K143570 · Gardia Medical · NTE · Jun 4, 2015 · Cardiovascular
Device Facts
Record ID
K143570
Device Name
WIRION
Applicant
Gardia Medical
Product Code
NTE · Cardiovascular
Decision Date
Jun 4, 2015
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.1250
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The WIRION™ is indicated for use as an embolic protection system to contain and remove embolic material (thrombus/debris) while performing angioplasty and stenting procedures in carotid arteries. The diameter of the vessel at the site of filter basket placement should be between 3.5 mm to 6.0 mm. WIRION™ may be used with commercially available 0.014" guide wires.
Device Story
WIRION is an embolic protection system; consists of an independent filter unit; delivered, locked, and deployed on commercially available 0.014" guide wires; rapid exchange system for single-operator use. Used in carotid artery angioplasty and stenting procedures; physician-operated in clinical settings. Captures and removes embolic debris during stenting; prevents distal embolization; improves procedural safety. Benefits patient by reducing risk of neurological events during carotid interventions.
Clinical Evidence
Prospective, multicenter, open-label, single-arm WISE study (n=120). Primary endpoint: MACCE rate compared to historic control (6.3%). Results: WIRION MACCE rate 3.3% (p=0.0008), meeting performance goal. Secondary endpoints (device, clinical, angiographic, and procedural success) all >95%. Complication rates: access site 1.7%, neurological events 4.1%.
Technological Characteristics
Embolic protection system; independent filter unit; compatible with 0.014" guide wires. Materials biocompatible per ISO 10993-1:2009. Sterilization: EtO (SAL 10^-6). Mechanical features include filter basket, rapid exchange delivery mechanism, and radiopaque markers. No software or electronic components.
Indications for Use
Indicated for patients undergoing angioplasty and stenting procedures in carotid arteries requiring embolic protection. Vessel diameter at filter placement site must be 3.5 mm to 6.0 mm. Compatible with 0.014" guide wires.
Regulatory Classification
Identification
A percutaneous catheter is a device that is introduced into a vein or artery through the skin using a dilator and a sheath (introducer) or guide wire.
Predicate Devices
FilterWire EZ™ Embolic Protection System (K063313)
AngioGuard RX Emboli Capture Guide wire System (K101651)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, stacked on top of each other.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
June 4, 2015
Gardia Medical Ltd. Ohad Haas VP Quality Assurance & Regulatory Affairs 2 Ha-Eshel Street, PO Box 3081 Caesarea Industrial Park 38900, Israel
Re: K143570
Trade/Device Name: WIRION Regulation Number: 21 CFR 870.1250 Regulation Name: Percutaneous Catheter Regulatory Class: Class II Product Code: NTE Dated: May 7, 2015 Received: May 11, 2015
#### Dear Ohad Haas:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{1}------------------------------------------------
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Kenneth J. Cavanaugh -S
for
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known)
K143570
Device Name WIRIONTM
#### Indications for Use (Describe)
The WIRION™ is indicated for use as an embolic protection system to contain and remove embolic material (thrombus/ debris) while performing angioplasty and stenting procedures in carotid arteries.
The diameter of the vessel at the site of filter basket placement should be between 3.5 mm to 6.0 mm. WIRION™ may be used with commercially available 0.014" guide wires.
| Type of Use (Select one or both, as applicable) | |
|---------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------|
| <div> <span> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | <div> <span> Over The Counter Use (21 CFR 801 Subpart C) </span> </div> |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for Gardia Medical. The logo features the word "GARDIA" in a bold, sans-serif font, with the word "MEDICAL" in a smaller font size underneath. Above the text is an abstract graphic element composed of two overlapping shapes. The top shape is a light blue, semi-circular pattern of dots, while the bottom shape is a red, triangular shape also composed of dots.
# TRADITIONAL 510(K) SUMMARY WIRION™ Embolic Protection System
| Date of summary: | June 1, 2015 |
|--------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k) Number: | K143570 |
| Applicant's Name: | Gardia Medical Ltd.<br>2 Ha-Eshel St. P.O. Box 3081<br>Caesarea Industrial Park 38900, Israel<br>Tel.: +972-4-6277166, Ext. 107<br>Fax: +972-4-6277266 |
| Contact Person: | Ohad Haas<br>VP QA & RA<br>Phone: (972) 4-6277166, Ext. 303<br>Fax: (972) -4-6277266<br>ohadh@gradiamedical.com |
| Trade Name: | WIRION™ |
| Common name: | Embolic Protection System |
| Classification: | Name: Embolic Protection System<br>Description: Cardiovascular Percutaneous catheter<br>Product Code: NTE<br>Regulation Number: 870.1250<br>Class: II<br>Classification Panel: Cardiovascular Percutaneous catheter |
| Predicate Devices: | Substantial equivalence to the following predicates is claimed:<br>1. FilterWire EZ™ Embolic Protection System; Boston<br>Scientific, cleared under 510(k) number K063313.<br>2. AngioGuard RX Emboli Capture Guide wire System<br>Cordis Corporation, a Johnson & Johnson Company<br>cleared under 510(k) number K101651. |
# Device Description:
WIRION™ is a embolic protection System comprised of an independent Filter Unit that can be delivered, locked and deployed on commercially marketed guide wires, according to physician preference, anywhere on the wire. The WIRION™ is a rapid exchange system for single use by a single operator.
