K143666 · Finemec Co., Ltd. · GEX · Apr 23, 2015 · General, Plastic Surgery
Device Facts
Record ID
K143666
Device Name
Noblex Long Pulse Alexandrite Laser
Applicant
Finemec Co., Ltd.
Product Code
GEX · General, Plastic Surgery
Decision Date
Apr 23, 2015
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4810
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Noblex Long Pulse Alexandrite Laser is indicated for stable long-term or permanent hair reduction. Permanent hair reduction is defined as the long-term, stable reduction in the number of hairs regrowing when measured at 6, 9 and 12 months after the completion of a treatment regime. It is used for all skin types (Fitzpatrick I - VI) including tanned skin. It is also indicated for the treatment of benign vascular lesions, benign pigmented lesions and wrinkles.
Device Story
Noblex Long Pulse Alexandrite Laser emits coherent 755nm near-infrared light; energy delivered via fiber-optic coupled hand-piece to target chromophores in human tissue. Rapid localized heating causes thermal destruction of targets for hair removal, vascular/pigmented lesion treatment, and wrinkle reduction. Operated by physicians in clinical settings; system controlled by microprocessor interfaced with LCD touch screen. Physician adjusts laser pulse energy and spot size to optimize treatment based on application. Device consists of main body, hand-piece, and protective goggles. Benefits include non-invasive, targeted thermal therapy for dermatological conditions.
Clinical Evidence
Bench testing only. Device conforms to EN 60825-1:2007 (laser safety), EN 60601-2-22:1996 (therapeutic laser safety), EN 60601-1:2006 (general safety), and EN 60601-1-2:2007 (electromagnetic compatibility).
Technological Characteristics
Near-infrared 755nm pulsed laser; fiber-optic delivery; microprocessor-controlled with LCD touch screen interface. Conforms to EN 60825-1, EN 60601-2-22, EN 60601-1, and EN 60601-1-2 standards. System includes main body and hand-piece.
Indications for Use
Indicated for stable long-term or permanent hair reduction in all skin types (Fitzpatrick I-VI), including tanned skin. Also indicated for treatment of benign vascular lesions, benign pigmented lesions, and wrinkles.
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized caduceus symbol, which is often associated with medicine and healthcare. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the symbol.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
April 23, 2015
FineMEC Company, Ltd. % Shin Kuk Yoo LSK Biopartners Incorporated 8 East Broadway, Suite 611 Salt Lake City, Utah 84111
Re: K143666
Trade/Device Name: Noblex Long Pulse Alexandrite Laser Regulation Number: 21 CFR 878.4810 Regulation Name: Laser surgical instrument for use in general and plastic surgery and in dermatology Regulatory Class: Class II Product Code: GEX Dated: March 12, 2015 Received: March 20, 2015
Dear Shin Kuk Yoo:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations. Title 21. Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical
{1}------------------------------------------------
device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# Jennifer R. Stevenson -S
For Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K143666
Device Name Noblex Long Pulse Alexandrite Laser
## Indications for Use (Describe)
The Noblex Long Pulse Alexandrite Laser is indicated for stable long-term or permanent hair reduction. Permanent hair reduction is defined as the long-term, stable reduction in the number of hairs regrowing when measured at 6, 9 and 12 months after the completion of a treatment regime. It is used for all skin types (Fitzpatrick I - VI) including tanned skin. It is also indicated for the treatment of benign vascular lesions, benign pigmented lesions and wrinkles.
