← Product Code [LYZ](/productcode/LYZ) · K143340

# Powder Free Vinyl Patient Examination Gloves, Yellow Color (K143340)

_Zibo Ling Yun Medical Products Co, Ltd. · LYZ · Jul 9, 2015 · General Hospital · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K143340

## Device Facts

- **Applicant:** Zibo Ling Yun Medical Products Co, Ltd.
- **Product Code:** [LYZ](/productcode/LYZ.md)
- **Decision Date:** Jul 9, 2015
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 880.6250
- **Device Class:** Class 1
- **Review Panel:** General Hospital

## Indications for Use

A patient examination glove is a disposable device intended for medical purposes that is worn upon the examiner's hands or finger to prevent contamination between patient and examiner.

## Device Story

Powder-free vinyl patient examination gloves; disposable; worn on examiner's hands/fingers. Function: physical barrier to prevent contamination between patient and examiner. Used in clinical settings by healthcare providers. Non-sterile; limited surface contact (<24 hours). Benefits: provides protection against contamination during medical examinations.

## Clinical Evidence

No clinical data. Bench testing only. Testing included physical dimensions, tensile strength, elongation, pinhole water-fill test (AQL 2.5), residual powder (<2 mg/glove), primary skin irritation, and dermal sensitization. All results met ASTM D5250-06 and ASTM D6124-06 standards.

## Technological Characteristics

Vinyl material; powder-free (<2 mg/glove). Dimensions and physical properties conform to ASTM D5250-06. Residual powder conforms to ASTM D6124-06. Non-sterile. Class I device. No software or electronic components.

## Regulatory Identification

A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.

## Predicate Devices

- Shijiazhuang Hongxiang Plastic Products Ltd. Synthetic Powder-Free Vinyl Patient Examination Gloves ([K992821](/device/K992821.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol that resembles a stylized human figure.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

July 9, 2015

Zibo Ling Yun Medical Products Co., Ltd. C/O Mr. Ray Zhou Official Correspondent Basic Medical Industries, Inc. 12390 East End Ave Chino, CA 91710

Re: K143340

Trade/Device Name: Powder-free Vinyl Patient Examination Gloves, Yellow Color Regulation Number: 21 CFR 880.6250 Regulation Name: Patient Examination Glove Regulatory Class: I Product Code: LYZ Dated: June 2, 2015 Received: June 5, 2015

Dear Mr. Zhou:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Page 2 - Mr. Zhou

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Tejashri Purohit-Sheth, M.D.

Tejashri Purohit-Sheth, M.D. Clinical Deputy Director DAGRID/ODE/CDRH FOR

Erin I. Keith, M.S. Division Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

## Indications for Use

510(k) Number (if known) K143340

Device Name

Powder-free Vinyl Patient Examination Gloves, Yellow Color

Indications for Use (Describe)

A patient examination glove is a disposable device intended for medical purposes that is worn upon the examiner's hands or fingers to prevent contamination between patient and examiner.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

# PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.

### FOR FDA USE ONLY

Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)

This section applies only to requirements of the Paperwork Reduction Act of 1995.

## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW."

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate of any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(K) Summary Attachment

{4}------------------------------------------------

# 510 (K) SUMMARY

This summary of 510(K) safety and effectiveness information is being submitted in accordance with requirements of SMDA 1990 and 21CFR 807.92.

#### 1. Submitter's Identification:

Zibo Ling Yun Medical Products Co. Ltd NO.318 Yuming Road Zibo Industrial Park Zibo City, Shandong China

Date summary prepared: July 8, 2015

#### 2. Name of the Device:

Zibo Ling Yun Medical Products Co. Ltd Powder-free Vinyl Patient Examination Gloves, Yellow Color

#### 3. Common name/classification name of the Device:

Powder-free Vinyl Patient Examination Gloves, Yellow Color

#### 4. Contact Person:

Ray Zhou, Tel: 909-548-4828 Email:rayzhou@basicmedical.com

#### ર. Predicate Device Information:

Shijiazhuang Hongxiang Plastic Products Ltd. Synthetic Powder-Free Vinyl Patient Examination Gloves (K992821)

#### 6. Device Description:

Subject device: Powder-free Vinyl Patient Examination Gloves, Yellow Color

i. Principal operation of the glove A disposable device intended for medical purposes that is worn upon the examiner's hands or finger

ii. Mechanism of action for achieving the intended effect Prevent contamination between patient and examiner

{5}------------------------------------------------

iii. Reference to the standards

Physical and Dimensions Testing conducted on gloves, Inspection Level S-2, AQL 2.5, meet ASTM-D-5250-06. The residual powder testing conforms with standards of ASTM D-6124-06.

iv. Biocompatibility results Not an irritant; Not a sensitizer

v. Bench testing N/A

#### 7. Intended Use:

A patient examination glove is a disposable device intended for medical purposes that is worn upon the examiner's hands or finger to prevent contamination between patient and examiner.

