K143175 · Medtronic Advanced Energy · GEI · Dec 31, 2014 · General, Plastic Surgery
Device Facts
Record ID
K143175
Device Name
AEx Generator; PlasmaBlade T
Applicant
Medtronic Advanced Energy
Product Code
GEI · General, Plastic Surgery
Decision Date
Dec 31, 2014
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4400
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
AEx Generator: The AEx Generator is a radio frequency (RF) electrosurgical generator capable of simultaneously powering specified monopolar and bipolar electrosurgical instruments. It is intended to be used for delivery of RF energy to instruments indicated for cutting and coagulation of soft tissue and for delivery of RF energy concurrent with saline to instruments indicated for hemostatic sealing and coagulation of soft tissue and bone. It is intended for, but not limited to, General, Plastic and Reconstructive (including but not limited to skin incisions and development of skin flaps), ENT, Gynecologic, Orthopaedic, Arthroscopic, Spinal and Neurological, Thoracic, and Open Abdominal Surgery procedures. The device is not intended for contraceptive tubal coagulation (Permanent Female Sterilization). PlasmaBlade T: The PlasmaBlade T is a monopolar, single use, sterile, disposable device intended for use with the AEX Generator. The device delivers RF energy concurrent with saline and coagulation of soft tissue and bone and RF energy for cutting and coagulation of soft tissue. It is intended for, but not limited to, General, Plastic and Reconstructive (including but not limited to skin incisions and development of skin flaps), ENT, Gynecologic, Orthopaedic, Arthroscopic, Spinal, Thoracic, and Open abdominal surgery procedures. The device is not intended for contraceptive tubal coagulation (permanent female sterilization).
Device Story
AEx Generator is a shelf-top electrosurgical unit combining functions of Pulsar and Aquamantys generators; provides monopolar/bipolar RF energy for soft tissue resection/coagulation and RF-saline delivery for hemostatic sealing of tissue and bone. PlasmaBlade T is a sterile, single-use monopolar handpiece with an enamel-coated insulated blade, telescoping shaft, and integrated controls. System includes a peristaltic pump for saline delivery. Used in OR by surgeons; generator features touchscreen LCD for power/saline settings. Generator monitors patient return circuit for safety. Output allows surgeons to perform precise cutting and hemostasis; benefits include reduced thermal damage and effective sealing during various surgical procedures.
Clinical Evidence
Bench testing only. Evaluated via side-by-side in-vivo and ex-vivo animal (porcine) studies comparing thermal effects on tissue against predicate generators and handpieces. Results showed no clinically significant differences in thermal damage zones. Software verification, electromagnetic compatibility (IEC 60601-1-2), and electrical safety (IEC 60601-1, IEC 60601-2-2) testing performed.
Indicated for patients undergoing general, plastic, reconstructive, ENT, gynecologic, orthopaedic, arthroscopic, spinal, neurological, thoracic, or open abdominal surgery requiring soft tissue cutting, coagulation, or hemostatic sealing of soft tissue and bone. Contraindicated for contraceptive tubal coagulation (permanent female sterilization).
Regulatory Classification
Identification
An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is an abstract symbol resembling a stylized caduceus, featuring three human profiles facing right, suggesting a sense of community and well-being.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
December 31, 2014
Medtronic Advanced Energy Mr. Deep Pal Principal Regulatory Affairs Specialist 180 International Drive Portsmouth, New Hampshire 03801
Re: K143175 Trade/Device Name: AEx Generator PlasmaBlade T Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical cutting and coagulation device and accessories Regulatory Class: Class II Product Code: GEI Dated: November 4, 2014 Received: November 5, 2014
Dear Mr. Pal:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21
{1}------------------------------------------------
Page 2 - Mr. Deep Pal
CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# David Krause -S
for
Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the Medtronic logo. On the left side of the logo is a circle with three figures inside. The figures are in different positions, suggesting movement. To the right of the circle is the word "Medtronic" in a bold, sans-serif font.
# SECTION 8: INDICATIONS FOR USE STATEMENT
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.
