Dental Blanks made from StarCeram® Z-Nature are indicated for crowns, multi-unit bridges, inlay bridges and all-ceramic restoration. Applications include both anterior and posterior bridges. Dental Blanks made from StarCeram® Z-Smile are indicated for crowns and bridges, multi-unit bridges, inlay bridges and all-ceramic restoration as max three-unit bridge at the anterior and posterior area.
Device Story
Semi-finished yttrium-stabilized pre-sintered zirconium dioxide dental blanks; used for milled production of crowns, bridges, and all-ceramic restorations. Operated by dental professionals using commercial CAD/CAM systems or hand-operated copy-milling machines. Device serves as raw material for custom dental prosthetics; final shape determined by user based on patient-specific needs. Benefits include provision of durable, biocompatible ceramic material for restorative dentistry.
Clinical Evidence
No clinical testing was performed. Biocompatibility and cytotoxicity testing conducted per ISO 10993-1 and ISO 10993-5.
Technological Characteristics
Yttrium-stabilized pre-sintered zirconium dioxide dental blanks. Form factor: semi-finished blocks for milling. Complies with DIN EN ISO 6872. Biocompatibility per ISO 10993-1 and ISO 10993-5.
Indications for Use
Indicated for patients requiring crowns, multi-unit bridges, inlay bridges, or all-ceramic restorations. Z-Nature is for anterior and posterior bridges; Z-Smile is for anterior and posterior restorations up to max three-unit bridges.
Regulatory Classification
Identification
Porcelain powder for clinical use is a device consisting of a mixture of kaolin, felspar, quartz, or other substances intended for use in the production of artificial teeth in fixed or removable dentures, of jacket crowns, facings, and veneers. The device is used in prosthetic dentistry by heating the powder mixture to a high temperature in an oven to produce a hard prosthesis with a glass-like finish.
{0}------------------------------------------------
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
July 9, 2015
H.C. Starck Ceramics GmbH c/o Ms. Maureen O'Connell O'Connell Regulatory Consultants, Inc. 5 Timber Lane North Reading, MA 01864
Re: K143656
Trade/Device Name: StarCeram® Z-Nature, StarCeram® Z-Smile Regulation Number: 21 CFR 872.6660 Regulation Name: Porcelain powder for clinical use Regulatory Class: II Product Code: EIH Dated: June 10, 2015 Received: June 11, 2015
Dear Ms. O'Connell:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
#### Page 2 - Ms. Maureen O'Connell
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Tina
Kiang-S
for Erin I. Keith, M.S. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K143656
Device Name StarCeram® Z-Nature StarCeram® Z-Smile
#### Indications for Use (Describe)
Dental Blanks made from StarCeram® Z-Nature are indicated for crowns, multi-unit bridges and allceramic restoration. Applications include both anterior and posterior bridges.
Dental Blanks made from StarCeram® Z-Smile are indicated for crowns and bridges, inlay bridges and all-ceramic restoration as max three-unit bridge at the anterior and posterior area.
| Type of Use (Select one or both, as applicable) | |
|---------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------|
| <div style="display:flex; align-items:center;"><input checked="" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D)</div> | <div style="display:flex; align-items:center;"><input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C)</div> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
#### 510(k) SUMMARY K143656
## H.C. Starck Ceramics GmbH StarCeram®
## Submitter's Name. Address. Telephone Number, Contact Person and Date Prepared
Maureen O'Connell O'Connell Regulatory Consultants, Inc. 5 Timber Lane North Reading, MA 01864 (978) 207-1245 Phone: Facsimile: (978) 824-2541
Date Prepared: July 6, 2015
Name of Device and Name/Address of 510(k) Owner
StarCeram® Z-Nature StarCeram® Z-Smile
H.C. Starck Ceramics GmbH Lorenz-Hutschenreuther-Str. 81 95100 Selb, Germany
Common or Usual Name
Powder, Porcelain
Classification Name 21 C.F.R. 872.6660 Porcelain powder for clinical use
Predicate Devices Primary Predicate StarCeram® products cleared in K133213
Reference Predicates StarCeram® Z-Med TransColour Red (K140924) DD cubeX2 (K150196)
{4}------------------------------------------------
# Intended Use / Indications for Use
Dental Blanks made from StarCeram® Z-Nature are indicated for crowns, multi-unit bridges, inlay bridges and all-ceramic restoration. Applications include both anterior and posterior bridges.
Dental Blanks made from StarCeram® Z-Smile are indicated for crowns and bridges, multi-unit bridges, inlay bridges and all-ceramic restoration as max three-unit bridge at the anterior and posterior area.
