K143318 · Dr Systems, Inc. · LLZ · Mar 10, 2015 · Radiology
Device Facts
Record ID
K143318
Device Name
DR Systems Zero Download 3D Viewer (Z3D)
Applicant
Dr Systems, Inc.
Product Code
LLZ · Radiology
Decision Date
Mar 10, 2015
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.2050
Device Class
Class 2
Attributes
Software as a Medical Device
Indications for Use
DR Systems Z3D is intended to provide reading physicians, referring physicians, and other appropriate healthcare professionals tools to aid in interpreting medical images, including: Displaying DICOM compliant medical image volumes, such as CT, MRI, and PET. Reformatting images, including creation of MPRs, MIPS, MinIPs, color/monochrome 3D volume rendered images. Manipulating displayed images via control of slice thickness, slice interval, obliquity, perspective, rotation, window/level, crop, zoom, color/monochrome transformations, segmentation, sculpting, straightening the display of curved structures, and creating images perpendicular to a curvilinear path. Creating series of DICOM images and individually captured images that can be displayed and stored in a PACS. Measuring coronary calcium, which is intended for non-invasive identification and quantification of calcified atherosclerotic plaques in the coronary arteries using tomographic medical image data and clinically accepted calcium scoring algorithms.
Device Story
DR Systems Z3D is a server-based, browser-accessible advanced imaging solution for PACS. It ingests DICOM-compliant medical image volumes (CT, MRI, PET). The system performs image processing on a secure server and renders outputs for display in a web browser. Features include multiplanar reconstruction (MPR), maximum/minimum intensity projections (MIP/MinIP), 3D volume rendering, and cardiac calcium scoring. Physicians use the tool to manipulate image orientation, slice thickness, and window/level settings, and to perform measurements. The device aids in clinical interpretation and decision-making by providing enhanced visualization and quantification of calcified plaques. It is intended for use by clinicians in clinical settings to facilitate diagnostic workflows.
Clinical Evidence
No clinical data. The device underwent nonclinical testing including verification, validation, risk analysis, design reviews, performance testing, and simulated use testing for human factors and usability.
Technological Characteristics
Server-based rendering with browser-based client viewing. Supports DICOM image volumes (CT, MRI, PET). Processing performed on secure image server. Minimum requirements: Intel Core i5 3GHz, 4GB RAM, 160GB HDD, 100Mb/s Ethernet. Compatible with Windows 7/8.1 and Mac OS X v10.6-10.8. Requires web browser with Adobe Flash Player 10+ and Silverlight 4+.
Indications for Use
Indicated for reading physicians, referring physicians, and other healthcare professionals to aid in interpreting medical images (CT, MRI, PET) and performing coronary calcium scoring for identification/quantification of calcified atherosclerotic plaques.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the seal for the Department of Health & Human Services - USA. The seal is circular, with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. In the center of the seal is an abstract symbol that appears to be three stylized human profiles facing to the right.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
March 10, 2015
DR Systems, Inc. % Ms. Kimberly J. Meade Manager, Regulatory Affairs/Quality Assurance 10140 Mesa Rim Road SAN DIEGO CA 92121
Re: K143318
Trade/Device Name: DR Systems Zero Download 3D Viewer (Z3D) Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: II Product Code: LLZ Dated: February 18, 2015 Received: February 18, 2015
Dear Ms. Meade:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21. Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Robert A Ochs
Robert Ochs. Ph.D. Acting Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
### Indications for Use
510(k) Number (if known) K143318
Device Name DR Systems Zero Download 3D Viewer (Z3D)
#### Indications for Use (Describe)
DR Systems Z3D is intended to provide reading physicians, and other appropriate healthcare professionals tools to aid in interpreting medical images, including:
· Displaying DICOM compliant medical image volumes, such as CT, MRI, and PET.
· Reformatting images, including creation of MPRs, MinIPs, color/monochrome 3D volume rendered images
· Manipulating displayed images via control of slice interval, obliquity, perspective, rotation, window/ level, crop, zoom, color/monochrome transformation, sculpting, straightening the display of curved structures, and creating images perpendicular to a curvilinear path.
· Creating series of DICOM images and individually captured images that can be displayed and stored in a PACS.
• Measuring coronary calcium, which is intended for non-invasive identification of calcified atherosclerotic plaques in the coronary arteries using tomographic medical image data and clinically accepted calcium scoring algorithms.
Type of Use (Select one or both, as applicable)
| <span style="font-size: 12pt;"> Prescription Use (Part 21 CFR 801 Subpart D) </span> |
|------------------------------------------------------------------------------------------|
| <span style="font-size: 12pt;"> Over-The-Counter Use (21 CFR 801 Subpart C) </span> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for DR Systems. The logo consists of the letters "D" and "R" in white on a blue square. To the right of the square, the word "Systems" is written in blue.
