The Digi-One System is indicated for non-invasive evaluation of intracranial and extracranial vascular flow velocity irregularities in adults and in children. It is not intended for fetal use. It is not intended for neonatal use.
Device Story
Digi-One System is a portable digital Transcranial Doppler (TCD) system; connects to external Windows-based PC. Inputs: ultrasound signals from 1MHz, 2MHz, 4MHz, or 8MHz transducers. Transforms inputs via Doppler ultrasound processing to measure blood flow velocity in cerebral arteries and detect High Intensity Transient Signals (HITS). Outputs: M-mode display, spectral data, and flow velocity metrics. Used in clinical settings by healthcare providers. Output allows clinicians to diagnose circulation irregularities. Benefits: non-invasive assessment of cerebral vascular flow.
Clinical Evidence
Bench testing only. Testing included software validation per FDA guidelines, validation of output energies, parameter calculation accuracy, electrical/mechanical safety (IEC 60601-1), and electromagnetic compatibility (IEC 60601-1-2).
Indicated for non-invasive evaluation of intracranial and extracranial vascular flow velocity irregularities in adults and children. Not for fetal or neonatal use.
Regulatory Classification
Identification
An ultrasonic pulsed doppler imaging system is a device that combines the features of continuous wave doppler-effect technology with pulsed-echo effect technology and is intended to determine stationary body tissue characteristics, such as depth or location of tissue interfaces or dynamic tissue characteristics such as velocity of blood or tissue motion. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the seal of the Department of Health & Human Services - USA. The seal is circular, with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the top half of the circle. The bottom half of the circle contains a stylized image of three human profiles facing to the right. The profiles are connected by a flowing line.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
June 23, 2015
Rimed Ltd. % Ms. Ahava Stein Regulatory Consultant 20 Hata'as St. (POB 124) Kfar Saba 44425 ISRAEL
Re: K143476
Trade/Device Name: Digi-One System Regulation Number: 21 CFR 892.1550 Regulation Name: Ultrasonic pulsed doppler imaging system Regulatory Class: II Product Code: IYN, ITX Dated: May 22, 2015 Received: May 28, 2015
Dear Ms. Stein:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Robert A. Ochs
Robert Ochs, Ph.D. Acting Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K143476
Device Name Digi-One System
## Indications for Use (Describe)
The Digi-One System is indicated for non-invasive evaluation of intracranial vascular flow velocity irregularities in adult and in children. It is not intended for fetal use. It is not intended for neonatal use.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.
{3}------------------------------------------------
### Indications for Use Form Fill out one form for each ultrasound system and each transducer. Digi-One System:
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical<br>Application | A | B | C | PWD | CWD | Color<br>Doppler | Power<br>(Amplitude)<br>Doppler | Color<br>Velocity<br>Imaging | Combined<br>(Specify) | Other<br>(Specify) |
|----------------------------------|---|---|---|-----|-----|------------------|---------------------------------|------------------------------|-----------------------|--------------------|
| Ophthalmic | | | | X | X | | | | | |
| Fetal | | | | | | | | | | |
| Abdominal | | | | | | | | | | |
| Intra-operative<br>(Specify) | | | | | | | | | | |
| Intra-operative<br>Neurological | | | | | | | | | | |
| Pediatric | | | | | | | | | | |
| Small Organ<br>(Specify) | | | | | | | | | | |
| Neonatal | | | | | | | | | | |
| Cephalic | | | | | | | | | | |
| Adult Cephalic | | | | X | | | | | | |
| Cardiac | | | | | | | | | | |
| Trans-esophageal | | | | | | | | | | |
| Trans-Rectal | | | | | | | | | | |
| Trans-Vaginal | | | | | | | | | | |
| Trans-Urethral | | | | | | | | | | |
| Intra-Luminal | | | | | | | | | | |
| Peripheral | | | | | | | | | | |
| Vascular | | | | X | X | | | | | |
| Laparoscopic | | | | | | | | | | |
| Musculo-Skeletal<br>Conventional | | | | | | | | | | |
