K143430 · Easy Med Instrument Co., Ltd. · NUH · May 29, 2015 · Neurology
Device Facts
Record ID
K143430
Device Name
SmartTENS
Applicant
Easy Med Instrument Co., Ltd.
Product Code
NUH · Neurology
Decision Date
May 29, 2015
Decision
SESE
Submission Type
Abbreviated
Regulation
21 CFR 882.5890
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
This device is intended for the relief of pain associated with sore or aching muscles of the lower back, arms, or legs due to strain from exercise or normal household and work activities.
Device Story
SmartTENS is a portable, battery-powered T.E.N.S. device for OTC use in home, clinic, or hospital settings. It delivers electrical stimulation to muscles via button-affixed electrodes with hydrogel pads. The device features three onboard control buttons (ON/OFF, intensity adjustment) and supports wireless control via Bluetooth 4.0 (2402-2480 MHz) using a dedicated mobile application on a smartphone or tablet. A 6-digit security code pairs the device to the user's smartphone to prevent unauthorized access. The device offers seven treatment modes for the shoulder/arm, lower back, and leg/foot. It includes automatic overload, no-load, and shut-off safety features. The healthcare provider or patient uses the output to manage muscle pain; the device provides symptomatic relief through electrical stimulation.
Clinical Evidence
Bench testing only. Performance testing included electrical safety (AAMI/ANSI ES60601-1), electromagnetic compatibility (IEC 60601-1-2), home healthcare environment requirements (IEC 60601-1-11), and nerve/muscle stimulator safety (IEC 60601-2-10). A usability study was conducted, demonstrating that users could operate the device correctly and safely.
Technological Characteristics
Materials: TPE and ABS. Sensing/Actuation: Electrical nerve stimulation via biphasic/monophasic rectangular waveforms. Energy: 3.7V rechargeable Li-Ion battery. Connectivity: Bluetooth 4.0 BLE. Dimensions: 155.4 x 64.6 x 19.1 mm. Weight: 64g. Standards: IEC 60601-1, IEC 60601-1-2, IEC 60601-1-11, IEC 60601-2-10, ISO 14971, IEC 62304.
Indications for Use
Indicated for adult patients seeking relief of pain associated with sore or aching muscles of the lower back, arms, or legs due to strain from exercise or normal household and work activities.
Regulatory Classification
Identification
A transcutaneous electrical nerve stimulator for pain relief is a device used to apply an electrical current to electrodes on a patient's skin to treat pain.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized image of three human profiles facing right, with flowing lines suggesting hair or movement. The profiles are arranged in a way that they appear to be connected. Encircling the image is the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" in a circular arrangement.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
May 29, 2015
Easymed Instruments Co., Ltd Mr. Jeffery Wu (Wu Tingjie) General Manager 5/F - 6/F, Block A, Gupo Gongmao Building, Fengxin Road, Fengxiang Industrial District, Daliang, 528300 Shunde, Foshan, Guangdong, CHINA
Re: K143430
Trade/Device Name: SmartTENS Regulation Number: 21 CFR 882.5890 Regulation Name: Stimulator, Nerve, Transcutaneous, Over-The-Counter Regulatory Class: Class II Product Code: NUH Dated: April 30, 2015 Received: May 4, 2015
Dear Mr. Tingjie Wu:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{1}------------------------------------------------
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
# Felipe Aquel -S
Carlos L. Peña, PhD, MS for Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K143430
Device Name SmartTENS
Indications for Use (Describe)
This device is intended for the relief of pain associated with sore or aching muscles of the lower back, arms, or legs due to strain from exercise or normal household and work activities.
Type of Use (Select one or both, as applicable)
| <input type="checkbox"/> Registration under Part 21 CFR 601 Subpart E |
|--------------------------------------------------------------------------------------|
| <input checked="" type="checkbox"/> Request for Exception under 21 CFR 601 Subpart G |
| | Prescription Use (Part 21 CFR 801 Subpart D)
|X | Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the Easymed logo. The logo is in blue and has a red line underneath it. Below the red line is the text "Easy for the relief of your suffering".
