← Product Code [OLO](/productcode/OLO) · K143668

# Navio (K143668)

_Blue Belt Technologies, Inc. · OLO · Mar 20, 2015 · Neurology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K143668

## Device Facts

- **Applicant:** Blue Belt Technologies, Inc.
- **Product Code:** [OLO](/productcode/OLO.md)
- **Decision Date:** Mar 20, 2015
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 882.4560
- **Device Class:** Class 2
- **Review Panel:** Neurology
- **Attributes:** Therapeutic

## Indications for Use

The Navio system is intended to assist the surgeon in providing software-defined spatial boundaries for orientation and reference information to anatomical structures during orthopedic procedures. The Navio system is indicated for use in surgical knee procedures, in which the use of stereotactic surgery may be appropriate, and where reference to rigid anatomical bony structures can be determined. These procedures include unicondylar knee replacement and patellofemoral arthroplasty. The Navio system is indicated for use with cemented implants only.

## Device Story

Navio is a computer-assisted orthopedic surgical navigation and burring system; used by orthopedic surgeons in surgical suites. System utilizes passive infrared tracking camera (Northern Digital Polaris) and reflective trackers to create intraoperative, image-free bone surface models. Surgeon defines surgical plan based on bone landmarks and implant configuration. During bone preparation, system provides software-defined spatial boundaries; controls surgical bur (eMax 2 Plus System) via two modes: bur retraction into guard or speed control as target surface is approached. System provides real-time visual (color-coded GUI) and audio feedback to surgeon. Surgeon maintains full control via footswitch; can override or move handpiece at any time. Device logs patient/procedure data for archival. Benefits include precise bone preparation for cemented knee implants, reducing risk of over-cutting.

## Clinical Evidence

No human clinical tests conducted. Evidence consists of bench verification testing, software database reviews, and clinical simulation (usability testing) using sawbones and cadavers. Testing confirmed surgeons, PAs, and technical staff could successfully place inlay patellofemoral implants per specifications. Accuracy metrics for onlay/inlay patellofemoral component placement (RMS error) were provided as validation of the existing algorithm's performance.

## Technological Characteristics

System includes IR tracking camera, reflective trackers, computer, GUI, probes, surgical bur, bur guards, bone screws, and clamps. Uses intraoperative image-free data collection. Bur control via mechanical retraction or speed modulation. Connectivity: data logging to CD. Software-defined spatial boundaries. No specific ASTM materials cited.

## Regulatory Identification

A stereotaxic instrument is a device consisting of a rigid frame with a calibrated guide mechanism for precisely positioning probes or other devices within a patient's brain, spinal cord, or other part of the nervous system.

## Predicate Devices

- Blue Belt Technologies Navio System ([K140596](/device/K140596.md))

## Reference Devices

- eMax 2 Plus System ([K080802](/device/K080802.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is an abstract image of three faces in profile, stacked on top of each other, resembling a stylized human figure.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

March 20, 2015

Blue Belt Technologies, Incorporated Mr. Richard G. Confer Vice President of Regulatory Affairs 2828 Liberty Avenue, Suite 100 Pittsburgh, Pennsylvania 15222

Re: K143668

Trade/Device Name: Navio™M Regulation Number: 21 CFR 882.4560 Regulation Name: Stereotaxic instrument Regulatory Class: Class II Product Code: OLO, HSX, HRY, KRR, NPJ Dated: December 22, 2014 Received: December 24, 2014

Dear Mr. Confer:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set

{1}------------------------------------------------

Page 2 - Mr. Richard G. Confer

forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely vours.

# Lori A. Wiggins -S

for Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

### Indications for Use

510(k) Number (if known)

K143668

Device Name Navio

#### Indications for Use (Describe)

The Navio system is intended to assist the surgeon in providing software-defined spatial boundaries for orientation and reference information to anatomical structures during orthopedic procedures.

The Navio system is indicated for use in surgical knee procedures, in which the use of stereotactic surgery may be appropriate, and where reference to rigid anatomical bony structures can be determined. These procedures include unicondylar knee replacement and patellofemoral arthroplasty.

The Navio system is indicated for use with cemented implants only.

Type of Use (Select one or both, as applicable)

X Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.

