The Option(TM) ELITE Vena Cava Filter 100cm System is intended for the prevention of recurrent pulmonary embolism (PE) via placement in the vena cava in the following conditions: - Pulmonary thromboembolism when anticoagulant therapy is contraindicated; - Failure of anticoagulant therapy in thromboembolic diseases; - Emergency treatment following massive pulmonary embolism when anticipated benefits of conventional therapy are reduced: - Chronic, recurrent pulmonary embolism when anticoagulation therapy has failed or is contraindicated The filter may be retrieved according to the instructions supplied in section entitled "Optional Procedure for Filter Retrieval" in patients who no longer require a filter can be performed only by jugular approach
Device Story
Intravascular filter system for percutaneous delivery in inferior vena cava (IVC) to prevent recurrent pulmonary embolism. Device consists of laser-cut Nitinol filter with shape-memory struts and retention anchors for vessel wall fixation; retrieval hook included. Delivery kit includes filter cartridge, 5F catheter sheath introducer, angiographic vessel dilator with radiopaque markers for linear measurement, and pusher. Operated by physicians via popliteal or antecubital approach. Filter expands upon deployment, exerting outward radial force for stability and central filtration. System allows for optional retrieval via jugular approach. Benefits include mechanical clot capture in patients where anticoagulation is contraindicated or ineffective.
Clinical Evidence
Bench testing only. Simulated use model evaluated deployment from antecubital/popliteal approach, measuring deployment forces, filter tilt angles, and occurrence of system damage or placement failures. Testing performed per ISO 10555-1 and ISO 10993-1 standards.
Indicated for patients requiring prevention of recurrent pulmonary embolism (PE) due to contraindication or failure of anticoagulant therapy, or following massive PE where conventional therapy benefits are reduced. Filter retrieval is indicated for patients who no longer require the device, via jugular approach.
Regulatory Classification
Identification
A cardiovascular intravascular filter is an implant that is placed in the inferior vena cava for the purpose of preventing pulmonary thromboemboli (blood clots generated in the lower limbs and broken loose into the blood stream) from flowing into the right side of the heart and the pulmonary circulation.
Special Controls
*Classification.* Class II. The special controls for this device are:(1) “Use of International Standards Organization's ISO 10993 ‘Biological Evaluation of Medical Devices Part I: Evaluation and Testing,’ ” and
(2) FDA's:
(i) “510(k) Sterility Review Guidance and Revision of 2/12/90 (K90-1)” and
(ii) “Guidance for Cardiovascular Intravascular Filter 510(k) Submissions.”
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized caduceus symbol, which is a staff with two snakes coiled around it. The symbol is surrounded by the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" in a circular arrangement. The logo is simple and recognizable, representing the department's mission to protect the health of all Americans.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
April 16, 2015
Rex Medical, LP % Susan Goldstein-Falk MDI Consultants, Inc. 55 Northern Blvd., Suite 200 Great Neck, New York 11021
Re: K143405
Trade/Device Name: Option™ Elite Vena Cava Filter 100cm System Regulation Number: 21 CFR 870.3375 Regulation Name: Cardiovascular Intravascular Filter Regulatory Class: Class II Product Code: DTK Dated: March 13, 2015 Received: March 17, 2015
Dear Susan Goldstein-Falk:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in
{1}------------------------------------------------
the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Kenneth J. Cavanaugh -S
for
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K143405
Device Name
Option(TM) ELITE Vena Cava Filter 100cm System
Indications for Use (Describe)
The Option(TM) ELITE Vena Cava Filter 100cm System is intended for the prevention of recurrent pulmonary embolism (PE) via placement in the vena cava in the following conditions:
- Pulmonary thromboembolism when anticoagulant therapy is contraindicated;
- Failure of anticoagulant therapy in thromboembolic diseases;
- Emergency treatment following massive pulmonary embolism when anticipated benefits of conventional therapy are reduced:
- Chronic, recurrent pulmonary embolism when anticoagulation therapy has failed or is contraindicated
The filter may be retrieved according to the instructions supplied in section entitled "Optional Procedure for Filter Retrieval" in patients who no longer require a filter can be performed only by jugular approach
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
|-------------------------------------------------|--|
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for REX MEDICAL. Below the logo is the address 555 North Ln, Suite 5035, Conshohocken, PA 19428. The logo is a stylized text with two curved lines above it.
Traditional 510(k): Device Submission Option™ ELITE Vena Cava Filter 100cm System
## 510(k) Summary
This summary of 510(k) is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR \$807.92.
The assigned 510(k) number is: K143405
| Submitter: | Rex Medical, L.P.<br>555 North Ln, Suite 5035<br>Conshohocken, PA 19428 |
|-----------------|--------------------------------------------------------------------------------------------------------------------------|
| Contact Person: | Erich Bressler<br>Chief Engineer<br>Phone: (610) 940-0665 x107<br>Fax: (610) 940-1590<br>Email: ebressler@rexmedical.com |
| Date Prepared: | April 15, 2015 |
| Trade Name: | OptionTM ELITE Vena Cava Filter 100cm System |
| Common Name: | Cardiovascular Intravascular Filter<br>(21 CFR 870.3375, Product Code DTK) |
# Predicate Device(s):
Option™ ELITE Vena Cava Filter System K133243
#### Device Description
The Option™ ELITE Vena Cava Filter is designed for the prevention of recurrent pulmonary embolism via percutaneous delivery in the inferior vena cava (IVC).
