K143319 · Siemens Medi Cal Solutions, Inc. · LLZ · Feb 24, 2015 · Radiology
Device Facts
Record ID
K143319
Device Name
Syngo X-workplace SW VD10
Applicant
Siemens Medi Cal Solutions, Inc.
Product Code
LLZ · Radiology
Decision Date
Feb 24, 2015
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.2050
Device Class
Class 2
Indications for Use
The syngo X Workplace is a medical workstation for real-time viewing, image manipulation, 3D-visualization, communication, and storage of medical images and data on exchange media. It is used for diagnostic image viewing and post processing and for viewing and post processing during interventional procedures. The syngo X Workplace can be configured as stand-alone diagnostic review, post-processing, and reporting workstation. It can also be configured within a network to send and receive DICOM data. Furthermore, the X Workplace can be combined with systems of the Artis family. In combination with an Artis system, it provides image guided solutions in the operating room, for image guided surgery, by Image Fusion and by navigation systems, image guided solutions in interventional cardiology and electrophysiology and image guided solutions for interventional oncology, interventional radiology, and interventional neuroradiology. The syngo X Workplace can also be combined with fluoroscopy systems or Radiographic systems. The syngo X Workplace can be configured with a variety of syngo or Windows based software options, which are intended to assist the physician in diagnosis, treatment planning and treatment control. It includes commercially available postprocessing techniques and OEM options. Procedures that can be performed include: minimally invasive surgical procedures and minimally invasive tumor treatment.
Device Story
Medical workstation for real-time viewing, manipulation, 3D-visualization, and storage of radiological images; used in clinical settings (OR, interventional suites) by physicians/technicians. Inputs include DICOM medical images from networked modalities (Artis, fluoroscopy, radiography). Transforms data via syngo/Windows-based software applications (e.g., 4D-Viewer, Dyna4D, 3D Wizard, 2D/3D fusion) to provide enhanced visualization and post-processing. Outputs displayed on workstation monitors to assist in diagnosis, treatment planning, and interventional guidance. Facilitates minimally invasive surgical and tumor treatment procedures. Benefits include improved clinical workflow and optimized diagnostic evaluation through advanced image processing tools.
Clinical Evidence
Bench testing only. Verification and validation testing performed to assess functionality of software modifications. Clinical images provided to support equivalence. Conformance to standards including IEC 60950-1, IEC 60601-1-2, ISO 14971, IEC 62304, and NEMA PS3/ISO 12052.
Technological Characteristics
Medical workstation running Windows 7 OS. Connectivity via DICOM network. Supports Blu-Ray disc archiving. Software-based image processing applications. Conforms to IEC 60950-1, 60601-1-2, 60601-1-6, 62304, and NEMA PS3/ISO 12052. Moderate level of concern software.
Indications for Use
Indicated for diagnostic image viewing, post-processing, and interventional procedure support. Intended for use by healthcare professionals in clinical settings, including operating rooms, for procedures such as minimally invasive surgery and tumor treatment. Compatible with Artis, fluoroscopy, and radiographic systems.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, with flowing lines connecting them.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
February 24, 2015
Siemens Medical Solutions, Inc. % Ms. Patricia Jones Regulatory Affairs Specialist 51 Valley Stream Parkway MALVERN PA 19355
Re: K143319
Trade/Device Name: syngo X-Workplace SW VD10 Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: II Product Code: LLZ Dated: January 29, 2015 Received: January 30, 2015
Dear Ms. Jones:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21. Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Robert A Ochs
Robert Ochs. Ph.D. Acting Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Image /page/2/Picture/0 description: The image shows the word "SIEMENS" in a sans-serif font. The letters are a light blue color, and there is a black line underneath the word. The logo is simple and modern.
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
### Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.
510(k) Number (if known)
K143319
Device Name syngo X-Workplace SW VD10
Indications for Use (Describe)
The syngo X Workplace is a medical workstation for real-time viewing, image manipulation, 3D-visualization, communication, and storage of medical images and data on exchange media. It is used for diagnostic image viewing and post processing and for viewing and post processing during interventional procedures.
