← Product Code [MUD](/productcode/MUD) · K143219

# NIRO-200NX (K143219)

_Hamamatsu Photonics K.K. · MUD · Jul 31, 2015 · Cardiovascular · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K143219

## Device Facts

- **Applicant:** Hamamatsu Photonics K.K.
- **Product Code:** [MUD](/productcode/MUD.md)
- **Decision Date:** Jul 31, 2015
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 870.2700
- **Device Class:** Class 2
- **Review Panel:** Cardiovascular
- **Attributes:** Real-World Evidence

## Real-World Evidence

| Submission | Device | Sponsor | RWD Sources | RWE Use Summary | Key Tags |
| --- | --- | --- | --- | --- | --- |
| K143219 · Jul 31, 2015 | NIRO-200NX | Hamamatsu Photonics K.K. | Published clinical literature; Clinical setting performance data | Clinical literature was used to demonstrate substantial equivalence by comparing the performance of the NIRO-200NX to the predicate INVOS device and blood draw results in a clinical setting, as well as to support the device's successful clinical history. | Clinical literature; Comparative performance; Clinical setting |

### Clinical Evidence

| Study Design | Population | Comparator | Key Endpoints |
| --- | --- | --- | --- |
| Bickler et al.; Comparative clinical study | 23 individuals (including light, intermediate, and dark skin); Sample Size: 23 individuals; 181 measurements for INVOS, 179 for NIRO-200NX | INVOS 5100C and blood draw results | Regional hemoglobin oxygen saturation |

## Indications for Use

The NIRO-200NX is intended for use as an adjunct trend monitor of regional hemoglobin oxygen saturation and relative level of oxygenated hemoglobin and deoxygenated hemoglobin of blood in brain or in other tissue beneath the probes in any individual. The clinical value of trend data has not been demonstrated in disease states. The NIRO-200NX should not be used as the sole basis for diagnosis or therapy.

## Device Story

NIRO-200NX is a reusable near-infrared oximeter for non-invasive monitoring of regional hemoglobin oxygen saturation and relative hemoglobin levels. Probes applied to skin contain a light source and two photodiodes at varying distances. Device measures light transmitted through tissue; subtracts signal from closer photodiode from farther photodiode to calculate hemoglobin oxygen saturation; calculates relative levels of oxygenated/deoxygenated hemoglobin via changes in detected light. Used in clinical settings; provides trend data to healthcare providers as an adjunct to other diagnostic information. Benefits include non-invasive tissue oxygenation monitoring.

## Clinical Evidence

Clinical evidence includes published literature (Bickler et al.) comparing NIRO-200NX and INVOS 5100C in 23 individuals (181 INVOS measurements, 179 NIRO-200NX measurements) across varying skin tones. Results showed substantially equivalent performance in measuring regional hemoglobin oxygen saturation. Additional evidence includes 4 years of clinical use in Japan and Europe without reported adverse events and literature confirming successful use in infants.

## Technological Characteristics

Non-invasive near-infrared oximeter. Light source: 3-wavelength LED. Detector: Photodiode. Measurement method: 2-point detection for hemoglobin oxygen saturation (TOI) and relative total hemoglobin (nTHI); 1-point detection for relative hemoglobin levels. Safety: Class I LED product. Compliance: IEC 60601-1 (electrical), IEC 60601-1-2 (EMC), IEC 60825-1 (laser/light safety).

## Regulatory Identification

An oximeter is a device used to transmit radiation at a known wavelength(s) through blood and to measure the blood oxygen saturation based on the amount of reflected or scattered radiation. It may be used alone or in conjunction with a fiberoptic oximeter catheter.

## Predicate Devices

- Somanetics INVOS 5100C ([K082327](/device/K082327.md))
- Hitachi ETG-100 ([K011320](/device/K011320.md))
- Hitachi ETG-4000 ([K042501](/device/K042501.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" around the perimeter. Inside the circle is an emblem consisting of three stylized human profiles facing to the right, with a design above them that resembles a bird in flight.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

Hamamatsu Photonics K.K. % Mr. Jeffrey K. Shapiro Hyman, Phelps and McNamara 700 Thirteenth Street, North West, Suite 1200 Washington, District of Columbia 20005

Re: K143219

Trade/Device Name: NIRO-200NX Regulation Number: 21 CFR 870.2700 Regulation Name: Oximeter Regulatory Class: Class II Product Code: MUD, DQA Dated: July 6, 2015 Received: July 6, 2015

Dear Mr. Shapiro:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set

July 31, 2015

{1}------------------------------------------------

forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

For

Sincerely yours.

