Signos RT Bladder scanner projects ultrasound energy through the lower abdomen of a patient to obtain an image of the bladder that is used to non-invasively determine bladder volume. Users must have ultrasound training before using the device.
Device Story
Hand-held ultrasound system; projects 3MHz ultrasonic energy through lower abdomen; receives echoes to generate B-mode images on integrated LCD. Used by trained clinicians for non-invasive bladder volume measurement. Device calculates volume from single or dual scans using embedded algorithms. Output displayed on-screen; supports external USB PictBridge printing or image transfer to PC via SigViewer software. Benefits include non-invasive assessment of bladder volume, aiding clinical decision-making regarding urinary retention or catheterization. Operates on Li-Ion battery; portable form factor.
Clinical Evidence
Bench testing only. Phantom studies demonstrated bladder volume measurement accuracy within ±9.6% (specification ±15%). Imaging performance (lateral/axial resolution) verified as identical to predicate. Compliance with IEC 60601-1, IEC 60601-2-37, IEC 60601-1-2, NEMA UD-2, NEMA UD-3, and ISO 10993 biocompatibility standards confirmed.
Technological Characteristics
Hand-held B-mode ultrasound; 3MHz annular sector transducer. Materials: Sabic Cycoloy HC1204HF, Mitsui TPX-MED18, Sabic Versollen OMX1255NX-1. Dimensions: 6.8 x 11.5 x 2.0 cm; 0.13 kg. Power: Li-Ion battery or 100-240 VAC external. Connectivity: USB for printing/PC data transfer. Safety: Class II, Type BF, IPX7 transducer lens. Software developed per IEC 62304.
Indications for Use
Indicated for non-invasive bladder volume determination in patients via percutaneous abdominal ultrasound scanning. Contraindicated for pregnant patients and intracavity imaging (trans-esophageal, trans-rectal).
Regulatory Classification
Identification
An ultrasonic pulsed echo imaging system is a device intended to project a pulsed sound beam into body tissue to determine the depth or location of the tissue interfaces and to measure the duration of an acoustic pulse from the transmitter to the tissue interface and back to the receiver. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
Special Controls
*Classification.* Class II (special controls). A biopsy needle guide kit intended for use with an ultrasonic pulsed echo imaging system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
{0}------------------------------------------------
January 14, 2015
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
Signostics Limited % Mr. Mark Job Responsible Third Party Official Regulatory Technology Services LLC 1394 25th Street. NW BUFFALO MN 55313
Re: K143493
Trade/Device Name: SignosRT Bladder Regulation Number: 21 CFR 892.1560 Regulation Name: Ultrasonic pulsed echo imaging system Regulatory Class: II Product Code: IYO ITX Dated: December 28, 2014 Received: January 7, 2015
Dear Mr. Job:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Robert A. Ochs
Robert Ochs, Ph.D. Acting Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
#### Indications for Use
510(k) Number (if known)
#### K143493
Device Name Signos RT Bladder
#### Indications for Use (Describe)
Signos RT Bladder scanner projects ultrasound energy through the lower abdomen of a patient to obtain an image of the bladder that is used to non-invasively determine bladder volume. Users must have ultrasound training before using the device.
#### Contraindications
The SignosRT Bladder Scanner is designed for percutaneous scanning only. Do not attempt intracavity imaging; in particular, trans-esophageal, trans-rectal scans are contraindicated.
The SignosRT Bladder Scanner is contraindicated for pregnant patients.