The syngo X Workplace can be configured as stand-alone diagnostic review, post-processing, and reporting workstation. It can also be configured within a network to send and receive DICOM data. Furthermore, the X Workplace can be combined with systems of the Artis family. In combination with an Artis system, it provides image guided solutions in the operating room, for image guided surgery, by Image Fusion and by navigation systems, image guided solutions in interventional cardiology and electrophysiology and image guided solutions for interventional radiology, and interventional neuroradiology.
The syngo X Workplace can also be combined with fluoroscopy systems or Radiographic systems.
The syngo X Workplace can be configured with a variety of syngo or Windows-based software options, which are intended to assist the physician in diagnosis, treatment control. It includes commercially available post-processing techniques and OEM options.
Procedures that can be performed include: minimally invasive surgical procedures and minimally invasive tumor treatment.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
FORM FDA 3881 (8/14)
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image displays the word "SIEMENS" in a bold, sans-serif font. The letters are a light blue color, and there is a thin black line underneath the word. The logo is simple and modern, with a clean design.
## 510(k) Summary: synqo X Workplace
Company: Siemens Medical Systems, Inc. 51 Valley Stream Parkway Malvern, PA 19355
Date Prepared: November 18, 2014
This 510(k) summary of safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR § 807.92.
- 1. General Information: Importer / Distributor: Siemens Medical Solutions, USA, Inc. 51 Valley Stream Parkway Malvern, PA 19355 Establishment Registration Number: 2240869
Manufacturing Site: SIEMENS AG Sector Healthcare Siemensstraße 1 D-91301 Forchheim, Germany Establishment Registration Number: 3004977335
#### 2. Contact Person:
Ms. Patricia D Jones Technical Specialist, Regulatory Submissions Siemens Medical Solutions USA, Inc. 51 Valley Stream Parkway D-02 Malvern, PA 19355 Phone: (610) 448 -3536 Fax: (610) 448-1787 Email: patricia.d.jones@siemens.com
### 3. Device Name and Classification: Trade Name:
svngo X Workplace SW VD10 Classification Name: Picture Archiving and Communications svstem Classification Panel: Radiology 21 CFR §892. 2050 Classification Requlation: Device Class: Class II Product Code: LLZ
{4}------------------------------------------------
# SIEMENS
- 4. Leqally Marketed Predicate Device Trade Name: X-LEONARDO Workstation 510(k) Clearance K042995 Clearance Date November 24, 2004 Classification Name: Picture Archiving and Communications System Classification Panel: Radiology CFR Section: 21 CFR §892. 2050 Device Class: Class II Product Code: LLZ Recall Information: This predicate device has not been the subject of any design related recalls.
### 5. Device Description:
SIEMENS Medical Solutions USA, Inc. intends to market the syngo X Workplace (XWP). SW Version VD10. This 510(k) submission describes several modifications to the previously cleared predicate device the X-LEONARDO Workstation (K042995) cleared on November 24. 2004. The modifications were made to the cleared X-LEONARDO Workstation SW VC10 that created the syngo X Workplace (XWP), SW Version VD10:
- 1). Updated Indications for Use Statement:
- 2). Updated Operating system, Windows 7
- 3). Updated SW version VD10 includes the following software features:
- > New 4D-Viewer
- > QuickZoom
- syngo Dyna4D A
- 3D Wizard A
- > syngo 2D/3D fusion
- > syngo DynaCT SMART
- > Dedicated graphical User Interface for the examination room
- 4). Image Archiving on Blu-Ray Disc, optional
- 5). Proposed product claims associated with the above device modifications (see Section 15)
- 6) Provides update 510(k) information (see Appendix E).
The X Workplace is a medical diagnostic workstation for real-time viewing, image manipulation, communication, and storage of medical images and data on exchange media.