### FOR FDA USE ONLY

Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)

{3}------------------------------------------------

# Section 6: 510(k) Summary

# 510(k) Summary

| 510(k) Owner             | Blue Belt Technologies<br>2905 Northwest Blvd Ste. 40<br>Plymouth, MN 55441<br>UNITED STATES<br>Tel: (763) 452-4905<br>Fax: (763) 452-4675          |
|--------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Person           | Richard G. Confer<br>Vice President of Regulatory Affairs and Quality<br>Assurance<br>Tel: (412) 683-3844 x 4106<br>Email: rconfer@bluebelttech.com |
| Date of Submission       | December 22, 2014                                                                                                                                   |
| Classification Reference | 21 CFR 882.4560                                                                                                                                     |
| Product Code             | OLO                                                                                                                                                 |
| Supported Codes          | HSX, HRY, KRR, NPJ                                                                                                                                  |
| Common/Usual Name        | Orthopedic Sterotaxic Instrument                                                                                                                    |
| Trade/Proprietary Name   | NavioTM                                                                                                                                             |
| Predicate Device(s)      | Blue Belt Technologies Navio System (K140596)                                                                                                       |
| Reason for Submission    | Expanded Indications to add NPJ as a supported<br>product code                                                                                      |

{4}------------------------------------------------

## Intended Use

The Navio system is intended to assist the surgeon in providing software-defined spatial boundaries for orientation and reference information to anatomical structures during orthopedic procedures.

The Navio system is indicated for use in surgical knee procedures, in which the use of stereotactic surgery may be appropriate, and where reference to rigid anatomical bony structures can be determined. These procedures include unicondylar knee replacement and patellofemoral arthroplasty.

The Navio system is indicated for use with cemented implants only.

This intended use statement is the same as the predicate, Navio System per K140596, and adds product code NPJ per regulation 888.3560 as a supported product code.

# Device Description

The Navio system is a computer-assisted orthopedic surgical navigation and surgical burring system. The system uses established technologies of navigation via a passive infrared tracking camera to aid the surgeon in establishing a bone surface model for the target surgery and to plan the surgical implant location based on predefined bone landmarks and known configuration of the surgical implant. The Navio system then aids the surgeon in executing the surgical plan by using a standard off-the-shelf surgical drill motor and bur (eMax 2 Plus System -K080802), which has been adapted using a tracking system. The surgical bur is located in a handpiece, which allows the bur to move within the handpiece. In the Navio system the software controls the position of the tip of the surgical bur relative to the end of a guard attached to the handpiece and prohibits the bur from cutting bone as it approaches the planned target surface. As the planned surface is reached, the tip of the bur is fully retracted within the guard.

An alternate mode of operation is the speed control mode. In this mode the speed of the bur is controlled and the bur stops as the planned surface is reached. In this mode of operation the bur does not retract into the guard. This mode of operation is useful in shaping surfaces of the condyle as well as placing post holes.

The Navio computer system maintains a log of the patient data and procedure data. Each entry is date and time stamped. Data log entries include date and time stamp for data line entry, patient and procedure ID, implant ID, step in process, and error messages. This data can be archived to a CD upon demand at the end of the procedure.

The following diagram shows the primary workflow steps in each application, UKR and PFA. Though the two procedures are very similar, they are mutually independent and cannot be planned or completed in parallel.

Note: The workflow has not changed from the predicate, Navio per K140596.

{5}------------------------------------------------

#### Figure 6-1

Image /page/5/Figure/2 description: The image shows two flowcharts outlining the steps for two different operative procedures: Navio-UKR for Unicondylar Knee Replacement and Navio-PFA for Patellofemoral Arthroplasty. Both procedures start with hardware and handpiece connections, followed by bur/control selection, handpiece retraction, calibration, and homing validation. The UKR procedure includes steps like malleoli point collection, femur neutral position, tibia landmarks collection, gap planning, and evaluate knee ROM. The PFA procedure includes hip center calculation and skips the tibia-related steps and gap planning.

