K143161 · Covidien · GEI · Dec 19, 2014 · General, Plastic Surgery
Device Facts
Record ID
K143161
Device Name
Force FX-8(S), Force FX-8C(S), Force FX-8CA(S)
Applicant
Covidien
Product Code
GEI · General, Plastic Surgery
Decision Date
Dec 19, 2014
Decision
SESE
Submission Type
Special
Regulation
21 CFR 878.4400
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Force FX™ Electrosurgical Generator is an electrosurgical generator containing monopolar and bipolar technology. It is intended for use with accessories during surgical procedures where the surgeon requires electrosurgical cutting (resecting, dividing, or separating) and coagulating (hemostasis).
Device Story
Microcontroller-based isolated output generator; provides RF power for cutting, desiccating, and fulgurating tissue. Inputs: tissue impedance (sensed via electrodes), user settings (mode/power), activation signals (handswitch/footswitch). Processing: Instant Response™ technology adjusts output voltage based on tissue resistance to maintain consistent effect; Autobipolar function (on specific models) uses impedance sensing to auto-start/stop RF activation. Outputs: RF energy for surgical application; audible activation tones. Used in OR/surgical settings by surgeons. Benefits: consistent tissue effect across varying densities; REM™ system monitors return electrode contact to prevent pad site burns; supports simultaneous activation.
Clinical Evidence
No clinical data. Substantial equivalence demonstrated via design verification testing, software regression/validation, and conformance to IEC 60601-1:2005, IEC 60601-2-2:2009, and IEC 60601-1-2:2007 standards.
Technological Characteristics
Microcontroller-based isolated output generator. Features: Instant Response™ technology, REM™ contact quality monitoring, monopolar/bipolar modes. Connectivity: RF activation port, RS-232 serial port, expansion port. Standards: IEC 60601-1:2005, IEC 60601-2-2:2009, IEC 60601-1-2:2007. Modifications include five PCBAs and firmware/software updates.
Indications for Use
Indicated for use in surgical procedures requiring electrosurgical cutting and coagulation (hemostasis) via monopolar or bipolar technology.
Regulatory Classification
Identification
An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized image of a human profile in triplicate, suggesting a sense of community and care. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the central image, emphasizing the organization's name and national scope. The logo is presented in black and white, giving it a clean and official appearance.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
December 19, 2014
Covidien Ms. Sharon McDermott Senior Product Regulatory Specialist 5920 Longbow Drive Boulder, Colorado 80301
Re: K143161 Trade/Device Name: Force FX Generators Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical cutting and coagulation device and accessories Regulatory Class: Class II Product Code: GEI Dated: December 4, 2014 Received: December 5, 2014
Dear Ms. McDermott:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical
{1}------------------------------------------------
device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# Binita SDAshar -S
Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K143161
Device Name Force FX Generators
#### Indications for Use (Describe)
The Force FX™ Electrosurgical Generator is an electrosurgical generator containing monopolar and bipolar technology. It is intended for use with accessories during surgical procedures where the surgeon requires electrosurgical cutting (resecting, dividing, or separating) and coagulating (hemostasis).
| Type of Use (Select one or both, as applicable) | |
|--------------------------------------------------------------------------|-------------------------------------------------------------------------|
| <div> <span>☑ Prescription Use (Part 21 CFR 801 Subpart D)</span> </div> | <div> <span>☐ Over-The-Counter Use (21 CFR 801 Subpart C)</span> </div> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image contains the word "COVIDIEN" in blue font. To the left of the word is a blue square with a white rectangle in the center and a light blue vertical line through the center. The word "COVIDIEN" is in a sans-serif font and is slightly offset to the right of the square. The image is likely a logo for a medical device company.
# 510(k) Summary
Date summary prepared: December 18, 2014
#### 510(k) Submitter/Holder
Covidien (formerly Valleylab, Inc.) 5920 Longbow Drive Boulder, CO 80301
#### Contact
Sharon McDermott Sr. Regulatory Affairs Product Specialist Telephone: 303-581-6789 Fax: 303-530-6313 Email: sharon.mcdermott@covidien.com
#### Name of Device
| Trade Name: | Force FX TM electrosurgical generators |
|----------------------|---------------------------------------------------------------------------------------------------|
| Common Name: | Electrosurgical Generator |
| Classification Name: | Electrosurgical cutting and coagulation device and accessories (21 CFR § 878.4400, class II, GEI) |
# Predicate Device
| Trade Name: | Force FX |
|-----------------|----------------------------|
| Common Name: | Electrosurgical Generator |
| Catalog Number: | Force FX |
| 510(k) Number: | K944602 (cleared 6/5/1995) |
| Manufacturer: | Covidien |
#### Device Description
The Force FX™ Electrosurgical Generators are microcontroller-based isolated output generators that provide power for cutting, desiccating, and fulgurating tissue during bipolar and monopolar surgery. Features included are:
- . Instant Response™ Technology - Automatically senses resistance and adjusts the output voltage to maintain a consistent effect across different tissue density.
- Three bipolar modes: precise, standard, and macro Low voltage, continuous current provides faster . desiccation without sparking.
- Three monopolar Cut modes: low, pure, blend Allows a wide range of power settings necessary to ● perform diverse surgical procedures.
- . Three monopolar Coag modes: desiccate, fulgurate, and spray - Helps control the size of the area and the depth of penetration during tissue coagulation.
- Support for Simultaneous Activation ●
- Covidien Energy-based Devices REM™ Contact Quality Monitoring System Monitors the quality . of electrical contact between the return electrode and the patient to eliminate risk of pad site burns.
- Handswitch or footswitch activation
- . Recall of most recently used mode and power settings
{4}------------------------------------------------
- Adjustable activation tone volume
- An RF activation port, RS-232 serial port, and expansion port .
The precise and standard bipolar modes of the Force FX™-8CA(S) models are equipped with an Autobipolar function. The autobipolar feature senses tissue impedance between the two bipolar electrodes and then uses the impedance information to automatically start or stop bipolar RF activation. Optionally, the user may choose between footswitch start and auto start, or program a delay between auto start and RF activation.
#### Indications for Use
The Force FX™ Electrosurgical Generator is an electrosurgical generator containing monopolar and bipolar technology. It is intended for use with accessories during surgical procedures where the surgeon requires electrosurgical cutting (resecting, dividing, or separating) and coagulating (hemostasis).
#### Technological Characteristics
The Force FX generators have the same technological and performance characteristics as the predicate, K944602. This Special 510(k) proposes changes to five printed circuit board assemblies (PCBAs), firmware/software, and a minor modification to the simultaneous (coag) activation tone. The function of the devices has not changed.
# Performance Characteristics
The modifications made to the Force FX generator were found to not affect safety or performance through design verification testing, which continued conformance to applicable technical design specifications and performance requirements, including requirements associated with industry safety and performance standards, as follows:
- Basic safety and essential performance in accordance with IEC 60601-1:2005 and IEC 60601-2-. 2:2009
- Electromagnetic compatibility in accordance with IEC 60601-1-2:2007
- . Design verification testing including application of multiple AC mains voltages and temperatures to ensure the proposed modifications perform within design parameters under varying environmental conditions.
- . Software regression/validation
# Clinical Studies
This premarket submission did not rely on the assessment of clinical performance data to demonstrate substantial equivalence.
# Conclusions
The intended use of the Force FX generator was not altered. In addition, the results of testing demonstrate that the modifications to the Force FX generator do not affect the safety or performance of the generator which is substantially equivalent to the predicate Force FX.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.