ANGIO-PRESS DVT Compression Device Model Name: IPCS
K143202 · Pegasus Medical Supply, Inc. · JOW · Jul 7, 2015 · Cardiovascular
Device Facts
Record ID
K143202
Device Name
ANGIO-PRESS DVT Compression Device Model Name: IPCS
Applicant
Pegasus Medical Supply, Inc.
Product Code
JOW · Cardiovascular
Decision Date
Jul 7, 2015
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.5800
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Pegasus Medical Supply Inc. ANGIO-PRESS IPCS Deep Vein Thrombosis (DVT) Compression Devices are intended to increase venous blood flow in patients in order to help prevent deep vein thrombosis.
Device Story
Non-invasive mechanical prophylactic system; counteracts blood stasis and coagulation changes. Device inflates sleeves around limbs periodically; massages legs in wavelike, milking motion; promotes blood flow; empties pooled/static blood from femoral vein valve cusps; increases fibrinolytic activity. Used to prevent blood pooling. Complements other prophylactic measures like anti-embolic stockings and anticoagulants. Operated via AC power; no battery pack. Intended for clinical or home use (excluding outdoor/transportation settings). Healthcare providers/patients use device to reduce DVT risk; output is mechanical compression; clinical decision-making supported by prophylactic therapy.
Clinical Evidence
No clinical data. Safety and effectiveness supported by bench testing, including biocompatibility (ISO 10993), electrical safety/EMC (IEC 60601-1, IEC 60601-1-2), software (IEC 62304), and performance testing (vibration, shock, temperature).
Technological Characteristics
Compressible limb sleeve system; AC-powered pneumatic pump. Materials evaluated per ISO 10993. Software developed per IEC 62304. Dimensions: 12.6" x 4.3" x 7.7". Weight: 3.1 kg. Intermittent pneumatic compression mode. Compliance with IEC 60601-1, IEC 60601-1-2, IEC 60601-1-6, and IEC 62366.
Indications for Use
Indicated for patients at risk of deep vein thrombosis (DVT) to increase venous blood flow and prevent DVT.
Regulatory Classification
Identification
A compressible limb sleeve is a device that is used to prevent pooling of blood in a limb by inflating periodically a sleeve around the limb.
Predicate Devices
VasoPress DVT Supreme Mini, Pump Model No. VP500DM (K101915)
Submission Summary (Full Text)
{0}------------------------------------------------
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
July 7, 2015
Pegasus Medical Supply, Inc. % Michael Lee President AcmeBiotechs Co., Ltd. No. 45, Minsheng Rd. Danshui Town New Taipei City, 251 TW
Re: K143202
Trade/Device Name: ANGIO-PRESS DVT Compression Device Model Name: IPCS Regulation Number: 21 CFR 870.5800 Regulation Name: Compressible Limb Sleeve Regulatory Class: Class II Product Code: JOW Dated: June 4, 2015 Received: June 8, 2015
Dear Michael Lee,
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply
{1}------------------------------------------------
with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
M.A. Hilleman
for
for Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
ANGIO-PRESS DVT Compression Device Model: IPCS
#### Indications for Use
## 510(k) Number (if known): K143202
Device Name: ANGIO-PRESS DVT Compression Device Model Name: IPCS
#### Indications for Use:
The Pegasus Medical Supply Inc. ANGIO-PRESS IPCS Deep Vein Thrombosis (DVT) Compression Devices are intended to increase venous blood flow in patients in order to help prevent deep vein thrombosis.
Prescription Use _____________________________________________________________________________________________________________________________________________________________ (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use_ (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Page 1 of ____________________________________________________________________________________________________________________________________________________________________
{3}------------------------------------------------
ANGIO-PRESS DVT Compression Device Model: IPCS
## 510(k) Summary
| 5.1 | Type of Submission: | Traditional |
|-----|------------------------|------------------------------------------------------------------------------------------------------------------------------------|
| 5.2 | Preparation Date: | 29th October 2014 |
| 5.3 | Submitter:<br>Address: | Pegasus Medical Supply, Inc.<br>No. 27-1, Ln. 473, Sec. 2, Hezun N. Rd.,<br>Zhongli City, Taoyuan County 32060,<br>Taiwan (R.O.C.) |
