K143475 · Depuy Mitek · GEI · Apr 3, 2015 · General, Plastic Surgery
Device Facts
Record ID
K143475
Device Name
VAPR Tripolar Suction Electrode
Applicant
Depuy Mitek
Product Code
GEI · General, Plastic Surgery
Decision Date
Apr 3, 2015
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4400
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The VAPR Electrodes, for use with the VAPR VUE Radiofrequency System, are intended for resection, ablation and excision of soft tissue, hemostasis of blood vessels and coagulation of soft tissues in patients requiring arthroscopic surgery.
Device Story
VAPR Tripolar 90 Suction Electrode is a single-use, bipolar RF probe for arthroscopic surgery. Device features three electrodes (one active, two return) in probe tip to optimize ablation and coagulation. Integrated vacuum fluid extraction removes bubbles and debris from operating site; flow rate is adjustable. Device connects to VAPR VUE Radiofrequency Generator; internal identification code automatically configures generator to optimal default power and waveforms. Surgeon operates device via ergonomic hand-control buttons. Used in surgical settings by trained arthroscopic surgeons. Output is RF energy delivered to joint space for tissue management. Benefits include efficient tissue resection/hemostasis and improved visualization via debris removal.
Clinical Evidence
Bench testing only. No clinical or animal studies were conducted. Performance testing included tissue tests, saline outflow temperature, suction control, thermal margin, clogging frequency, tip compression/side load, three-point bend, fluid ingress, system compatibility, thermal shock, and in-joint/surface contact temperature tests. All results were acceptable.
Technological Characteristics
Bipolar RF electrosurgical probe; single-use; 155mm working length. Features three-electrode configuration (1 active, 2 return). Integrated suction channel with variable flow. Connects to VAPR VUE generator via proprietary plug with internal ID code for automatic parameter setting. Sterilized via gamma irradiation. Complies with IEC 60601-1 and IEC 60601-2-2.
Indications for Use
Indicated for patients requiring arthroscopic surgery for soft tissue resection, ablation, excision, hemostasis of blood vessels, and coagulation of soft tissues.
Regulatory Classification
Identification
An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized image of three human profiles facing right, stacked slightly offset from each other. The profiles are simple and abstract, with a flowing design. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the image.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
April 3, 2015
DePuy Mitek Incorporated Ms. Susan Kagan Project Manager 325 Paramount Drive Raynham, Massachusetts 02067
Re: K143475
Trade/Device Name: VAPR® Tripolar 90" Suction Electrode Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical cutting and coagulation device and accessories Regulatory Class: Class II Product Code: GEI Dated: March 5, 2015 Received: March 6, 2015
Dear Ms. Kagan:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{1}------------------------------------------------
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
# Jennifer R. Stevenson -S
For Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
# Indications for Use
510(k) Number (if known) K143475
Device Name VAPR® Tripolar90™ Suction Electrode
Indications for Use (Describe)
The VAPR Electrodes, for use with the VAPR VUE Radiofrequency System, are intended for resection, ablation and excision of soft tissue, hemostasis of blood vessels and coagulation of soft tissues in patients requiring arthroscopic surgery.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
] Over-The-Counter Use (21 CFR 801 Subpart C)
### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
### FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
{3}------------------------------------------------
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
### 510(k) SUMMARY
#### l. SUBMITTER
DePuy Mitek, Inc. a Johnson & Johnson company 325 Paramount Drive Raynham, MA 02767
Telephone: 508-880-8097 Facsimile: 508-977-6955 e-mail: skagan@its.jnj.com
Susan Kagan Project Manager, Requlatory Affairs
Date Prepared: December 4, 2014
#### II. DEVICE
| Name of Device: | VAPR® TRIPOLAR 90™ Suction Electrode |
|----------------------|------------------------------------------------------------------------------------|
| Common/Usual Name: | Electrosurgical cutting and coagulation device and accessories |
| Classification Name: | Electrosurgical cutting and coagulation device and accessories:<br>21 CFR 878.4400 |
| FDA Classification: | II |
| FDA product code: | GEI |
#### PREDICATE lll.
Primary Predicate: CoolPulse 90 Electrode-K113545/ K100638 Reference Device: ArthroCare Super Turbovac-K120791
These predicate devices have not been the subject of a design related recall.