{6}------------------------------------------------

| Devices                                       | Blue Belt Technologies<br>Navio                                                                                                                                                                                                                                                                                                                                       | Predicate A<br>Blue Belt Technologies<br>Navio (K14059)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               |
|-----------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications for use                           | The same as Predicate                                                                                                                                                                                                                                                                                                                                                 | The Navio system is intended to assist<br>the surgeon in providing software-<br>defined spatial boundaries for<br>orientation and reference information<br>to anatomical structures during<br>orthopedic procedures.<br><br>The Navio System is indicated for use in<br>surgical knee procedures in which the<br>use of stereotactic surgery may be<br>appropriate, and where reference to<br>rigid anatomical bony structures can be<br>determined.<br><br>These procedures include unicondylar<br>knee replacement and patellofemoral<br>arthroplasty.<br><br>The Navio system is indicated for use<br>with cemented implants only. |
| Supported Product<br>Code                     | Supported Product<br>Codes:<br>HSX, HRY, KRR, NPJ                                                                                                                                                                                                                                                                                                                     | Supported Product Codes:<br>HSX, HRY, KRR                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             |
| Environment of Use                            | Same as Predicate                                                                                                                                                                                                                                                                                                                                                     | Used by trained orthopedic surgeons in<br>an Orthopedic surgical suite.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               |
| Technological<br>Characteristics              | Same as Predicate                                                                                                                                                                                                                                                                                                                                                     | The Navio system uses established<br>technologies to prepare bone for<br>attachment of implant components.<br>Navio uses intraoperative data<br>collection (image- free or non-CT data<br>generation) to create a model of the<br>patient's femur and/or tibia,<br>dependent on the procedure being<br>performed, and allows the surgeon to<br>prepare a surgical plan.<br><br>The Navio system uses predefined<br>boundaries generated during the<br>planning process to control the motion<br>of the surgical bur and limit the amount<br>of bone removed in order to shape the                                                     |
| Devices                                       | Blue Belt Technologies<br>Navio                                                                                                                                                                                                                                                                                                                                       | Predicate A<br>Blue Belt Technologies<br>Navio (K14059)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               |
| Construction                                  | Same as Predicate                                                                                                                                                                                                                                                                                                                                                     | patellofemoral joint in preparation for<br>placement of the surgical implant.<br><br>Bur cutting is controlled either by<br>retracting the bur in a guard, or by<br>controlling the speed of the bur as the<br>target surface is approached.                                                                                                                                                                                                                                                                                                                                                                                          |
| Pre-Surgical Planning<br>Method               | Same as Predicate                                                                                                                                                                                                                                                                                                                                                     | Consists of an IR image system<br>(Northern Digital Polaris), reflective<br>trackers, computer, user interface<br>display, various probes, a surgical bur,<br>bur guards, bone screws, and clamps.                                                                                                                                                                                                                                                                                                                                                                                                                                    |
| Imaging<br>Requirements                       | Same as Predicate                                                                                                                                                                                                                                                                                                                                                     | Uses data collected intra-operatively by<br>surgeon during the initial surgical<br>procedure to generate a real-time plan<br>of cut surfaces.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |
| Tool manipulation                             | Same as Predicate                                                                                                                                                                                                                                                                                                                                                     | None preoperative.<br>Possible post-operative to verify<br>implant placement after surgeon<br>finalizes placement.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |
| Feedback that<br>boundary has been<br>reached | Same as Predicate                                                                                                                                                                                                                                                                                                                                                     | Controlled by retracting surgical bur<br>into a guard as a function of proximity<br>of bur to planned cut surface. Also,<br>speed of bur changes as a function of<br>proximity to planned cut surface.<br><br>The surgeon has control of the tool at<br>all times. Position of the bur is based<br>on navigated guidance and planning of<br>desired cut surface.                                                                                                                                                                                                                                                                      |
| Visual indicator                              | Same as Predicate                                                                                                                                                                                                                                                                                                                                                     | Visual (color-coded surface on GUI) and<br>audio indicators as feedback that the<br>boundary has been reached.<br>Withdrawal of bur or reduced bur<br>speed.                                                                                                                                                                                                                                                                                                                                                                                                                                                                          |
|                                               | Same as Predicate                                                                                                                                                                                                                                                                                                                                                     | Color changes as surgical bur<br>approaches planned cut boundary.<br>Magenta (purple) is the starting bone<br>surface (>3 mm from target surface).<br>Intermediate colors (blue and green)<br>indicate bone to be removed. Blue<br>indicates 2 mm to target surface.<br>Green indicates 1 mm to target surface.<br>Colors are correlated to the distance.                                                                                                                                                                                                                                                                             |
| Devices                                       | Blue Belt Technologies<br>Navio                                                                                                                                                                                                                                                                                                                                       | Predicate A<br>Blue Belt Technologies<br>Navio (K14059)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               |
|                                               |                                                                                                                                                                                                                                                                                                                                                                       | from planned cut surface in 1mm<br>increments. Red indicates over cut<br>exceeding 0.5 mm.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            |
| User Override                                 | Same as Predicate                                                                                                                                                                                                                                                                                                                                                     | Surgeon has full control of handpiece<br>at all times. Cutting bur operation<br>speed is enabled by surgeon operated<br>footswitch. The computer program<br>only modifies the speed up to the limit<br>set by the surgeon pressing the<br>footswitch.<br>The surgeon always has the ability to<br>move the handpiece away from the cut<br>surface or disable the footswitch.                                                                                                                                                                                                                                                          |
| Accuracy of<br>placement of<br>implant        | Same placement<br>accuracies as Predicate<br>for patellofemoral knee<br>replacements<br>(algorithms have not<br>changed).<br>Onlay Patellofemoral<br>component RMS<br>placement error along<br>any single axis averaged<br>0.884 mm and 1.013°.<br>Inlay Patellofemoral<br>component RMS<br>placement error along<br>any single axis averaged<br>0.462 mm and 1.028°. | Onlay Patellofemoral component RMS<br>placement error along any single axis<br>averaged 0.884 mm and 1.013°.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          |

