K143270 · Ge Hualun Medical Systems Co. , Ltd. · KPR · Feb 18, 2015 · Radiology
Device Facts
Record ID
K143270
Device Name
Optima XR646
Applicant
Ge Hualun Medical Systems Co. , Ltd.
Product Code
KPR · Radiology
Decision Date
Feb 18, 2015
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1680
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The Optima XR646 is intended to generate digital radiographic images of the skull, spinal column, chest, abdomen, extremities, and other body parts in patients of all ages. Applications can be performed with the patient sitting, standing, or lying in the prone or supine position and is intended for use in all routine radiography exams. The device is not intended for mammographic applications.
Device Story
Stationary digital radiographic X-ray system; generates images of human anatomy. Components: WallStand, elevating table, overhead tube support, X-ray tube, collimator, system controller, X-ray generator, wireless/tethered flat-panel digital detector (amorphous silicon/cesium iodide scintillator). Used in clinical settings by healthcare professionals. Inputs: X-ray radiation attenuated by patient anatomy. Transformation: Detector converts X-rays to digital signals; system processes data for display, storage, and printing via DICOM network. Output: Digital radiographic images. Clinical utility: Assists physicians in diagnostic evaluation of skeletal and soft tissue structures. Benefits: Enables routine diagnostic imaging across diverse patient populations.
Clinical Evidence
No clinical data. Substantial equivalence supported by bench testing, including module verification, system integration testing, and simulated use validation.
Indicated for patients of all ages requiring routine digital radiographic imaging of the skull, spinal column, chest, abdomen, and extremities. Not for mammographic use.
Regulatory Classification
Identification
A stationary x-ray system is a permanently installed diagnostic system intended to generate and control x-rays for examination of various anatomical regions. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
Special Controls
*Classification.* Class II (special controls). A radiographic contrast tray or radiology diagnostic kit intended for use with a stationary x-ray system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized depiction of three human profiles facing right, arranged in a stacked formation. The profiles are rendered in a dark color, creating a silhouette effect. The logo is encircled by the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" in all capital letters.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
January 26, 2016
GE Hualun Medical Systems Co., Ltd. Kennv Ma Regulatory Affairs Manager - X ray No.1, Yong Chang North Road Economic & Technology Development Zone Beijing, 100176 CHINA
Re: K143270
Trade/Device Name: Optima XR646 Regulation Number: 21 CFR 892.1680 Regulation Name: Station X-ray System Regulatory Class: Class II Product Code: KPR, MQB Dated (February 3, 2015): Received (February 5, 2015):
Dear Kenny Ma:
This letter corrects our substantially equivalent letter of February 3, 2015.
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must
{1}------------------------------------------------
comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Astest Oclo
Robert Ochs, Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
#### Indications for Use
510(k) Number (if known) K143270
Device Name Optima XR646
Indications for Use (Describe)
The Optima XR646 is intended to generate digital radiographic images of the skull, spinal column, chest, abdomen, extremities, and other body parts in patients of all ages. Applications can be performed with the patient sitting, standing, or lying in the prone or supine position and is intended for use in all routine radiography exams.
The device is not intended for mammographic applications.
| Type of Use (Select one or both, as applicable) | |
|------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------|
| <span style="padding-right: 5px;"> ☑ </span> Prescription Use (Part 21 CFR 801 Subpart D) | <span style="padding-right: 5px;"> ☐ </span> Over-The-Counter Use (21 CFR 801 Subpart C) |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for General Electric (GE). The logo consists of the letters "GE" written in a stylized script, enclosed within a blue circle. The circle has a white border and a few white swirls around the edge, giving it a dynamic and recognizable appearance.