Joshua C. Nipper -S

Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

## Indications for Use

510(k) Number (if known) K143219

Device Name Hamamatsu NIR0-200NX

#### Indications for Use (Describe)

The NIRO-200NX is intended for use as an adjunct trend monitor of regional hemoglobin oxygen saturation and relative level of oxygenated hemoglobin and deoxygenated hemoglobin of blood in brain or in other tissue beneath the probes in any individual. The clinical value of trend data has not been demonstrated in disease states. The NR0-200NX should not be used as the sole basis for diagnosis or therapy.

| Type of Use (Select one or both, as applicable)                                  |                                                                                 |
|----------------------------------------------------------------------------------|---------------------------------------------------------------------------------|
| <div> <span>      Prescription Use (Part 21 CFR 801 Subpart D)    </span> </div> | <div> <span>      Over-The-Counter Use (21 CFR 801 Subpart C)    </span> </div> |

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

### K143219

Page 1 of 6

### 510(k) Summary NIRO-200NX

Submitter Name: Hamamatsu Photonics K.K.

Submitter Address: 812 Joko-cho, Higashi-ku, Hamamatsu City, 431-3196, JAPAN

Contact Person: Susumu Suzuki

Phone Number: 81-53-431-0124

Fax Number: 81-53- 431-0148

Date Prepared: July 24, 2015

Device Trade Name: NIRO-200NX

Device Common Name: Oximeter

Product Code: MUD

Subsequent Product Code: DQA

Classification: Class II per 21 C.F.R. § 870.2700

Predicate Device:

Somanetics INVOS 5100C (K082327), Hitachi ETG-100 (K011320) and ETG-4000 (K042501)

The NIRO-200NX is a reusable piece of equipment that uses near Device Description: infrared light for non-invasive measurement of hemoglobin oxygen saturation and relative levels of oxygenated hemoglobin and deoxygenated hemoglobin of blood in brain or in other tissue beneath the probes. The patient probes are applied to the skin over the tissue of interest. The probes have a light source and 2 photodiodes, one closer to the light source and one further away from the light source. The 2 photodiodes detect the light transmitted through the patient's tissue. The detected light is analyzed with the known light absorption characteristics of oxyhemoglobin and deoxyhemoglobin. The amount of light detected by the photodiode closer to the light source is subtracted from the light detected by the farther photodiode. The result is then used to calculate the hemoglobin oxygen saturation. Also, by measuring the changes in light detected from one of the

{4}------------------------------------------------

photodiodes, the relative levels of oxygenated hemoglobin and deoxygenated hemoglobin are calculated.

Intended Use: The NIRO-200NX is intended for use as an adjunct trend monitor of regional hemoglobin oxygen saturation and relative level of oxygenated hemoglobin and deoxygenated hemoglobin of blood in brain or in other tissue beneath the probes in any individual. The clinical value of trend data has not been demonstrated in disease states. The NIRO-200NX should not be used as the sole basis for diagnosis or therapy.

Performance data: The following electrical, performance, and clinical literature has been conducted with and reported using the NIRO-200NX and is described in the 510(k) submission. All tests and literature demonstrate that the device functions as intended and performs equivalently to the predicate devices.

- 1. Electrical per IEC 60601-1
- 2. Electromagnetic Compatibility per IEC 60601-1-2
- 3. Light Emitting LED Product per IEC 60825-1 (Class1 LED product)

4. Bench data demonstrating that NIRO-200NX performs equivalently in tissue phantoms compared to the predicate devices.

- 5. The NIRO-200NX has been sold and used clinically for more than 4 years in Japan and Europe without any reported adverse events. A review of the published literature concludes that the device worked as intended by safely being used as an adjunct monitor of regional hemoglobin oxygen saturation and relative level of oxygenated hemoglobin and deoxygenated hemoglobin of blood in brain or in other tissue beneath the probes. Specifically, clinical literature demonstrated that NIRO-200NX and the predicate INVOS performed equivalently in a clinical setting. In addition, clinical literature was provided to show that NIRO-200NX has been used successfully in infants.
Substantial Equivalence:

The predicate devices are ETG-100 (K011320), ETG-4000 (K042501), and INVOS 5100C (K082327).

