K143254 · Siemens Medical Solutions USA, Inc. · LLZ · Dec 10, 2014 · Radiology
Device Facts
Record ID
K143254
Device Name
eSie Apps Suite
Applicant
Siemens Medical Solutions USA, Inc.
Product Code
LLZ · Radiology
Decision Date
Dec 10, 2014
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.2050
Device Class
Class 2
Attributes
Software as a Medical Device, 3rd-Party Reviewed
Indications for Use
eSie Apps Suite software is a software-only product to be run on a user's PACS (Picture Archiving and Communication System) workstation. It is intended to launch Siemens CAPs (Clinical Application Packages) for image processing, including the acceptance, transfer, display and digital processing of ultrasound images. Digital processing includes image manipulation and quantification on a workstation. Use of a clinical application package by a qualified clinician can add information to the study to be used for a clinical diagnosis. The software supports the following clinical application packages: eSie Volume Viewer eSie LVA eSie PISA eSie Valves
Device Story
eSie Apps Suite is a software-only PACS plug-in; hosts Clinical Application Packages (CAPs) for ultrasound image analysis. Input: ultrasound image data sets from PACS. Operation: launched via command line on PACS workstation; shares patient context; utilizes SC2000 renderer for volume display/manipulation. Output: processed images and quantification results appended to patient study. Used in clinical settings by clinicians. Benefits: enables advanced viewing, manipulation, and quantification of ultrasound data (e.g., volume, LVA, PISA, valves) within existing PACS workflow; supports clinical decision-making by providing quantitative anatomical information.
Clinical Evidence
No clinical data. Bench testing only; verification and validation performed per design control processes, including DICOM and IEC 62304 compliance.
Indicated for use by qualified clinicians to perform image processing, manipulation, and quantification of ultrasound images on a PACS workstation to assist in clinical diagnosis.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
ACUSON SC2000 Diagnostic Ultrasound System Software (K142628)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized symbol featuring three human profiles facing to the right, arranged in a cascading manner.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
Siemens Medical Solutions USA, Inc. % Mr. Mark Job Responsible Third Party Official 1394 25th Street, NW BUFFALO MN 55313
December 10, 2014
Re: K143254
Trade/Device Name: eSie Apps Suite Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: II Product Code: LLZ Dated: November 12, 2014 Received: November 13, 2014
Dear Mr. Job:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21. Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Robert A Ochs
for
Janine M. Morris Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
**Indications for**
510(k) Number (if known) K143254
Device Name eSie Apps Suite
Indications for Use (Describe)
eSie Apps Suite software is a software-only product to be run on a user's PACS (Picture Archiving and Communication System) workstation. It is intended to launch Siemens CAPs (Clinical Application Packages) for image processing, including the acceptance, transfer, display and digital processing of ultrasound images. Digital processing includes image manipulation and quantification. Use of a clinical application package by a qualified clinician can add information to the study to be used for a clinical diagnosis.
The software supports the following clinical application packages:
· eSie Volume Viewer
· eSie LVA
· eSie PISA
· eSie Valves
Type of Use (Select one or both, as applicable)
> Prescription Use {Part 21 CFR 801 Subpart D}
Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
### 5. 510(K) SUMMARY
| Submission Date | October 30, 2014 |
|-----------------|-------------------------------------------------------------------------------------------------------------------------|
| Sponsor | Siemens Medical Solutions, Inc.,<br>Ultrasound Division<br>685 East Middlefield Road<br>Mountain View, California 94043 |
| Contact Person | Nancy Burke<br>Telephone: (425) 295-8665<br>Fax: (425) 391-9161 |
| Device Name | eSie Apps Suite. Note that for marketing purposes |
, the names eSie Apps Suite, syngo® US Apps Suite or syngo® Ultrasound Apps Suite may be used.
Common Name System, Image Processing, Radiological
#### Classification
Regulatory Class: Review Category: Tier II Classification Panel: Radiology
Picture Archiving and Communications System FR #892.2050 Product Code 90-LLZ
### Legally Marketed Predicate Devices
The eSie Apps Suite as described in this 510k is substantially equivalent to the company's current legally marketed devices shown below:
| Predicate Device | 510(k) Number | Clearance Date |
|--------------------------------------------------------|---------------|----------------|
| eSie Apps Suite | K141554 | 7/11/2014 |
| ACUSON SC2000 Diagnostic<br>Ultrasound System Software | K142628 | 10/10/2014 |
### 4. Device Description:
eSie Apps Suite is intended to be the Clinical Application Package (CAP) host for 2D and volume imaging applications on a PACS workstation. It is intended to maximize the reuse of the SC2000 renderer for volume display and manipulation. Additionally, the imaging applications from the SC2000 will be redeployed on a PACS workstation for the 2D and volume imaging analysis.
eSie Apps Suite is intended to have a simple basic configuration as a PACS plug-in by utilizing the third party launching capability of the host PACS. On the customer's workstation a command line will launch the eSie Apps Suite application - patient context will be shared between the PACS and eSie Apps Suite. Results created by the respective CAPs will be sent back to the PACS for appending to the patient study.
{4}------------------------------------------------
The software level of concern for the eSie Apps Suite is considered moderate.
### 5. Intended Use
eSie Apps Suite software is a software-only product to be run on a user's PACS (Picture Archiving and Communication System) workstation. It is intended to launch Siemens CAPs (Clinical Application Packages) for image processing, including the acceptance, transfer, display and digital processing of ultrasound images. Digital processing includes image manipulation and quantification on a workstation. Use of a clinical application package by a qualified clinician can add information to the study to be used for a clinical diagnosis.