510(k) Number (if known) K143175 Device Name
AEx Generator PlasmaBlade T
Indications for Use (Describe)
AEx Generator:
The AEx Generator is a radio frequency (RF) electrosurgical generator capable of simultaneously powering specified monopolar and bipolar electrosurgical instruments. It is intended to be used for delivery of RF energy to instruments indicated for cutting and coagulation of soft tissue and for delivery of RF encrgy concurrent with saline to instruments indicated for hemostatic sealing and coagulation of soft tissue and bone. It is intended for, but not limited to. General. Plastic and Reconstructive (including but not limited to skin incisions and development of skin flaps), ENT, Gyneologic, Orthopaedic. Arthroscopic, Spinal and Neurological, Thoracic, and Open Abdominal Surgery procedures. The device is not intended for contraceptive tubal coagulation (Permanent Female Sterilization).
PlasmaBlade T:
The PlasmaBlade T is a monopolar, single use, sterile, disposable device intended for use with the AEX Generator. The device delivers RF energy concurrent with saling and coagulation of soft tissue and bone and RF energy for cutting and coagulation of soft tissue. It is intended for, but not limited to, General, Plastic and Reconstructive (including but not limited to skin incisions and development of skin flaps), ENT, Gynecologic, Orthopaedic, Arthroscopic, Spinal, Thoracic, and Open abdominal surgery procedures. The device is not intended for contraceptive tubal coagulation (permanent female sterilization).
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW."
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number.
FORM FDA 3881 (8/14)
Page 1 of 1
PSC Publishing Services (301) 443-6748 171
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the Medtronic logo. The logo consists of a circular graphic on the left and the word "Medtronic" on the right. The graphic shows three figures in a circle, with one figure standing upright and the other two figures lying horizontally. The word "Medtronic" is written in a bold, sans-serif font.
# SECTION 9: 510(k) SUMMARY
This summary of substaintial equivalence is submitted in accordance with the requirements of 21CFR807.92.
#### DATE PREPARED 9.1
November 3, 2014
#### 9.2 NAME AND ADDRESS OF THE 510(k) OWNER
Medtronic Advanced Energy 180 International Drive Portsmouth, NH 03801
#### 9.3 CONTACT PERSON
# Deep Pal
Principal Regulatory Affairs Specialist Telephone: 817.788.6685 Facsimile: 817.788.6222 E-Mail: deep.pal@medtronic.com
#### 9.4 PROPRIETARY NAME OF THE PROPOSED DEVICES
AEX Generator (Device Catalog Number: 40-405-1); and PlasmaBlade T Handpiece (Device Catalog Number: 27-101-1)
#### COMMON/USUAL NAME 9.5
Electrosurgical Device and Accessories
#### 9.6 DEVICE CLASSIFICATION NAME
Electrosurgical Cutting and Coagulation Device and Accessories (21 CFR 878.4400, Product Code GEI, Class II)
#### 9.7 PREDICATE DEVICE IDENTIFICATION
The devices that are within the scope of this 510(k) Submission, the proposed AEX Generator and PlasmaBlade T Handpiece devices, are substantially equivalent to the following predicates:
| PROPOSED DEVICE | | CLAIMING SUBSTANTIAL EQUIVALNCE TO | |
|-----------------|---------------|-----------------------------------------------------------------------|------------------|
| AEX | PlasmaBlade T | PREDICATE DEVICE | 510(k) NUMBER |
| ✓ | N/A | Aquamantys System<br>- Aquamantys Generator and Aquamantys Handpieces | K052859 |
| ✓ | ✓ | PEAK Surgery System<br>- PULSAR Generator and PlasmaBlade Handpieces | K082786 |
| N/A | ✓ | Aquamantys SBS 5.0 Handpiece | K111732, K132974 |
| D DEVICE CLAIMING SUBSTANTIAL EQUIVALNCE TO | |
|---------------------------------------------|--|
|---------------------------------------------|--|
Table 2: Predicate device claiming equivalence to
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the Medtronic logo. On the left side of the logo is a circular graphic with a person in three different positions, showing movement. To the right of the graphic is the company name, "Medtronic," in a bold, sans-serif font.
#### 9.8 DEVICE DESCRIPTION
Currently included as part of the system in this 510(k) submission are the AEX Generator, and the PlasmaBlade T Handpiece which is a one hand-held disposable, sterile, single use disposable electrosurgical device that is compatible only with the AEX Generator.
The information on the previously cleared hand-held disposable electrosurgical handpieces that are compatible with the AEX Generator is provided in section 9.8.3.
#### 9.8.1 AEX GENERATOR
In one compact generator, the proposed AEX system provides the functional capabilities available in two separate previously marketed generators: the Medtronic Pulsar Generator (Previously cleared under K082786) and the Aguamantys Generator (Previously cleared under K052859). The AEX Generator applies similar fundamental technologies and methods of operation as the predicate Generators, delivering bipolar and monopolar RF energy for resection and coagulation of soft tissue, and RF based hemostatic sealing concurrent with saline delivery for hemostatic and coagulation sealing of soft tissue and bone at the operative site.
