# Euphora Rapid Exchange Balloon Dilatation Catheter (K143480)

_Medtronic, Inc. · LOX · Apr 2, 2015 · Cardiovascular · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K143480

## Device Facts

- **Applicant:** Medtronic, Inc.
- **Product Code:** [LOX](/productcode/LOX.md)
- **Decision Date:** Apr 2, 2015
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 870.5100
- **Device Class:** Class 2
- **Review Panel:** Cardiovascular
- **Attributes:** Therapeutic

## Indications for Use

The balloon dilatation catheter is indicated for balloon dilatation of the stenotic portion of a coronary artery or bypass graft stenosis for the purpose of improving myocardial perfusion. The balloon dilatation catheter (balloon models 2.00 mm to 4.00 mm) is also indicated for post deployment expansion of balloon expandable stents.

## Device Story

Euphora RX Balloon Dilatation Catheter is a percutaneous transluminal coronary angioplasty (PTCA) system; used in healthcare facilities by physicians. Device consists of a catheter with a distal balloon and proximal female luer for inflation device attachment. Guidewire lumen enables positioning; radiopaque marker bands and shaft markers facilitate placement. Principle of operation: balloon inflated to defined diameter at specific pressure to dilate stenotic vessels or expand stents. Output is mechanical force applied to vessel wall or stent. Benefits include improved myocardial perfusion and stent apposition. Single-use, sterile, hydrophilic-coated device.

## Clinical Evidence

No clinical investigation performed. Substantial equivalence supported by extensive in-vitro bench testing and biocompatibility studies per ISO 10993-1:2009. Bench tests included effective length, catheter profile, pressure capability, balloon fatigue, compliance, inflation/deflation time, bond strength, coating integrity/particulates, flexibility, and torque strength. Biocompatibility testing included cytotoxicity, sensitization, reactivity, systemic toxicity, pyrogenicity, thromboresistance, hemolysis, PTT, and complement activation assays.

## Technological Characteristics

PTCA rapid exchange catheter; hydrophilic coating; radiopaque marker bands. Nominal pressure 8atm; rated burst pressure 14atm. External communicating device with limited blood contact (≤ 24 hours). Sterilized via EtO (ISO 11135, EN 556) to SAL 10^-6. Materials are identical to predicate. No software or electronic components.

## Regulatory Identification

Standard PTCA Catheter: A PTCA catheter is a device that operates on the principle of hydraulic pressurization applied through an inflatable balloon attached to the distal end. A PTCA balloon catheter has a single or double lumen shaft. The catheter features a balloon of appropriate compliance for the clinical application, constructed from a polymer. The balloon is designed to uniformly expand to a specified diameter and length at a specific pressure as labeled, with well characterized rates of inflation and deflation and a defined burst pressure. The device generally features a type of radiographic marker to facilitate fluoroscopic visualization of the balloon during use. A PTCA catheter is intended for balloon dilatation of a hemodynamically significant coronary artery or bypass graft stenosis in patients evidencing coronary ischemia for the purpose of improving myocardial perfusion. A PTCA catheter may also be intended for the treatment of acute myocardial infarction; treatment of in-stent restenosis (ISR) and/or post-deployment stent expansion. Cutting/scoring PTCA Catheter: A cutting/scoring PTCA catheter is a balloon-tipped catheter with cutting/scoring elements attached, which is used in those circumstances where a high pressure balloon resistant lesion is encountered. A cutting/scoring PTCA catheter is intended for the treatment of hemodynamically significant coronary artery stenosis for the purpose of improving myocardial perfusion. A cutting/scoring PTCA catheter may also be indicated for use in complex type C lesions or for the treatment of in-stent restenosis.

## Special Controls

*Classification.* Class II (special controls). The special control for this device is “Class II Special Controls Guidance Document for Certain Percutaneous Transluminal Coronary Angioplasty (PTCA) Catheters.” See § 870.1(e) for the availability of this guidance document.(b)
*Cutting/scoring PTCA Catheter* —(1)*Identification.* A cutting/scoring PTCA catheter is a balloon-tipped catheter with cutting/scoring elements attached, which is used in those circumstances where a high pressure balloon resistant lesion is encountered. A cutting/scoring PTCA catheter is intended for the treatment of hemodynamically significant coronary artery stenosis for the purpose of improving myocardial perfusion. A cutting/scoring PTCA catheter may also be indicated for use in complex type C lesions or for the treatment of in-stent restenosis.(2)
*Classification.* Class III (premarket approval). As of May 28, 1976, an approval under section 515 of the act is required before this device may be commercially distributed. See § 870.3.

