K143356 · Ceterix Orthopaedics, Inc. · GAT · Jul 21, 2015 · General, Plastic Surgery
Device Facts
Record ID
K143356
Device Name
NovoStitch Plus Meniscal Repair System
Applicant
Ceterix Orthopaedics, Inc.
Product Code
GAT · General, Plastic Surgery
Decision Date
Jul 21, 2015
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.5000
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The NovoStitch Plus Meniscal Repair System is intended for approximation of soft tissue in meniscal repair procedures.
Device Story
NovoStitch Plus Meniscal Repair System is a handheld, manually operated, disposable arthroscopic surgical instrument; used by surgeons to pass size 2-0 braided, non-absorbable, polyethylene surgical suture through soft tissue. System consists of a polymer delivery handle and sterile suture cartridges. Surgeon operates device to retain tissue between upper and lower jaws; flexible nitinol needle facilitates suture placement. Device allows for multiple suture deployments (up to six stitches) via reloadable cartridges. Output is the placement of suture for manual knot tying to approximate meniscal tissue. Benefits include reliable suture delivery and tissue approximation during arthroscopic repair.
Clinical Evidence
Bench testing only. Testing included simulated use environment, device strength, reliability, insertion, suture deployment, removal, and reload capability. Verification confirmed device can reliably deliver up to six stitches. Suture tensile strength and diameter tested per USP requirements. Biocompatibility testing conducted per ISO 10993-1 for permanent implants. Sterilization, packaging, and shelf life validated.
Technological Characteristics
Hand-held, disposable arthroscopic instrument. Materials: polymer handle, stainless steel distal end, nitinol needle, polyethylene suture. Energy source: manual. Connectivity: none. Sterilization: ethylene oxide. System is reloadable for up to six suture deployments.
Indications for Use
Indicated for approximation of soft tissue in meniscal repair procedures.
Regulatory Classification
Identification
Nonabsorbable poly(ethylene terephthalate) surgical suture is a multifilament, nonabsorbable, sterile, flexible thread prepared from fibers of high molecular weight, long-chain, linear polyesters having recurrent aromatic rings as an integral component and is indicated for use in soft tissue approximation. The poly(ethylene terephthalate) surgical suture meets U.S.P. requirements as described in the U.S.P. Monograph for Nonabsorbable Surgical Sutures; it may be provided uncoated or coated; and it may be undyed or dyed with an appropriate FDA listed color additive. Also, the suture may be provided with or without a standard needle attached.
Special Controls
*Classification.* Class II (special controls). The special control for this device is FDA's “Class II Special Controls Guidance Document: Surgical Sutures; Guidance for Industry and FDA.” See § 878.1(e) for the availability of this guidance document.
Teleflex Force Fiber® Polyethylene Non-Absorbable Surgical Suture (K063778)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized image of an eagle with three human profiles incorporated into its design. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the eagle.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
July 21, 2015
Ceterix Orthopaedics Incorporated Mr. Scott King Senior Director, Regulatory Affairs, Quality Assurance and Compliance 959 Hamilton Avenue Menlo Park. California 94025
Re: K143356
Trade/Device Name: NovoStitch Plus Meniscal Repair System Regulation Number: 21 CFR 878.5000 Regulation Name: Nonabsorbable poly(ethylene terephthalate) surgical suture Regulatory Class: Class II Product Code: GAT Dated: June 12, 2015 Received: June 15, 2015
Dear Mr. King:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{1}------------------------------------------------
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
For
Sincerelv vours.
Joshua C. Nipper -S
Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known)
K143356
# Device Name
NovoStitch Plus Meniscal Repair System
#### Indications for Use (Describe)
The NovoStitch Plus Meniscal Repair System is intended for approximation of soft tissue in meniscal repair procedures.
Type of Use (Select one or both, as applicable)
2 Prescription Use (Part 21 CFR 801 Subpart D)
_ Over-The-Counter Use (21 CFR 801 Subpart C)
## PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
#### FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for "CETERIX ORTHOPAEDICS". The logo consists of an orange square with a white curved line inside, followed by the word "CETERIX" in red. Below the word "CETERIX" is the word "ORTHOPAEDICS" in a smaller, lighter orange font.
