The EZDilate 3-Stage Balloon Dilatation Catheters are indicated for use in adult populations to endoscopically dilate strictures of the alimentary tract. It is also indicated in adults for endoscopic dilatation of the Sphincter of Oddi with or without prior sphincterotomy.
Device Story
The EZDilate 3-Stage Balloon Dilatation Catheter is a wire-guided, dual-lumen catheter featuring a semi-compliant nylon balloon at the distal end. It is designed for use by physicians during endoscopic procedures to dilate strictures in the alimentary tract or the Sphincter of Oddi. The device is operated via an external inflation device (e.g., Atrion K032840) which delivers fluid through the catheter shaft to expand the balloon. The balloon is engineered for progressive radial expansion to three distinct target sizes. A silicone coating on the balloon facilitates insertion, positioning, and removal. The device is supplied sterile for single-patient use. By providing controlled mechanical dilation, the catheter assists in restoring patency to narrowed anatomical structures, potentially improving patient symptoms related to strictures.
Clinical Evidence
Bench testing only. No clinical data provided. Design verification included first article inspection, tensile, fatigue, luer gauging, balloon working length, tip stiffness, compliance, burst, insertion force, friction, deflation, and endoscope compatibility testing.
Technological Characteristics
Dual-lumen catheter; nylon balloon; Pebax balloon tube, guidewire tube, and bifurcation hub; silicone balloon coating; PTFE-coated guidewire. 5.5cm balloon length, 240cm catheter length. Compatible with 0.035 in. guidewire. Sterilized via ethylene oxide. Complies with ISO 594-1/2, ISO 10993-1/5/7/10, ISO 11135-1, ISO 1138-2, ISO 14971, ISTA P2A, and ASTM D4169, F2096, F88, F1886, F1980.
Indications for Use
Indicated for adult patients requiring endoscopic dilation of alimentary tract strictures or the Sphincter of Oddi (with or without prior sphincterotomy).
Regulatory Classification
Identification
A biliary catheter and accessories is a tubular flexible device used for temporary or prolonged drainage of the biliary tract, for splinting of the bile duct during healing, or for preventing stricture of the bile duct. This generic type of device may include a bile collecting bag that is attached to the biliary catheter by a connector and fastened to the patient with a strap.
Special Controls
*Classification.* Class II (special controls). The device, when it is a bile collecting bag or a surgical biliary catheter that does not include a balloon component, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.
Predicate Devices
Boston Scientific CRE Dilatation Balloon (K112994)
{0}------------------------------------------------
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
March 27, 2015
Olympus Surgical Technologies America Gyrus ACMI, Inc. Mary Anne Patella Senior Specialist, Regulatory Affairs 136 Turnpike Road Southborough, MA 01772-2104
Re: K143609
> Trade/Device Name: EZDilate 3-Stage Balloon Dilatation Catheters Regulation Number: 21 CFR§ 876.5365 Regulation Name: Esophageal dilator Regulatory Class: II Product Code: KNO, FGE Dated: January 16, 2015 Received: January 20, 2015
Dear Mary Anne Patella.
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.
{1}------------------------------------------------
You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# Benjamin R. Fisher -S
Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K143609
Device Name EZDilate 3-Stage Balloon Dilatation Catheters
#### Indications for Use (Describe)
The EZDilate 3-Stage Balloon Dilatation Catheters are in adult populations to endoscopically dilate strictures of the alimentary tract. It is also indicated in adults for endoscopic dilatation of Oddi with or without prior sphincterotomy.