| Muscolo-Skeletal<br>Superficial | | | | | | | | | | |
| Other (Specify) | | | | | | | | | | |
Mode of Oneratio
{4}------------------------------------------------
## Indications for Use Form Fill out one form for each ultrasound system and each transducer. 1 MHZ PW HAND-HELD TRANSDUCER
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Mode of Operation | | | | | | | | | | |
|----------------------------------|---|---|---|-----|-----|------------------|---------------------------------|------------------------------|-----------------------|--------------------|
| Clinical<br>Application | A | B | C | PWD | CWD | Color<br>Doppler | Power<br>(Amplitude)<br>Doppler | Color<br>Velocity<br>Imaging | Combined<br>(Specify) | Other<br>(Specify) |
| Ophthalmic | | | | | | | | | | |
| Fetal | | | | | | | | | | |
| Abdominal | | | | | | | | | | |
| Intra-operative<br>(Specify) | | | | | | | | | | |
| Intra-operative<br>Neurological | | | | | | | | | | |
| Pediatric | | | | | | | | | | |
| Small Organ<br>(Specify) | | | | | | | | | | |
| Cephalic | | | | | | | | | | |
| Adult Cephalic | | | | X | | | | | | |
| Cardiac | | | | | | | | | | |
| Trans-esophageal | | | | | | | | | | |
| Trans-Rectal | | | | | | | | | | |
| Trans-Vaginal | | | | | | | | | | |
| Trans-Urethral | | | | | | | | | | |
| Intra-Luminal | | | | | | | | | | |
| Peripheral<br>Vascular | | | | X | | | | | | |
| Laparoscopic | | | | | | | | | | |
| Musculo-Skeletal<br>Conventional | | | | | | | | | | |
| Muscolo-Skeletal<br>Superficial | | | | | | | | | | |
| Other (Specify) | | | | | | | | | | |
{5}------------------------------------------------
## Indications for Use Form Fill out one form for each ultrasound system and each transducer. 2 MHZ PW HAND-HELD TRANSDUCER
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| | | | | | | Mode of Operation | | | | |
|----------------------------------|---|---|---|-----|-----|-------------------|---------------------------------|------------------------------|-----------------------|--------------------|
| Clinical<br>Application | A | B | C | PWD | CWD | Color<br>Doppler | Power<br>(Amplitude)<br>Doppler | Color<br>Velocity<br>Imaging | Combined<br>(Specify) | Other<br>(Specify) |
| Ophthalmic | | | | X | | | | | | |
| Fetal | | | | | | | | | | |
| Abdominal | | | | | | | | | | |
| Intra-operative<br>(Specify) | | | | | | | | | | |
| Intra-operative<br>Neurological | | | | | | | | | | |
| Pediatric | | | | | | | | | | |
| Small Organ<br>(Specify) | | | | | | | | | | |
| Neonatal | | | | | | | | | | |
| Cephalic | | | | | | | | | | |
| Adult Cephalic | | | | X | | | | | | |
| Cardiac | | | | | | | | | | |
| Trans-esophageal | | | | | | | | | | |
| Trans-Rectal | | | | | | | | | | |
| Trans-Vaginal | | | | | | | | | | |
| Trans-Urethral | | | | | | | | | | |
| Intra-Luminal | | | | | | | | | | |
| Peripheral<br>Vascular | | | | X | | | | | | |
| Laparoscopic | | | | | | | | | | |
| Musculo-Skeletal<br>Conventional | | | | | | | | | | |
| Muscolo-Skeletal<br>Superficial | | | | | | | | | | |
| Other (Specify) | | | | | | | | | | |
{6}------------------------------------------------
## Indications for Use Form Fill out one form for each ultrasound system and each transducer. 4 MHZ TRANSDUCER
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical<br>Application | A | B | C | PWD | CWD | Color<br>Doppler | Power<br>(Amplitude)<br>Doppler | Color<br>Velocity<br>Imaging | Combined<br>(Specify) | Other<br>(Specify) |
|----------------------------------|---|---|---|-----|-----|------------------|---------------------------------|------------------------------|-----------------------|--------------------|
| Ophthalmic | | | | | | | | | | |
| Fetal | | | | | | | | | | |
| Abdominal | | | | | | | | | | |
| Intra-operative<br>(Specify) | | | | | | | | | | |
| Intra-operative<br>Neurological | | | | | | | | | | |
| Pediatric | | | | | | | | | | |
| Small Organ<br>(Specify) | | | | | | | | | | |
| Neonatal | | | | | | | | | | |
| Cephalic | | | | | | | | | | |
| Adult Cephalic | | | | | | | | | | |
| Cardiac | | | | | | | | | | |
| Trans-esophageal | | | | | | | | | | |
| Trans-Rectal | | | | | | | | | | |
| Trans-Vaginal | | | | | | | | | | |