Type of Use (Select one or both, as applicable)
🇿 Prescription Use (Part 21 CFR 801 Subpart D)
□ Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwoork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
#### 510(k) Indications for Use Form
#### TABLE 1 - SIGNOSRT BLADDER INDICATIONS FOR USE FORM
System: SignosRT Bladder Intended Use: Diagnostic ultrasound imaging of the human body as follows: Clinical Application Mode of Operation PWD General Specific B M CWD Color Combined Other* (Track I Only) (Tracks I & III) Doppler (Specify) (Specify) Ophthalmic Ophthalmic Fetal Abdominal N Intra-operative (Specify) Intra-operative(Neuro) Laparoscopic Pediatric Small Organ (Specify) Neonatal Cephalic Adult Cephalic Fetal Imaging Trans-rectal & Other Trans-vaginal Trans-urethral Trans-esoph. (non-Card.) Musculo-skeletal (Conventional) Musculo-skeletal (Superficial) Intravascular Other (Specify) Cardiac Adult Cardiac Pediatric Intravascular (Cardiac) Cardiac Trans-esoph. (Cardiac) Intra-cardiac Other (Specify) Peripheral Peripheral vessel Vessel Other (Specify)
N= new indication; P=previously cleared by FDA; E=added under AppendixE Additional Comments:
Type of Use (Select one or both, as applicable)
区 Prescription Use (Part 21 CFR 801 Subpart D)
□ Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
{4}------------------------------------------------
## TABLE 2 - SIGNOSRT BLADDER SCANNER INDICATIONS FOR USE FORM
System: SignosRT Bladder
Transducer: S3 (P03479)
Diagnostic ultrasound imaging of the human body as follows : Intended Use:
| Clinical Application | | Mode of Operation | | | | | | |
|---------------------------|------------------------------------|-------------------|---|-----|-----|------------------|-----------------------|---------------------|
| General<br>(Track I Only) | Specific<br>(Tracks I & III) | B | M | PWD | CWD | Color<br>Doppler | Combined<br>(Specify) | Other*<br>(Specify) |
| Ophthalmic | Ophthalmic | | | | | | | |
| Fetal Imaging<br>& Other | Fetal | | | | | | | |
| | Abdominal | N | | | | | | |
| | Intra-operative (Specify) | | | | | | | |
| | Intra-operative(Neuro) | | | | | | | |
| | Laparoscopic | | | | | | | |
| | Pediatric | | | | | | | |
| | Small Organ (Specify) | | | | | | | |
| | Neonatal Cephalic | | | | | | | |
| | Adult Cephalic | | | | | | | |
| | Trans-rectal | | | | | | | |
| | Trans-vaginal | | | | | | | |
| | Trans-urethral | | | | | | | |
| | Trans-esoph. (non-Card.) | | | | | | | |
| | Musculo-skeletal<br>(Conventional) | | | | | | | |
| | Musculo-skeletal<br>(Superficial) | | | | | | | |
| | Intravascular | | | | | | | |
| | Other (Specify) | | | | | | | |
| Cardiac | Cardiac Adult | | | | | | | |
| | Cardiac Pediatric | | | | | | | |
| | Intravascular (Cardiac) | | | | | | | |
| | Trans-esoph. (Cardiac) | | | | | | | |
| | Intra-cardiac | | | | | | | |
| | Other (Specify) | | | | | | | |
| Peripheral | Peripheral vessel | | | | | | | |
| Vessel | Other (Specify) | | | | | | | |
N= new indication; P=previously cleared by FDA; E=added under AppendixE Additional Comments:
Type of Use (Select one or both, as applicable)
区 Prescription Use (Part 21 CFR 801 Subpart D)
□ Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
{5}------------------------------------------------
## 510(k) Summary
#### 1. Sponsor:
Signostics Ltd PO Box 1048, Pasadena 1284 South Road Clovelly Park, SA 5042 Australia
## 2. Contact Person: Stewart Bartlett Quality and Regulatory Manager Telephone: +61 (8) 7424 0600
- 3. Date Prepared: December 28, 2014
- 4. De vice Name : SignosRT Bladder
- 5. Proprietary/Marketed Names: SignosRT Bladder
## 6. Common/Usual Name:
Diagnostic ultrasound system and transducer
### 7. Classification
Regulatory Class: II
Review Category: Tier II
Ultrasonic Pulsed Echo Imaging System (21 CFR 892.1560, 90-IYO)
(21 CFR 892.1570, 90-ITX)
Diagnostic Ultrasound Transducer
Classification Panel: Radiology
### 8. Predicate Devices
Portascan Bladder Scanner (K033906)
SignosRT Ultrasound System (K130659)
{6}------------------------------------------------
## 9. Risk Analysis Method Used
Signostics Ltd applied ISO-14971 to the design and development of the SignosRT Bladder. The conclusion from the risk analysis was the device was safe for its intended use and does not pose any unacceptable risks.
#### 10. Basis for Substantial Equivalence - Technological Characteristics
Signostics Ltd believes the SignosRT Bladder described in this Submission is substantially equivalent to the predicate devices as follows:
- a. Portascan Bladder Scanner (K033906)
- b. SignosRT Ultrasound System (K130659)
The SignosRT Bladder is substantially equivalent to the predicate devices listed above. All systems transmit ultrasonic energy into patients, then process received echoes to produce on-screen images of anatomy. All systems allow for B-mode imaging and measurements of volume of a bladder. The indication for use statement of the SignosRT Bladder is identical to the Portascan. The SignosRT Personal Ultrasound system (K130659) is a general ultrasound imaging device with an intended use statement covering a number of areas, including the abdomen which includes the bladder.
All systems enable an estimated bladder volume to be calculated from a single scan or from two scans for more accurate results. The algorithms used to calculate bladder volume are identical for the SignosRT Bladder and the predicate SpeqRT Personal Ultrasound System.
