Dynamic Compression Bending per ASTM F1717
- Static Axial Slip per ASTM F1798 ●
- Static and Dynamic Flexion -Extension moment per ASTM F1798 ●
- Static Tulip Pull-off ●

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image shows the Nuvasive logo. The logo consists of a purple and gray abstract shape on the left, followed by the word "NUVASIVE" in gray. Below the word "NUVASIVE" is the phrase "Speed of Innovation" in a smaller, blurred font. The logo is simple and modern, with a focus on the company name and tagline.

The results demonstrate that the subject NuVasive Reline System is substantially equivalent to the predicate.

#### Conclusions H.

The subject NuVasive Reline System has been shown to be substantially equivalent to legally marketed predicate devices for its intended use.

---

**Source:** [https://fda-staging.innolitics.com/device/K143684](https://fda-staging.innolitics.com/device/K143684)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