# Section 5 – 510(k) Summary
| Submitter: | DR Systems, Inc.<br>10140 Mesa Rim Road<br>San Diego, CA 92121<br>Phone: (858) 625-3344<br>FAX: (858) 625-3335 |
|-------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact: | Kimberly J. Meade, MSQA, ASQ CQA<br>Manager, Regulatory Affairs/Quality Assurance<br>Phone: (858) 625-3344, x2260<br>kmeade@drsys.com |
| Date Prepared: | November 17, 2014 |
| Trade Name: | DR Systems Zero Download 3D Viewer (Z3D) |
| Common Name: | Digital Image Processing |
| Classification Name: | Systems, Image processing, Radiological |
| Classification<br>Regulation: | 21 CFR 892.2050, Class II |
| Product Code: | 90 LLZ |
| Predicate Device: | K060505 - Voxar 3D Product Family |
| Device Description: | DR Systems Z3D (Zero Download 3D) advanced imaging solution supports<br>MIP (maximum intensity project), MPR (multiplanar reconstruction), 3D<br>volume rendering, and cardiac calcium scoring. The system employs<br>server-based rendering and browser-based viewing. Reading physicians,<br>referring physicians, and other appropriate healthcare personnel and can<br>employ advanced image processing and display from local or remote<br>locations.<br><br>Viewing Modes:<br>Multiplanar reconstruction Two 3-D viewing modes, one showing 4 images, and one showing 6 images Cardiac calcium scoring DR Systems Z3D is similar to the predicate device, Voxar 3D. Both are<br>PACS system accessories that allow the user to view and manipulate 3D<br>image data sets. DR Systems Z3D does not have an AvelP analysis tool.<br>The principles of operation and technological characteristics of the new and<br>predicate device are the same. |
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for DR Systems. The logo consists of the letters "D" and "R" inside of a blue square. To the right of the square, the word "Systems" is written in blue.
## Technical Specifications
| Specifications | |
|-------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Processor | o Intel® Core™ i5 3 GHz or faster:<br>o Quad-core minimum. |
| Operating<br>system | ■ Microsoft® Windows® 7/8.1<br>■ Apple® Mac OS ® X v10.6, v10.7, or v10.8 |
| Hard drive | 160 GB 7200 RPM minimum |
| Available hard<br>drive space | 20 GB with a minimum free disk space of at least 15%, for program files<br>and to copy exams to the comnputer's local drive |
| RAM | 4 GB (minimum) |
| Keyboard and<br>mouse | Standard or USB keyboard and a 5-button optical mouse is required. |
| NIC | Integrated. Ethernet data rate:<br>■ Minimum: 100 Mb/s<br>■ Recommended: 1 Gb/s |
| Monitor | 17" 1280 x 1024 resolution (minimum)<br>20" 1600 x 1200 resolution<br>24" 1920 x 1200 resolution |
| Network<br>Interface | For connection to the PACS local area network (LAN)<br>100 Mbit connection recommended |
| Internet<br>browser | ■ Standard web browser that supports Adobe Flash Player 10 or greater and<br>Silverlight 4 or greater:<br>o Internet Explorer® starting with 8.<br>o Mozilla Firefox® starting with 17<br>o Google Chrome™ starting with version 32<br>o Apple Safari® starting with 5.0 |
| Access | ■ Application Gateway Server (AGS)<br>■ Web address. For example: http://<AGS>/DRPacs/zda/ |
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for DR Systems. The logo consists of the letters "D" and "R" in white, set against a blue square. To the right of the blue squares, the word "Systems" is written in blue.
Intended Use/ DR Systems Z3D is intended to provide reading physicians, referring Indications for Use: physicians, and other appropriate healthcare professionals tools to aid in interpreting medical images, including: Displaying DICOM compliant medical image volumes, such as ● CT, MRI, and PET. . Reformatting images, including creation of MPRs, MIPS, MinIPs, color/monochrome 3D volume rendered images. ● Manipulating displayed images via control of slice thickness, slice interval, obliquity, perspective, rotation, window/level, crop, zoom, color/monochrome transformations, segmentation, sculpting, straightening the display of curved structures, and creating images perpendicular to a curvilinear path. Creating series of DICOM images and individually captured ● images that can be displayed and stored in a PACS. ● Measuring coronary calcium, which is intended for non-invasive identification and quantification of calcified atherosclerotic plaques in the coronary arteries using tomographic medical image data and clinically accepted calcium scoring algorithms. Z3D receives and processes image volumes from cross-sectional DICOM Technological medical imaging examinations such as CT, MRI, and PET. The system Characteristics: provides a graphical user interface to enable healthcare professional to execute the functions supporting the intended use. Image processing is performed on a secure image server, and images are then displayed using a browser-based viewer. Nonclinical Testing: DR Systems Z3D was designed, developed and tested according to DR System's written procedures. Testing included verification, validation and evaluation of human factors and usability. The following quality assurance measures were applied to the development of Z3D: . Risk analysis ● Requirements reviews Design reviews ● Performance testing (verification) ● Simulated use testing (validation and human factors and ● usability)
Clinical Testing: N/A
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for DR Systems. The letters "D" and "R" are in white and are inside of blue squares. The word "Systems" is in blue and is to the right of the blue squares.
Conclusion:
The information in this 510(k) submission demonstrates that based on the nonclinical testing, the subject, Zero Download 3D Viewer is substantially equivalent to the predicate device with respect to safety, effectiveness, and performance.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.