The X Workplace is a further development of the X-LEONARDO Workstation. The X Workplace offers a comprehensive solution to view, optimize, and postprocess diagnostic information and to aid the doctors in the evaluation of digital radiological examinations and patient information.
Due to special customer requirements based on the modality image type and the clinical focus, the X Workplace can be configured with different combinations of clinical applications. Applications can be added to the X Workplace either individually or as clinical focus packages.
{5}------------------------------------------------
# SIEMENS
The X Workplace can be configured with a variety of syngo- or Windows based software options, which are intended to assist the physician in diagnosis or treatment planning. This includes commercially available OEM applications.
The syngo X Workplace (XWP) is within the same classification regulation, but with a more descriptive indication for use from the primary predicate device; however the intended use and the general Indications for Use Statement for Siemens' Picture Archiving and Communications System remains the same.
Provided in Sections 5 and 14 of this 510(k) submission, is a completed "510(k) Decision Making FlowChart". We believe these modifications are eligible for the Traditional 510(k) process since the subject device has the same fundamental scientific technology, Intended Use and general Indications for Use as the predicate device svstem. The new descriptive Indications for Use statement is supported with non-clinical testing in support of a claim of substantial equivalence to Siemens' predicate device the X-LEONARDO Workstation (K042995).
### 6. Indications for Use:
The X Workplace is a medical workstation for real-time viewing, image manipulation, 3D-visualization, communication, and storage of medical images and data on exchange media. It is used for diagnostic image viewing and post processing and for viewing and post processing during interventional procedures.
The syngo X Workplace can be configured as stand-alone diagnostic review, post-processing, and reporting workstation. It can also be configured within a network to send and receive DICOM data. Furthermore, the X Workplace can be combined with systems of the Artis family. In combination with an Artis system, it provides image guided solutions in the operating room, for image guided surgery, by Image Fusion and by navigation systems, image guided solutions in interventional cardiology and electrophysiology and image guided solutions for interventional oncology, interventional radiology, and interventional neuroradiology.
The syngo X Workplace can also be combined with fluoroscopy systems or Radiographic systems.
The syngo X Workplace can be configured with a variety of syngo or Windowsbased software options, which are intended to assist the physician in diagnosis, treatment planning and treatment control. It includes commercially available postprocessing techniques and OEM options.
Procedures that can be performed include: minimally invasive surgical procedures and minimally invasive tumor treatment.
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the word "SIEMENS" in a sans-serif font. The letters are a light blue color, and there is a black line underneath the word. The word is likely a logo for the company Siemens.
- Summary of Technological Characteristics of the Subject Device as 7. Compared with the Predicate Device:
The syngo X Workplace offers a comprehensive solution to view, optimize, and post-process diagnostic information and can be configured with different applications. These syngo applications can be added to the svngo X Workplace either individually or as clinical focus packages. For the viewing of 3D image data the Inspace viewer of the predicate device X-Leonardo is replaced with a new syngo 4D viewer.
The subject device (syngo X Workplace) is comparable to general Indications for Use for medical workstation for real-time viewing, image manipulation, 3Dvisualization, communication, and storage of medical images and data on exchange media. The subject device has the same intended use, design, material, functionality, technology, energy source and is considered substantially equivalent to the commercially available Siemens' X-Leonardo workstation.
