← Product Code [GDW](/productcode/GDW) · K143129

# easyEndo Linear Cutting Stapler and Loading Unit for Single Use(Stapler;Loading Unit) (K143129)

_Ezisurg Medical Co., Ltd. · GDW · Jan 29, 2015 · General, Plastic Surgery · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K143129

## Device Facts

- **Applicant:** Ezisurg Medical Co., Ltd.
- **Product Code:** [GDW](/productcode/GDW.md)
- **Decision Date:** Jan 29, 2015
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 878.4750
- **Device Class:** Class 2
- **Review Panel:** General, Plastic Surgery
- **Attributes:** Therapeutic, Pediatric

## Indications for Use

The easyEndo® Linear Cutting Stapler and Loading Unit for Single Use have applications in abdominal, gynecologic and pediatric surgery for resection, transection, and creation of anastomoses.

## Device Story

The easyEndo® Linear Cutting Stapler is a manual, single-use surgical instrument used in abdominal, gynecologic, and pediatric surgeries. It functions by placing two, triple-staggered rows of titanium staples into tissue while simultaneously dividing the tissue along a central line. The device is operated manually by a surgeon. It features a safety mechanism (green button) to prevent accidental mis-firing. The stapler can be reloaded and fired up to 25 times during a single procedure. The device provides mechanical tissue closure and transection, facilitating surgical procedures and potentially reducing operative time compared to manual suturing.

## Clinical Evidence

No clinical data. Bench testing only.

## Technological Characteristics

Manual linear cutting stapler; titanium staples (ASTM F67) and polyamide resin construction; 45mm and 60mm suture lengths; 1.0mm, 1.5mm, 2.0mm closed staple heights; irradiation sterilization (SAL 10^-6); safety mechanism (green button). Complies with ASTM F88/F88M-09, ASTM 1929-12, ISO 10993-5, and ISO 10993-10.

## Regulatory Identification

An implantable staple is a staple-like device intended to connect internal tissues to aid healing. It is not absorbable.

## Predicate Devices

- Endo GIA™ Universal Staplers with Duet TRS™ reloads ([K120258](/device/K120258.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is an emblem featuring a stylized image of three human profiles facing to the right, stacked one behind the other.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

January 29, 2015

EziSurgMedical Company, Ltd. % Ms. Diana Hong Mid-Link Consulting Company, Ltd. P.O. Box 120-119 200120 Shanghai China

Re: K143129

Trade/Device Name: easyEndo® Linear Cutting Stapler and Loading Unit for Single Use Regulation Number: 21 CFR 878.4750 Regulation Name: Implantable staple Regulatory Class: Class II Product Code: GDW Dated: October 29, 2014 Received: October 31, 2014

Dear Ms. Hong:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21

{1}------------------------------------------------

CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

# David Krause -S

for

Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

## Indications for Use

510(k) Number (if known)

K143129

Device Name

easyEndo® Linear Cutting Stapler and Loading Unit for Single Use

Indications for Use (Describe)

The easyEndo® Linear Cutting Stapler and Loading Unit for Single Use have applications in abdominal, gynecologic and pediatric surgery for resection, transection, and creation of anastomoses.

|  | <span> <span style="font-size: 12px;"> <span style="text-decoration: underline;"></span> </span> <span style="font-size: 12px;"> <span style="font-family: Symbol;"></span> </span> </span> <span style="font-size: 12px;"> Prescription Use (Part 21 CFR 801 Subpart D)</span> | <span> <span style="font-size: 12px;"> <span style="text-decoration: underline;"></span> </span> <span style="font-size: 12px;"> <span style="font-family: Symbol;">□</span> </span> </span> <span style="font-size: 12px;"> Over-The-Counter Use (21 CFR 801 Subpart C)</span> |
|--|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|--|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

***DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.***

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services  

Food and Drug Administration  

Office of Chief Information Officer

Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

Type of Use (Select one or both, as applicable)

{3}------------------------------------------------

## Tab #7 510(k) Summary

This 510(k) Summary of 510(k) safety and effectiveness information is being submitted in accordance with requirements of SMDA 1990 and Title 21, CFR Section 807.92.