## Predicate Devices

- Sprinter Legend Balloon Dilatation Catheter ([P790017](/device/P790017.md)/S096)

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features the department's name encircling a symbol. The symbol is a stylized representation of a human figure, with three profiles overlapping to suggest a sense of community and care. The profiles are depicted in a flowing, abstract manner.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

April 2, 2015

Medtronic, Inc. c/o Elaine Gullane Senior Regulatory Affairs Specialist, Medtronic Ireland Parkmore Business Park West Galway, Ireland

Re: K143480

Trade/Device Name: Euphora Rapid Exchange Balloon Dilatation Catheter Regulation Number: 21 CFR 870.5100 Regulation Name: Percutaneous Transluminal Coronary Angioplasty (PTCA) Catheter Regulatory Class: Class II Dated: February 17, 2015 Received: February 19, 2015

Dear Ms. Elaine Gullane:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set

{1}------------------------------------------------

forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerelv vours.

M.A. Hilleman

for Bram Zuckerman, M.D. Director, Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

|                                                                         | K143480<br>Page 1 of 1                                                                                                                                                                                                                                                                                                                                                                                                     |
|-------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| DEPARTMENT OF HEALTH AND HUMAN SERVICES<br>Food and Drug Administration | Form Approved: OMB No. 0910-0120<br>Expiration Date: January 31, 2017                                                                                                                                                                                                                                                                                                                                                      |
| Indications for Use                                                     | See PRA Statement below.                                                                                                                                                                                                                                                                                                                                                                                                   |
| 510(k) Number ( <i>if known</i> )                                       | K143480                                                                                                                                                                                                                                                                                                                                                                                                                    |
| Device Name                                                             | Euphora™ Rapid Exchange Balloon Dilatation Catheter                                                                                                                                                                                                                                                                                                                                                                        |
| Indications for Use ( <i>Describe</i> )                                 | The balloon dilatation catheter is indicated for balloon dilatation of the stenotic portion of a coronary artery or bypass graft stenosis for the purpose of improving myocardial perfusion.<br>The balloon dilatation catheter (balloon models 2.00 mm to 4.00 mm) is also indicated for post deployment expansion of balloon expandable stents.                                                                          |
| Type of Use ( <i>Select one or both, as applicable</i> )                |                                                                                                                                                                                                                                                                                                                                                                                                                            |
|                                                                         | <label><input checked="" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D)</label> <label><input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C)</label>                                                                                                                                                                                                                                        |
|                                                                         | CONTINUE ON A SEPARATE PAGE IF NEEDED.                                                                                                                                                                                                                                                                                                                                                                                     |
|                                                                         | This section applies only to requirements of the Paperwork Reduction Act of 1995.<br>*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*                                                                                                                                                                                                                                                               |
|                                                                         | The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: |
|                                                                         | Department of Health and Human Services<br>Food and Drug Administration<br>Office of Chief Information Officer<br>Paperwork Reduction Act (PRA) Staff<br><a href="mailto:PRAStaff@fda.hhs.gov">PRAStaff@fda.hhs.gov</a>                                                                                                                                                                                                    |
|                                                                         | "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."                                                                                                                                                                                                                                                           |
| FORM FDA 3881 (8/14)                                                    | Page 1 of 1                                                                                                                                                                                                                                                                                                                                                                                                                |
|                                                                         | PSC Publishing Services (301) 443-6740 EF                                                                                                                                                                                                                                                                                                                                                                                  |