The proposed NIRO-200NX and the predicate devices all have the same intended use: as Oximeters. The proposed and predicate devices also have similar indications for use. The proposed and predicate devices all have the same principles of operation: they all measure levels of oxygen based on light measurements taken from the surface of intact tissue on the outside of a patient's body. The NIRO-200NX and the predicate

{5}------------------------------------------------

devices have similar technological characteristics, and any minor differences do not raise different questions of safety or efficacy, as confirmed by Hamamatsu's testing and validation activities described in this submission, including EMC, electrical safety, and laser safey testing in accordance with IEC 60601-1-2 (2008), IEC 60601-1 (2005), and IEC 60825-1 (1993 + A1 (1997) + A2 (2001). respectively. Hamamatsu followed the FDA guidance document, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices May 11, 2005," to classify the NIRO-200NX software as a "moderate level of concern." The software was verified and validated, and the software verification and validation documents were prepared and presented in accordance with FDA's guidance document.

Further, NIRO-200NX is at least as safe and effective as the predicate devices as demonstrated by the results of bench and clinical data reported in literature, including the phantom study and the published clinical literature described below.

#### Phantom Study

Hamamatsu performed a phantom study to compare performance of the proposed and predicate devices side by side in a simulated model. The results of the study demonstrated that performance of the NIRO-200NX is substantially equivalent to the performance of INVOS at measuring regional hemoglobin oxygen saturation; and ETG at measuring relative levels of oxygenated hemoglobin and deoxygenated hemoglobin.

### Published Clinical Literature

Bickler et al reported on a clinical study that directly compared the performance of NIRO-200NX and INVOS to each other (along with several other cleared devices) and to results of a blood draw in a clinical setting. The evaluation was performed in 23 individuals with 181 individual measurements taken for INVOS and 179 individual measurements taken for NIRO-200NX. The study subjects included those with light, intermediate and dark skin. The study results showed substantially equivalent performance between NIRO-200NX and INVOS in a clinical setting at measuring regional hemoglobin oxygen saturation.

The above information leads to the conclusion that the NIRO-200NX is substantially equivalent to the predicate devices. The Substantial Equivalence comparison chart is found below in Table 1.

{6}------------------------------------------------

## Table 1

|                        | NIRO-200NX<br>Hamamatsu Photonics K.K.<br>(New Device)                                                                                                                                                                                                                                                                                                                                                                                                     | ETG-100 and ETG-4000<br>Hitachi Medical Systems America, Inc<br>(ETG-100 for K011320)<br>(ETG-4000 for K042501)                                                                                                                                                                                   | INVOS 5100C<br>Somanetics Corporation<br>(K082327)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                |
|------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| INTENDED<br>USE        | Oximeter (MUD, DQA)                                                                                                                                                                                                                                                                                                                                                                                                                                        | Oximeter (DQA)                                                                                                                                                                                                                                                                                    | Oximeter (MUD)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |
| INDICATIONS<br>FOR USE | The NIRO-200NX is intended for<br>use as an adjunct trend monitor of<br>regional hemoglobin oxygen<br>saturation and relative level of<br>oxygenated hemoglobin and<br>deoxygenated hemoglobin of blood<br>in brain or in other tissue beneath the<br>probes in any individual. The clinical<br>value of trend data has not been<br>demonstrated in disease states. The<br>NIRO-200NX should not be used as<br>the sole basis for diagnosis or<br>therapy. | A) K011320<br>The intended use of the ETG-100 is<br>the measurement of relative levels of<br>cerebral deoxy-hemoglobin and<br>oxyhemoglobin.<br><br>B) K042501<br>The intended use of the ETG-4000<br>is the measurement of relative levels<br>of cerebral deoxy-hemoglobin and<br>oxyhemoglobin. | The noninvasive INVOS 5100C is<br>intended for use as an adjunct<br>monitor or regional hemoglobin<br>oxygen saturation of blood in the<br>brain or in other tissue beneath the<br>sensor. It is intended for use in<br>individuals greater than 2.5kg at risk<br>for reduced-flow or no-flow ischemic<br>states.<br><br>It is also intended for use as an<br>adjunct trend monitor of regional<br>hemoglobin oxygen saturation of<br>blood in the brain or in other tissue<br>beneath the sensor in any individual.<br>The clinical value of trend data has<br>not been demonstrated in disease<br>states. The INVOS System should<br>not be used as the sole basis for<br>diagnosis or therapy. |

