← Product Code [ODP](/productcode/ODP) · K143450

# Ackermann Intervertebral Body Fusion Device (K143450)

_Ackermann Instrumente GmbH · ODP · Sep 16, 2015 · Orthopedic · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K143450

## Device Facts

- **Applicant:** Ackermann Instrumente GmbH
- **Product Code:** [ODP](/productcode/ODP.md)
- **Decision Date:** Sep 16, 2015
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 888.3080
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Indications for Use

When used as a cervical intervertebral body fusion device, the Ackermann Intervertebral Body Fusion Device cispine implants are indicated for spinal fusion procedures in skeletally mature patients. Cervical interbody fusion implants are intended for use at one level in the cervical spine, from C2 to T1, for treatment of cervical degenerative disc disease (defined as neck pain of discogenic origin with degeneration of the disc confirmed by history and radies). The cervical device is intended to be used in patients who have had six weeks of non-operative treatment. Additionally, the clspine implants are to be used with autogenous bone graft and supplemental fixation via an open, anterior approach. When used as a lumbar intervertebral body fusion device, the Ackermann Intervertebral Body Fusion Device tispine implants are indicated for spinal fusion procedures in skeletally mature patients. Lumbar interbody fusion implants are intended for use at either one level or two contiguous levels in the lumbar spine, from L2 to S1, for treatment of lumbar degenerative disc disease (DDD) with up to Grade 1 spondylolisthesis. DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. The lumbar device is intended to be used in patients who have had six months of non-operative treatment. The tlspine implants may be implanted via an open or a minimally invasive posterior approach. Alternatively, these implanted via a transforaminal approach. These implants are to be used with autogenous bone graft and supplemental fixation.

## Device Story

Intervertebral body fusion device (c|spine and t|spine implants) designed to improve spinal stability and support fusion. Implants are inserted into the vertebral body space; used with autogenous bone graft and supplemental fixation. Cervical implants utilize an open, anterior approach; lumbar implants utilize open, minimally invasive posterior, or transforaminal approaches. Components available in various shapes/sizes to match patient anatomy. Device provides mechanical support to the intervertebral space to facilitate bone fusion, potentially reducing pain associated with degenerative disc disease.

## Clinical Evidence

Bench testing only. Testing included static axial compression, dynamic axial compression, static torsion, dynamic torsion (ASTM 2077), and load-induced subsidence (ASTM 2267).

## Technological Characteristics

Implant grade PEEK (VESTAKEEP i4 R bv Evonik). Mechanical testing per ASTM 2077 (compression/torsion) and ASTM 2267 (subsidence). Form factor: intervertebral body spacers in various shapes/sizes.

## Regulatory Identification

An intervertebral body fusion device is an implanted single or multiple component spinal device made from a variety of materials, including titanium and polymers. The device is inserted into the intervertebral body space of the cervical or lumbosacral spine, and is intended for intervertebral body fusion.

## Special Controls

*Classification.* (1) Class II (special controls) for intervertebral body fusion devices that contain bone grafting material. The special control is the FDA guidance document entitled “Class II Special Controls Guidance Document: Intervertebral Body Fusion Device.” See § 888.1(e) for the availability of this guidance document.(2) Class III (premarket approval) for intervertebral body fusion devices that include any therapeutic biologic (e.g., bone morphogenic protein). Intervertebral body fusion devices that contain any therapeutic biologic require premarket approval.
(c)
*Date premarket approval application (PMA) or notice of product development protocol (PDP) is required.* Devices described in paragraph (b)(2) of this section shall have an approved PMA or a declared completed PDP in effect before being placed in commercial distribution.

## Predicate Devices

- Amedica Valeo II Interbody Fusion Device ([K142347](/device/K142347.md))
- Aesculap CeSpace PEEK Spinal Implant System ([K083311](/device/K083311.md))
- Aesculap PEEK Spinal Implant System ([K071983](/device/K071983.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is circular and features the department's name around the perimeter. In the center is an abstract image of a bird-like figure, composed of three human profiles facing to the right.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

September 16, 2015

Ackermann Instrumente GmbH % Mr. Frank Ferguson Chief Executive Officer Ferguson Medical International Device Consultants, LLC 332 Laskin Road, Suite 437 Virginia Beach, Virginia 23451

Re: K143450

Trade/Device Name: Ackermann Intervertebral Body Fusion Device Regulation Number: 21 CFR 888.3080 Regulation Name: Intervertebral body fusion device Regulatory Class: Class II Product Code: MAX, ODP Dated: August 14, 2015 Received: August 17, 2015

Dear Mr. Ferguson:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set

{1}------------------------------------------------

#### Page 2 - Mr. Frank Ferguson

forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

# Mark N. Melkerson -S

Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

### Indications for Use

510(k) Number (if known) K143450

#### Device Name

Ackermann Intervertebral Body Fusion Device

#### Indications for Use (Describe)

When used as a cervical intervertebral body fusion device, the Ackermann Intervertebral Body Fusion Device cispine implants are indicated for spinal fusion procedures in skeletally mature patients. Cervical interbody fusion implants are intended for use at one level in the cervical spine, from C2 to T1, for treatment of cervical degenerative disc disease (defined as neck pain of discogenic origin with degeneration of the disc confirmed by history and radies). The cervical device is intended to be used in patients who have had six weeks of non-operative treatment. Additionally, the clspine implants are to be used with autogenous bone graft and supplemental fixation via an open, anterior approach.