### Device Description
Dental blanks made from StarCeram® products are semi finished products made of yttrium stabilized pre-sintered zirconium dioxide for milled production of crowns and bridge framework on commercial CAD/CAM systems or hand-operated copy-milling. StarCeram® Z-Smile and StarCeram® Z-Nature are a modification to the StarCeram® products that have already been cleared by the Food and Drug Administration in K133213 and K140924. StarCeram® Z-Smile and StarCeram® Z-Nature have the same intended use and fundamental scientific technology as the StarCeram® products previously cleared by FDA. The only change between the cleared products and the new products is the addition of new color qualities.
# Non-clinical Testing
Biocompatibility and cytotoxicity testing was performed which showed that all versions of the product comply with ISO 10993-1 and ISO 10993-5. Biocompatibility testing was performed under Design Controls to show that the modified version of the product continued to comply with the recognized consensus standards.
# Clinical Testing
No clinical testing was performed.
| Product<br>Name | StarCeram®<br>Z-Nature and<br>Z-Smile | StarCeram® Z-<br>Med,<br>StarCeram® Z-<br>Al-Med HD,<br>StarCeram® Z-<br>Al-Med-HD<br>Colour,<br>StarCeram® Z-<br>Al-Med HD<br>Translucent,<br>StarCeram® Z-<br>Med<br>TransColour | StarCeram® Z-<br>Med<br>TransColour<br>Red | DD cubeX2 and<br>accessories |
|-----------------|---------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------|------------------------------|
|-----------------|---------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------|------------------------------|
# Substantial Equivalence
{5}------------------------------------------------
| 510(k) Number | K143656 | K133213 | K140924 | K150196 |
|---------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Product Code | EIH | EIH | EIH | EIH |
| Product Description | Semi finished products made of yttrium stabilized pre-sintered zirconium dioxide for milled production of crowns and bridge framework on commercial CAD/CAM systems or hand-operated copy-milling machines. | Semi finished products made of yttrium stabilized pre-sintered zirconium dioxide for milled production of crowns and bridge framework on commercial CAD/CAM systems or hand-operated copy-milling machines. | Semi finished products made of yttrium stabilized pre-sintered zirconium dioxide for milled production of crowns and bridge framework on commercial CAD/CAM systems or hand-operated copy-milling machines. | <i>DD cubeX2</i> is a semi-finished dental blank made of yttrium stabilized pre-sintered zirconium dioxide, which has a super high translucency. The ceramics is of type II (not powder), Class 5 according to DIN EN ISO 6872 (FDA Recognition Number 4-178). The <i>DD cubeX2</i> dental blanks are designed for milled production of crowns and bridge frameworks on commercial CAD/CAM systems or hand-operated copy milling machines. |
| Material | Yttrium stabilized zirconium dioxide | Yttrium stabilized zirconium dioxide | Yttrium stabilized zirconium dioxide | Yttrium stabilized zirconium dioxide |
| Availability | Various colors, translucencies and thicknesses | Various colors, translucencies and thicknesses | Various colors, translucencies and thicknesses | Various colors, translucencies and thicknesses |
| Flexural Strength | Z-Smile >500<br>Z-Nature 1,000 +/- 200 | 1,000 +/- 200 | 1,000 +/- 200 | >720 |
H.C. Starck's StarCeram® Z-Smile and StarCeram® Z-Nature are modifications to the StarCeram® products cleared in K133213 and K140924. StarCeram® Z-Nature has the same intended use and indications for use,
{6}------------------------------------------------
principles of operation, and similar technological characteristics as the previously cleared predicate devices. StarCeram® Dental Blanks are indicated for crowns, multi-unit bridges, inlay bridges and all-ceramic restoration. Applications include anterior and posterior bridges. This is the exact indications for use statement cleared for StarCeram® in K133213 and K140924. Thus, StarCeram® Z-Nature has the same intended use and may be substantially equivalent.
StarCeram® Z-Smile is also substantially equivalent to the DD cubeX2 cleared in K150196. StarCeram® Z-Smile has the same intended use also allowing a maximum of 3 bridges to be constructed using the product. The flexural strength of StarCeram Z-Smile is slightly lower than the predicate, but exactly as required by DIN EN ISO 6872 for Class 5 dental ceramics (>500MPa). Therefore, STARCERAM Z-Smile is substantially equivalent to the DD cubeX2 cleared in K150196.
StarCeram® Z-Smile and StarCeram® Z-Nature have the same technological characteristics as the predicate devices. All of the devices are yttrium stabilized pre-sintered zirconium dioxide to be used in dental restorations. The StarCeram® products cleared in K133213 and K140924, and StarCeram® Z-Smile and StarCeram® Z-Nature are all dental blanks which are fabricated to the desired shape by the user based on the specific needs of the patient.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.