The AEX Generator is a line powered, electrosurgical generator with monopolar and bipolar RF that is intended for use only with specific compatible Medtronic electrosurgical Handpieces; the AEX compatible Handpieces are identified in Section 9.8.3. The AEX Generator is a shelf-top unit consisting of a plastic, metal housing and a front LCD control panel. The Generator has a peristaltic pump outside the generator's housing, which is capable of transferring saline through the disposable accessory device concurrent with the generator's provision of RF energy. The LCD control panel is a touchscreen and serves as the user interface for power and saline settings. The AEX Generator has three-pin and seven-pin receptacles in its front panel that provides for the monopolar PlasmaBlade and bipolar Aquamantys disposable Handpieces connection to the Generator's RF power.
The proposed PlasmaBlade T Handpiece connection to the AEX Generator pump is configured to interface with the AEX Generator peristaltic pump. The AEX Generator accepts designated, commercially available, split-pad and single foil patient return electrode pads (neutral electrodes), non-REM neutral electrodes for monopolar applications <= 50 Watts, and provides monitoring of the patient return circuit for safety purposes.
#### 9.8.2 PLASMABLADE T HANDPIECE
The proposed single-use disposable accessory device, the PlasmaBlade T monopolar handpiece, provides the hemostatic capabilities of the Aquamantys SBS 5.0 Handpiece (Cleared under K111732, K132974) and the cutting and coagulative capabilities of the PEAK PlasmaBlade 4.0 (Previously cleared under K082786).
The proposed PlasmaBlade T Handpiece device is a monopolar, single use, disposable devices and is provided sterile. The devices are not intended for reuse or resterilization. The PlasmaBlade T handpieces consists of an enamel coated insulated blade electrode, an insulated telescoping shaft, handle with three integrated controls, and a co-extruded
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the Medtronic logo. On the left side of the logo, there is a circular graphic with three figures in different positions, representing movement or transformation. To the right of the graphic, the word "Medtronic" is written in a bold, sans-serif font. The logo is simple and professional, conveying a sense of innovation and progress.
cable assembly to provide both power and saline. The seven-pin electrical connector is designed to be plugged into the proposed AEx Generator.
The proposed PlasmaBlade T Handpiece is a disposable device, that when connected to the AEX Generator, uses monopolar RF energy for the resection and coagulation of soft tissue and bone. The proposed PlasmaBlade T Handpiece device shares similar handpiece designand tip configuration as its predicate device, the PlasmaBlade 4.0 Monopolar Handpiece (Cleared under K082786).
In addition to the monopolar cutting and coagulation capabilities the propsoed PlasmaBlade T Handpiece device also uses monopolar RF energy concurrent with saline delivery to provide a broader coagulative effect; similar to the predicate Aquamantys SBS 5.0 Handpiece (Cleared under K111732, K132974), this result is trademarked as Transcollation® Technology by Medtronic. The proposed PlasmaBlade T Handpiece device also has a pump header tubing segment and saline bag spike tubing allowing the user to manually connect the device to saline source in the OR as well as the AEX Generator's peristaltic pump. This ensures the device is connected for saline delivery at the same time as the device is connected electrically to the proposed AEX Generator.