# 510(k) Summary
Date of submission prepared: October 6, 2014
| <b>Submitter:</b> | EasyMed Instruments Co., Ltd. |
|-------------------|------------------------------------------------------------------------------------------------------------------------------------------------|
| <b>Address:</b> | 5/F – 6/F, Block A, Gupo Gongmao Building, Fengxin Road,<br>Fengxiang Industrial District, Daliang, 528300 Shunde, Foshan,<br>Guangdong, CHINA |
| <b>Tel:</b> | +86 757 2809 1300 |
| <b>Fax:</b> | +86 757 2772 7868 |
| <b>E-Mail:</b> | jeffery@easymed.com.cn |
- Official Contact: Jeffery Wu (Wu Tingjie)
Address of the manufacturing facility: The same as above
# SUBMITTED DEVICE:
| Generic Name: | Transcutaneous Electrical Nerve Stimulator (T.E.N.S.) |
|----------------------------|------------------------------------------------------------------------|
| Proprietary or Trade Name: | SmartTENS |
| Common/Usual Name: | Stimulator, nerve, transcutaneous, over-the-counter |
| Classification Name: | Stimulator, nerve, transcutaneous, over-the-counter<br>21 CFR 882.5890 |
| Product Code: | NUH |
| Device Panel: | Neurology |
| Device Classification: | Class II |
# PREDICATE DEVICES:
| Device Name: | EasyStim TN28_OTC |
|----------------|-------------------------------|
| Manufacturer: | EasyMed Instruments Co., Ltd. |
| 510(k) Number: | K140168 |
| Product Code: | NUH |
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the Easymed logo. The word "Easymed" is written in a large, bold, blue font. There is a red line underneath the word "Easymed". Below the red line, the words "Easy for the relief of your suffering" are written in a smaller, black font. There is a registered trademark symbol to the right of the word "Easymed".
WiTouch™ Pro Device Name: Manufacturer: Hollywog, LLC 510(k) Number: K120398 Product Code: GZJ. NYN
## INDICATIONS FOR USE:
This device is intended for the relief of pain associated with sore or aching muscles of the lower back, arms, or legs due to strain from exercise or normal household and work activities.
## DEVICE DESCRIPTION:
SmartTENS is a portable battery powered T.E.N.S. device with a wireless remote control feature for over-the-counter use, mainly operating in home, clinic, or hospital environments.
There are three control buttons on the unit itself: ON/OFF, intensity UP and intensity DOWM. The device is also designed to be controlled wirelessly through Bluetooth 4.0 connection (Frequency: 2402.0 - 2480.0 MHz) by a matched smart phone, which is featuring a mobile application (App) developed by EasyMed Instruments Co., Ltd.
A user entered 6-digit security code ensures the access to the device can only be the user's matched smart phone, thus avoiding unauthorized controls to the device.
There are totally seven (7) modes that are intended for pain relieving treatments to the following areas: Shoulder/ Arm, Lower Back and Leg/Foot. The specifications of each mode will be described in greater details in the User Manual.
#### Accessories:
| M-CBX-C-008 | Button-affixed Electrode |
|-------------|--------------------------------------------------|
| P-AE-002-A | Replaceable Hydrogel Pads (gel-pads) 73mm x 62mm |
| | for external use. Pack of 2* |
| CBX-057-B | Charger Cradle |
| M-PA-004-B | Charging Adaptor |
The accessories include an electrode and two hydrogel pads (gel-pads). The electrode is connected to the device by snapping the two buttons onto the device. One side of the adhesive gel-pads adheres to the electrode, and the other side adhere the device to the healthy intact skin of the user's body. Generally, the device produces either a fixed or modulated electrical signal through electrodes normally placed over the area of pain.
5
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for Easymed. The logo is in blue and has a red line underneath it. Below the red line is the text "Easy for the relief of your suffering".
# EasyMed Instruments Co., Ltd.
A built in Li-Ion battery is the power source of the device, which can be recharged for a limited large number of times. To charge the device, remove the electrode assembly by unsnapping it off the device, place the device on the charger cradle, insert the small pin of the charging adaptor into the charge cradle, and connect the charging adaptor to the electrical power supply.
## Smart Phone Hardware Requirement:
- 1) iPhone 4s or later models, or iPad 3 or later models, with iOS 6.0 or better system. Or,
- 2) Smart phone or Tablet supporting BT 4.0 BLE feature with Android 4.3 or better system.
As above, the device is battery powered; there is no connection to AC mains supply.
ENVIRONMENT OF USE: Clinics, hospital and home environments
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the brand name "Easymed" in a large, bold, blue font. Below the brand name, there is a thin red line. Underneath the line, the text "Easy for the relief of your suffering" is written in a smaller, blue font.