#### 8. Comparison to Predicate Devices:

Zibo Ling Yun Medical Products Co, Ltd Powder-free Vinyl Patient Examination Gloves, Yellow Color are similar to the Shijiazhuang Hongxiang Plastic Products Ltd. Synthetic Powder-Free Vinyl Patient Examination Gloves (K992821)

## Table 7-2. Side-by-Side Comparison of Intended Use, Design, Material, Physical, Biocompatibility, and Performance Testing

|                                                                                                         | Proposed Device (K143340)                                                                                                                                                                                                                                                                                                                                                                                           | Predicate Device (K992821)                                                                                                                                   |
|---------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device Description                                                                                      | Zibo Ling Yun Medical Products Co, Ltd<br>Powder-free Vinyl Patient Examination Gloves,<br>Yellow Color                                                                                                                                                                                                                                                                                                             | Shijiazhuang Hongxiang Plastic Products<br>Ltd.<br>Synthetic<br>Powder-Free<br>Vinyl<br>Patient Examination Gloves                                           |
| Indication for Use                                                                                      | Disposable device intended for medical<br>purposes that is worn on the examiner's hand or<br>finger to prevent contamination between patient<br>and examiner                                                                                                                                                                                                                                                        | Disposable device intended for medical<br>purposes that is worn on the examiner's<br>hand or finger to prevent contamination<br>between patient and examiner |
| Basic Design                                                                                            | Cover the hand and wrist area. Covers have<br>separate sheaths or openings for each finger and<br>the thumb.                                                                                                                                                                                                                                                                                                        | Cover the hand and wrist area. Covers<br>have separate sheaths or openings for<br>each finger and the thumb.                                                 |
| Regulation #                                                                                            | 21 CFR 880.6250                                                                                                                                                                                                                                                                                                                                                                                                     | 21 CFR 880.6250                                                                                                                                              |
| Device Class                                                                                            | Class I                                                                                                                                                                                                                                                                                                                                                                                                             | Class I                                                                                                                                                      |
| Product Code:                                                                                           | LYZ                                                                                                                                                                                                                                                                                                                                                                                                                 | LYZ                                                                                                                                                          |
| Labeling                                                                                                | Labels include: Product name; color; "single<br>use Only" size, piece count, lot number,<br>distributor name, and manufacturer address.                                                                                                                                                                                                                                                                             | Substantially equivalent                                                                                                                                     |
| Physical Properties:<br>Length on Large<br>Size                                                         | Confirmed with ASTM D5250-06                                                                                                                                                                                                                                                                                                                                                                                        | Confirmed with ASTM D5250-06                                                                                                                                 |
| Physical Properties:<br>Width of Palm on<br>Large Size                                                  | Confirmed with ASTM D5250-06                                                                                                                                                                                                                                                                                                                                                                                        | Confirmed with ASTM D5250-06                                                                                                                                 |
| Physical Properties:<br>Palm Thickness                                                                  | Confirmed with ASTM D5250-06                                                                                                                                                                                                                                                                                                                                                                                        | Confirmed with ASTM D5250-06                                                                                                                                 |
| Physical Properties:<br>Fingertip Thickness                                                             | Confirmed with ASTM D5250-06                                                                                                                                                                                                                                                                                                                                                                                        | Confirmed with ASTM D5250-06                                                                                                                                 |
| Non-Clinical results:<br>Residual Powder                                                                | <2 mg per glove and meet the requirement of<br>ASTM D6124-06.                                                                                                                                                                                                                                                                                                                                                       | Meet ASTM D6124-06                                                                                                                                           |
| Non-Clinical results:<br>Pinhole Results                                                                | Meet ASTM-D-5250-06                                                                                                                                                                                                                                                                                                                                                                                                 | Meet ASTM-D-5250-06                                                                                                                                          |
| Non-Clinical results:<br>Dermal<br>Sensitization                                                        | Not a sensitizer                                                                                                                                                                                                                                                                                                                                                                                                    | Not a sensitizer                                                                                                                                             |
| Non-Clinical results:<br>Primary Skin<br>Irritation                                                     | Not an irritant                                                                                                                                                                                                                                                                                                                                                                                                     | Not an irritant                                                                                                                                              |
| Dimensions with<br>Tolerances: Before<br>Aging: Tensile<br>Strength(Mpa) and<br>Ultimate<br>Elongations | Meet ASTM D5250-06                                                                                                                                                                                                                                                                                                                                                                                                  | Substantially equivalent                                                                                                                                     |
| Dimensions with<br>Tolerances: After<br>Aging: Tensile<br>Strength(Mpa) and<br>Ultimate<br>Elongations  | Meet ASTM D5250-06                                                                                                                                                                                                                                                                                                                                                                                                  | Substantially equivalent                                                                                                                                     |
| Summary of<br>comparison                                                                                | Zibo Ling Yun Medical Products Co, Ltd Powder-free Vinyl Patient Examination Gloves, Yellow Color (subject device) and Shijiazhuang Hongxiang Plastic Products Ltd. Synthetic Powder-Free Vinyl Patient Examination Gloves (predicate device) are substantially equivalent in all technological characteristics, has similar intended uses and technological characteristics and performed similar to the predicate |                                                                                                                                                              |

{6}------------------------------------------------

{7}------------------------------------------------

### 9. Discussion of Non-Clinical tests performed for Determination of Substantial Equivalence are as follows:

The standards used for Zibo Ling Yun Medical Products Co, Ltd glove production are based on ASTM-D-5250-06. All testing meets requirements for Physical and Dimensions Testing conducted on gloves, Inspection Level S-2, AQL 2.5.

The FDA 1000 ml. Water Fill Test was also conducted with samplings of AQL 2.5, Inspection Level I, meeting these requirements, Primary Skin irritation and Skin Sensitization testing was conducted with results showing no primary skin irritant or sensitization reactions.

No special labeling claims. Our gloves are "powder-free." They contain no more than 2 mg powder per glove.

#### 10. Sterilization

Not sterilized

#### 11. Patient Contact

Patient contact is limited for surface-contacting and has less than 24 hour duration.

#### 12. Discussion of Clinical Tests Performed:

Not Applicable

#### 13. Conclusions:

The conclusions drawn from the nonclinical tests demonstrate that the device is as safe and as effective, and performs as well as the legally marketed device.

---

**Source:** [https://fda-staging.innolitics.com/device/K143340](https://fda-staging.innolitics.com/device/K143340)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).