- ELECTROSURGICAL HANDPIECE DEVICES COMPATIBLE WITH THE AEX GENERATOR 9.8.3 The proposed AEX Generator will power the following proposed and the previously cleared hand-held disposable electrosurgical handpieces currently marketed by Medtronic:
| HANDPIECE | | GENERATOR COMPATIBILITY | | | |
|----------------------------|--------------------------------|-------------------------|------|---------|------|
| 510(k) NUMBER | NAME | TYPE | AEX1 | PULSAR3 | AQM3 |
| Subject of this submission | PlasmaBlade T1 | Monopolar | ✓ | - | - |
| K082786 | PlasmaBlade 4.02 | Monopolar | ✓ | ✓ | - |
| K082786 | PlasmaBlade Needle | Monopolar | ✓ | ✓ | - |
| K083415 | PlasmaBlade Tonsil | Monopolar | ✓ | ✓ | - |
| K083415 | PlasmaBlade TnA | Monopolar | ✓ | ✓ | - |
| K083415 | PlasmaBlade Adenoid TIP | Monopolar | ✓ | ✓ | - |
| K093695 | PlasmaBlade 3.0S | Monopolar | ✓ | ✓ | - |
| K102709 | PlasmaBlade PLUS | Monopolar | ✓ | ✓ | - |
| K103775 | PlasmaBlade Suction Coagulator | Monopolar | ✓ | ✓ | - |
| K052859 | Aquamantys 6.0 | Bipolar | ✓ | - | ✓ |
| K052859, K111285, K132974 | Aquamantys 2.3 | Bipolar | ✓ | - | ✓ |
| K063639, K132974 | Aquamantys EVS 4.0 | Bipolar | ✓ | - | ✓ |
| K063639, K132974 | Aquamantys Mini EVS 3.4 | Bipolar | ✓ | - | ✓ |
| K073495 | Aquamantys MBS with Light | Bipolar | ✓ | - | ✓ |
| K101057 | Aquamantys 9.5XL | Bipolar | ✓ | - | ✓ |
| K111732, K132974 | Aquamantys SBS 5.02 | Bipolar | ✓ | - | ✓ |
| K123201 | Aquamantys Endo DBS 8.7 | Bipolar | ✓ | - | ✓ |
4Subject of this submission.
-Predicates for the proposed PlasmaBlade T Handpiece.
3Predicates for the proposed AEX Generator: AQM (K052859), Pulsar Generator (K082786)
Table 3: AEX Generator compatible Handpieces
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the Medtronic logo. The logo consists of a circular graphic on the left and the word "Medtronic" in bold, sans-serif font on the right. The graphic depicts three human figures in different positions within a circle, suggesting movement or transformation. The word "Medtronic" is written in a clean, modern typeface, conveying a sense of professionalism and innovation.
#### 9.9 INDICATIONS FOR USE
#### 9.9.1 AEX GENERATOR
The AEX Generator is a radio frequency (RF) electrosurgical generator capable of simultaneously powering specified monopolar and bipolar electrosurgical instruments. It is intended to be used for delivery of RF energy to instruments indicated for cutting and coagulation of soft tissue and for delivery of RF energy concurrent with saline to instruments indicated for hemostatic sealing and coagulation of soft tissue and bone. It is intended for, but not limited to, General, Plastic and Reconstructive (including but not limited to skin incisions and development of skin flaps), ENT, Gynecologic, Orthopaedic, Arthroscopic, Spinal and Neurological, Thoracic, and Open Abdominal Surgery procedures. The device is not intended for contraceptive tubal coagulation (Permanent Female Sterilization).
#### 9.9.2 PLASMABLADE T HANDPIECE
The PlasmaBlade T is a monopolar, single use, sterile, disposable device intended for use with the AEX Generator. The device delivers RF energy concurrent with saline for hemostatic sealing and coagulation of soft tissue and RF energy for cutting and coagulation of soft tissue. It is intended for, but not limited to, General, Plastic and Reconstructive (including but not limited to skin incisions and development of skin flaps), ENT, Gynecologic, Orthopaedic, Arthroscopic, Spinal, Thoracic, and Open abdominal surgery procedures. The device is not intended for contraceptive tubal coagulation (permanent female sterilization).
#### 9.10 SUMMARY OF TECHNOLOGICAL CHARACTERISTICS
The predicates and the proposed AEX generator and Plasmablade T are both intended for RF based resection and coagulation of soft tissue, and RF based hemostatic sealing concurrent with saline delivery for hemostatic sealing and coagulation of soft tissue and bone during various surgical procedures. The proposed devices share the same operational characteristics as the predicate platforms, comprised of a radio-frequency generator which supplies RF power to disposable electrode devices for electrosurgical procedures.
# 9.10.1 AEX GENERATOR
The AEx Generator applies the same fundamental technology as the predicate Aguamantys Generator (Previously cleared under K052859) and the Pulsar Generator (Previously cleared under K082786) in that it delivers bipolar and monpolar RF energy for resection and coagulation of soft tissue and bone, and RF based hemostatic sealing concurrent with saline delivery at the operative site.
# 9.10.2 PLASMABLADE T HANDPIECE
The proposed single-use disposable accessory device, the PlasmaBlade T monopolar handpiece, provides the hemostatic capabilities of the Aquamantys SBS 5.0 Handpiece (Cleared under K111732, K132974) and the cutting and coagulative capabilities of the PEAK PlasmaBlade 4.0 (Previously cleared under K082786).