# SUMMARY OF SUBSTANTIAL EQUIVALENCE
| Attribute | New Device | Predicate Device (1) | Predicate Device (2) |
|--------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Product Name | SmartTENS | EasyStim TN28_OTC | WiTouch™ Pro |
| 510(K) number | K143430 | K140168 | K120398 |
| Product Code | NUH | NUH | GZJ, NYN |
| Regulation No. | 21 CFR 882.5890 | 21 CFR 882.5890 | CFR Title 21, 882.5890 |
| Indications for Use | This device is intended for the relief of pain associated with sore or aching muscles of the lower back, arms, or legs due to strain from exercise or normal household and work activities. | This device is intended for the relief of pain associated with sore or aching muscles of the lower back, arms, or legs due to strain from exercise or normal household and work activities. | This device is used for the symptomatic relief and management of chronic intractable back pain and relief of pain of the upper and lower back associated with arthritis. It is also used for adjunctive treament fro post-surgical and post-trauma acute back pain. |
| Patient Population | Adult | Adult | Not specified |
| Prescriptive or OTC | OTC | OTC | Prescription |
| Wireless Control | Yes | No | Yes |
| Number of output modes | 7 | 8 | 4 |
| Number of output channels | 1 | 2 | 1 |
| Waveform | Biphasic rectangular<br>Monophasic rectangular | Biphasic rectangular<br>Monophasic rectangular | Plused biphasic rectangular |
| Maximum Output Voltage(max) | | | |
| 500 ohm | 68V @500Ω | 68V @500Ω | 35.2 V @500Ω |
| 2k ohm | 102V @ 2 kΩ | 102V @ 2 kΩ | 52V @ 2 kΩ |
| 10k ohm | 110V @10 kΩ | 110V @10 kΩ | 60.8V @10 kΩ |
| Maximum Output Current(max) | | | |
| 500 ohm | 133mA | 133mA | 70mA @500Ω |
| 2k ohm | 51mA | 51mA | 26mA @ 2 kΩ |
| 10k ohm | 11mA | 11mA | 6mA @10 kΩ |
| Maximum Phase | 20.02µC | 20.02µC | 11.6µC |
| charge (500 ohm) | | | |
| Maximum Average<br>Current (500 ohm) | 3.0375mA | 3.0375mA | 0.63mA |
| Maximum Current<br>Density (500 ohm) | 0.066mA/cm² | 0.188mA/cm² | 0.00149 mA/cm² |
| Maximum Average<br>Power Density<br>(500 ohm) | 2.66mW/ cm² | 7.52mW/ cm² | 0.0313mW/cm² |
| Frequency (Hz) | From 1Hz to 150Hz | From 1Hz to 150Hz | From 5Hz to 120Hz |
| Pulse Duration (us) | 50-250μs | 50-250μs | 120-250μs, |
| Burst Mode | Yes | Yes | No |
| Timer range(min) | 20min, 25min, 30min, 40min depending on preset program | 20min, 25min, 30min, 40min depending on preset program | 30min |
| Power Source | 3.7V rechargeable lithium battery | 2 Alkaline AA 1.5V (LR6) Batteries | 2 Alkaline AAA 1.5V DC Batteries<br>1 CR2032 Lithium battery (Internal Power) |
| Dimensions (mm) | 155.4 x 64.6 x 19.1 | 66×136×30.7 | 18 x 191 x 90 |
| Weight | 64g | 146.5 grams | 4.8 oz |
| Housing material | TPE & ABS | ABS | Silicone & ABS |
| Microprocessor<br>control | Yes | Yes | Yes |
| Automatic<br>Overload trip | Yes | Yes | No |
| Automatic no-load<br>trip | Yes | Yes | No |
| Automatic shut-off | Yes | Yes | Yes |
| User override<br>control | Yes | Yes | Yes |
| Indicator Display:<br>- On/Off Status?<br>- Low Battery?<br>- Voltage/Current Level?<br>-Output mode<br>-Time to cut-off | Yes<br>Yes<br>Yes<br>Yes<br>Yes | Yes<br>Yes<br>Yes<br>Yes<br>Yes | Yes<br>Yes<br>No<br>No<br>No |
| Electrode<br>compliance with 21<br>CFR 898 | Yes | Yes | Yes |
| Electrode cable | No | Yes | No |
7
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the Easymed logo. The logo is in blue and has a registered trademark symbol next to it. Underneath the logo is a red line.
Easy for the relief of your suffering
EasyMed Instruments Co., Ltd.
Add: 5/F-6/F, Gupo Gongmao Building, Fengxin Road, Fengxiang Industrial District, Da Liang, 528300 ShunDe, GuangDong, China Fax: 86-757-27727868 Tel: 86-757-28091300
8
{8}------------------------------------------------
Image /page/8/Picture/0 description: The image contains the logo for Easymed. The logo is in blue and has a registered trademark symbol next to it. Below the logo is a red line, and below that is the text "Easy for the relief of your suffering".
## DIFFERENCES BETWEEN NEW DEVICE AND PREDICATE DEVICES
The technical characteristics of SmartTENS are similar to those of the predicate devices in design, intended use and function. Like the predicate device EasyStim TN28 OTC (K140168) and WiTouch™ Pro (K120398), the SmartTENS is a device used to apply an electrical current to electrodes on a patient's skin to relieve pain. The stimulation parameters of new device EasyStim TN28 OTC are same as those of predicate device EasyStim TN28 OTC (K140168) and similar to those of WiTouch™ Pro (K120398).