The SignosRT Bladder is identical in construction, materials, and controls to the SignosRT Personal Ultrasound (K130659), with the only difference being the indications for use statement and the software configuration. The patient contact materials are identical. The transducer and operating frequency (3.5MHz) are also identical. Both systems contain sector transducers with annular (circular) acoustic elements generating pie shaped images, have identical operating frequency, identical outer diameter of acoustic crystal, identical acoustic output, and both maintain MI and TI to be <1.0 at all times.
The predicate SignosRT Personal Ultrasound has a remote display mode to allow streaming of the video to a larger display. The SignosRT Bladder also has support to allow streaming of the video to a larger display.
The predicate Portascan has printer as part of the device. The SignosRT Bladder scanner and predicate SignosRT Ultrasound System both can be connected directly to a USB PictBridge printer for full-size printouts, or transfer images to a PC running SigViewer accessory software to print images.
{7}------------------------------------------------
# 11. Device Description
The Signostics Ltd SignosRT Bladder is a hand-held, diagnostic ultrasound system with an on-screen display. Its purpose is to acquire ultrasound echo data and display it in B-Mode on an LCD display to enable non-invasive volume measurement of a patients bladder.
Technical specifications for the Signostics SignosRT Bladder are as follows:
| System | | | | | |
|----------------------------|--------|------------------------------------------------------------------------------------------------|--------------------------|---------|------------|
| Transducer frequencies: | | 3MHz (S3 transducer) | | | |
| Frame rate: | | 8Fps or 16Fps (Imaging only) | | | |
| Ultrasound lines/frame: | | 128 lines for 90° frame at 10cm | | | |
| Fields of View: | | 1-18 cm for 3MHz | | | |
| External Video Output: | | No | | | |
| Liquid-Crystal Display: | | 18 bit, 262,000 Color, Active Matrix TFT LCD | | | |
| Materials | | Sabic Cycoloy HC1204HF, Mitsui TPX-MED18, Sabic Versollen<br>OMX1255NX-1 | | | |
| Size: - | | | | | |
| Width: | | 6.8 cm | | | |
| Height: | | 11.5 cm | | | |
| Depth: | | 2.0 cm | | | |
| Weight: | | 0.13 kg | | | |
| Electrical | | | | | |
| External Power: | Input: | 100-240 VAC,<br>50-60Hz | | Output: | 5 VDC @ 2A |
| Battery: | | Li-Ion battery pack (2 Whr) | | | |
| Leakage Current: | | 10 $ μ $ A maximum | | | |
| Primary Breakdown Voltage: | | 3000VAC | | | |
| Safety Standards: | | IEC 60601-1:2005, ES60601-1:2005, IEC 60601-2-37:2007, IEC<br>60601-1-2:2007, ISO 10993-1:2003 | | | |
| Protection Class: | | Class II: per IEC 60601-1 | | | |
| Degree of Protection: | | | Type BF: per IEC 60601-1 | | |
| Environmental | |
|--------------------------|---------------------------------------------------------------------|
| Mechanical Shock: | Drop and push testing per ES60601-1 |
| Mechanical Vibration: | Random Acceleration Profile per JIS Z 0232:2004 5Hz-300Hz |
| Drop Test (to concrete): | 1 meter |
| Operating Temperature: | 0 to 40 C |
| Humidity: | 20 to 80% RH, non-condensing |
| Water Resistance: | Transducer IPX7 lens, IPX1 probe degree of protection against water |
| Altitude: | 0.7 – 1.05 standard atmospheres (2500m or 8200 feet) operating |
| Storage | |
| Temperature: | -20 to 45 C |
| Humidity: | 10 to 90% RH, non-condensing |
| Altitude: | 0.5 – 1.05 standard atmospheres storage |
{8}------------------------------------------------
## 12. Non-clinical Performance Data
The SignosRT Bladder has been bench tested for imaging performance and measurement accuracy, with tests showing the SignosRT Bladder imaging performance and measurement accuracy to be substantially equivalent to the predicate devices. The measured lateral and axial resolution for the SignosRT Bladder is identical to the predicate SpeqRT Personal Ultrasound System. The measured volume accuracy on a phantom was a maximum of +-9.6%. well within the specified +-15%.
The SignosRT Bladder Scanner device has been tested by independent laboratories to IEC 60601-1:2005, IEC 60601-2-37:2007, IEC 60601-1-2:2007, ISO 10993-5:2009, ISO 10993-10:2002, ISO 10993-10:2010, ISO 10993-12:2007, NEMA UD-2-2004 (R2009), NEMA UD-3-2004(R2009) and found to comply with all standards.
The software and firmware in the SignosRT Bladder has been developed and verified according to IEC 62304:2006. The verification reports (Appendix R and V), traceability (Appendix Q), and risk analysis (Appendix A) demonstrate the SignosRT Bladder operates as intended and risks mitigated in firmware have been verified.
The conclusion from the testing is the device is safe and effective for its intended use, and performs as well or better than the predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.