{7}------------------------------------------------

| NIRO-200NX<br>Hamamatsu Photonics K.K.<br>(New Device) | ETG-100 and ETG-4000<br>Hitachi Medical Systems America, Inc<br>(ETG-100 for K011320)<br>(ETG-4000 for K042501)                                                                                                                                                                                                                | INVOS 5100C<br>Somanetics Corporation<br>(K082327)                                                                                                                                                                                                                                       |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  |
|--------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| PRINCIPLE<br>OF<br>OPERATION                           | The patient probe consists of a light<br>source (2 wavelengths) and 1<br>detector, and it is applied to the skin<br>over the tissue of interest.<br>Relative levels of oxy-hemoglobin,<br>deoxy-hemoglobin and total<br>hemoglobin are calculated from the<br>changes in the amount of the<br>detected lights by one detector. | The patient probe consists of a light<br>source (2 wavelengths) and 2<br>detectors, and it is applied to the skin<br>over the tissue of interest.<br>Hemoglobin oxygen saturation<br>(rSO2) is calculated from the<br>difference of the amount of detected<br>lights by the 2 detectors. | The patient probe consists of a light<br>source (3 wavelengths) and 2<br>detectors, and it is applied to the skin<br>over the tissue of interest.<br>Hemoglobin oxygen saturation (TOI)<br>is calculated from the differences of<br>the amount of detected lights by the 2<br>detectors. Relative value of the total<br>hemoglobin (nTHI) is calculated by<br>the same methods as that of TOI.<br>Relative levels of oxygenated,<br>deoxygenated and total<br>hemoglobines (△O2Hb, △HHb and<br>△cHb) are calculated from the<br>changes in the amount of the<br>detected lights by one detector. |

{8}------------------------------------------------

# TECHNOLOGICAL CHARACTERISTICS

|                                                      | NIRO-200NX<br>Hamamatsu Photonics K.K.<br>(New Device)                                                                                                                                           | ETG-100 and ETG-4000<br>Hitachi Medical Systems America, Inc<br>(ETG-100 for K011320)<br>(ETG-4000 for K042501)       | INVOS 5100C<br>Somanetics Corporation<br>(K082327)                                                                            |
|------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------|
| Light Source<br>Device<br>Wavelength<br>Safety Class | LED (Light Emitting Diode)<br>3 wavelengths<br>Class I                                                                                                                                           | LD (Laser Diode)<br>2 wavelengths<br>Class I                                                                          | LED<br>2 wavelengths<br>Class I                                                                                               |
| Light Detector                                       | Photodiode                                                                                                                                                                                       | Photomultiplier Tube                                                                                                  | Photodiode                                                                                                                    |
| Measurement<br>Method                                | 2 Point Detection Method for<br>Hemoglobin Oxygen Saturation<br>(TOI) and Relative value of the total<br>hemoglobin (nTHI)<br><br>1 Point Detection Method for<br>Relative Levels of Hemoglobins | Not applicable. Not measured by<br>this device.<br><br>1 Point Detection Method for<br>Relative Levels of Hemoglobins | 2 Point Detection Method for<br>Hemoglobin Oxygen Saturation<br>(rSO2)<br><br>Not applicable. Not measured by<br>this device. |
| Patient<br>Contact                                   | Non-Invasive                                                                                                                                                                                     | Non-Invasive                                                                                                          | Non-Invasive                                                                                                                  |
| EMC and<br>Electrical<br>Safety                      | Passed applicable safety testing                                                                                                                                                                 | Passed applicable safety testing                                                                                      | Passed applicable safety testing                                                                                              |

---

**Source:** [https://fda-staging.innolitics.com/device/K143219](https://fda-staging.innolitics.com/device/K143219)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
