← Product Code [FTL](/productcode/FTL) · K143386

# Progrip Laparoscopic Self-Fixating Mesh (K143386)

_Sofradim Production · FTL · Mar 13, 2015 · General, Plastic Surgery · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K143386

## Device Facts

- **Applicant:** Sofradim Production
- **Product Code:** [FTL](/productcode/FTL.md)
- **Decision Date:** Mar 13, 2015
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 878.3300
- **Device Class:** Class 2
- **Review Panel:** General, Plastic Surgery
- **Attributes:** Therapeutic

## Indications for Use

ProGrip™ laparoscopic self-fixating mesh is indicated for the reinforcement of soft tissues during repair of inguinal hernia defects by laparoscopic approach.

## Device Story

ProGrip™ is a surgical mesh used for inguinal hernia repair. It consists of knitted monofilament polyester with resorbable polylactic acid (PLA) grips on one side and a resorbable porcine collagen/glycerol film on the other. The grips provide tissue fixation for at least 8 weeks before resorbing via hydrolysis. A green band aids orientation. The device is used by surgeons in a laparoscopic setting. The mesh is inserted through a trocar; the collagen film facilitates handling and deployment, while the grips secure the mesh to surrounding tissue without requiring additional mechanical fixation. This submission adds three new sizes (13x9 cm, 16x12 cm, 16x14 cm) to the existing product family to meet clinical requirements.

## Clinical Evidence

Bench testing only. No clinical data presented. Performance evaluations verified trocar compatibility (10mm/12mm) and structural integrity of gripping points after passage through trocars. Biocompatibility is supported by compliance with ISO 10993-1.

## Technological Characteristics

Knitted monofilament polyester mesh; resorbable monofilament polylactic acid (PLA) grips; resorbable porcine collagen and glycerol film. Biocompatible per ISO 10993-1. Dimensions: 13x9 cm, 16x12 cm, 16x14 cm. Non-electronic, mechanical device.

## Regulatory Identification

Surgical mesh is a metallic or polymeric screen intended to be implanted to reinforce soft tissue or bone where weakness exists. Examples of surgical mesh are metallic and polymeric mesh for hernia repair, and acetabular and cement restrictor mesh used during orthopedic surgery.

## Predicate Devices

- Progrip™ Laparoscopic Self-Fixating Mesh ([K142900](/device/K142900.md))

## Reference Devices

- Progrip™ Laparoscopic Self-Fixating Mesh ([K120897](/device/K120897.md))
- Progrip™ Laparoscopic Self-Fixating Mesh ([K123479](/device/K123479.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image is a black and white logo for the U.S. Department of Health and Human Services. The logo features a stylized caduceus symbol, which is a staff with two snakes coiled around it. The caduceus is surrounded by the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" in a circular arrangement.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

March 18, 2015

Sofradim Production % Ms. Mary Mellows Surgical devices, a global business unit of Covidien 60 Middleton Avenue North Haven, Connecticut 06473

Re: K143386

Trade/Device Name: ProGrip™ laparoscopic self-fixating mesh Regulation Number: 21 CFR 878.3300 Regulation Name: Surgical mesh Regulatory Class: Class II Product Code: FTL Dated: February 10, 2015 Received: February 11, 2015

Dear Ms. Mellows:

This letter corrects our substantially equivalent letter of March 13, 2015.

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must

{1}------------------------------------------------

Page 2 - Ms. Mary Mellows

comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours.

# David Krause -S

- for Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K143386

Device Name ProGrip™ laparoscopic self-fixating mesh

Indications for Use (Describe)

ProGrip™ laparoscopic self-fixating mesh is indicated for the reinforcement of soft tissues during repair of inguinal hernia defects by laparoscopic approach.

Type of Use (Select one or both, as applicable)

2 Prescription Use (Part 21 CFR 801 Subpart D)

_ Over-The-Counter Use (21 CFR 801 Subpart C)

### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.

### FOR FDA USE ONLY

Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)

This section applies only to requirements of the Paperwork Reduction Act of 1995.

*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

EF

{3}------------------------------------------------

# 510(k) Summary

This 510(k) summary information is submitted in accordance with the requirements of the Safe Medical Devices Act (SMDA) of 1990. The contents of the 510(k) summary have been provided in conformance with 21 CFR §807.92.

## Submitter Information

| Name:                                                           | Sofradim Production (subsidiary of Covidien LLC)                                                          |
|-----------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------|
| Address:                                                        | 116, avenue du formans<br>01600 Trevoux, France                                                           |
| Phone number                                                    | +33 (0)4 74 08 90 00                                                                                      |
| Fax number:                                                     | +33 (0) 4 74 08 90 02                                                                                     |
| Establishment<br>Registration:                                  | 9615742                                                                                                   |
| Name of contact person:                                         | Mary Mellows<br>Covidien<br>Senior Regulatory Specialist<br>60 Middletown Avenue<br>North Haven, CT 06473 |
| Phone:                                                          | 203-492-5284                                                                                              |
| Date prepared:                                                  | November 25, 2014                                                                                         |
| Trade or proprietary name:                                      | ProGrip™ Laparoscopic Self-Fixating Mesh                                                                  |
| Common or usual name:                                           | Surgical Mesh                                                                                             |
| Classification name:                                            | Mesh, Surgical, Polymeric                                                                                 |
| Classification panel:                                           | General and Plastic Surgery (79)                                                                          |
| Regulation:                                                     | 21 CFR 878.3300                                                                                           |
| Product Code:                                                   | FTL                                                                                                       |
| Legally marketed devices<br>to which equivalence is<br>claimed: | Progrip™ Laparoscopic Self-Fixating Mesh (K142900)                                                        |
| Referenced devices noted<br>in this submission:                 | Progrip™ Laparoscopic Self-Fixating Mesh (K120897)<br>Progrip™ Laparoscopic Self-Fixating Mesh (K123479)  |

