Fujifilm Double Balloon Endoscopes Models EN-530T and EN-580T
K143556 · Fujifilm Medical Systems U.S.A, Inc. · FDA · Aug 20, 2015 · Gastroenterology, Urology
Device Facts
Record ID
K143556
Device Name
Fujifilm Double Balloon Endoscopes Models EN-530T and EN-580T
Applicant
Fujifilm Medical Systems U.S.A, Inc.
Product Code
FDA · Gastroenterology, Urology
Decision Date
Aug 20, 2015
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 876.1500
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Fujifilm Double Balloon Endoscopes are intended for the upper and lower digestive tracts. Specifically, for the observation, diagnosis, and endoscopic treatment of the esophagus, stomach, duodenum, small intestine, large intestine, and rectum.
Device Story
Fujifilm Double Balloon Endoscopes (EN-530T, EN-580T) are flexible endoscopes used for visualization and treatment of the gastrointestinal tract. The system includes an endoscope, over-tube, balloon, and balloon controller (PB-20). The endoscope features a CCD image sensor, glass fiber illumination, and multiple channels (air/water, suction, balloon, forceps). The balloon controller inflates/deflates balloons on the endoscope and over-tube to facilitate navigation. Used in clinical settings by physicians, the device connects to a video processor and light source to display images on a monitor. The physician uses the visual output to perform diagnostic or therapeutic procedures. The double-balloon mechanism allows for deep enteroscopy, potentially benefiting patients by enabling access to areas of the small intestine otherwise difficult to reach.
Clinical Evidence
Bench testing only. Performance evaluated against consensus standards including ANSI/AAMI ES60601-1, IEC 60601 series, ISO 10993 (biocompatibility), and ISO 8600 (endoscopes). Testing confirmed specifications for inflation time, operating pressure, frictional resistance, field of view, bending capability, flow rates (air/water/suction), and resolution.
Technological Characteristics
Materials: EPDM, Silicone rubber (TSE260-5U), JSR EP21. Sensing: CCD image sensor (1/9.7"). Illumination: Glass fiber bundles. Connectivity: 500 series connector to video processor/light source. Sterilization: Compliant with AAMI/ANSI/ISO 11135-1 and 17665-1. Dimensions: 9.4mm distal end, 9.3mm flexible portion, 3.2mm forceps channel. Energy: 120V 60Hz (balloon controller).
Indications for Use
Indicated for patients requiring visualization, diagnosis, and endoscopic treatment of the upper and lower digestive tracts, including esophagus, stomach, duodenum, small intestine, large intestine, and rectum.
Regulatory Classification
Identification
An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.
Special Controls
*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.
In combination with the general controls of the FD&C Act, the integrated operating table-electromechanical surgical system is subject to the following special controls:
1. (1) Premarket clinical performance testing, or a combination of premarket clinical performance testing and postmarket surveillance (in accordance with special control (2)), must include the following:
1. (i) Objective performance measures (e.g., rate and number of conversions to other surgical modalities, rate of device related adverse events (including tissue injury, hematoma, and increased blood loss), and their severity, cause, and outcomes) must be reported with relevant descriptive comparator performance measures.
2. (ii) The data must demonstrate the performance of the device for providing accurate and precise control of attached surgical instruments in range of clinical conditions relevant to the device's intended use.
3. (iii) The test dataset must include data collected from a patient population representative of the intended patient population under anticipated conditions of use.
2. (2) Data obtained from postmarket surveillance must demonstrate, in consideration of the premarket data obtained in accordance with special control (1), that the device performs in accordance with special control (1), unless FDA determines, based on the totality of the premarket data, that data from postmarket surveillance is not required to demonstrate that the device performs as intended. Such postmarket surveillance must be conducted per a protocol determined appropriate by FDA to demonstrate that the device performs as intended (in consideration of the premarket data obtained in accordance with special control (1)), and must include initiation, enrollment, and reporting requirements to ensure timely periodic updates to FDA on post-market surveillance progress and outcomes.