| Trans-Urethral | | | | | | | | | | |
| Intra-Luminal | | | | | | | | | | |
| Peripheral<br>Vascular | | | | X | X | | | | | |
| Laparoscopic | | | | | | | | | | |
| Musculo-Skeletal<br>Conventional | | | | | | | | | | |
| Muscolo-Skeletal<br>Superficial | | | | | | | | | | |
| Other (Specify) | | | | | | | | | | |
Mode of Oneration
{7}------------------------------------------------
## Indications for Use Form Fill out one form for each ultrasound system and each transducer. 8 MHZ TRANSDUCER
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Mode of Operation | | | | | | | | | | |
|----------------------------------|---|---|---|-----|-----|------------------|---------------------------------|------------------------------|-----------------------|--------------------|
| Clinical<br>Application | A | B | C | PWD | CWD | Color<br>Doppler | Power<br>(Amplitude)<br>Doppler | Color<br>Velocity<br>Imaging | Combined<br>(Specify) | Other<br>(Specify) |
| Ophthalmic | | | | | | | | | | |
| Fetal | | | | | | | | | | |
| Abdominal | | | | | | | | | | |
| Intra-operative<br>(Specify) | | | | | | | | | | |
| Intra-operative<br>Neurological | | | | | | | | | | |
| Pediatric | | | | | | | | | | |
| Small Organ<br>(Specify) | | | | | | | | | | |
| Neonatal | | | | | | | | | | |
| Cephalic | | | | | | | | | | |
| Adult Cephalic | | | | | | | | | | |
| Cardiac | | | | | | | | | | |
| Trans-esophageal | | | | | | | | | | |
| Trans-Rectal | | | | | | | | | | |
| Trans-Vaginal | | | | | | | | | | |
| Trans-Urethral | | | | | | | | | | |
| Intra-Luminal | | | | | | | | | | |
| Peripheral<br>Vascular | | | | X | X | | | | | |
| Laparoscopic | | | | | | | | | | |
| Musculo-Skeletal<br>Conventional | | | | | | | | | | |
| Muscolo-Skeletal<br>Superficial | | | | | | | | | | |
| Other (Specify) | | | | | | | | | | |
Mode of Operation
{8}------------------------------------------------
# SUMMARY OF SAFETY AND EFFECTIVENESS
K143476 (Premarket Notification [510(k)] Number)
### 1. Submitter Information
| <b>Manufacturer Name and Address</b> | Rimed Ltd. |
|--------------------------------------|-----------------------------------------------|
| | 25 Hacharoshet St., |
| | PO Box 2402 |
| | Industrial Park Raanana 4365613, |
| | Israel |
| <b>Official Correspondent</b> | Ahava Stein |
| | A. Stein - Regulatory Affairs Consulting Ltd. |
| | 20 Hata'as St. (Beit Hapaamon, Suite 102) |
| | Kfar Saba 44425, |
| | Israel |
2. **Date Prepared:** May 18, 2015
3. **Device Name** Digi-One System
**Proprietary Name:** Digi-One System
Common Name: Transcranial Doppler (TCD) Device
FDA Classification 21 CFR 892.1550; Ultrasonic Pulsed Doppler Imaging System Name:
FDA Classification: Class II, Product Code IYN; ITX
#### 4. Predicate Devices
The Digi-One System is substantially equivalent to the following devices:
| Manufacturer | Device | 510(k) | Date Cleared |
|--------------|------------------|---------|-----------------|
| Rimed Ltd. | Digi-Lite System | K062578 | August 23, 2006 |
### ട. Device Description
The Digi-One System is a small, lightweight, portable digital Transcranial Doppler (TCD) system with an advanced M-mode display which can be connected to any external Windows based PC.
{9}------------------------------------------------
It measures the blood flow velocity in the main cerebral arteries non-invasively for circulation diagnosis and HITS (High Intensity Transient Signals) detection using the same Rimed TCD software application as the predicate device.
The Digi-One System is a modification of the Digi-Lite System. The hardware was modified using current electronic components, and the software was modified to allow installation on a dedicated PC (such as a laptop device). Otherwise, the functionality of the system remains identical to that of the Digi-Lite system in all aspects, including acoustic power outputs, clinical parameter measurement accuracy, examination modes, available transducers, software menus and commands (with some minor user interface modifications). As the specifications remain identical, the safety and efficacy of the device are not affected when used as labeled. Therefore, the Digi-One system remains substantially equivalent to the predicate, un-modified Digi-Lite System.