The components of the subject device have many of the same technological characteristics as the one from the predicate device. There are several technological characteristics that differ slightly as shown in the comparison table below which summarizes differences betwwen the Subject Device and Legally Marketed Predicate device:
| Technological Characteristics of the Subject Device as Compared with the Predicate Device: | | |
|--------------------------------------------------------------------------------------------|------------------------------------------------------------------|------------------------------------------------------------------------------|
| Property | Subject Device: syngo X Workplace<br>SW VD10 | Predicate Device: X-Leonardo<br>SW VC10 |
| New<br>Operating<br>System | Updated Operating system<br>Windows 7 | Operating system Windows XP |
| New SW VD10<br>/ SW<br>Applications | New syngo 4D viewer: | InSpace viewer (K042995): |
| | Quick Zoom – one button | Zoom with mouse interactions |
| | syngo Dyna4D - 3D rotational | InSpace 3D rotational |
| | angiography with Improved<br>angulation range | angiography |
| | 3D Wizard - Electronic Protocol<br>book | Printed protocol book<br>Manual selection of organ program |
| | syngo DynaCT S.M.A.R.T.<br>Dedicated Graphical User<br>Interface | syngo DynaCT<br>1:1 mirroring of control room<br>display in examination room |
| | syngo 2/D/3D Fusion | 3D/3D registration based on syngo<br>InSpace 3D/3D Fusion |
| | | |
| | | |
| Image<br>Archiving | Archiving Blu-Ray disc | Archiving DVD |
Table 1: Predicate and Subject Device Compared Technological Characteristics
The subject device (syngo X Workplace) modifications do not alter its fundamental scientific technology from the 510(k) cleared predicate device
{7}------------------------------------------------
# SIEMENS
Siemens' X-Leonardo workstation. The interface/usability has been optimized to improve clinical workflows.
### 8. Nonclinical Performance Testing:
Non-clinical tests were conducted for the syngo X Workplace during product development. The modifications described in this Premarket Notification were supported with verification and validation testing. and provided clinical images.
Siemens claims conformance to the following performance standards: 60950-1+A1; 94 A/B; 60601-1-2; 14971; 60601-1-6; 62304; 60038; CAN 7CSA-C22.2 No. 60950-1-08; NEMA PS3 and following / ISO 12052 as provided in Section 11.
## Verification and Validation:
Software Documentation for a Moderate Level of Concern software per FDA's Guidance Document "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices" issued on May 11, 2005 is also included as part of this submission. The performance data demonstrates continued conformance with special controls for medical devices containing software. Nonclinical tests (Section 21) were conducted on the syngo X Workplace during product development.
The Risk analysis was completed and risk control implemented to mitigate identified hazards. The testing results supports that all the software specifications have met the acceptance criteria. Testing for verification and validation for the device was found acceptable to support the claims of substantial equivalence.
Siemens conforms to the Cybersecurity requirements by implementing a process of preventing unauthorized access, modifications, misuse or denial of use, or the unauthorized use of information that is stored, accessed, or transferred from a medical device to an external recipient. Contained in Attachment 8 of this submission is a cybersecurity statement that considers IEC 80001-1:2010. The responsibility for compliance with IEC 8001-1-2010 is the hospital. Provided in the Software section, (Section 18) and the Operator's Manual (Appendix B) is the required cybersecurity information.
## Summary:
Performance tests were conducted to test the functionality of the syngo X Workplace workstation. These tests have been performed to assess the functionality of the subject device. Results of all conducted testing and clinical images were found acceptable in supporting the claim of substantial equivalence.
#### 9. General Safety and Effectiveness Concerns:
Instructions for use are included within the device labeling, and the information provided will enable the user to operate the device in a safe and effective manner.
{8}------------------------------------------------
Image /page/8/Picture/1 description: The image contains the word "SIEMENS" in a sans-serif font. The word is written in a light blue color. There is a black line underneath the word.
Risk management is ensured via a hazard analysis, which is used to identify potential hazards. These potential hazards are controlled via software development, verification and validation testing. To minimize electrical hazards, Siemens adheres to recognized and established industry practice, and all equipment is subject to final performance testing. Furthermore the operators are health care professionals familiar with and responsible for the evaluating and post processing of medical images.
#### 10. Conclusion as to Substantial Equivalence:
The predicate device was cleared based on non-clinical supportive information and clinical images. The results of these tests demonstrate that syngo X-Workplace acceptance criteria are adequate for this intended use. The comparison of technological characteristic, non-clinical performance data, clinical images, and software validation data demonstrates that the subject device is as safe, and effective when compared to the predicate device that is currently marketed for the same intended use.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.