| | Phone: | +886-3-2753341 |
| | Fax: | +886-3-4616802 |
| | Contact: | PAO-MING SHIH<br>(borisshih@pegasus-pms.com) |
| | Registration number: | - |
#### 5.4 Identification of the Device:
| Proprietary/ | ANGIO-PRESS DVT Compression Device |
|------------------------|------------------------------------|
| Trade name: | Model Name: IPCS |
| Classification Name: | Sleeve, Limb, Compressible |
| Device Classification: | II |
| Regulation Number: | 870.5800 |
| Panel: | Cardiovascular |
| Product Code: | JOW |
#### 5.5 Identification of the Predicate Device:
| Predicate Device Name: | VasoPress DVT Supreme Mini, Pump Model |
|------------------------|----------------------------------------|
| | No. VP500DM |
| Manufacturer: | Compression Therapy Concepts, Inc. |
| Regulation number: | 870.5800 |
| Product Code: | JOW |
| 510(k) Number: | K101915 |
{4}------------------------------------------------
### 5.6 Intended Use and Indications for Use of the subject device.
The Pegasus Medical Supply Inc. ANGIO-PRESS IPCS Deep Vein Thrombosis (DVT) Compression Devices are intended to increase venous blood flow in patients in order to help prevent deep vein thrombosis.
# 5.7 Device Description
ANGIO-PRESS DVT Compression Device that counteracts blood stasis and coagulation changes - two of the three major factors that promote deep vein thrombosis (DVT). ANGIO-PRESS DVT Compression Device is a non-invasive mechanical prophylactic system that massages the legs in a wavelike, milking motion that promotes blood flow and deters thrombosis, helping to empty pooled or static blood from the valve cusps of the femoral vein. Fibrinolytic activity is increased, stimulating the release of a plasminogen activator. This therapy typically complements other prophylactic measures, such as ant embolic stockings and anticoagulants.
ANGIO-PRESS DVT Compression Device is used to prevent pooling of blood in a limb by inflating periodically a sleeve around the limb. Therefore, ANGIO-PRESS DVT Compression Device is identified as a compressible limb sleeve.
# 5.8 Non-clinical Testing
A series of safety tests were performed to assess the safety and effectiveness of the ANGIO-PRESS DVT Compression Device.
| Testing Item | Standard and regulations applied |
|--------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Biocompatibility | ISO 10993-1:2009/Cor. 1:2010(E) Biological evaluation of medical<br>devices - Part 1: Evaluation and testing with a risk management<br>process. |
| | ISO 10993-5:2009 (E) Biological evaluation of medical devices –<br>Part 5: Tests for <i>in vitro</i> cytotoxicity. |
| | ISO 10993-10:2010 (E) Biological evaluation of medical devices – |
| Part 10: Tests for Tests for irritation and skin sensitization. | |
| ISO 10993-12: 2012 (E) Biological evaluation of medical devices – | |
| Part 12: Sample Preparation And Reference Materials. | |
| ISO/IEC 17025:2005 General Requirements for the Competence of<br>Testing and Calibration Laboratories. | |
| USP Biological reactivity tests, in vitro | |
| Software | IEC 62304 First Edition 2006-05, Medical Device Software -<br>Software Life Cycle Processes. (Software/Informatics) |
| Electromagnetic<br>Compatibility &<br>Electrical Safety | EN 60601-1-2 : 2007/AC:2010 Medical electrical equipment - Part<br>1-2: General requirements for basic safety and essential<br>performance - Collateral standard: Electromagnetic compatibility -<br>Requirements and tests. |
| | IEC 60601-1-2 : 2007 Medical Electrical Equipment - Part 1-2:<br>General Requirements For Basic Safety And Essential Performance<br>- Collateral Standard: Electromagnetic Compatibility –<br>Requirements And Tests. (General I (QS/RM)) |
| | IEC 60601-1 Medical Electrical Equipment – Part 1:General<br>requirements for basic safety and essential performance |
| Performance | IEC 60068-2-6 Environmental testing - Part 2-6: Tests - Test Fc:<br>Vibration (sinusoidal) |
| | IEC 60068-2-27 Environmental testing - Part 2-27: Tests - Test Ea<br>and guidance: Shock |
| | ISTA 2A Standard |
| | IEC 60068-2-14: 2009 Environmental testing - Part 2-14: Tests -<br>Test N: Change of temperature. |
| Risk Management | ISO 14971 Medical Devices - Application Of Risk Management<br>To Medical Devices. |
| | IEC 60812 Analysis Techniques For System Reliability –<br>Procedure For Failure Mode And Effects Analysis (Fmea). |
| Usability | IEC 60601-1-6 Edition 3.1 2013-10, Medical Electrical Equipment<br>- Part 1-6: General Requirements For Basic Safety And Essential<br>Performance - Collateral Standard: Usability. |
| IEC 62366 Edition 1.1 2014-01, Medical Devices - Application Of | |
| Usability Engineering To Medical Devices. | |
{5}------------------------------------------------
{6}------------------------------------------------
All the test results demonstrate ANGIO-PRESS DVT Compression Device meet the requirements of its pre-defined acceptance criteria and intended uses.