3. (3) Animal performance testing must evaluate the extent of port site trauma due to repositioning of table during surgical procedures when utilizing robotic minimally invasive and laparoscopic approaches
4. (4) The device manufacturer must develop, and update as necessary, a device-specific use training program that ensures proper device setup/use/shutdown, accurate control of instruments to perform the intended surgical procedures, troubleshooting and handling during unexpected events or emergencies, and safe practices to mitigate use error.
5. (5) The device manufacturer may only distribute the device to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program.
(6) Human factors assessment must demonstrate that the user can correctly use the device system across all intended use environments with the provided instructions and training materials, including patient access during normal operating conditions and emergency situations, and effects arising from the integrated nature of the operating table and robotic surgical arms.
(7) Labeling must include:
(i) A detailed summary of clinical performance testing conducted with the device, including study population, results, adverse events, and comparisons to any comparator groups identified;
(ii) A statement in the labeling that the safety and effectiveness for the representative specific procedures was based on evaluation of the device as a surgical tool and did not include evaluation of outcomes related to the treatment of the patient's underlying disease or condition, unless FDA determines that it can be removed or modified based on clinical performance data submitted to FDA;
(iii) Identification of compatible devices;
(iv) The list of surgical procedures for which the device has been determined to be safe with clinical justification;
(v) Reprocessing instructions for reusable components;
(vi) A shelf life for any sterile components;
(vii) A description of the device-specific use training program;
(viii) A statement that the device is only for distribution to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program; and
(ix) A summary of any completed postmarket surveillance data collected as required by special control (2), including updated labeling to accurately reflect outcomes observed in postmarket surveillance.
(8) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use and must include:
(i) Device motion accuracy and repeatability;
(ii) System testing;
(iii) Instrument reliability;
(iv) Crosstalk;
(v) Table motion control;
(vi) Thermal effects on tissue;
(vii) User-device interface performance;
(viii) Workspace access testing; and
(ix) Performance testing with compatible devices.
(9) Software verification, validation, and hazard analysis must be performed.
(10) Electromagnetic compatibility and electrical, thermal, and mechanical safety testing must be performed.
(11) Performance data must demonstrate the sterility of all patient-contacting device components.
(12) Performance data must support the shelf life of the device components provided sterile by demonstrating continued sterility and package integrity over the labeled shelf life.
(13) Performance data must validate the reprocessing instructions for the reusable components of the device.
(14) Performance data must demonstrate that all patient-contacting components of the device are biocompatible.
(15) Performance data must demonstrate that all patient-contacting components of the device are non-pyrogenic.
(16) The device manufacturer must submit a report to the FDA annually on the anniversary of initial marketing authorization for the device, until such time as FDA may terminate such reporting, which comprises the following information:
(i) Cumulative summary, by year, of complaints and adverse events since date of initial marketing authorization; and
(ii) Identification and rationale for changes made to the device, labeling, or device specific use training program, which did not require submission of a premarket notification during the reporting period.