Table 6-1: Summary of Technological Similarities with Predicates

{7}------------------------------------------------

{8}------------------------------------------------

### Nonclinical Testing

Design verification tests were performed on the Blue Belt Technologies Navio system to support inlay patellofemoral implant placement as a result of the risk analysis and product requirements. Testing included software data base reviews, bench verification testing, user manual/labeling inspection, drawing inspections, and a clinical simulation (usability testing). Simulated-use testing included testing in simulated knees (sawbones) and cadaver lab testing. Users included surgeons, physician's assistants, and technical support personnel who were able to successfully use the Navio system and place inlay patellofemoral implants per Blue Belt Technologies' specifications and implant manufacturer's specifications after being adequately trained.

{9}------------------------------------------------

## Discussion of Similarities and Differences

The Navio system uses established technologies to prepare bone for attachment of implant components. Navio uses intraoperative data collection (image-free or non-CT data generation) to create a model of the patient's femur and tibia and allow the surgeon to prepare a surgical plan. The Navio uses predefined boundaries generated during the planning process to control the motion of the surgical bur and limit the amount of bone removed in order to shape the condyles, tibial plateau, or patellofemoral joint in preparation for placement of the surgical implant. This is equivalent to the methodology used by the NavioPFS™ system except for the Navio's additional capability to prepare the patellofemoral joint for implant.

The Navio uses predefined boundaries generated during the planning process to control the motion of the surgical bur and limit the amount of bone removed in order to shape the condyles, tibial plateau, and/or the patellofemoral joint in preparation for placement of the surgical implant.

This submittal supports the expanded indication for use of the Navio system to place inlay patellofemoral implants using the same techniques used in the predicate Navio system.

Though the UKR and PFA procedures are very similar, they are mutually independent and cannot be planned or completed in parallel. If the user is completing a bicompartmental knee joint replacement in which a patellofemoral arthroplasty and a unicondylar knee replacement are both being performed, preparation of the patellofemoral joint must be completed independently of the preparations of the femoral condyle and tibial plateau surfaces.

### Clinical Testing

No human clinical tests were conducted to determine safety and effectiveness of the Navio system.

# Summary and Conclusions

The Navio system described in this submittal has the same intended use and the same technological characteristics as the Navio system (K140596). Non-clinical testing was completed to verify that the use of the Navio system to assist with the placement of inlay patellofemoral implants does not raise any new issues of safety or effectiveness. The information presented in this 510(k) notification demonstrates that the Navio, when used to place an inlay patellofemoral implant, is as safe and effective as Navio™ (K140596).

---

**Source:** [https://fda-staging.innolitics.com/device/K143668](https://fda-staging.innolitics.com/device/K143668)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).