BHV VIVID (etafilcon A) Soft (Hydropholic) Contact Lens
Applicant
Brien Holden Vision Pty , Ltd.
Product Code
LPL · Ophthalmic
Decision Date
Apr 23, 2015
Decision
SESE
Submission Type
Abbreviated
Regulation
21 CFR 886.5925
Device Class
Class 2
Indications for Use
The BHV VIVID (etafilcon A) Soft (Hydrophilic) Contact Lens is indicated for daily wear for the correction of distance, intermediate and near vision in presbyopic, phakic or pseudo-phakic persons with non-diseased eyes who can have up to approximately 1.50 diopters (D) of astigmatism. The BHV VIVID (etafilcon A) Soft (Hydrophilic) Contact Lens helps protect against transmission of harmful UV radiation to the cornea and into the eye. Eye care practitioners may prescribe the lenses either for single use disposable wear or frequent/planned replacement wear with cleaning, disinfection, and scheduled replacement (see "Wearing Schedule"). When prescribed for single-use disposable wear (see "Wearing Schedule") the BHV VIVID (etafilcon A) Soft (Hydrophilic) Contact Lens is to be discarded after each removal. Therefore no cleaning or disinfection is required. When prescribed for frequent/planned replacement wear the BHV VIVID (etafilcon A) Soft (Hydrophilic) Contact Lens is to be cleaned, rinsed, and disinfected each time the lens is removed. The BHV VIVID (etafilcon A) Soft (Hydrophilic) Contact Lens contact lens is to be discarded after the recommended wearing period as prescribed by the eye care practitioner. When prescribed for frequent/planned replacement wear, the lenses should be disinfected using a chemical disinfection system only (i.e., not heat).
Device Story
BHV VIVID is an extended depth-of-focus (EDOF) soft contact lens; daily wear; corrects presbyopia; provides UV protection. Lens material is etafilcon A (HEMA/MAA copolymer); tinted with Reactive Blue No. 19 for visibility; contains benzotriazole UV blocker. Prescribed by eye care practitioners for single-use disposable or frequent/planned replacement. If frequent replacement, requires chemical disinfection. Supplied sterile in blister packs with borate buffered saline. Benefits include vision correction across multiple distances and protection against harmful UV radiation.
Clinical Evidence
No clinical data required; substantial equivalence established via bench testing, biocompatibility, microbiology, bacteriostatic validation, and leachables testing.
Technological Characteristics
Material: etafilcon A (HEMA/MAA copolymer); Water content: 58%; Refractive index: 1.402; Production: Cast molding; UV blocker: Benzotriazole; Standards: ISO 18369-1/2/3/4, ISO 10993-1/5/10, ISO 11607-1/2, ISO 11737-2, ISO 17665-1, ASTM F1980.
Indications for Use
Indicated for daily wear correction of distance, intermediate, and near vision in presbyopic, phakic, or pseudo-phakic patients with non-diseased eyes and up to 1.50D of astigmatism. Provides UV protection.
Regulatory Classification
Identification
A soft (hydrophilic) contact lens is a device intended to be worn directly against the cornea and adjacent limbal and scleral areas of the eye to correct vision conditions or act as a therapeutic bandage. The device is made of various polymer materials the main polymer molecules of which absorb or attract a certain volume (percentage) of water.
Predicate Devices
VISTAKON® (etafilcon A) Soft (Hydrophilic) Contact Lens, Clear and Tinted (Visibility and/or Cosmetically) with UV Blocker for Daily Wear (K062614)
Reference Devices
PEGAVISION CORPORATION Aquamax (Etafilcon A) Bi-Weekly and Daily Disposable Soft (Hydrophilic) Contact Lenses (K120028)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, with flowing lines connecting them.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
April 23, 2015
Brien Holden Vision Pty Ltd. Mr. Randall May Vice President, Regulatory and Ouality Affairs ABN 85 138 617 062 Level 3 North Wing, Rupert Myers Building Gate 14 Barker Street, UNSW Sydney NSW 2052, Australia
Re: K143280
Trade/Device Name: BHV VIVID (etafilcon A) Soft (hydrophilic) Contact Lens Regulation Number: 21 CFR 886.5925 Regulation Name: Soft (Hydrophilic) Contact Lens Regulatory Class: Class II Product Code: LPL. MVN Dated: March 23, 2015 Received: March 27, 2015
Dear Mr. May:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.
