K143253 · Philips Medical Systems Nederland B.V. · LNH · Apr 9, 2015 · Radiology
Device Facts
Record ID
K143253
Device Name
O-MAR
Applicant
Philips Medical Systems Nederland B.V.
Product Code
LNH · Radiology
Decision Date
Apr 9, 2015
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1000
Device Class
Class 2
Indications for Use
O-MAR is a combination of an acquisition technique and post-processing software intended for use on Achieva and Ingenia, 1.5T & 3T MR Systems. O-MAR is suitable for use on all patients with passive MR Conditional orthopedics implants that are scanned according to the conditions of safe use for the specific MR Conditional implant being scanned. In addition O-MAR is suitable for use on patients without implants that are cleared for MR exams. O-MAR helps reduce artifacts caused by presence of metal in both in-plane and through-plane dimensions compared to conventional MR imaging techniques. Thus O-MAR improves visualization of more tissue in the vicinity of MR Conditional orthopedic implants. When interpreted by a trained physician, images generated by O-MAR provide information that can be useful in determining a diagnosis.
Device Story
O-MAR is an MR acquisition and post-processing software feature for Philips Achieva and Ingenia 1.5T/3T systems. It utilizes two primary methods: SEMAC+VAT (Slice Encoding for Metal Artifact Correction + View Angle Tilting) and MARS+VAT (Metal Artifact Reduction Sequence + View Angle Tilting). SEMAC+VAT uses multiple z-encodings per excited slice to recover off-resonant signals and reduce both in-plane and through-plane distortions; MARS+VAT uses high-bandwidth Turbo Spin Echo (TSE) to reduce in-plane distortions. The system employs a TSE-based SENSE reference scan for robustness against metal-induced field inhomogeneities. Operated by radiology staff, the device integrates into standard MR workflows via specific ExamCards. The output consists of distortion-corrected MR images, which are viewed and interpreted by a radiologist to improve tissue visualization near orthopedic implants, aiding in clinical diagnosis.
Clinical Evidence
Clinical validation involved volunteers with knee, hip, or lumbar spine implants. Comparison of MARS+VAT and SEMAC+VAT against high-bandwidth TSE sequences demonstrated improved tissue visualization and reduced artifact size. A board-certified radiologist confirmed superior performance of O-MAR compared to the cleared high-bandwidth TSE sequence. No new hazards or product defects were identified.
Technological Characteristics
Software-based MR acquisition and reconstruction feature. Operates on HP Z420 host computer (Windows 7, 64-bit). Employs TSE-based pulse sequences, VAT gradient control, and multi-volume frequency-shifted signal combination. Compatible with 1.5T and 3T Achieva/Ingenia systems. No hardware changes required. Not designed for external network connectivity.
Indications for Use
Indicated for patients with passive MR Conditional orthopedic implants or patients without implants undergoing MR exams on 1.5T or 3T Achieva and Ingenia systems.
Regulatory Classification
Identification
A magnetic resonance diagnostic device is intended for general diagnostic use to present images which reflect the spatial distribution and/or magnetic resonance spectra which reflect frequency and distribution of nuclei exhibiting nuclear magnetic resonance. Other physical parameters derived from the images and/or spectra may also be produced. The device includes hydrogen-1 (proton) imaging, sodium-23 imaging, hydrogen-1 spectroscopy, phosphorus-31 spectroscopy, and chemical shift imaging (preserving simultaneous frequency and spatial information).
Special Controls
*Classification.* Class II (special controls). A magnetic resonance imaging disposable kit intended for use with a magnetic resonance diagnostic device only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the seal of the U.S. Department of Health and Human Services. The seal features a stylized caduceus, a symbol often associated with medicine and healthcare, consisting of a staff with two snakes coiled around it. The seal is encircled by the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" in a circular arrangement. The image is in black and white.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
April 9, 2015
Philips Medical Systems, Nederland B.V. % Ms. Susan Ouick Regulatory Affairs Specialist 595 Miner Road CLEVELAND OH 44143
Re: K143253
Trade/Device Name: O-MAR Regulation Number: 21 CFR 892.1000 Regulation Name: Magnetic resonance diagnostic device Regulatory Class: II Product Code: LNH Dated: March 23, 2015 Received: March 24, 2015
Dear Ms. Quick:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21. Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
Michael D'Hara
for
Robert Ochs, Ph.D. Acting Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
| DEPARTMENT OF HEALTH AND HUMAN SERVICES<br>Food and Drug Administration<br><b>Indications for Use</b> | Form Approved: OMB No. 0910-0120<br>Expiration Date: January 31, 2017<br>See PRA Statement below. |
|-------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k) Number (if known) | K143253 |
| Device Name | O-MAR |
| Indications for Use (Describe) | O-MAR is a combination of an acquisition technique and post-processing software intended for use on Achieva and Ingenia, 1.5T & 3T MR Systems. O-MAR is suitable for use on all patients with passive MR Conditional orthopedics implants that are scanned according to the conditions of safe use for the specific MR Conditional implant being scanned. In addition O-MAR is suitable for use on patients without implants that are cleared for MR exams. O-MAR helps reduce artifacts caused by presence of metal in both in-plane and through-plane dimensions compared to conventional MR imaging techniques. Thus O-MAR improves visualization of more tissue in the vicinity of MR Conditional orthopedic implants. When interpreted by a trained physician, images generated by O-MAR provide information that can be useful in determining a diagnosis. |
| Type of Use (Select one or both, as applicable) | <label><input checked="checked" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D)</label> <label><input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C)</label> |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
***DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.***
The burden time for this collection of information is estimated to average 79 hours per response, including the
time to review instructions, search existing data sources, gather and maintain the data needed and complete
and review the collection of information. Send comments regarding this burden estimate or any other aspect
of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of
information unless it displays a currently valid OMB number."