#### DEVICE DESCRIPTION IV.
The VAPR TRIPOLAR 90 Suction Electrode is a single use, one-piece bipolar RF suction electrode for use in a surgical setting by a surgeon trained in arthroscopic surgery.
The VAPR TRIPOLAR 90 Suction Electrode is an RF device which incorporates 3 electrodes in the probe tip: one active electrode and two return electrodes. This configuration optimizes ablation and coagulation in a single device.
{5}------------------------------------------------
It has been designed to facilitate access and control the delivery of RF energy to the joint space. The connector plug of the VAPR TRIPOLAR 90 Suction Electrode is designed to interface only with the VAPR VUE Radiofrequency Generator. It has an internal identification code which automatically adjusts the VAPR VUE Generator to the optimal default and accessible powers and waveforms. If required, the settings for the VAPR TRIPOLAR 90 Suction Electrode can be modified within safe pre-determined limits by accessing the generator or footswitch control. It is intended to be only operated with the VAPR VUE generator at pre-determined default settings specific for this device.
The VAPR TRIPOLAR 90 Suction Electrode has three sets of buttons to address surgeon preference when holding the electrode. Each pair of buttons offers the same functionality and are ergonomically positioned for surgeon comfort.
In addition the VAPR TRIPOLAR 90 Suction Electrode features vacuum fluid extraction which enhances the efficiency of the electrode and assists in the removal of bubbles and debris created during activation within the operating site. The design includes the ability to vary flow rate to allow the surgeon to make adjustments depending on particular procedural requirements.
#### V. INDICATION FOR USE
The VAPR Electrodes, for use with the VAPR VUE Radiofrequency System, are intended for resection, ablation and excision of soft tissue, hemostasis of blood vessels and coaqulation of soft tissues in patients requiring arthroscopic surgery.
The Indication for Use statement for the VAPR TRIPOLAR 90 Suction Electrode is not identical to the predicate devices however both the subject device and the predicates devices are intended for resection, ablation and excision of soft tissue, hemostasis of blood vessels and coagulation of soft tissues in patients requiring arthroscopic surgery.
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image displays the logo for DePuy Synthes Mitek Sports Medicine. The logo features a circular graphic in gold on the left, followed by the text "DePuy Synthes" in a dark blue sans-serif font. Below that, "MITEK SPORTS MEDICINE" is written in smaller, uppercase letters. At the bottom, it says "COMPANIES OF Johnson & Johnson" in a smaller font.
### VI. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PRIMARY PREDICATE DEVICE
| Component | VAPR CoolPulse90 Electrode | VAPR TRIPOLAR 90<br>Suction Electrode |
|------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| FDA Clearance | Predicate Device K100638/K113545 | Subject Device |
| Indication for Use | The VAPR Electrodes for use with the<br>VAPR VUE Radiofrequency System<br>are intended for resection, ablation<br>and excision of soft tissue,<br>hemostasis of blood vessels and<br>coagulation of soft tissue in patients<br>requiring arthroscopic surgery of the<br>knee, shoulder, hip, ankle, elbow,<br>and wrist. | The VAPR Electrodes, for use<br>with the VAPR VUE<br>Radiofrequency System, are<br>intended for resection,<br>ablation and excision of soft<br>tissue, hemostasis of blood<br>vessels and coagulation of<br>soft tissues in patients<br>requiring arthroscopic<br>surgery. |
| Working Length | 160mm | 155mm |
| Tip Geometry | oval tip w/suction slot & partial<br>peripheral suction | near oval tip w/suction &<br>partial peripheral suction |