{1}------------------------------------------------
You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# Deborah L. Falls -S
for Malvina B. Eydelman, MD Director Division of Ophthalmic and Ear, Nose, and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K143280
### Device Name
BHV VIVID (etafilcon A) Soft (Hydrophilic) Contact Lens
Indications for Use (Describe)
The BHV VIVID (etafilcon A) Soft (Hydrophilic) Contact Lens is indicated for daily wear for the correction of distance, intermediate and near vision in presbyopic, phakic or pseudo-phakic persons with non-diseased eyes who can have up to approximately 1.50 diopters (D) of astigmatism.
The BHV VIVID (etafficon A) Soft (Hydrophilic) Contact Lens helps protect against transmission of harmful UV radiation to the cornea and into the eye.
Eye care practitioners may prescribe the lenses either for single use disposable wear or frequent/planned replacement wear with cleaning, disinfection, and scheduled replacement (see "Wearing Schedule").
When prescribed for single-use disposable wear (see "Wearing Schedule") the BHV VIVID (etafilcon A) Soft (Hydrophilic) Contact Lens is to be discarded after each removal. Therefore no cleaning or disinfection is required. When prescribed for frequent/planned replacement wear the BHV VIVID (etafilcon A) Soft (Hydrophilic) Contact Lens is to be cleaned, rinsed, and disinfected each time the lens is removed. The BHV VIVID (etafilcon A) Soft (Hydrophilic) Contact Lens contact lens is to be discarded after the recommended wearing period as prescribed by the eye care practitioner. When prescribed for frequent/planned replacement wear, the lenses should be disinfected using a chemical disinfection system only (i.e., not heat).
Type of Use (Select one or both, as applicable)
2 Prescription Use (Part 21 CFR 801 Subpart D)
_ Over-The-Counter Use (21 CFR 801 Subpart C)
### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
### FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Druq Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for Brien Holden Vision. The logo consists of a blue gemstone on the left, followed by the letters "BHV" in a bold, sans-serif font. Below the letters, the words "brien holden vision" are written in a smaller, lighter font. The gemstone is a multifaceted, round shape with varying shades of blue.
# 510(k) Summary
- l. Submitter
ll.
| Manufacturer Name: | Brien Holden Vision Pty Ltd. |
|--------------------|------------------------------------------------------------------------------------------------------------------------------------|
| Address: | Level 3, North Wing, Rupert Myers Building<br>Gate 14 Barker Street<br>University of New South Wales<br>Sydney, NSW 2052 Australia |
| Phone: | +61 2 9385 7410 |
| Fax: | +61 2 9385 7401 |
| Contact Person | Randall May |
| Date Prepared: | 11 November 2014 |
| Device | |
| Trade Name: | BHV VIVID (etafilcon A) Soft (Hydrophilic)<br>Contact Lens |
| Common Name: | Daily Wear Soft (Hydrophilic) Contact Lens |
| Device Class | Class II; Ophthalmic therapeutic device |
| Classification: | 21 CFR 886.5925 |
| Product Codes: | LPL, MVN |
#### III. Predicate Device
Primary Predicate
VISTAKON® (etafilcon A) Soft (Hydrophilic) Contact Lens, Clear and Tinted (Visibility and/or Cosmetically) with UV Blocker for Daily Wear (K062614)
This predicate has not been subject to a design-related recall.
Reference Device
PEGAVISION CORPORATION Aquamax (Etafilcon A) Bi-Weekly and Daily Disposable Soft (Hydrophilic) Contact Lenses (K120028)
This reference device has not been subject to a design-related recall.
Brien Holden Vision Pty Ltd ABN 85 138 617 062 Level 3 North Wing, Rupert Myers Building Gate 14 Barker Street, UNSW
Sydney NSW 2052 Australia Tel +61 2 9385 7410 www.brienholdenvision.com
Image /page/3/Picture/13 description: The image contains a logo with the word "vision" in a large, bold, blue font. Below "vision" is the phrase "for everyone...everywhere" in a smaller, gray font. The word "everywhere" is in blue, and there is a trademark symbol to the right of the word.