FORM FDA 3881 (8/14) the provided text is not in English, so I cannot correct it. I will mark it as
{3}------------------------------------------------
# PHILIF
# Philips Medical Systems Nederland B.V.
## 510(k) Summary
## O-MAR
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of 21 CFR 807.92.
- I. General Information 21 CFR 807.92 (a)(1), (2)
| Company Name: | Philips Medical Systems Nederland B.V. |
|-----------------------|-------------------------------------------------------------------------------------------------------------------------------------|
| Address: | Veenpluis 4-6<br>5684 PC Best<br>The Netherlands |
| Registration No .: | 3003768277 |
| Contact Person: | Susan Quick<br>595 Miner Rd<br>Cleveland, Oh 44143<br>Tel: (440)-483-2291<br>Fax: (440)-483-4799<br>E-mail: susan.quick@philips.com |
| Prepared (date): | 2014 November XX |
| Trade Name of Device: | O-MAR |
| Classification: | Class II |
| Regulatory Section: | Magnetic Resonance Diagnostic Device.<br>892.1000 |
| Product Code: | 90LNH |
{4}------------------------------------------------
510(k) Summary O-MAR
### 21 CFR 807.92(a)(3):Legally marketed predicate device to which substantial equivalence is claimed:
| 1. Primary Predicate Device: | MAVRIC SL |
|------------------------------|---------------|
| Manufacturer: | GE Healthcare |
| Predicate Device k#: | K113394 |
#### 21 CFR 807.92(a)(4): Description of the device that is the subject of this premarket notification:
#### Summary of functions of the device and its major components
The O-MAR feature has two components. The SEMAC+VAT feature and the MARS+VAT feature. SEMAC+VAT is a Turbo Spin Echo method in combination with VAT (View Angle Tilting) and with multiple z-encodings per excited slice (aka SEMAC) to reduce in-plane and through-plane distortions caused by magnetic field inhomogeneities. MARS is high band width TSE. MARS+VAT can also be referred to as high band width TSE+VAT.
A difference between SEMAC+VAT and MARS+VAT is that SEMAC+VAT also provides through plane as well as in plane artifact reduction, MARS+VAT only provides in plane artifact reduction. SEMAC uses a slice selective TSE acquisition. Multiple z-encodings per excited slice are used to recover off-resonant signal caused by magnetic field inhomogeneities. The output image for each slice represents a combination of the signal acquired at different off-resonant frequencies. SEMAC takes care of corrections in the through-slice direction. The VAT (View angle tilting) technique is used to reduce in-plane distortions. For this, the gradient applied during slice selection is reapplied during the signal readout.
The feature consists of:
- Specific imaging sequence based on multiple overlapping 3D volumes, where the 3D volume aims at capturing the different frequencies caused by the distortion.
- A new calculation function to combine different frequency MR signals into a single . undistorted slice.
- A TSE-based SENSE reference scan, which is more robust towards the metal distortions than standard FFE reference scans
- VAT gradient control in sequences.
- . VAT is combined with SEMAC or MARS. MARS (Metal artifact reduction sequence) is a slice selective high bandwidth TSE sequence which can be achieved with standard settings of the TSE sequences.
O-MAR is supported on the following systems:
- 3.0T Ingenia
- 3.0T Achieva ●
- 1.5T Ingenia ●
- 1.5T Achieva ●
{5}------------------------------------------------
The functionality is supported on all available gradient performance levels. Optimized protocols will be provided for the different performance points. O-MAR is supported on the centralized data acquisition systems of the Achieva systems as well as the digitally networked data acquisition system of the Ingenia systems. The data acquisition system is fully transparent to the O-MAR pulse sequences reconstructions.