43-6740

81

{3}------------------------------------------------

## 510(k) Summary per 21 CRF 807.92

December 4th, 2014 Date Prepared: Applicant: Medtronic Ireland Parkmore Business Park West Galway Ireland Official Elaine Gullane Senior Regulatory Affairs Specialist Correspondent: Medtronic Ireland Parkmore Business Park West Galway Ireland Phone: (353) 91 708682 Fax: (353) 91 708672 Email: elaine.gullane@medtronic.com Proprietary Euphora™ Rapid Exchange Balloon Dilatation Catheter Name: Percutaneous Transluminal Coronary Angioplasty (PTCA) Catheter Common Name: Device Class II (special controls) Classification: Regulation 21 CFR 870.5100 Number: Percutaneous Transluminal Coronary Angioplasty (PTCA) Catheter Classification Name: LOX Product Code:

{4}------------------------------------------------

The Euphora™ RX Balloon Dilatation catheter is a Percutaneous Device Transluminal Coronary Angioplasty (PTCA) Rapid Exchange System. The Description: balloon at the distal end of the catheter can be inflated to a defined diameter at a specific pressure as defined in the product labelling. The proximal end of the catheter has a female luer for attachment to an inflation device. The catheter provides a lumen which enables the use of a guidewire to position the catheter. Radiopaque balloon marker bands enable accurate placement. Shaft markers for brachial and femoral techniques are in place.

| Diameters<br>(mm) | Lengths (mm) |           |           |           |           |           |          |
|-------------------|--------------|-----------|-----------|-----------|-----------|-----------|----------|
|                   | 6            | 10        | 12        | 15        | 20        | 25        | 30       |
| 1.50              | EUP1506X     | EUP1510X  | EUP1512X  | EUP1515X  | EUP1520X  | -         | -        |
| 2.00              | EUP2006X     | EUP2010X  | EUP2012X  | EUP2015X  | EUP2020X  | EUP2025X  | EUP2030X |
| 2.25              | EUP22506X    | EUP22510X | EUP22512X | EUP22515X | EUP22520X | EUP22525X | -        |
| 2.50              | EUP2506X     | EUP2510X  | EUP2512X  | EUP2515X  | EUP2520X  | EUP2525X  | EUP2530X |
| 2.75              | EUP27506X    | EUP27510X | EUP27512X | EUP27515X | EUP27520X | EUP27525X | -        |
| 3.00              | EUP3006X     | EUP3010X  | EUP3012X  | EUP3015X  | EUP3020X  | EUP3025X  | EUP3030X |
| 3.25              | EUP32506X    | EUP32510X | EUP32512X | EUP32515X | EUP32520X | EUP32525X | -        |
| 3.50              | EUP3506X     | EUP3510X  | EUP3512X  | EUP3515X  | EUP3520X  | EUP3525X  | EUP3530X |
| 3.75              | EUP37506X    | EUP37510X | EUP37512X | EUP37515X | EUP37520X | EUP37525X | -        |
| 4.00              | EUP4006X     | EUP4010X  | EUP4012X  | EUP4015X  | EUP4020X  | EUP4025X  | EUP4030X |

- . The Euphora device is available in the following model numbers:
The models differ only in the balloon diameter and the balloon length across the size matrix.

- The Euphora device has a nominal pressure of 8atm and a rated burst pressure of 14atm.
- The Euphora device is single use, sterile, EtO sterilised device. It has a hydrophilic coating and is classified as an external communicating device with limited exposure i.e. whose contact with circulating blood is ≤ 24 hours.
- . The Euphora device environment of use is a healthcare facility.
- Indications For The Euphora™ RX Balloon Dilatation catheter is indicated for balloon dilatation of the stenotic portion of a coronary artery or bypass graft Use: stenosis for the purpose of improving myocardial perfusion. The balloon dilatation catheter (balloon models 2.00mm to 4.00mm) is also indicated for post deployment expansion of balloon expandable stents.

{5}------------------------------------------------

The Euphora™ RX Balloon Dilatation Catheter is similar to the following Substantially Equivalent predicate device with respect to intended use, design and technology: Sprinter Legend® Device: Medtronic Balloon Dilatation Catheter (P790017/S096, approved October 31, 2008) This predicate has not been subject to a design-related recall. Reference devices were used in this submission.