## 510(k) Summary
#### Ceterix Orthopaedics
# NovoStitch Plus Meniscal Repair System
| ADMINISTRATIVE INFORMATION | |
|------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k) Number: | K143356 |
| 510(k) Owner: | Ceterix Orthopaedics, Inc.<br>959 Hamilton Avenue<br>Menlo Park, CA 94025<br>Telephone (650) 316-8660<br>Fax (650) 323-2209 |
| Contact Person: | Scott King<br>959 Hamilton Avenue<br>Menlo Park, CA 94025<br>sking@ceterix.com<br>Telephone (650) 316-8660<br>Fax (650) 323-2209 |
| DATE PREPARED | July 14, 2015 |
| DEVICE NAME | |
| Trade/Proprietary Name:<br>Common Name:<br>Classification Names:<br>Classification Regulation: | NovoStitch Plus Meniscal Repair System<br>Suture, Nonabsorbable, Synthetic, Polyethylene<br>Suture, Nonabsorbable, Synthetic, Polyethylene<br>21CFR 878.5000 |
| Device Class:<br>Panel:<br>Product Code:<br>PREDICATE DEVICES | Class II<br>General & Plastic Surgery<br>GAT |
| Delivery Device Predicate<br>Suture Implant Predicate | Cayenne CrossFix II® Meniscal Repair Device (K121413)<br>Teleflex Force Fiber® Polyethylene Non-Absorbable Surgical Suture (K063778) |
| | |
#### DEVICE DESCRIPTION
The NovoStitch Plus Meniscal Repair System passes size 2-0 braided, non-absorbable, polyethylene surgical suture through soft tissue in arthroscopic surgery. It is comprised of a handheld surgical instrument to which cartridges preloaded with polyethylene suture are attached.
#### INDICATIONS FOR USE
The NovoStitch Plus Meniscal Repair System is intended for approximation of soft tissue in meniscal repair procedures.
#### PERFORMANCE DATA
Product performance testing for NovoStitch Plus Meniscal Repair System was performed in a simulated use environment and device strength and reliability, device insertion, suture deployment, device removal and ability to reload the device with a new suture cartridge and perform additional device insertions, deployments and removals was validated. Bench testing was performed to verify that the implantable suture
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for "Ceterix Orthopaedics". The logo consists of a stylized letter "C" in orange on the left, followed by the word "CETERIX" in red. Below "CETERIX" is the word "ORTHOPAEDICS" in a lighter shade of orange. The logo is simple and modern, with a focus on the company name.
meet all pre-established acceptance criteria. The verification testing has demonstrated that the device can reliably deliver up to six stitches. This includes the cartridge provided with the handle assembly and up to five additional cartridges. The verification testing included device joint strength, delivery device, suture tensile strength and suture diameter. Suture tensile strength and suture diameter testing was performed per USP requirements for non-absorbable surgical suture. Sterilization, packaging and shelf life were validated. In addition to these performance tests, full biocompatibility testing was conducted to demonstrate compliance with requirements for biocompatibility of permanent implants in contact with tissue and/or bone in accordance with ISO 10993-1.
# TECHNOLOGICAL CHARACTERISTICS
- The NovoStitch Plus Meniscal Repair System is a hand-held, disposable surgical instrument coupled with an implantable suture for the approximation of soft tissue in meniscal repair procedures. The system is comprised of a delivery handle and a suture cartridge, both of which are provided sterile.
- . Comparison of technological characteristics to the delivery device predicate device
| Characteristic | NovoStitch Plus Meniscal Repair System | CrossFix II Meniscal Repair Device |
|------------------------|-------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------|
| Method of use | Handheld, manually operated, single procedure, arthroscopic suture placement system | Handheld, manually operated, single procedure, arthroscopic suture placement system |
| Handle material | Polymer delivery handle | Polymer delivery handle |
| Distal end material | Stainless steel distal end | Stainless steel distal end |
| Tissue interaction | Tissue retained by upper and lower jaws | Tissue pierced with two hypotubes |
| Needle characteristics | Flexible, nitinol needle | Flexible, nitinol needle |
| Suture material | Polyethylene, nonabsorbable surgical suture. | Polyethylene, nonabsorbable surgical suture. |
| Sterilization | Sterile, ethylene oxide | Sterile, ethylene oxide |
| Knot tying method | Manually tied knot | Pre-tied knot plus additional manual throws |
- . The subject device has the same knot tying method as the suture implant predicate Force Fiber® Polyethylene Nonabsorbable Surgical Suture
- The suture used in the NovoStitch Plus Meniscal repair system is the same suture as the predicate suture cleared under K063778 and the same suture used in the predicate delivery device cleared under K121413
## EQUIVALENCE TO MARKETED PRODUCT
The NovoStitch Plus Meniscal Repair Device has demonstrated equivalence to the unmodified predicate devices in the following areas:
- Has the same intended use for approximation of soft tissues in meniscal repair procedures
- Has equivalent technological characteristics
- Has equivalent labeling .
- . Has demonstrated equivalent usability with respect to handle deployment force
- Has demonstrated product reliability and biocompatibility
In conclusion, the NovoStitch Plus Meniscal Repair System described in this submission is substantially equivalent to the predicate devices with respect to safety, effectiveness and reliability.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.