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image is a logo for Brien Holden Vision. The logo features a blue gemstone on the left, followed by the letters "BHV" in a bold, sans-serif font, also in blue. Below the letters, the words "brien holden vision" are written in a smaller, lighter blue font.
#### Device Description IV.
The BHV VIVID (etafilcon A) Soft (Hydrophilic) Contact Lens is an extended depth-of-focus (EDOF) contact lens for the daily wear correction of distance, intermediate and near vision in presbyopic persons.
The lens material, etafilcon A. is a random co-polymer of 2-hydroxyethyl methacrylate (HEMA) and methacrylic acid (MAA), which were cross-linked with ethylene qlycol dimethacrylate (EGDMA) and 1,1,1 -trimethylolpropane trimethacrylate (TMPTMA) via photo-polymerization. These lenses are tinted blue using color additive Reactive Blue No. 19 to make them more visible for handling. These lenses contain UV blocker, a benzotriazole UV absorbing monomer to block UV radiation. The average transmittance characteristics of theses lenses are less than 5 % in the UVB range of 280-315 nm and less than 30% in the UVA range of 315-380 nm. Lenses are supplied sterile in sealed blister package containing sterile isotonic borate buffered saline solution.
#### V. Indications for Use
The BHV VIVID (etafilcon A) Soft (Hydrophilic) Contact Lens is indicated for daily wear for the correction of distance, intermediate and near vision in presbyopic, phakic or pseudo-phakic persons with non-diseased eyes who can have up to approximately 1.50 diopters (D) of astigmatism.
The BHV VIVID (etafilcon A) Soft (Hydrophilic) Contact Lens helps protect against transmission of harmful UV radiation to the cornea and into the eye.
Eve care practitioners may prescribe the lenses either for single use disposable wear or frequent/planned replacement wear with cleaning, disinfection, and scheduled replacement (see "Wearing Schedule").
When prescribed for single-use disposable wear (see "Wearing Schedule") the BHV VIVID (etafilcon A) Soft (Hydrophilic) Contact Lens is to be discarded after each removal. Therefore no cleaning or disinfection is required.
When prescribed for frequent/planned replacement wear the BHV VIVID (etafilcon A) Soft (Hydrophilic) Contact Lens is to be cleaned, rinsed, and disinfected each time the lens is removed. The BHV VIVID (etafilcon A) Soft (Hydrophilic) Contact Lens contact lens is to be discarded after the recommended wearing period as prescribed by the eye care practitioner. When prescribed for frequent/planned replacement wear, the lenses should be disinfected using a chemical disinfection system only (i.e., not heat).
Image /page/4/Picture/10 description: The image shows the word "vision" in a large, bold, blue font. Below the word "vision" is the phrase "for everyone...everywhere" in a smaller, gray font. The phrase is spread out horizontally, taking up about the same width as the word "vision" above it. The overall design is simple and modern.
Brien Holden Vision Pty Ltd _evel 3 North Wing. Rupert Myers Buildin Sydney NSW 2052 Australia Tel +61 2 9385 7410 ww.brienholdenvision.co
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image is a logo for Brien Holden Vision. The logo consists of a blue diamond on the left, followed by the letters "BHV" in blue. Below the letters is the text "brien holden vision" in a smaller blue font. The logo is clean and modern, and the use of blue gives it a sense of trust and professionalism.
#### VI. Predicate Device Comparison
The BHV VIVID (etafilcon A) Soft (Hydrophilic) Contact Lenses have similar intended use indications and technological characteristics of the primary predicate device and reference device listed above.