Covidien

{4}------------------------------------------------

.

| Reason for 510(k) Submission: | The purpose of this 510(k) is to notify the Agency of the addition of additional sizes 13x9 cm, 16x12 cm and 16x14 cm to the Progrip™ Laparoscopic Self-Fixating Mesh product family (current size is 15x10 cm). These three sizes are being added to the product family to satisfy customer requirements.<br><br>These proposed sizes will result in the addition of 6 new reorder codes to the predicate product, Progrip™ Laparoscopic Self-Fixating Mesh (K142900):<br><br>1. The proposed size 13x9 will be available in two configurations:<br>Right anatomical mesh,Left anatomical mesh<br>2. The proposed size 16x12 will be available in three configurations:<br>Right anatomical mesh,Left anatomical mesh,Rectangular mesh<br>3. The proposed size 16x14 will be available in one configuration:<br>Rectangular mesh |
|-------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device description:           | Both the predicate and the proposed mesh are made of knitted monofilament polyester with monofilament polylactic acid resorbable grips on one side and a resorbable film made of collagen from porcine origin and glycerol, on the other side. The grips allow positioning and fixation of the mesh to the surrounding tissue, while the collagen film facilitates mesh handling and deployment. The mesh incorporates a green band to facilitate orientation.<br><br>The monofilament polylactic acid grips are bioresorbable and provide the fixation of the mesh to surrounding tissue for at least 8 weeks. The polylactic acid grips degrade and resorb in vivo by hydrolysis and are metabolized by the body into CO2 and H2O.                                                                                              |
| Intended use of the device:   | Both the predicate and the proposed ProGrip™ Laparoscopic Self-Fixating Mesh are intended for the reinforcement of tissue during surgical repair.<br><br>No change to the intended use has been made in this submission.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          |
| Indications for use:          | Both the predicate and the proposed ProGrip™ Laparoscopic Self-Fixating Mesh are indicated for the reinforcement of soft                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          |

Covidien

{5}------------------------------------------------

tissues during repair of inguinal hernia defects by laparoscopic approach.

No change to the indication for use has been made in this submission.

### Summary comparing the technological characteristics of the subject and predicate devices:

Both the predicate and the proposed ProGrip™ Laparoscopic Self-Fixating Mesh are knitted monofilament polyester with monofilament polylactic acid resorbable grips on one side and an absorbable film made of collagen from porcine origin and glycerol. on the other side. The intended use and technology of the three additional sizes of ProGrip™ Laparoscopic Self-Fixating Mesh are identical to that of the predicate ProGrip™ Laparoscopic Self-Fixating Mesh.

### Performance data:

The materials of the new ProGrip™ Laparoscopic Self-Fixating Mesh references are identical to the predicate ProGrip™ Laparoscopic Self-Fixating Mesh and therefore, the performance data submitted in the predicate ProGrip™ Laparoscopic Self-Fixating Mesh 510k (K142900) applies to the three additional sizes.

ProGrip™ Laparoscopic Self-Fixating Mesh new sizes are composed of biocompatible materials that are in compliance with ISO 10993-1 standard.

As materials and manufacturing processes have not changed, the stability data of the predicate ProGrip™ Laparoscopic Self-Fixating Mesh (K142900) is applicable for the additional sizes of ProGrip™ Laparoscopic Self-Fixating Mesh.

In addition to tests referenced in K142900, bench performance evaluations were completed to verify that the proposed additional sizes of ProGrip™ Laparoscopic Self-Fixating Mesh (16x12cm and 16x14cm) are compatible with 10mm and/or 12mm trocars. Mesh sizes 16x12cm and 16x14cm were chosen for testing due to their worst case configuration regarding trocar passage compared to 15x10cm size.

The tests performed were:

- Passage through 10mm and/or 12mm trocar ●
- . Visual inspection after trocar passage
- Gripping point measurement and comparison . with specification (≤27 gripping points for 16x12cm and16x14cm) after trocar passage.

{6}------------------------------------------------

ProGrip™ Laparoscopic Self-Fixating Mesh

Conclusion:

Bench testing results demonstrate that the proposed additional sizes of Progrip™ Laparoscopic Self-Fixating Mesh are substantially equivalent to the predicate Progrip™ Laparoscopic Self-Fixating Mesh (K142900).

---

**Source:** [https://fda-staging.innolitics.com/device/K143386](https://fda-staging.innolitics.com/device/K143386)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
