FOXTROT NC PTCA Balloon Catheter
Device Facts
| Record ID | K143160 |
|---|---|
| Device Name | FOXTROT NC PTCA Balloon Catheter |
| Applicant | Shanghai MicroPort Medical (Group) Co., Ltd. |
| Product Code | LOX · Cardiovascular |
| Decision Date | Mar 5, 2015 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 870.5100 |
| Device Class | Class 2 |
| Attributes | Therapeutic |
Indications for Use
The FOXTROT™ NC PTCA Balloon Catheters are indicated for the balloon catheter dilatation of the stenotic portion of a native coronary artery or bypass graft stenosis for the purpose of improving myocardial perfusion. FOXTROT™ NC PTCA Balloon Catheters are also indicated for the post-delivery expansion of balloon expandable stents.
Device Story
FOXTROT NC PTCA Balloon Catheter is a sterile, single-use, rapid-exchange catheter for coronary angioplasty. Device features integrated shaft system with distal balloon; two platinum-iridium radiopaque markers for fluoroscopic positioning; hydrophilic coating on distal shaft for lubrication. Compatible with 0.014" guide wires. Used by physicians in clinical settings to dilate stenotic coronary arteries/bypass grafts or expand stents. Physician advances catheter over guide wire to target site; inflates balloon to specific diameter/length at recommended pressures. Output is mechanical dilation of vessel or stent. Benefits include improved myocardial perfusion and successful stent deployment.
Clinical Evidence
Bench testing only. No clinical data. Mechanical testing performed per FDA Class II Special Controls Guidance Document for PTCA Catheters, including dimensional verification, balloon fatigue, burst pressure, inflation/deflation time, bond strength, torque, and coating integrity. Biocompatibility testing conducted per current standards.
Technological Characteristics
Rapid-exchange PTCA balloon catheter; 145 cm effective length; 0.014" guide wire compatibility; 2.5-4.0 mm balloon diameters; 10-15 mm balloon lengths. Materials include platinum-iridium radiopaque markers and hydrophilic coating. Nominal pressure 12 atm; rated burst pressure 20 atm. Sterile, single-use.
Indications for Use
Indicated for balloon catheter dilatation of stenotic native coronary arteries or bypass grafts to improve myocardial perfusion and for post-delivery expansion of balloon-expandable stents.
Regulatory Classification
Identification
Standard PTCA Catheter: A PTCA catheter is a device that operates on the principle of hydraulic pressurization applied through an inflatable balloon attached to the distal end. A PTCA balloon catheter has a single or double lumen shaft. The catheter features a balloon of appropriate compliance for the clinical application, constructed from a polymer. The balloon is designed to uniformly expand to a specified diameter and length at a specific pressure as labeled, with well characterized rates of inflation and deflation and a defined burst pressure. The device generally features a type of radiographic marker to facilitate fluoroscopic visualization of the balloon during use. A PTCA catheter is intended for balloon dilatation of a hemodynamically significant coronary artery or bypass graft stenosis in patients evidencing coronary ischemia for the purpose of improving myocardial perfusion. A PTCA catheter may also be intended for the treatment of acute myocardial infarction; treatment of in-stent restenosis (ISR) and/or post-deployment stent expansion. Cutting/scoring PTCA Catheter: A cutting/scoring PTCA catheter is a balloon-tipped catheter with cutting/scoring elements attached, which is used in those circumstances where a high pressure balloon resistant lesion is encountered. A cutting/scoring PTCA catheter is intended for the treatment of hemodynamically significant coronary artery stenosis for the purpose of improving myocardial perfusion. A cutting/scoring PTCA catheter may also be indicated for use in complex type C lesions or for the treatment of in-stent restenosis.
Special Controls
*Classification.* Class II (special controls). The special control for this device is “Class II Special Controls Guidance Document for Certain Percutaneous Transluminal Coronary Angioplasty (PTCA) Catheters.” See § 870.1(e) for the availability of this guidance document.(b) *Cutting/scoring PTCA Catheter* —(1)*Identification.* A cutting/scoring PTCA catheter is a balloon-tipped catheter with cutting/scoring elements attached, which is used in those circumstances where a high pressure balloon resistant lesion is encountered. A cutting/scoring PTCA catheter is intended for the treatment of hemodynamically significant coronary artery stenosis for the purpose of improving myocardial perfusion. A cutting/scoring PTCA catheter may also be indicated for use in complex type C lesions or for the treatment of in-stent restenosis.(2) *Classification.* Class III (premarket approval). As of May 28, 1976, an approval under section 515 of the act is required before this device may be commercially distributed. See § 870.3.
Predicate Devices
- NC Quantum Apex™ Monorail PTCA Dilatation Catheter (K121667)