The software supports the following clinical application packages:
- . eSie Volume Viewer
- eSie LVA ●
- eSie PISA
- eSie Valves ●
### 6. Technological Characteristics as Compared to Predicate Devices
The eSie Apps Suite software in this submission is essentially equivalent to the cleared eSie Apps Suite software (K141554). eSie Apps Suite software is activated on a workstation connected to a cardiology PACS that includes a third-party application launcher. It is a graphical launch pad designed to provide viewing, manipulation and quantification functionality for ACUSON SC2000 ultrasound image data sets using Clinical Application Packages (CAPS).
The analysis packages included in this release of eSie Apps Suite are essentially the same as those available in the software on the ACUSON SC2000 Ultrasound System (K142628), except that the eSie Apps Suite CAPS are limited to eSie Volume Viewer, eSie LVA, eSie PISA, and with this submission, eSie Valves.
| Feature /<br>Characteristic | eSie Apps Suite | eSie Apps Suite<br>(Primary Predicate) | ACUSON SC2000<br>Ultrasound System |
|----------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k) Reference | This submission | K141554 | K142628 |
| Product Code(s) | LLZ | LLZ | IYN, IYO, ITX, OBJ |
| Feature /<br>Characteristic | eSie Apps Suite | eSie Apps Suite<br>(Primary Predicate) | ACUSON SC2000<br>Ultrasound System |
| Indications for<br>Use Statement | eSie Apps Suite software is<br>a software-only product to<br>be run on a user's PACS<br>(Picture Archiving and<br>Communication System)<br>workstation. It is intended<br>to launch Siemens CAPs<br>(Clinical Application<br>Packages) for image<br>processing, including the<br>acceptance, transfer,<br>display and digital<br>processing of ultrasound<br>images. Digital processing<br>includes image<br>manipulation and<br>quantification on a<br>workstation. Use of a<br>clinical application package<br>by a qualified clinician can<br>add information to the study<br>to be used for a clinical<br>diagnosis.<br>The software supports the<br>following clinical application<br>packages:<br>eSie Volume<br>Viewer eSie LVA eSie PISA eSie Valves | eSie Apps Suite software is<br>a software-only product to<br>be run on a user's PACS<br>(Picture Archiving and<br>Communication System)<br>workstation. It is intended<br>to launch Siemens CAPs<br>(Clinical Application<br>Packages) for image<br>processing, including the<br>acceptance, transfer,<br>display and digital<br>processing of ultrasound<br>images. Digital processing<br>includes image<br>manipulation and<br>quantification on a<br>workstation. Use of a<br>clinical application package<br>by a qualified clinician can<br>add information to the study<br>to be used for a clinical<br>diagnosis.<br>The software supports the<br>following clinical application<br>packages:<br>eSie Volume<br>Viewer eSie LVA eSie PISA | The SC2000 ultrasound<br>imaging system is<br>intended for the following<br>applications: Cardiac,<br>Neo-natal and Fetal<br>Cardiac, Pediatric,<br>Transesophageal, Adult<br>Cephalic, Peripheral<br>Vessel, Abdominal,<br>Intraoperative<br>Abdominal, Musculo-<br>skeletal Conventional,<br>and Musculo-skeletal<br>Superficial applications.<br>The system also<br>provides the ability to<br>measure anatomical<br>structures and<br>calculation packages that<br>provide information to the<br>clinician that may be<br>used adjunctively with<br>other medical data<br>obtained by a physician<br>for clinical diagnosis<br>purposes.<br>The system has Cardiac<br>Measurements and<br>Calculation Packages<br>that provide information<br>that may be used<br>adjunctively with other<br>medical data obtained by<br>a physician for clinical<br>diagnosis purposes. |
| Display, Measurement and Calculation Packages | | | |
| Viewing and<br>manipulation of<br>volume datasets | eSie Volume Viewer | eSie Volume Viewer | Yes |
| 2D quantitative<br>tool for<br>assessment of<br>global and<br>regional<br>myocardial<br>mechanics | N/A | N/A | eSie VVI |
| 3D Left Ventricle<br>volume<br>quantitative<br>analyses (Single +<br>Multibeat) | eSie LVA | eSie LVA | eSie LVA |
| Feature /<br>Characteristic | eSie Apps Suite | eSie Apps Suite<br>(Primary Predicate) | ACUSON SC2000<br>Ultrasound System |
| Quantification of<br>proximal<br>isovelocity surface<br>area | eSie PISA | eSie PISA | eSie PISA 3D semi<br>automatic |
| 2D Automated<br>tool to identify and<br>measure contours<br>of left ventricle<br>and atrium from<br>transthoracic<br>exams | N/A | N/A | eSie Left Heart |
| Visualize and<br>quantify mitral and<br>aortic valve<br>anatomy | eSie Valves | N/A | eSie Valves |
(Table continued on next page)
{5}------------------------------------------------
{6}------------------------------------------------
# 7. A brief discussion of nonclinical tests submitted, referenced, or relied on in the 510(k) for a determination of substantial equivalence.
The eSie Apps Suite is designed, verified, and validated according to the company's design control process and has been subjected to extensive safety and performance testing before release. Final testing of the eSie Apps Suite included various safety and performance testing designed to ensure the device meets all of its specifications including:
- . DICOM (Digital Imaging and Communications in Medicine)
- . IEC 62304 Medical device software - Software Life Cycle Process
# 8. A summary discussion of the clinical tests submitted, referenced, or relied on for a determination of substantial equivalence.
Since the eSie Apps Suite in this submission use the same technology and principles as existing devices, clinical data is not required.
# 9. Conclusion
As shown by the device comparison table above, the modified eSie Apps Suite software has the same intended use as the predicate device, incorporates technological features of the predicate devices cleared through premarket notification and testing indicates that no new issues of safety or effectiveness are raised.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.