K143368 · Ge Medical Systems, LLC · LLZ · Jul 29, 2015 · Radiology
Device Facts
Record ID
K143368
Device Name
GenIQ
Applicant
Ge Medical Systems, LLC
Product Code
LLZ · Radiology
Decision Date
Jul 29, 2015
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.2050
Device Class
Class 2
Attributes
Software as a Medical Device
Indications for Use
GenIQ is an automated post-processing software option that is indicated for use on dynamic magnetic resonance imaging data sets to generate parametric image intensity variations over time. This dynamic change in signal intensity is used to calculate functional parameters related to tissue flow and leakage of the contrast agent from the intravascular to the extracellular space. GenIQ provides information that when interpreted by a trained physician, can be useful for assessing tissue vascular properties.
Device Story
GenIQ is an automated post-processing software application for Dynamic Contrast Enhanced (DCE) MRI data. It performs pharmacokinetic analysis using a General Kinetic Model (GKM) to transform signal intensity variations over time into parametric images. Outputs include functional parameters describing tissue vascular properties, such as forward/backward transfer constants, plasma volume, and extracellular space volume. Used in clinical settings by trained physicians to interpret tissue vascularity; output assists in clinical decision-making regarding tissue characterization. Benefits include quantitative assessment of vascular properties derived from dynamic imaging.
Clinical Evidence
No clinical trials performed. Validation included bench testing using digital phantom data referenced by the Quantitative Imaging Biomarkers Alliance (QIBA) to demonstrate correct implementation of the General Kinetic Model. Additionally, anonymized MR contrast-enhanced clinical datasets were used for validation.
Technological Characteristics
Software-only post-processing application. Implements General Kinetic Model (GKM) for pharmacokinetic analysis. Complies with ISO 13485, ISO 14971, IEC 62304, and IEC 62366 standards. Operates on dynamic MRI data sets.
Indications for Use
Indicated for use on dynamic magnetic resonance imaging (MRI) data sets to calculate functional parameters related to tissue flow and contrast agent leakage for assessment of tissue vascular properties by a trained physician.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, stacked on top of each other.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
July 29, 2015
GE Medical Systems, LLC % Ms. Jenny Wong Regulatory Affairs Leader, Magnetic Resonance 3200 N. Grandview Blvd. WAUKESHA WI 53188
Re: K143368
Trade/Device Name: GenIO Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: II Product Code: LLZ Dated: July 7, 2015 Received: July 8, 2015
Dear Ms. Wong:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21. Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
Michael D'Hara
For
Robert Ochs, Ph.D. Acting Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K143368
Device Name GenIQ
Indications for Use (Describe)
GenIQ is an automated post-processing software option that is indicated for use on dynamic magnetic resonance imaging data sets to generate parametric image intensity variations over time. This dynamic change in signal intensity is used to calculate functional parameters related to tissue flow and leakage of the contrast agent from the intravascular to the extracellular space.
GenIQ provides information that when interpreted by a trained physician, can be useful for assessing tissue vascular properties.
Type of Use (Select one or both, as applicable)
2 Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
#### FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Druq Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for General Electric (GE). The logo is a blue circle with the letters "GE" in a stylized script inside. There are three white teardrop shapes evenly spaced around the inside of the circle, creating a swirling effect. The logo is simple and recognizable, representing the company's long history and global presence.
Section 5: 510(k) Summary
GenIQ
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for General Electric (GE). The logo is a blue circle with the letters "GE" in a stylized font in the center. There are decorative swirls around the letters. The logo is simple and recognizable, and it is often used to represent the company's brand.
## 510(k) Summary
In accordance with 21 CFR 807.92 the following summary of information is provided:
Date: November 24, 2014
| Submitter: | GE Medical Systems SCS<br>283 Rue de la Miniere<br>Buc, France 78530<br>FDA Registration Number: 9611343 |
|---------------------------|------------------------------------------------------------------------------------------------------------------------|
| Primary Contact Person: | Jenny Wong<br>Regulatory Affairs Leader, MR<br>GE Medical Systems, LLC<br>Phone (262) 521-6102<br>Fax (262) 546-0902 |
| Secondary Contact Person: | Glen Sabin<br>Regulatory Affairs Director, MR<br>GE Medical Systems, LLC<br>Phone (262) 521-6848<br>Fax (262) 364-2785 |
| Device: | GenIQ |
| Trade Name: | GenIQ |
| Common/Usual Name: | System, image processing, radiological- Picture archiving<br>and communications system. |
| Classification Names: | 21CFR 892.2050 |
| Product Code: | LLZ |
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'GE' written in a stylized, cursive font, enclosed within a circular frame. The color of the logo is a light blue.
| | Predicate Device(s): Predicate Device Name: MR Permeability Software |
|----------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | Predicate 510k Number: K130278 |
| | Predicate Manufacturer: Philips Medical Systems<br>Nederland B.V. |
| Device Description: | GenIQ is a software application used for the<br>pharmacokinetic analysis of Dynamic Contrast Enhanced<br>(DCE) MRI data sets. The application is used to perform a<br>General Kinetic Model (GKM)-based pharmacokinetic<br>modeling of DCE-MRI data. The goal of GenIQ is to<br>extract functional parameters describing tissue vascular<br>properties such as forward and backward transfer<br>constants, plasma volume, and volume of extra-cellular<br>space. |
| Indications for Use: | GenIQ is an automated post-processing software option<br>that is indicated for use on dynamic magnetic resonance<br>imaging data sets to generate parametric images from the |
image intensity variations over time. This dynamic change in signal intensity is used to calculate functional parameters related to tissue flow and leakage of the contrast agent from the intravascular to the extracellular space.
> GenIQ provides information that when interpreted by a trained physician, can be useful for assessing tissue vascular properties.
Comparison of The proposed medical device, GenIQ, employs the same Technological fundamental scientific technology as its predicate device, Characteristics with MR Permeability. The proposed device (GenIQ) is Predicate Device: substantially equivalent to the predicate device because it is a post-processing software option for use on dynamic contrast enhanced (DCE) MR image datasets.
{6}------------------------------------------------
#### Summary of Non-Clinical and Clinical Tests: Determination of
Substantial Equivalence:
The GenIO and its applications comply with voluntary standards:
- . ISO 13485
- ISO 14971
- IEC 62304 ●
- IEC 62366
The following quality assurance measures were applied to the development of the system:
- Risk Analysis
- Requirements Reviews
- Design Reviews
- Testing on unit level (Module verification)
- Integration testing (System verification)
- Performance testing (Verification) ●
- Safety testing (Verification) ●
- Simulated use testing (Validation)
Simulated use testing was performed on digital phantom data referenced by Quantitative Imaging Biomarkers Alliance (QIBA). This validation demonstrated good implementation of the General Kinetic Model. In addition, anonymized MR contrast-enhanced images were used as clinical datasets to validate the GenIQ application.
- GE Healthcare considers the GenIQ application to be as Conclusion: safe, as effective, and performance is substantially equivalent to the predicate device.
Image /page/6/Picture/19 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'g' and 'e' intertwined in a stylized script, enclosed within a blue circle. The circle has four stylized flourishes extending from it, one at each cardinal direction.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.