The assigned 510(k) Number: ________

- 1. Date of Preparation: 10/24/2014
- 2. Sponsor Identification

## EziSurg Medical Co., Ltd.

Room 312, Bldg. 2, No. 1690, Cailun Road Zhangjiang High-Tech Park Pudong, Shanghai, 201203, P.R. China

Establishment Registration Number: Not yet registered

Contact Person: Johnny Yan Position: System Submission Specialist Tel: 86-21-50456156 Fax: 86-21-50456176 Email: johnnyyan@ezisurg.com

- 3. Designated Submission Correspondent
Ms. Diana Hong (Primary Contact Person) Mr. Lee Fu (Alternative Contact Person)

## Mid-Link Consulting Co., Ltd

P.O. Box 120-119, Shanghai, 200120, China

Tel: +86-21-22815850, Fax: 240-238-7587 Email: info@mid-link.net

{4}------------------------------------------------

#### 4. Identification of Proposed Device

Trade Name: easyEndo® Linear Cutting Stapler and Loading Unit for Single Use Common Name: Stapler Model: Stapler F12S, F12M, F12L Loading Unit FDU 45W, FDU 45B, FDU 45G, FDU 60B, FDU 60G

#### Regulatory Information

Classification Name: Staple, Implantable; Classification: II; Product Code: GDW; Regulation Number: 21 CFR 878.4750; Review Panel: General & Plastic Surgery;

### Intended Use Statement:

The easyEndo® Linear Cutting Stapler and Loading Unit for Single Use have applications in abdominal, gynecologic and pediatric surgery for resection, and creation, and creation of anastomoses.

### Device Description

The proposed devices, easyEndo® Linear Cutting Stapler and Loading Unit for Single Use, are sterilized and disposable surgical instruments, which have applications in abdominal, gynecologic and pediatric surgery for resection, transection, and creation of anastomoses.

It places two, triple-staggered rows of titanium staples and simultaneously divides the tissue from central line. The easyEndo® Linear Cutting Stapler may be reloaded and fired no more than 25 times in a single procedure.

#### 5. Identification of Predicate Device

510(k) Number: K120258 Product Name: Endo GIA™ Universal Staplers with Duet TRS™ reloads

#### 6. Non-Clinical Test Conclusion

Non clinical tests were conducted to verify that the proposed device met all design specifications as was Substantially Equivalent (SE) to the predicate device. The test results demonstrated that the proposed device complies with the following standards:

ASTM F88/F88M-09, Standard Test Method for Seal Strength of Flexible Barrier Materials. >

{5}------------------------------------------------

- > ASTM 1929-12, Standard Test Method For Detecting Seal Leaks In Porous Medical Packaging By Dye Penetration.
- > USP 37-NF32:2014, <85> Bacterial Endotoxins Test
- > USP 37-NF32 <151> Pyrogen Test (USP Rabbit Test)
- > ISO 10993-5,2009/(R) 2014:Biological evaluation of medical devices - Part 5: Tests for in vitro Cytotoxicity
- > ISO 10993-10:2010, Biological evaluation of medical devices - Part 10: Tests for irritation and skin sensitization.
- 7. Clinical Test Conclusion

No clinical study is included in this submission.

- 8. Substantially Equivalent (SE) Comparison

| ITEM                        | Proposed Device                                                                                                                                                                                             | Predicate Device |
|-----------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------|
| Product Code                | GDW                                                                                                                                                                                                         | Same             |
| Regulation No.              | 21 CFR 878.4750                                                                                                                                                                                             | Same             |
| Class                       | II                                                                                                                                                                                                          | Same             |
| Intended Use                | The easyEndo® Linear Cutting Stapler and<br>Loading Unit for Single Use have<br>applications in abdominal, gynecologic and<br>pediatric surgery for resection, transection,<br>and creation of anastomoses. | Similar          |
| Cutting Mechanism           | Linear                                                                                                                                                                                                      | Same             |
| Operation Principle         | Manual                                                                                                                                                                                                      | Same             |
| Safety Machanism            | Green button for preventing from mis-firing.                                                                                                                                                                | Same             |
| Suture Length               | 45,60 mm                                                                                                                                                                                                    | Same             |
| Closed Staples<br>Height    | 1.0, 1.5, 2.0 mm                                                                                                                                                                                            | Same             |
| Closed Staples<br>Form      | Image: Staples Form                                                                                                                                                                                         | Same             |
| Patient-contact<br>Material | Unalloyed Titanium conforms to ASTM F<br>67, Polyamide resin                                                                                                                                                | Similar          |
| Sterilization               | Irradiation Sterilized, SAL: 10-6                                                                                                                                                                           | Similar          |
| Endotoxin Limit             | 20 EU per Product                                                                                                                                                                                           | Same             |
| Labeling                    | Conforms to 21 CFR part 801                                                                                                                                                                                 | Same             |

Table 1 Comparison of Technology Characteristics

#### 9. Substantially Equivalent (SE) Conclusion

Based on the comparison and analysis above, the proposed devices are determined to be Substantially Equivalent (SE) to the predicate devices.

---

**Source:** [https://fda-staging.innolitics.com/device/K143129](https://fda-staging.innolitics.com/device/K143129)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).