Predicate Devices
Fujinon, Inc., Double Balloon Enteroscope, Model EN-450P5/20 (K040048)
Fujinon, Inc., Double Balloon Enteroscope, Model EC-450B15 (K090116)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol that resembles a stylized human figure or a bird in flight.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
August 20, 2015
Fujifilm Medical Systems, U.S.A., Inc. Mary Moore Senior Director, Regulatory Affairs and Ouality Assurance, Endoscopy 10 High Point Drive Wayne, NJ 07470
Re: K143556
> Trade/Device Name: Fujifilm Double Balloon Endoscopes Models EN-530T and EN-580T Regulation Number: 21 CFR 876.1500 Regulation Name: Endoscope and Accessories Regulatory Class: II Product Code: FDA, FDF Dated: July 13, 2015 Received: July 13, 2015
Dear Mary Moore,
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR
{1}------------------------------------------------
Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
Herbert P. Lerner -S
for Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K143556
#### Device Name
Fujifilm Double Balloon Endoscopes Models EN-530T and EN-580T
Indications for Use (Describe)
The Fujifilm Double Balloon Endoscopes are intended for the upper and lower digestive tracts. Specifically, for the observation, diagnosis, and endoscopic treatment of the esophagus, stomach, duodenum, small intestine, large intestine, and rectum.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------|
| <label><input checked="" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D)</label> | <label><input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C)</label> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## 510(k) SUMMARY
## Fujifilm Double Balloon Endoscope Models EN-530T and EN-580T
Date: December 15, 2014
### Submitter's Information
FUJIFILM Medical Systems U.S.A., Inc., Endoscopy Division 10 High Point Drive Wayne, NJ 07470 USA FDA Establishment Registration Number: 2431293
## Contact Person:
Mary K. Moore Senior Director, Requlatory Affairs and Quality Assurance Telephone: (800) 385-4666 ext. 522498 Facsimile: (973) 686-2616 E-Mail: mkmoore@fujifilm.com
## ldentification of the Subject Device:
| Proprietary/Trade Name: | Fujifilm Double Balloon Endoscopes<br>Models EN-530T and EN-580T |
|-------------------------|------------------------------------------------------------------|
| Common Name: | Enteroscope |
| Device Class: | Class II |
| Review Panel: | Gastroenterology/Urology |
#### Classification
Enteroscope and accessories, 21 CFR § 876.1500, Product Codes: FDA, FDF
#### Predicate Devices
- · Fujinon, Inc., Double Balloon Enteroscope, Model EN-450P5/20 (K040048)
- · Fujinon, Inc., Double Balloon Enteroscope, Model EC-450B15 (K090116)
#### Intended Use / Indications for Use
The Fujifilm Double Balloon Endoscopes, Models EN-580T are intended for the visualization of the upper and lower digestive tracts. Specifically, for the observation, diagnosis, and endoscopic treatment of the esophagus, stomach, duodenum, small intestine, large intestine, and rectum.
{4}------------------------------------------------
## Device Description
The Fujifilm Double Balloon Endoscopes, Models EN-530T and EN-580T and related accessories consist of double balloon endoscopes (models EN-530T and EN-580T), a hood (model DH-17EN), a balloon (model BS-2), an over-tube (model TS-13140), and a balloon controller (model PB-20).
The endoscopes are comprised of three general sections; an operation section, an insertion portion and an umbilicus. The operation section controls the angulation (up/down/left/right) of the distal end of the endoscope. The insertion contains glass fiber bundles, several channels and a charged couple device (CCD) image sensor. The glass fiber bundles allow light to travel through the endoscope and emit light from the tip of the insertion portion to illuminate the gastrointestinal cavity. This provides adequate light to the CCD image sensor to capture an image and display it on the monitor. The endoscope also contains several channels such as an air/water channel, a suction channel, a balloon channel and a forceps channel. The balloon controller is used to inflate or deflate the balloons on the distal end of the endoscope and on the over-tube through the balloon channel. The forceps channel is used to introduce endoscope accessories such as biopsy forceps for endoscopic procedures. The umbilicus section consists of electronic components needed to operate the endoscope when plugged to the video processor and the light source. The subject models are used in combination with Fujifilm's video processor, light source and peripheral devices (water tank, endoscope accessories, monitor, printer, DVD recorder, electrosurgical instruments, foot switch, and cart). The hood is an optional accessory that is attached to the distal end of the endoscopes and is intended to maintain the field of view during the endoscopic procedure.
The minor modifications to the endoscopes were made for the purpose of overall product enhancement and general technological advancement.