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for Gardia Medical. The logo features the company name in blue, with "GARDIA" in a larger, bolder font above "MEDICAL" in a smaller font. Above the text is a graphic element composed of two overlapping shapes made of small circles. The shape on the left is blue, and the shape on the right is red.
## Indication for Use Statement:
The WIRION™ is indicated for use as an embolic protection system to contain and remove embolic material (thrombus/debris) while performing angioplasty and stenting procedures in carotid arteries.
The diameter of the vessel at the site of filter basket placement should be between 3.5 mm to 6.0 mm. WIRION™ may be used with commercially available 0.014" guide wires.
# Technological characteristics and Substantial Equivalence:
A comparison between the WIRION™ Embolic Protection System and its predicates was performed in order to support substantial equivalence determination.
The substantial equivalency was concluded based on the device's intended use, system components, functional characteristics, principle of operation, technological characteristics, sterilization, materials and biocompatibility.
The conclusion of the comparison analysis is that the WIRION™ Embolic Protection System is Substantially Equivalent to the predicate devices.
## Performance Testing:
Clinical and non-clinical tests were conducted. The appropriate tests performed in order to determine substantial equivalency were completed.
These include testing in accordance with FDA Guidance for coronary and carotid embolic protection devices - premarket notification submissions (February 15, 2008) and additional tests as determined during WIRION™s V&V activities. All tests met their predetermined acceptance criteria and the device was found substantially equivalent to the predicate devices for its intended use.
### Non-clinical Data:
The following tests have been performed:
- a) Embolic Capture Efficiency and Retrieval Ability
- b) Stent Compatibility
- c) Simulated Use
- d) Resistance to filter rupture during removal of a fully loaded filter
- e) Flow Characteristics
- f) Radial Outward Force
- g) Tip Flexibility
- h) Tensile Strength
- Torque Strength and Response i)
- j) Dimensional Verification
- k) Catheter Coating Integrity
- I) Corrosion Resistance
- m) Fatigue
- n) Radio-opacity
- o) EtO Sterilization Validation for at least SAL 10-6.
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for Gardia Medical. The logo features a stylized graphic element above the company name. The graphic element is composed of two overlapping shapes: a blue, semi-transparent shape on the left and a red shape on the right, both with a dotted pattern. Below the graphic, the word "GARDIA" is written in a bold, blue sans-serif font, and below that, the word "MEDICAL" is written in a smaller, lighter blue font.
- p) Shelf Life/Package Integrity Tests
- q) Biocompatibility according to ISO 10993-1:2009 and FDA Blue Book Memorandum, G95-1.
# Animal Data:
Two animal studies were conducted using swine model.
Safety parameters such as vessel injury, thrombus formation, neurological effects, device integrity and hemolysis were evaluated.
Performance parameters including usability, navigation, delivery, placement and retrieval of the system and the ability to capture embolic particles were evaluated as well. The device performed comparably to the predicate device. The animal studies results support the substantial equivalence of the WIRION EPS.
# Clinical Data
Clinical data from the WIRION™ was collected from clinical studies and commercial use.
Open label single arm feasibility study aimed at demonstrating the safety and performances of the device used during treatment of carotid artery lesions undergoing stenting was conducted.
The WISE study: Multicenter non-randomized prospective, open label, single arm study with comparison to historic control was conducted on 120 patients.
The primary endpoint was MACCE rate and the secondary endpoint evaluated the device performances.
The data demonstrated that the observed MACCE rate was 3.3% while the MACCE rate of the historic control was 6.3%. The obtained P value was 0.0008, less than the 0.0015 which is the performance goal of the study thus the WIRION system met the primary end point.
Secondary endpoint (performances) included device success, clinical success, access site complications, angiographic and procedural success rate. All functions had high success rate of more than 95%. The endpoints related to incidents and complications, were low resulting in 1.7% and 4.1% as summarized below: Device success: 99.2%; Clinical success: 97.5%; Angiographic success: 99.2%; Procedural success: 98.3%; Access site complications: 1.7%; Neurological events: 4.1%. The device performed well and met all specifications and intended use requirements therefore secondary endpoint was met as well.
# Conclusions:
The evaluation of the WIRION™ Embolic Protection System performance tests and biocompatibility studies demonstrated that the device performs as intended and that it is as safe and as effective as the predicate devices.
In light of the above, we believe it is substantially equivalent to the FilterWire EZ™ Embolic Protection System (K063313) and to the AngioGuard RX Emboli Capture Guidewire System (K101651) and that it is appropriate for its intended use.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.