### Table 8.1 Subject and Primary Predicate Device Comparison
| Characteristic | BHV VIVID (etafilcon A)<br>Soft (Hydrophilic)<br>Contact Lenses | VISTAKON® (etafilcon<br>A) Soft (hydrophilic)<br>Contact Lens, Clear and<br>Tinted (Visibility and/or<br>Cosmetically) with UV<br>Blocker for Daily Wear<br>[Primary Predicate] |
|---------------------------|-----------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k) Number | Subject Device | K062614 |
| FDA Group | Group IV > 50% water,<br>ionic polymer | Group IV > 50% water,<br>ionic polymer |
| USAN Name | etafilcon A | etafilcon A |
| Production Method | Cast Molding | Cast Molding |
| Water Content | 58% | 58% |
| Refractive Index | 1.402 | 1.40 |
| Power (Diopter) | +20.00D to -20.00D | +20.00D to -20.00D |
| Base Curve Radius<br>(mm) | 7.85 to 10.0 | 7.85 to 10.0 |
| Diameter (mm) | 12.0 to 15.0 | 12.0 to 15.0 |
| Centre Thickness<br>(mm) | Varies with power: 0.060<br>to 0.500 mm | Varies with power: 0.060<br>to 1.00 mm |
The subject device complies with the following recognised consensus standards:
- . ISO 18369-1 First edition 2006-08-15 Ophthalmic optics -- Contact lenses -- Part 1: Vocabulary, classification system and recommendations for labeling specifications [Including: Amendment 1 (2009)]
- ISO 18369-2 Second edition 2012-12-01 Ophthalmic optics -- Contact lenses -- Part 2: Tolerances
- ISO 18369-3 First edition 2006-08-15 Ophthalmic optics Contact lenses ● - Part 3: Measurement Methods
Image /page/5/Picture/11 description: The image contains a logo for "vision for everyone...everywhere". The word "vision" is in a large, bold, blue font. Below it, the phrase "for everyone...everywhere" is in a smaller, gray font. The TM symbol is in the upper right corner of the word "everywhere".
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for Brien Holden Vision. The logo consists of a blue gemstone on the left, followed by the letters "BHV" in blue. Below the letters is the text "brien holden vision" in a smaller font, also in blue. The logo is clean and modern, with a focus on the company's name and a visual representation of clarity and precision through the gemstone.
- ISO 18369-4 First edition 2006-08-15 Ophthalmic optics contact lenses - Part 4: Physicochemical properties of contact lens materials
- ASTM F1980 07 (Reapproved 2011) Standard Guide for Accelerated . Aging of Sterile Barrier Systems for Medical Devices
- . ISO 10993-1:2009 Biological evaluation of medical devices – Part 1: Evaluation and testing within a risk management process
- . ISO 10993-5:2009 Biological evaluation of medical devices – Part 5: Tests for in vitro cytotoxicity
- . ISO 10993-10:2010 Biological evaluation of medical devices – Part 10: Tests for irritation and skin sensitization
- . ISO 11607-1:2006 Packaging for terminally sterilized medical devices -Part 1: Requirements for materials, sterile barrier systems and packaging systems
- . ISO 11607-2:2006 Packaging for terminally sterilized medical devices -Part 2: Validation requirements for forming, sealing and assembly processes
- . ISO 11737-2:2009 Sterilization of medical devices - Microbiological methods - Part 2: Test of sterility performed in the definition, validation and maintenance of a sterilization process
- . ISO 17665-1:2006 Sterilization of health care products - Moist heat -Part 1: Requirements for the development, validation and routine control of a sterilization process for medical devices
#### Summary of Non-clinical Testing VII.
All tests were conducted in accordance with the FDA Premarket Notification (510(k)) Guidance Document for Daily Wear Contact lenses, May 12, 1984:
- Bench Testing
- . Biocompatibility
- . Microbiology
- Bacteriostatic Validation
- Leachables
## VIII. Clinical Testing
The technological characteristics, formulation, manufacturing and sterilization processes are substantially equivalent to the predicate devices, therefore no clinical data is required.
#### IX. Conclusions
The BHV VIVID (etafilcon A) Soft (Hydrophilic) Contact Lens is substantially equivalent to the predicate devices for the labeled indications for use. Successful results from chemical/physical, stability, and toxicology tests further confirmed the lenses are within established finished product specifications, remain stable, and are non-toxic and biocompatible with the ocular environment.
Image /page/6/Picture/23 description: The image contains the word "vision" in a large, bold, blue font. Below the word "vision" is the phrase "for everyone...everywhere" in a smaller, gray and blue font. The word "for everyone" is in gray, while the word "everywhere" is in blue. There is a trademark symbol next to the word "everywhere".
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.