The Euphora™ RX Balloon Dilatation Catheter is offered in more balloon Summary of sizes compared to its predicate and the 2.0mm diameter size is indicated Technological for post deployment expansion of balloon expandable stents. The Differences to the Euphora™ RX Balloon Dilatation Catheter has a consistent nominal Predicate Device: (8atm) and rated burst pressure (14atm) across all balloon diameters, and has a higher nominal and rated burst pressure for the 1.5mm balloon diameter when compared to the predicate Sprinter Legend® RX Balloon Dilatation Catheter. The EuphoraTM RX Balloon Dilatation Catheter incorporates the identical balloon material, identical radiopaque marker band material, identical hydrophilic coating and identical outer shaft material to the predicate device. All other materials utilized are deemed substantially equivalent to the predicate device Sprinter Legend® RX Balloon Dilatation Catheter.

Design Verification in-vitro testing: Summary of Non-

Clinical Data:

1- Design verification in-vitro testing:

The following in-vitro bench tests were completed on the Euphora™ RX Balloon Dilatation Catheter in accordance with the requirements of Guidance for Industry and FDA Staff - Class II Special Controls Guidance Document for Certain Percutaneous Transluminal Coronary Angioplasty (PTCA) Catheters, September 8, 2010 and verify that it meets the required performance specifications:

- Effective length ●
- Catheter Profile ●
- System Pressure Capability & System Pressure Capability (in Stent)
- Balloon Fatigue & Balloon Fatigue (in Stent)
- Balloon Compliance ●
- Balloon Inflation and Deflation Time ●
- Catheter Bond Strength
- Coating Integrity ●
- Coating Particulate Evaluation ●
- Flexibility and Kink ●
- Torque Strength ●

## Pre-Clinical Study (non-GLP):

Medtronic conducted pre-clinical in-vivo (non-GLP) studies for evaluation of the Radiopacity attribute of the Euphora™ RX Balloon Dilatation 

{6}------------------------------------------------

Catheter.

## Biocompatibility Testing:

Biocompatibility testing for the Euphora™ RX Balloon Dilatation Catheter has been completed in accordance with the recommendations of Guidance for Industry and FDA Staff - Class II Special Controls Guidance Document for Certain Percutaneous Transluminal Coronary Angioplasty (PTCA) Catheters, September 8, 2010 and International Standard ISO10993-1:2009, Biological Evaluation of Medical devices- Part 1: Evaluation and Testing for an external communicating device with limited exposure i.e. whose contact with circulating blood is ≤ 24 hours.

The following Biocompatibility tests were performed to support the Euphora™ RX Balloon Dilatation Catheter:

- Cytotoxicity Study using ISO MEM Elution method ●
- ISO Maximisation Sensitisation Study ●
- ISO Acute Intracutaneous Reactivity ●
- ISO Acute Systemic Toxicity ●
- USP Material Mediated Pyrogen Study in Rabbits ●
- In Vivo Thromboresistance Study ●
- ASTM In-vitro Haemolysis .
- ASTM Partial Thromboplastin Time (PTT) Coagulation Testing ●
- C3a Complement Activation Assay Study ●
- Sc5b-9 Complement Activation Assay Study ●
- USP Physicochemical Tests ●

The Euphora" RX Balloon Dilatation Catheter met all specified design and performance requirements. No new safety or effectiveness issues were raised during the testing. The bench testing qualification and biocompatibility testing demonstrated that the subject device Euphora™ RX Balloon Dilatation Catheter is substantially equivalent in terms of safety and effectiveness to the predicate device.

Summary of No clinical investigation has been performed for this device.

Clinical Data:

{7}------------------------------------------------

| Sterilization | The Euphora™ RX Balloon Dilatation Catheter will be sterilized and                                                                                                                |
|---------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Validation:   | validated for EtO sterilization in accordance with ISO11135 and EN556 to<br>achieve a minimum Sterility Assurance Level (SAL) of 10°.                                             |
| Conclusion:   | Through the data and information presented, Medtronic Ireland considers<br>the Euphora™ RX Balloon Dilatation Catheter to be substantially<br>equivalent to the predicate device. |

---

**Source:** [https://fda-staging.innolitics.com/device/K143480](https://fda-staging.innolitics.com/device/K143480)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).