510(k) Premarket Notification Submission
#### 510(k) Summary
In accordance with 21 CFR 807.92 the following summary of information is provided:
| Date: | October 29, 2014 |
|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Submitter: | GE Healthcare, (GE HUALUN MEDICAL SYSTEMS CO. Ltd) |
| | No.1 Yong Chang North Road, Beijing Economic<br>Technological Development Zone BEIJING 100176 CHINA |
| Primary Contact Person: | Kenny Ma |
| | Regulatory Affairs Manager |
| | GE Healthcare, (GE HUALUN MEDICAL SYSTEMS CO. Ltd) |
| | Office: +8610 59347159<br>eFax: +8610 67882574<br>Email: Kenny.Ma@ge.com |
| Secondary Contact<br>Person: | Christopher Paulik |
| | RA Leader - Product |
| | GE Healthcare, (GE Medical Systems, LLC) |
| | +1 262 548 2010<br>Email: Christopher.A.Paulik@ge.com |
| Device Trade Name: | Optima XR646 |
| Common/Usual Name: | Digital Radiographic X-Ray System |
| Classification Names:<br>Product Code: | Stationary X-Ray System (21CFR § 892.1680), solid state x-<br>ray imager (21CFR § 892.1680)<br>KPR, MQB |
| Predicate Device(s): | Stationary X-Ray System, MODEL: Discovery XR656<br>(K132261) and Revolution XR/d with Image Pasting<br>(K050704) |
| Device Description: | The Optima XR646 remains a radiographic X-ray system<br>capable of generating radiographic images of human<br>anatomy. |
| | The Optima XR646 is designed to handle radiographic<br>applications using GE's flat-panel wireless digital detector.<br>The digital detector is comprised of amorphous silicon and<br>cesium iodide scintillator. The resulting digital image can<br>be sent through a DICOM network for applications such as |
| printing, viewing and storage. | |
| The Optima XR646 Digital Radiographic Imaging system<br>consists of a WallStand, elevating table, overhead Tube<br>support, X-ray tube, collimator, system controller, X-ray<br>generator, and wireless or tethered digital detector.<br>Various configurations such as Table only, WallStand only,<br>or OTS only are available to meet customer radiographic<br>requirements. | |
| Intended Use: | General Purpose Digital Radiographic Imaging System |
| Indication for Use: | The Optima XR646 is intended to generate digital<br>radiographic images of the skull, spinal column, chest,<br>abdomen, extremities, and other body parts in patients of<br>all ages. Applications can be performed with the patient<br>sitting, standing, or lying in the prone or supine position<br>and is intended for use in all routine radiography exams. |
| | The device is not intended for mammographic<br>applications. |
| Technology: | The Optima XR646 employs the same fundamental<br>scientific technology as its predicate device. |
| Determination of<br>Substantial Equivalence: | Summary of Non-Clinical Tests:<br>The Optima XR646 is based on the software and Hardware<br>platform of the predicate device. It was designed and is<br>manufactured under GE's quality system that meets the<br>Quality System Regulations of 21 CFR 820 and ISO 13485.<br>Optima XR646 is certified to comply with the X-ray<br>requirements of 21 CFR Subchapter J - Radiological<br>Health, as well as safety requirements of IEC 60601-1 and<br>associated collateral and particular standards.<br>The Optima XR646 and its applications comply with<br>voluntary standards. The following quality assurance<br>measures were applied to the development of the system:<br>■ Risk Analysis<br>■ Requirements Reviews<br>■ Design Reviews<br>■ Testing on unit level (Module verification)<br>■ Integration testing (System verification)<br>■ Performance testing (Verification) |
| Safety testing (Verification) Simulated use testing (Validation) | |
| Minor obsolescence and feature changes have occurred in the evolution of the device for which a 510(k) was not required based on evaluation using the FDA "Deciding When to Submit a 510(k) for a change to an Existing Device". These include changes in operator display and the reduction of option in overhead rails. The changes that precipitated this 510(k) submission were the reduction in auto positioning axes of the Overhead Tube Support that resulted in a change in the sequence of acquisitions for the Image Pasting feature. | |
| The Image Pasting feature change does not affect the functional equivalence of the product today. These differences do not raise new issues of safety and effectiveness. The Overhead Tube Support hardware change and the Overhead Tube Support movement control software change for Image Pasting function were verified and validated through bench testing. All the design inputs requirements were successfully tested for Optima XR646 with Image Pasting and meet the design specifications and user requirements. This Verification and Validation bench testing demonstrated that no adverse effects have been introduced by the changes. The control mechanism, operating principle, energy type and Intended use have not changed. The indication for use is slightly modified to reflect the removal of the VolumeRad feature from this version of the product also due to the reduction of auto positioning axes on the Overhead Tube Support. | |
| Based on the successful verification, validation, and bench testing and the above comparison, GE Healthcare believes that the Optima XR646 is as safe and effective and performs in a substantially equivalent manner to the predicate device. | |
| Summary of Clinical Tests: The subject of this premarket submission, Optima XR646, did not require clinical studies to support substantial equivalence. All changes were verified and validated during bench testing and did not reveal any new questions of safety and effectiveness. | |
| Conclusion: | GE Healthcare considers the Optima XR646 to be as safe<br>and effective, and performance is substantially equivalent<br>to the predicate device. |
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the General Electric (GE) logo. The logo consists of the letters "GE" in a stylized, cursive font, enclosed within a blue circle. The letters are white, and the circle has a white outline, creating a contrast that makes the logo stand out.
# 510(k) Premarket Notification Submission
{5}------------------------------------------------
510(k) Premarket Notification Submission
Image /page/5/Picture/2 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'GE' in a stylized, cursive font, enclosed within a blue circle. There are three white, flame-like shapes surrounding the circle, giving the impression of movement or energy.
{6}------------------------------------------------
Image /page/6/Picture/1 description: The image shows the General Electric (GE) logo. The logo consists of the letters "GE" in a stylized, cursive font, enclosed within a blue circle. There are three wave-like shapes emanating from the circle, positioned at the top, bottom, and right side of the circle. The logo is simple and recognizable, representing the multinational conglomerate corporation.
# 510(k) Premarket Notification Submission
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.