| Integrated Cable and<br>Plug | Yes | Yes |
| Integrated Hand<br>Control | Yes | Yes |
| Integrated Suction | Yes | Yes |
| RF Energy | Bipolar | Bipolar |
| Sterilization | Gamma Irradiation | Gamma Irradiation |
#### PERFORMANCE DATA VII.
Biocompatibility
The biocompatibility evaluation of the VAPR TRIPOLAR 90 Suction Electrode was conducted in accordance with FDA Blue Book Memorandum # G95-1 "Use of International Standard ISO-10993, Biological Evaluation of Medical Devices: Part 1: Evaluation and Testing," May 1, 1995 and International Standard ISO 10993-1 "Biological Evaluation of Medical Devices Part 1: Evaluation and Testing Within a Risk Management Process," as recognized by FDA. The battery of testing included the following tests:
- Cytotoxicity
- . Sensitization
- . Irritation
- Systemic toxicity ●
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image displays the logo for DePuy Synthes Mitek Sports Medicine, a company of Johnson & Johnson. The DePuy Synthes logo is in a dark blue font, with the words "Mitek Sports Medicine" in a smaller font below. The Johnson & Johnson logo is in red and is located below the DePuy Synthes logo. To the left of the DePuy Synthes logo is a circular graphic with overlapping shapes.
The VAPR TRIPOLAR 90 Suction Electrode is considered an external communicating device used for a duration of less than 24 hours.
# Electrical safety
Electrical Safety was conducted on the VAPR TRIPOLAR 90 Suction Electrode when used with the VAPR VUE generator. The system complies with IEC 60601-1 "Medical electrical equipment -- Part 1: General requirements for safety and IEC 60601-2-2 "Medical electrical equipment -Part 2-2: Particular requirements for the basic safety and essential performance of high frequency surgical equipment and high frequency surgical accessories"
### Software Status
The VAPR TRIPOLAR 90 Suction Electrode is compatible only with the VAPR VUE generator. The generator was previously programed to recognize the VAPR TRIPOLAR 90 Suction Electrode so no software update was required.
### Performance Data
The following performance data were provided in support off the substantial equivalence determination.
| TEST | RESULTS |
|------------------------------------|------------|
| • Tissue Tests | Acceptable |
| • Saline Outflow Temperature Test | Acceptable |
| • Fingerswitching Tests | Acceptable |
| • Suction Control | Acceptable |
| • Area of Thermal Margin | Acceptable |
| • Clogging Frequency | Acceptable |
| • Active Tip Compression | Acceptable |
| • Active Tip Side Load | Acceptable |
| • Three Point Bend Test | Acceptable |
| • Fluid Ingress | Acceptable |
| • System Compatibility | Acceptable |
| • Thermal Shock | Acceptable |
| • In-joint Temperature Test | Acceptable |
| • Surface Contact Temperature Test | Acceptable |
### Animal Study
No animal studies are required to demonstrate safety and efficacy of the device in support of an application for premarket clearance. The VAPR TRIPOLAR 90 Suction Electrode does not differ from the predicate device in fundamental scientific technology or intended use.
### Clinical Study
No clinical studies are required to demonstrate safety and efficacy of the device in support of an application for premarket clearance. The VAPR TRIPOLAR 90 Suction Electrode does not differ from the predicate device in fundamental scientific technology or intended use.
{8}------------------------------------------------
Image /page/8/Picture/0 description: The image contains the logo for DePuy Synthes Mitek Sports Medicine. The logo consists of a circular design on the left, followed by the text "DePuy Synthes" in a bold, sans-serif font. Below that, the text "MITEK SPORTS MEDICINE" is displayed in a smaller, sans-serif font. At the bottom, the text "COMPANIES OF Johnson & Johnson" is displayed.
# VIII. CONCLUSION
Results of performance and safety testing have demonstrated that the modified device is suitable for its intended use.
Based on the indications for use, fundamental scientific technology, and comparison to the predicate devices, the VAPR TRIPOLAR 90 Suction Electrode is shown to be substantially equivalent to the predicate devices under the Federal Food, Drug and Cosmetic Act.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.