#### 21 CFR 807.92(a)(5): Intended Use
O-MAR is a combination of an acquisition technique and post-processing software intended for use on Achieva and Ingenia. 1.5T & 3T MR Systems. O-MAR is suitable for use on all patients with passive MR Conditional orthopedics implants that are scanned according to the conditions of safe use for the specific MR Conditional implant being scanned. In addition O-MAR is suitable for use on patients without implants that are cleared for MR exams. O-MAR helps reduce artifacts caused by presence of metal in both in-plane and through-plane dimensions compared to conventional MR imaging techniques. Thus O-MAR improves visualization of more tissue in the vicinity of MR Conditional orthopedic implants. When interpreted by a trained physician, images generated by O-MAR provide information that can be useful in determining a diagnosis.
#### 21 CFR 807.92(a)(6): Technological Characteristics:
The main functional units in the software are:
- Methods (acquisition of MR signals by means of MR pulse sequences) -
- Reconstruction (transforming the MR signals to images) -
- Patient Administration (storing of the images in the database and providing access) -
- -Viewing (display of images)
The technical impact of the feature O-MAR comprises:
- Methods: Modify an existing 2D sequence to acquire multiple 3D volumes for near metal imaging, within the cleared limits provided by the basic MR system. Add an additional gradient during readout for VAT. The TSE SENSE reference scan and the MARS protocol are achieved with cleared settings within the existing capabilities of the MRI system.
- Reconstruction: Add a new calculation function that combines the frequency shifted slices into distortion corrected slices.
No off-the-shelf software is used for the feature near metal imaging. The off-the-shelf software used in the basic MR system is cleared. O-MAR is not designed to be connected to an external network.MR system is cleared.
O-MAR does not require any change of the hardware platform. The extension introduced by O-MAR, are in methods pulse sequence code, and in reconstruction for the new calculation. Those run on the host computer:
Computer characteristics:
- Manufacturer: HP; Model: Z420; Processor clock: 3.5 GHz; RAM: 64 GB RAM; -Processors: six core with hyper threading
- Operating system: Windows 7, 64 bits -
Changes to operator workflow for O-MAR are:
- -Protocol selection: The operator selects an ExamCard with O-MAR protocols
- -Planscan phase: Optionally the operator may want to change the SEMAC factor.
{6}------------------------------------------------
All other steps are not changed. The generated image types can be viewed, post-processed, printed and archived as any other image type.
#### 21 CFR 807.92(b)(1): Brief discussion of nonclinical tests submitted, referenced or relied on in this premarket notification:
O-MAR has been verified to function with the Achieva and Ingenia 1.5T and 3.0T MR systems. The verification testing showed the examcards could be loaded, a Sense reference scan could be added and all scans ran properly. It was also verified that the IFU contained the appropriate warning messages for persons with metallic implants.
The conclusion from Verification Report DHF218318 is:
All the tests performed for O-MAR were successful. Workflow was smooth and no problems occurred.
Additional testing was conducted using phantoms that individually contained three total hip implants with varying materials, one total knee implant and two spine implants (screws and fixation rod, and screws and fixation plate). A comparison between high bandwidth TSE scans, MARS+VAT scans and SEMAC+VAT scans of the phantoms was conducted based on a reduction of artifact size. All testing showed a better artifact reduction using either the MARS+VAT scan or the SEMAC+VAT scan versus the high bandwidth TSE scan. This testing is found in OMAR Test Data DHF218534.
#### 21 CFR 807.92(b)(2): Brief discussion of clinical tests submitted, referenced or relied on in this premarket notification:
Validation testing was conducted to show improved visualization of tissue around the orthopedic implant and reduced orthopedic implant artifacts. This was based on the comparison of a high bandwidth TSE scans to MARS+VAT scans and SEMAC+VAT scans. The MARS+VAT or SEMAC+VAT scans should show better tissue visualization and reduced artifacts compared to the high bandwidth TSE scans. This testing was performed on volunteers that had a knee, hip, or lumbar spine implant. Written justification of the use of these implants in testing can be found in the Test Report O-MAR DHF218534 and is based on size and material of construction.
The conclusion from Validation Report DHF217689 is:
The clinical validation of O-MAR has completed successfully. Either the MARS+VAT or SEMAC+VAT scans showed improved tissue visualization and reduced artifacts in comparison to the high bandwidth TSE scan in all test areas.
The conclusion from Exteranl Image Evaluation report DHF229248 is:
A board certified radiologist confirmed that the artifact reduction and more tissue visualization were better with O-MAR (MARS+VAT or SEMAC+VAT) when compared to the FDA cleared sequence (High band width TSE).
No product defects were submitted during validation and no new hazards were identified.
{7}------------------------------------------------
#### 21 CFR 807.92(b)(3): The conclusions drawn from the nonclinical and clinical tests that demonstrate that the device is as safe, as effective, and performs as well as or better than the legally marketed device identified in paragraph (a)(3) of this section:
The nonclinical and clinical tests have demonstrated that the device is safe and works according to its intended use.
O-MAR software does not introduce new indications for use, nor does the use of the device result in any new potential hazard. Philips Medical Systems considers O-MAR Software to be substantially equivalent to the above mentioned predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.