The predicate device EasyStim TN28_OTC (K140168) utilizes flexible wires between the electrodes and the electical stimulus generator. However, the new device SmartTENS utilize wireless remote controls for the considerations of a more user friendly design, which is similar to WiTouch™ Pro (K120398).
The intended design of SmartTENS limits the application for use to sites of lower back, arms and legs. The design includes Thermal Plastic Elastomer (TPE) electrodes that are intended for reuse and are button-affixed to the electrical stimulus generator. The unique connection makes the electrodes integral to the generator. The connection and shape of the electrodes limit application of the device. These unique characteristics of the integral electrodes are similar to those of WiTouch™ Pro (K120398) and insignificant as it relates to safety and effectiveness, which is not critical to the intended use between the referenced predicate devices.
The predicate device EasyStim TN28 OTC (K140168) utilizes affixed buttons as the sole method to control the electrical stimulus generator. The new device SmartTENS utilizes an additional method of a wireless remote control which similar to the predicate device WiTouch™ Pro (K120398). The difference is WiTouch™ Pro (K120398) utilizes radio frequency transveiver to control the electrical stimulus generator, while the SmartTENS ultilzes Bluetooth 4.0 as the wireless connecting technology to control the electrical stimulus generator. Furthermore, the remote controller is the Mobile Medical Application that need to be installed in the smart phone. This uniqueness of controlling the electrical stimulus generator by utilizing a Bluetooth wireless connecting technology is insignificant as it relates to safety and effectiveness, and is not critical to the intended use between the device and the referenced predicate devices.
Technological characteristics, features, specifications and intended uses of the new device SmartTENS are substantially equivalent to the referenced predicate devices. The differences that exist between SmartTENS and predicate devices are insignificant in the terms of safety or effectiveness.
{9}------------------------------------------------
## PERFORMANCE TESTS
| FDA recognition No. | Standard Title |
|---------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 19-5 | AAMI/ANSI ES60601-1:2005/(R)2012 And A1:2012, C1:2009/(R)2012<br>And A2:2010/(R)2012 (Consolidated Text) Medical Electrical<br>Equipment - Part 1: General Requirements For Basic Safety And<br>Essential Performance (IEC 60601-1:2005, Mod). (General I (QS/RM)) |
| 19-2 | AAMI / ANSI / IEC 60601-1-2:2007/(R)2012, Medical Electrical<br>Equipment - Part 1-2: General Requirements For Basic Safety And<br>Essential Performance - Collateral Standard: Electromagnetic<br>Compatibility - Requirements And Tests (Edition 3). (General II (ES/EMC)) |
| 19-6 | IEC 60601-1-11 Edition 1.0 2010-04, Medical Electrical Equipment -<br>Part 1-11: General Requirements For Basic Safety And Essential<br>Performance - Collateral Standard: Requirements For Medical<br>Electrical Equipment And Medical Electrical Systems Used In The<br>Home Healthcare Environment [Including: Technical Corrigendum 1<br>(2011)]. (General II (ES/EMC)) |
| 17-11 | IEC 60601-2-10 Edition 2.0 2012-06, Medical Electrical Equipment --<br>Part 2-10: Particular Requirements For The Basic Safety And Essential<br>Performance Of Nerve And Muscle Stimulators. (Neurology) |
| 5-40 | ISO 14971: Second Edition 2007-03-01 Medical devices- Application of<br>Risk Management To Medical Devices. (General I (QS/RM)) |
| 5-85 | IEC 60601-1-6 Edition 3.0 2010-01, Medical Electrical Equipment-Part<br>1-6: General Requirements For Basic Safety and Essential<br>Performance- Collateral Standard: Usability(General I (QS/RM)) |
| 13-8 | IEC 62304 First Edition 2006-05, Medical devices software- Software<br>life cycle processes (Software/ Informatics) |
#### USABILITY STUDY:
A usability study was conducted and showed that users were able to use the device correctly and safety.
Add: 5/F-6/F, Gupo Gongmao Building, Fengxin Road, Fengxiang Industrial District, Da Liang, 528300 ShunDe, GuangDong, China Tel:
{10}------------------------------------------------
Image /page/10/Picture/0 description: The image shows the logo for Easymed. The logo is in blue and has a red line underneath it. Below the red line is the text "Easy for the relief of your suffering".
# CONCLUSION:
The new device SmartTENS has the same indications for use and similar technological characteristics to the predicate device EasyStim TN28 OTC (K140168) and WiTouch™ Pro (K120398). The stimulation parameters of new device SmartTENS are all in the same range of those of predicate devices. Thus, the new device SmartTENS is substantially equivalent to the predicate devices and any differences between the devices do not pose any new questions of safety and effectiveness.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.