Type of Use (Select one or both, as applicable)
| <span style="font-size: 100%;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) |
|--------------------------------------------------------------------------------------|
| □ Over-The-Counter Use (21 CFR 801 Subpart C) |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## 510(k) Summary Gyrus ACMI, Inc. EZDilate 3-Stage Balloon Dilatation Catheter
#### General Information
| Contract Manufacturer: | Vention Medical, Inc.<br>261 Cedar Hill Drive<br>Marlborough, MA 01752<br>Phone: 508-481-6233 |
|-------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------|
| Establishment Registration Number: | 3004734318 |
| 510(k) Submitter: | Gyrus ACMI, Inc.<br>136 Turnpike Rd.<br>Southborough, MA 01772-2104 |
| Establishment Registration Number: | 3003790304 |
| Contact Person: | Mary Anne Patella<br>Senior Specialist Regulatory Affairs<br>508-804-2771<br>Maryanne.patella@olympus-osta.com |
| Date Prepared: | December 16, 2014 |
| Device Description | |
| Classification Name:<br>CFR citation<br>Procode<br>Classification<br>Classification Panel | Catheter, Biliary, Diagnostic<br>Dilator, Esophageal<br>21 CFR 876.5010, 876.5365<br>FGE, KNQ<br>Class II<br>Gastroenterology/Urology |
| Trade Name: | EZDilate 3-Stage Balloon Dilatation<br>Catheter |
| Generic/Common Name: | Balloon Dilation Catheter |
| Predicate Devices | |
Boston Scientific CRE Dilatation Balloon
K112994
{4}------------------------------------------------
## Product Description
The EZDilate 3-Stage Balloon Dilatation Catheter consists of a semi-compliant nylon blow molded balloon; and, a shaft that communicates a fluid passage for expansion and collapse of the balloon portion, operated by an inflation device. The EZDilate Wire Guided Balloon is designed to be used with a 0.035 in. (0.89mm) guidewire and has a wire lumen sized appropriately. Each balloon catheter is packaged with a soft tip 0.035 in. (0.89mm) guidewire pre-loaded into the guidewire lumen. All wire-guided balloons with have lengths of 5.5cm and a catheter length of 240cm. The balloon is inflated with a 60cc inflation device, which will be sold separately. The suggested inflation device is manufactured by Atrion (K032840).
## Technological Characteristics
The EZDilate 3-StageWire Guided Balloon Dilatation Catheter is a reinforced catheter attached to a distal dilatation balloon. The semi-compliant balloon allows for a progressive radial expansion through three target sizes using one balloon. A silicone coating applied to the balloon increases ease of insertion, positioning of the balloon within the alimentary tract, and device removal.
# Material
The EZDilate - Wire Guided 3-stage balloon dilatation catheter is a dual lumen catheter with a distal end comprising a nylon balloon and a proximal end comprising an inflation port and guidewire port. . The balloon tube, guidewire tube, and bifurcation hub are all constructed of Pebax. A silicone coating is applied to the balloon. A PTFE coated guidewire is pre-loaded into the guidewire lumen.
# Intended Uses
The EZDilate 3-Stage Balloon Dilatation Catheters are indicated for use in adult populations to endoscopically dilate strictures of the alimentary tract. It is also indicated in adults for endoscopic dilatation of the Sphincter of Oddi with or without prior sphincterotomy.
## Compliance to Voluntary Standards
The design of the proposed device complies with the following standards:
ISO 594-1:1986 ISO 594-2: 1998 ISO 10993-1, 2009 ISO 10993-5, 2009
{5}------------------------------------------------
Gyrus ACMI - EZDilate 3-Stage Balloon Dilatation Catheter Gyrus ACMI, Inc.
- ISO 10993-7:2008(R)2012 ISO 10993-10, 2010 ANSI/AAMI/ISO 11135-1, 2007 ISO 1138-2:2006 ISO 14971, 2012 ISTA P2A, 2011 ASTM D4169-09 ASTM F2096-11 ASTM F88/F88M:2009 ASTM F1886/F1886M:2009 ASTM F1980-07 (2011)
#### Summary of Sterilization and Shelf Life Discussion
The EZDilate 3-Stage Balloon Dilatation Catheter is delivered in a sterile state and is intended for single patient use only. The sterilization method used is ethylene oxide. Six month shelf life data will be submitted to support the 510(k) submission; the product will be launched with a shelf life of three (3) years.
#### Summary of Performance Testing
During design verification, the output of the design process was evaluated against the physical and performance specifications. In general, the evaluation compared the function of the EZDilate 3-Stage Balloon Dilatation Catheter against these specified requirements.
- First Article Inspection (Dimensional Measurements) ●
- Tensile Testing
- Fatigue Testing ●
- Luer Gauging Test ●
- Balloon Working Length
- Tip Stiffness Testing
- Compliance Testing
- Balloon Burst Testing
- Balloon Insertion Force Testing
- Balloon Friction Testing ●
- Balloon Deflation Testing
- Balloon Endoscope Compatibility Testing ●
## Substantial Equivalence
The proposed EZDilate 3-Stage Balloon Dilatation Catheter has the same intended use, design, and scientific technology as the Predicate Boston Scientific CRE
{6}------------------------------------------------
Dilatation Balloon (K112994). Both devices are of similar design and there were no new issues of safety or effectiveness with the proposed device.
## Conclusion:
In summary, the EZDilate 3-Stage Balloon Dilatation Catheter is substantially equivalent to the predicate devices and presents no new questions of safety or efficacy.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.