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the Medtronic logo. The logo consists of a circular graphic on the left and the word "Medtronic" on the right. The graphic depicts three human figures in different positions, arranged in a circular pattern. The word "Medtronic" is written in a bold, sans-serif font.
#### 9.11 SUMMARY OF NON-CLINICAL TESTING
### 9.11.1 THERMAL EFFECTS ON TISSUE
The representative legally marketed Aguamantys and the PlasmaBlade Handpieces with the proposed AEX Generator were evaluated side by side against the predicate Aquamantys Generator (Cleared under K052859) and Pulsar Generator (Cleared under K082786) devices, in the in-vivo, ex-vivo animal (porcine) studies.
Thermal Effects on Tissue of the proposed PlasmBlade T Handpiece with the proposed AEX Generator was also evaluated side by side against the performance of the predicate Handpieces with the AEX Generator, in the in-vivo, ex-vivo animal (porcine) studies.
These studies demonstrated that the comparisons of the data show no clinically significant difference where the proposed AEX Generator exhibited a comparable zone of thermal damage as the predicate Aquamantys and Pulsar Generators. Furthermore, the proposed PlasmaBlade T tip assembly and performance exhibited a comparable zone of thermal damage as the predicate Aquamantys SBS 5.0 (Previously cleared under K111732, K132974) and PlasmaBade 4.0 (Previously cleared under K082786) Handpieces.
### 9.11.2 SOFTWARE VERIFICATION
The software development and testing was executed in compliance to the following FDA Recognized Consensus Standard:
| RECOGNITION NUMBER | STANDARD | TITLE OF STANDARD |
|--------------------|---------------------------|------------------------------------------------------------|
| 13-8 | IEC 62304 Edition 1: 2006 | Medical Device Software - Software Life Cycle<br>Processes |
Table 4: Software Verification
### 9.11.3 ELECTOMAGNETIC COMPATIBILITY
The proposed AEX Generator was tested in compliance to the following FDA Recognized Consensus Standards:
| RECOGNITION NUMBER | STANDARD | TITLE OF STANDARD |
|--------------------|-----------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 19-1 | IEC 60601-1-2 Edition 3: 2007-03 | Medical electrical equipment Part 1-2: General requirements for basic safety<br>and essential performance - Collateral<br>standard: Electromagnetic compatibility -<br>Requirements and tests |
| 6-228 | IEC 60601-2-2 Edition 5.0 2009-02 | Medical electrical equipment Part 2-2: Particular requirements for the basic<br>safety and essential performance of high<br>frequency surgical equipment and high<br>frequency surgical accessories |
Table 5: Electromagnetic Compatibility
{8}------------------------------------------------
Image /page/8/Picture/0 description: The image contains the Medtronic logo. The logo consists of a circular emblem on the left and the word "Medtronic" on the right. The emblem features three stylized human figures in different poses within a circle. The word "Medtronic" is written in a bold, sans-serif font.
# 9.11.4 ELECTRICAL SAFETY
The proposed AEX Generator and the PlasmaBlade T Handpiece devices were tested in compliance to the following FDA Recognized Consensus Standards:
| RECOGNITION NUMBER | STANDARD | TITLE OF STANDARD |
|--------------------|---------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 19-4 | IEC 60601-1:2005 3rd Edition And<br>A1:2012 | Medical electrical equipment-Part 1<br>General requirements for safety and<br>Essential Performance |
| 6-228 | IEC 60601-2-2 Edition 5.0 2009-02 | Medical electrical equipment Part 2-2<br>Particular requirements for the basic<br>safety and essential performance of high<br>frequency surgical equipment and high<br>frequency surgical accessories |
Table 6: Electrical Safety of the AEX Generator and the PlasmaBlade T Handpiece
### 9.11.5 BENCH-TOP PERFORMANCE TESTING
General performance verification testing of the proposed AEX Generator and the PlasmaBlade T devices were also performed to verify the peformance and output characteristics.
#### 9.12 CONCLUSION
Based on the various testing performed, including the ex-vivo, in-vivo animal study, sofware verification testing, electrical safety and electromagnetic compatibility testing, it can be concluded that the proposed devices do not introduce considerations for safety and efficacy different from the considerations of their predicate devices. The similar indications for use, technology and performance characteristics of the proposed AEX Generator and the PlasmaBlade T Handpiece are assessed to be substantially equivalent to the predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.