# 5.9 Clinical Testing
No clinical test data was used to support the decision of safety and effectiveness.
# 5.10 Substantial Equivalence Determination
The ANGIO-PRESS DVT Compression Device is substantially equivalent in intended use, design, technology/principles of operation, materials and performance to the cleared VasoPress DVT Supreme Mini, Pump Model No. VP500DM (K101915). Differences between the devices cited in this section do not raise any new issues of safety or effectiveness.
| Item | ANGIO-PRESS DVT<br>Compression Device | VasoPress DVT Supreme<br>Mini, Pump Model No.<br>VP500DM (K101915) |
|---------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Regulation Number | 870.5800 | 870.5800 |
| Classification | Class II | Class II |
| Product Code | JOW | JOW |
| Prescription Use | Yes | Yes |
| Indications for Use | The Pegasus Medical<br>Supply Inc. ANGIO-PRESS<br>IPCS Deep Vein<br>Thrombosis (DVT)<br>Compression Devices are<br>intended to increase venous<br>blood flow in patients in<br>order to help prevent deep<br>vein thrombosis. | The VasoPress Pump (Model<br>VP500DM) is an external<br>pneumatic compression device<br>for use with VasoPress DVT<br>garments, and is intended to<br>lower the risk of deep vein<br>thrombosis (DVT) in patients<br>who may be at risk. |
{7}------------------------------------------------
| Model | IPCS | VP500DM |
|----------------------------------|---------------------|---------------------------|
| Dimension | 12.6" x 4.3" x 7.7" | 9.45" x 4.53" x 6.1" |
| Weight | 3.1 kg | 2 kg (without Power Cord) |
| Fuse Rating | 1A/250V | 1A or T1AH 250V |
| Pressure Range | Same | 40 and 80 mmHg |
| Input Rating | Same | AC 100-240V, 50/60Hz |
| Operating Humidity | Same | 30 - 75% |
| Operation Temperature | Same | 15°C - 35°C |
| Applied Part | Same | Garment and Air Hose |
| Applied Mode of<br>Pressure | Same | Intermittent |
| Number of Chambers in<br>Garment | Same | No |
| Inflation time per<br>chamber | Same | 12 seconds |
| Deflation time per<br>chamber | Same | 48 seconds |
| Pressure Range<br>Calf/Thigh | Same | 40 mmHg |
| Pressure Range Foot | Same | 80 mmHg |
| Battery Pack | No | Yes |
### 5.11 Similarity and difference
The difference between the proposed device and the predicate device is the battery pack. For predicate device, VasoPress DVT Supreme Mini, Pump Model No. VP500DM (K101915), the original intention to include battery pack function was to offer user a power solution when device was being used outdoor or during transportation. The system will automatically shift to battery power when the device which has been installed with the battery pack be disconnected from AC power source. Battery pack is an optional feature to VasoPress DVT and AC power source is the main power source.
{8}------------------------------------------------
The main power source of the proposed device, ANGIO-PRESS DVT Compression Device is AC power, same with the predicate device. ANGIO-PRESS DVT Compression Device design does not include battery pack function, so it can't offer an outdoor or transpiration power solution as VasoPress DVT. But ANGIO-PRESS DVT Compression Device can offer the same function as VasoPress DVT which has not been installed with the battery pack. Also, ANGIO-PRESS DVT Compression Device will not change or affect the patient population who is intended to use this device. Therefore, the only difference of the proposed device without the battery pack to the predicate device is that the proposed device is not intended to be used outdoor or on transportation.
The proposed device has tested on safety and performance tests and the results were complied with the test requests. Therefore, the difference of proposed device and predicate device did not raise any problems of safety or effectiveness. The proposed device is substantially equivalent to the predicate device in intended use, design, and performance claims.
# 5.12 Conclusion
After analyzing bench tests, safety testing data, it can be concluded that ANGIO-PRESS DVT Compression Device is substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.