Type of Use (Select one or both, as applicable)
| Prescription Use (Part 21 CFR 801 Subpart D) | <span style="font-size: 20px;">☑</span> |
|----------------------------------------------|-----------------------------------------|
| Over-The-Counter Use (21 CFR 801 Subpart C) | ☐ |
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and comblete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
### 510(k) Summary
Prepared date: Mar. 5. 2015
New Device: Noblex Long Pulse Alexandrite Laser
- 1. Submitter and US Official Correspondent Submitter: FineMEC Co., Ltd. Address: #230, 7, Galhyeon-ro 11-gil, Eunpyeong-gu Seoul, 122-902, Korea Tel: +82-2-309-1882 Fax: +82-2-376-1884 Contact: Mr. Ghun Hee Lee (General Manager) E-mail: ghlee@finemec.co.kr
| US Official Correspondent: | Shin Kuk Yoo, Consultant<br>Telephone No.: 714-313-7442<br>Fax No.: 801-303-7455<br>Email: skyone@LSKBioPartners.com |
|----------------------------|----------------------------------------------------------------------------------------------------------------------|
|----------------------------|----------------------------------------------------------------------------------------------------------------------|
#### 2. Device Information
| Proprietary/Trade Name : | Noblex Long Pulse Alexandrite Laser |
|--------------------------|--------------------------------------|
| Common/Usual Name: | Medical Laser System |
| Classification Name: | Instrument, Surgical, Powered, Laser |
| Device Class: | 21 CFR 878.4810 GEX Class II |
#### 3. Description of Device
The Nobelx Long Pulsed Laser system produces a pulsed beam of coherent near infrared (755nm) light. This beam is directed to the treatment zone by fiber optic coupled to hand-piece. When the beam contacts human tissue, the energy in the beam is absorbed, resulting in a very rapid highly localized temperature increase to the target chromospheres such as Hair removal and Pigment treatment. This increases localized temperature of the chromospheres to smaller particles.
By directing the beam onto specific tissue locations, using different hand-pieces, and controlling the treatment fluence, the intensity of the temperature of the target can be varied. The physician can optimize the effect for different applications by controlling the energy of the laser pulse and the spot size of the treatment beam. This equipment is composed of the main body and a hand-piece which is an irradiation device and as an accessory part, protective goggles for protection of the worker. Operation principle: This equipment is controlled by a micro processor interfaced to a LCD touch screen control panel. The computer controls start and stop of the treatment. When the key switch of the system is turned clockwise, the main power will be inputted, which will be conveyed to the hand piece through the control board.
Meanwhile, the control board connected to the touch screen is connected to the lamp of the hand-piece and controls the same, and controls the whole system through the data connected to the touch screen control panel. When the switch of the hand piece is pressed, the lamp will laser.
{4}------------------------------------------------
#### 4. Indications for Use
The Noblex Long Pulse Alexandrite Laser is indicated for stable long-term or permanent hair reduction. Permanent hair reduction is defined as the long-term, stable reduction in the number of hairs regrowing when measured at 6, 9 and 12 months after the completion of a treatment regime. It is used for all skin types (Fitzpatrick I -VI) including tanned skin. It is also indicated for the treatment of benign vascular lesions, benign pigmented lesions and wrinkles.
#### 5. Performance Testing
The Noblex Long Pulse Alexandrite Laser is required to conform and does conform to the laser performance standards as follows;
- 1. EN 60825-1:2007, Safety of laser products-Part 1: Equipment classification and requirements
- 2. EN 60601-2-22: 1996 Medical electrical equipment- part 2: particular requirements for the safety of diagnostic and therapeutic laser equipment.
- 3. EN 60601-1: 2006 Medical electrical equipment Part 1 : General requirements for basic safety and essential performance
- 4. EN 60601-1-2: 2007 Medical electrical equipment Part 1-2 : General requirements for basic safety and essential performance- Collateral standard : Electromagnetic compatibility- requirements and tests
#### 6. Substantial Equivalence
Substantial Equivalence for the Noblex Long Pulse Alexandrite Laser is based in its similarities in indication for use, design features, operational principles and material composition when compared to the predicate devices cleared under the following submissions:
- K034030 - Cynosure Apogee Elite Laser, Manufactured by Cynosure, Inc.
- K024335 - GentleLASE Family of Laser Systems, Manufactured by Candela Corporation
#### 7. Conclusions
Based on the technical comparisons between the Noblex Long Pulse Alexandrite Laser and the predicate lasers above, these devices are Substantially Equivalent.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.