The Option ELITE Vena Cava Filter 100cm System is designed for IVC filter insertion, delivery, deployment and placement via the popliteal and antecubital approach.
The self-centering Option™ ELITE Filter is laser cut from nickel -titanium alloy (Nitinol) tubing. The Option™ ELITE Filter consists of shape memory Nitinol struts emanating from a central location and is designed for optimal clot capture. Retention anchors (retention hooks) are located at the caudal portion of the filter. These anchors are intended for filter fixation to the vessel wall. The Option™ ELITE Filter is intended to be used in caval diameters up to 30mm. A retrieval hook is centrally located at the cranial extremity.
The constrained Option™ ELITE Filter is flexible and expands to the internal diameter of the IVC upon deployment. The Option™ ELITE Filter imparts an outward radial force on the luminal surface of the vena cava to ensure proper positioning and stability. The Option™ ELITE Filter is designed to prevent pulmonary embolism while maintaining caval patency through central filtration.
The introduction kit is comprised of a filter housed in a filter cartridge, Catheter Sheath Introducer (5F ID),
{4}------------------------------------------------
Angiographic Vessel Dilator with an open end, and a Pusher with deployment marker. The Angiographic Vessel Dilator has side holes and 2 radiopaque markers, separated by 32mm (between the marker bands), that provide linear measurement of the inferior vena cava and assists in angiographic visualization when radiopaque contrast is delivered. The pusher advances the filter through the Catheter Sheath Introducer up to the deployment marker, and is then used to fix the filter in place during uncovering. The distal end of the Catheter Sheath Introducer can be controlled by rotating the entire device to position the Catheter Sheath Introducer in the center of the vena cava. The Filter Cartridge houses the Option™ ELITE Filter. The body of the Cartridge has text and colored arrows printed on it that identify assembly orientation; femoral is printed in green and jugular is printed in blue. The femoral orientation is used for the popliteal approach and the jugular orientation is used for the antecubital approach. The arrow of the desired access site will point into the Catheter Sheath Introducer hub. The Angiographic Vessel Dilator is designed to provide angiorraphic visualization and linear measurement of the vasculature when used in conjunction with the delivery of radiopaque contrast media to the vena cava.
### Intended Use:
The Option™ ELITE Vena Cava Filter 100cm System is intended for the prevention of recurrent pulmonary embolism (PE) via placement in the vena cava in the following conditions:
- Pulmonary thromboembolism when anticoagulant therapy is contraindicated; ●
- Failure of anticoagulant therapy in thromboembolic diseases;
- . Emergency treatment following massive pulmonary embolism when anticipated benefits of conventional therapy are reduced;
- . Chronic, recurrent pulmonary embolism when anticoagulation therapy has failed or is contraindicated
The filter may be retrieved according to the instructions supplied in section entitled "Optional Procedure for Filter Retrieval" in patients who no longer require a filter. Retrieval of the filter can be performed only by jugular approach.
This is the same intended use as previously cleared for the Option™ ELITE Vena Cava Filter System (K133243).
#### Technological Characteristics:
The Option"10 ELITE Vena Cava Filter 100cm System is similar with regards to materials, intended use, principles of operation and technological characteristics to the predicate device. Results of bench testing demonstrate Option 11 ELITE Vena Cava Filter 100cm System is as safe and effective as the legally marketed predicate device.
#### Modifications to Existing Technology:
The effective length of the Option™ ELITE Vena Cava Filter 100cm System has been increased from 70cm to 100cm to enable filter insertion through either the antecubital approach or popliteal vein. The vessel dilator, sheath introducer and pusher have increased functional lengths. All other kit components including the Option™ ELITE filter are identical to the predicate device. The system packaging has been modified to accommodate the longer system. Identical materials are used for the new packaging as compared to the predicate device.
{5}------------------------------------------------
Non-Clinical Performance Testing:
All testing performed on the Option™ ELITE Vena Cava Filter 100cm System was derived from the risk assessment which evaluated the safety and performance of the design modifications. Test methodology and acceptance criteria were derived from Rex Medical and from related ISO standards for evaluation of this device.
- Simulated use testing .
A simulated use model that adequately represented the deployment from an antecubital/popliteal approach into a clinically challenging anatomy for this device was used for testing. Relevant deployment forces and filter tilt angles were measured. The occurrence of failures, filter or delivery system damage, excessive resistance, filter off the vessel wall, and tilting when removing the guidewire were recorded. A discussion regarding the testing for the insertion, deployment and placement of the filter using the antecubital/popliteal approach compared with the jugular/femoral approach was also provided.
. ISO Standards: ISO 10555-1, ISO 10993-1
### Conclusion:
Rex Medical considers the Option™ ELITE Vena Cava Filter 100cm System to be substantially equivalent to the predicate device listed above. The conclusions are based on performance testing and similarities in indications for use, materials, technological characteristics, principle of operation and design features.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.