## Technological Characteristics
A comparison of the technological characteristics between the subject and predicate devices is provided in the table below.
| DESCRIPTION | EN-450P5/20<br>(K040048) | EC-450BI5<br>(K090116) | EN-580T | EN-530T |
|---------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications for Use | The device is intended for<br>the Optical visualization<br>of the upper<br>gastrointestinal tract. This<br>includes the esophagus,<br>stomach, duodenum, and<br>small bowel. It is intended<br>for the observation,<br>diagnosis, and<br>endoscopic treatment. | The device is intended<br>for the optical<br>visualization of the<br>gastrointestinal tract.<br>This includes the<br>rectum, large and small<br>intestines. It is intended<br>for observation,<br>diagnosis, and<br>endoscopic treatment. | This device is intended for the visualization of the<br>upper and lower digestive tracts. Specifically, for<br>the observation, diagnosis, and endoscopic<br>treatment of the esophagus, stomach, duodenum,<br>small intestine, large intestine and rectum. | This device is intended for the visualization of the<br>upper and lower digestive tracts. Specifically, for<br>the observation, diagnosis, and endoscopic<br>treatment of the esophagus, stomach, duodenum,<br>small intestine, large intestine and rectum. |
| Insertion route | Per oral | Trans anal | Per oral or trans anal | Per oral or Trans anal |
| CCD Type | 1/6" | 1/6" | 1/9.7" double pixel | 1/9.7" |
| Field of view | 120 degrees | 140 degrees | 140 degrees | 140 degrees |
| Resolution | 0.09mm pitch chart is<br>recognizable.<br>(Distance:5mm) | 0.07mm pitch chart is<br>recognizable.<br>(Distance:4mm) | 0.056mm pitch chart is<br>recognizable.<br>(Distance:4mm) | 0.063mm pitch chart is<br>recognizable.<br>(Distance:4mm) |
{5}------------------------------------------------
| DESCRIPTION | EN-450P5/20<br>(K040048) | EC-450BI5<br>(K090116) | EN-580T | EN-530T |
|---------------------------------------------|-----------------------------------------------------------|------------------------------------------------------------|-----------------------------------------------------------|-----------------------------------------------------------|
| | 1.6mm pitch chart is<br>recognizable.<br>(Distance:100mm) | 1.25mm pitch chart is<br>recognizable.<br>(Distance:100mm) | 1.6mm pitch chart is<br>recognizable.<br>(Distance:100mm) | 1.4mm pitch chart is<br>recognizable.<br>(Distance:100mm) |
| F number | 6.21 | 7.05 | 7.2 | 7.2 |
| Distal end<br>diameter | 8.5mm | 9.4mm | 9.4mm | 9.4mm |
| Flexible portion<br>diameter | 8.5mm | 9.3mm | 9.3mm | 9.3mm |
| Maximum<br>diameter of<br>insertion portion | 9.0mm | 10mm | 10mm | 10mm |
| Number of Light<br>guide fiber | 900x2 | 1200x2 | 650x2 | 650x2 |
| Forceps channel<br>diameter | 2.2mm | 2.8mm | 3.2mm | 3.2mm |
| Forceps channel<br>length | 2100mm | 1620mm | 2100mm | 2100mm |
| Working length | 2000mm | 1520mm | 2000mm | 2000mm |
| Total length | 2300mm | 1820mm | 2300mm | 2300mm |
| Processor | VP-4400, VP-4400HD,<br>VP-4440HD | VP-4400, VP-4400HD,<br>VP-4440HD | VP-4440HD | VP-4400, VP-4400HD,<br>VP-4440HD |
| LG Connector | 400 series connector | 400 series connector | 500 series connector | 500 series connector |
| Video Connector | 400 series connector | 400 series connector | 500 series connector | 500 series connector |
| Location of<br>Balloon inlet | On the control portion | On the control portion | On the LG connector | On the LG<br>connector |
| DESCRIPTION | PB-10<br>(K040048) | PB-20 |
|----------------------------------------------|-----------------------|------------------------------------------------------------------------------|
| Compatible endoscope | EN-450P5/20 | EC-450BI5, EN-450P5/20, EN-<br>450T5, EN-530T, EN-580T |
| Compatible overtube | TS-12140 | TS-12140,TS-13140,TS-13101 |
| Compatible Balloon | BS-1 | BS-2 |
| Power supply | 120V 60Hz | 120V 60Hz |
| Current consumption | 0.37A | 0.58A |