| Technological<br>Characteristic | Digi-One System<br>Rimed Ltd. | Digi-Lite System<br>Rimed Ltd.<br>(K062578) |
|---------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Product Code,<br>Class | IYN, ITX<br>Class II | IYN, ITX<br>Class II |
| Indications for<br>Use | The Digi-One System is<br>indicated for non-invasive<br>evaluation of intracranial and<br>extracranial vascular flow<br>velocity irregularities in adults<br>and in children. It is not<br>intended for fetal use. It is not<br>intended for neonatal use. | The Digi-Lite System is<br>indicated for non-invasive<br>evaluation of intracranial and<br>extracranial vascular flow<br>velocity irregularities in adults<br>and in children. It is not<br>intended for fetal use. It is not<br>intended for neonatal use. |
| Energy Used /<br>Delivered | Ultrasound energy | Ultrasound energy |
| Design:<br>(Modification<br>from predicate) | The Digi-One System consists<br>of a small unit which connects<br>to a PC and to which probes are<br>connected. | The Digi-Lite System consists<br>of a unit with a display to<br>which probes are connected. |
| - Mechanism of<br>Action | Doppler Ultrasound, with the<br>following modes:<br>Unilateral<br>Bilateral<br>Multichannel<br>Monitoring<br>Multidepth (up to 8 spectrums<br>at various depths)<br>M-Mode (64 gates) | Doppler Ultrasound, with the<br>following modes:<br>Unilateral<br>Bilateral<br>Multifrequency<br>Multichannel<br>Monitoring<br>Multidepth (up to 8 spectrums<br>at various depths)<br>M-Mode (64 gates) |
Predicate Comparison Table:
{10}------------------------------------------------
| Technological<br>Characteristic | Digi-One System<br>Rimed Ltd. | Digi-Lite System<br>Rimed Ltd.<br>(K062578) |
|---------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| - Accessories | Probe types:<br>- 1MHz PW, 16mm<br>- 2MHz PW, 14mm<br>- 4MHz PW/CW, 8mm<br>- 8MHz PW/CW, 5mm<br>Remote control<br>Monitoring probe holder<br>Color printer Footswitch | Probe types:<br>- 1MHz PW, 16mm<br>- 2MHz PW, 14mm<br>- 4MHz PW/CW, 8mm<br>- 8MHz PW/CW, 5mm<br>Remote control<br>Monitoring probe holder<br>Color printer Footswitch<br>CD ROM backup<br>DAT Tape |
| Performance | | |
| Sample volume | (for 1 MHz) (for 2 MHz)<br>15-92 mm 15-146 mm | (for 1 MHz) (for 2 MHz)<br>15-92 mm 15-146 mm |
| Depth | | |
| Power (%) | 0-100% at 7% steps | 0-100% at 7% steps |
| Acoustic output | 1MHz PW Ispta=34.15mW/cm2<br>2MHz PW Ispta=181.3mW/cm2<br>4MHz PW Ispta=232.2mW/cm2<br>4MHz CW Ispta=183mW/cm2<br>8MHz PW Ispta=149.8mW/cm2<br>8MHz CW Ispta=414mW/cm2 | 1MHz PW Ispta=34.15mW/cm2<br>2MHz PW Ispta=181.3mW/cm2<br>4MHz PW Ispta=232.2mW/cm2<br>4MHz CW Ispta=183mW/cm2<br>8MHz PW Ispta=149.8mW/cm2<br>8MHz CW Ispta=414mW/cm2 |
### Indications for Use 6.
The Digi-One System is indicated for non-invasive evaluation of intracranial and extracranial vascular flow velocity irregularities in adults and in children. It is not intended for fetal use. It is not intended for neonatal use.
### 7. Performance Standards
There are no performance standards under the Federal Food, Drug and Cosmetic Act, for the Digi-One System.
### 8. Performance Testing
The following Software and Performance tests were performed on the Digi-One System:
- Software Validation according to FDA Guidelines (tested modified ● software)
- Bench testing to validate output energies and parameter calculation accuracies
- Electrical and Mechanical Safety Standard (IEC 60601-1)
- Electromagnetic Compatibility Standard (IEC60601-1-2)
### 9. Technological Characteristics Compared to Predicate Device
The original Digi-Lite device underwent modification to electronics and software. The safety and effectiveness questions that were raised by these changes were
{11}------------------------------------------------
equivalence of acoustic power outputs, equivalence of measurement accuracy, and safe use under low-power modes. Performance testing of the modified device included measurements of output power and measurement accuracy. The testing results demonstrated that the modified device is substantially equivalent to the predicate, unmodified device. The software functions and testing modes did not change, except for some User Interface improvements; the system defaults for low-power modes were reviewed for equivalence to the predicate device, and were demonstrated to remain identical. Software testing was performed on the modified software. The technological characteristics, e.g., overall design, materials, mechanism of action, mode of operation, performance characteristics, etc., and the indications for use of the modified Digi-One System are substantially equivalent to the predicate device cited above. Based on the results of performance testing, the Digi-One device is substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.