When used as a lumbar intervertebral body fusion device, the Ackermann Intervertebral Body Fusion Device tispine implants are indicated for spinal fusion procedures in skeletally mature patients. Lumbar interbody fusion implants are intended for use at either one level or two contiguous levels in the lumbar spine, from L2 to S1, for treatment of lumbar degenerative disc disease (DDD) with up to Grade 1 spondylolisthesis. DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. The lumbar device is intended to be used in patients who have had six months of non-operative treatment. The tlspine implants may be implanted via an open or a minimally invasive posterior approach. Alternatively, these implanted via a transforaminal approach. These implants are to be used with autogenous bone graft and supplemental fixation.

| Type of Use (Select one or both, as applicable)                                      |                                                                      |
|--------------------------------------------------------------------------------------|----------------------------------------------------------------------|
| <input checked="true" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D) | <input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C) |

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

finest surgical technology

# Ackermann®

## 510(k) SUMMARY

This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of 21 CFR Part 807.92 for the Ackermann Medical Intervertebral Body Fusion Device c|spine and t|spine implants.

DATE PREPARED: 13 August 2015

### APPLICANTS NAME AND ADDRESS:

Ackermann Instrumente GmbH Attn: Rolf Ackermann, CEO Eisenbahnstrasse 65-67 78604 Rietheim-Weilheim Germanv Phone: +49 7461 966170; FAX: +49 7461 9661770 RA@Ackermanninstrumente.de

#### APPLICANTS CONTACT PERSON IN THE USA:

Ferguson Medical International Device Consultants LLC Attn: Frank Ferguson, CEO 332 Laskin Road Suite 437 Virginia Beach, VA 23451 Phone: 757.286.4383; FAX: 801.749.0451 frankf@fergusonmedical-idc.com

{4}------------------------------------------------

#### DEVICE NAME:

Trade Name: Ackermann Intervertebral Body Fusion Device Common Name: Intervertebral Body Fusion Device Classification Name: Intervertebral Body Fusion Device, 21 CFR 888.3080 Product Code: MAX, ODP

## LEGALLY MARKETED DEVICES TO WHICH ACKERMANN IS CLAIMING SUBSTANTIAL EQUIVALENCE:

Primary Predicate Device: Amedica Valeo II Interbody Fusion Device (K142347) Additional Predicate Devices: Aesculap CeSpace PEEK Spinal Implant System (K083311) and Aesculap PEEK Spinal Implant System (K071983)

## DEVICE DESCRIPTION:

Ackermann Intervertebral Body Fusion Device c|spine and t|spine is an intervertebral body fusion device that is implanted into the vertebral body space to improve stability of the spine while supporting fusion. Components are offered in a variety of shapes and sizes to meet the requirements of the individual patient anatomy. Components are manufactured from implant grade PEEK (polyetheretherketone, VESTAKEEP® i4 R bv Evonik).

#### INTENDED USE:

When used as a cervical intervertebral body fusion device, the Ackermann Intervertebral Body Fusion Device c|spine implants are indicated for spinal fusion procedures in skeletally mature patients. Cervical interbody fusion implants are intended for use at one level in the cervical spine, from C2 to T1, for treatment of cervical degenerative disc disease (defined as neck pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies). The cervical device is intended to be used in patients who have had six weeks of non-operative treatment. Additionally, the c|spine implants are to be used with autogenous bone graft and supplemental fixation via an open, anterior approach.

When used as a lumbar intervertebral body fusion device, the Ackermann Intervertebral Body Fusion Device t|spine implants are indicated for spinal fusion procedures in skeletally mature patients. Lumbar interbody fusion implants are intended for use at either one level or two contiquous levels in the lumbar spine, from L2 to S1, for treatment of lumbar degenerative disc disease (DDD) with up to Grade 1 spondylolisthesis. DDD is defined as back pain of discogenic origin with degeneration of 

{5}------------------------------------------------

the disc confirmed by history and radiographic studies. The lumbar device is intended to be used in patients who have had six months of non-operative treatment. The t|spine implants may be implanted via an open or a minimally invasive posterior approach. Alternatively, these implants may be implanted via a transforaminal approach. These implants are to be used with autogenous bone graft and supplemental fixation.

## SUMMARY OF THE TECHNOLOGICAL CHARACTERISTICS OF THE DEVICE COMPARED TO THE PREDICATE DEVICE:

The components of the Ackermann Intervertebral Body Fusion Device c|spine and t|spine implants are offered in the same general range of sizes and shapes of the predicate devices. The material used for the Ackermann device is the same as that used to manufacture the reference predicate.

## PERFORMANCE DATA:

The Ackermann Intervertebral Body Fusion Device c|spine implants were tested for static axial compression, dynamic axial compression, static torsion, and dynamic torsion in accordance with ASTM 2077 and for load induced subsidence in accordance with ASTM 2267. The t|spine implants were tested for static axial compression and dynamic axial compression in accordance with ASTM 2077 and for load induced subsidence in accordance with ASTM 2267.

#### CONCLUSIONS:

Based upon the testing and comparison to the predicate device, the Ackermann Intervertebral Body Fusion Device c|spine and t|spine implants have the same intended use and similar technological characteristics. Therefore, the device is substantially equivalent to other legally marketed devices.

---

**Source:** [https://fda-staging.innolitics.com/device/K143450](https://fda-staging.innolitics.com/device/K143450)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com