| Dimensions | W300 x H200 X D300 mm | W350 x H130 X D420 mm |
| Weight | 8.7kg | 10kg |
| Set air supply pressure | 5.6kPa | 5.6kPa |
| Warning pressure | 8.2kPa | 8.2kPa |
| Tubing | TY-01 and TY-02 | TY-06D for use with EN-530T<br>and EN-580T<br>TY-04D for use with 400 Series |
| Trap for preventing fluid to invade the pump | Trap bottle (TP-01) | Membrane filter (for TY-06D,<br>TY-04D) |
{6}------------------------------------------------
| DESCRIPTION | PB-10<br>(K040048) | PB-20 |
|--------------------------------------------------------------------------------------------------|--------------------|-------|
| Foot switch terminal to connect the optional foot<br>switch FS1. | No | Yes |
| The Video output terminal to the external monitor<br>to graphically show the balloon conditions. | No | Yes |
| DESCRIPTION | DH-14EN<br>(K040048) | DH-17EN | |
|-----------------------------------------|----------------------|----------------|-----------------|
| Outer diameter | 10.8mm | 11.5mm | |
| Maximum diameter of attaching endoscope | 12.5mm | 13.5mm | |
| Total length | 8.0mm | 8.0mm | |
| Distance from the tip | 1.5mm | 1.5mm | |
| Inner diameter of attaching portion | 8.8mm | 9.5mm | |
| Inner diameter of distal end | 7.8mm | 8.5mm | |
| Sterility | Non Sterilized | Non Sterilized | |
| Reuse or single use | Single Use | Single Use | |
| Material | Distal portion | EPDM | Silicone rubber |
| | Attaching portion | JSR EP21 | TSE260-5U |
The subject devices are also used with the company's previously cleared balloon model BS-2 (K090116).
## Substantial Equivalence
The subject and predicate endoscopes have the same intended use/indications for use of visualization of the digestive tracts. Per oral and trans anal insertion routes were already cleared in the company's predicate devices K040048 and K090116, respectively. They also have similar technological characteristics as shown in the table above. Both use the Double Balloon Endoscope system to access the gastrointestinal tracts. Device dimensions and visualization technology are also similar. The technological differences between the subject endoscopes and the predicate devices raise no new questions of safety or effectiveness. Bench testing data demonstrated that the subject endoscopes have substantially equivalent performance to the predicates.
{7}------------------------------------------------
## Performance Data
EMC and electrical safety of the subject devices were evaluated using the following consensus standards: ANSI/AAMI ES60601-1:2005; IEC 60601-1-2:2007; IEC 60601-1-6:2010; and IEC 60601-2-18:2009.
Biocompatibility of the subject devices was evaluated using the following consensus standards: ISO 10993-1:2009; ISO 10993-5:2009; and ISO 10993-10:2010.
Cleaning, disinfection, and sterilization were evaluated according to the following consensus standards: AAMI TIR30:2011; AAMI TIR12:2010; AAMI/ANSI/ISO 11135-1:2007; and AAMI/ANSI/ISO 17665-1:2006.
Endoscope specific testing was conducted using the following consensus standards: ISO 8600-1:2013; ISO 8600-3:1997; and ISO 8600-4:1997.
Subject devices met performance specifications of the following additional testing:
- . Time for inflation
- Operating pressure .
- Frictional resistance ●
- Field of view
- Bending capability ●
- Rate of air supply ●
- Rate of water supply ●
- Rate of suction ●
- Viewing direction ●
- . Resolution
- Pump air feed pressure .
- Pump air evacuation pressure ●
- Pump air flow rate .
#### Conclusions
The subject devices are substantially equivalent to the predicates based on intended use/indications for use and technological characteristics. The minor technological differences between the propose endoscopes and its predicate devices raise no new issues of safety or effectiveness. Bench testing data demonstrated that the subject endoscopes have substantially equivalent performance to the predicates.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.