K143128 · Philips Medical Systems Nederlands B.V. · LNH · Mar 20, 2015 · Radiology
Device Facts
Record ID
K143128
Device Name
mDIXON XD
Applicant
Philips Medical Systems Nederlands B.V.
Product Code
LNH · Radiology
Decision Date
Mar 20, 2015
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1000
Device Class
Class 2
Indications for Use
mDIXON XD is a software option intended for use on Achieva, Ingenia 1.5T & 3T MR Systems. It is indicated for use in Magnetic Resonance Imaging of the human body. mDIXON XD is an enhanced 2-point fat-free mDIXON technology, available for gradient echo and spin echo acquisitions. mDIXON XD enables the combination of a fat-free technology with the MultiVane XD motion-free technique for anatomies such as head. mDIXON XD also enables subtractionless MR Angiography in areas of the body where exogenous contrast media has been approved for MR imaging.
Device Story
mDIXON XD is a software option for Philips Achieva and Ingenia 1.5T/3T MR systems; enhances 2-point fat-free imaging using 7-peak fat modeling, B0-correction, and fat-shift correction. Inputs include MR signals from gradient echo (FFE) or spin echo (TSE) pulse sequences. The software transforms these signals into four distinct image contrasts: Water, Fat, In-Phase, and Out-Phase. It integrates with MultiVane XD for motion-free head imaging and enables subtractionless MRA by eliminating the need for pre-contrast baseline scans. Operated by clinicians in a radiology setting, the output images are displayed on the system console for diagnostic review. Benefits include improved fat suppression, reduced motion artifacts, and streamlined MRA workflows by removing the requirement for pre-contrast image subtraction.
Clinical Evidence
Clinical validation compared mDIXON XD to e-Thrive. Results showed improved or equivalent image quality and fat suppression for total body and head/neck exams. Cardiac examinations showed superior image quality and fat suppression. MultiVane XD combined with mDIXON TSE demonstrated reduced motion artifacts compared to mDIXON TSE alone. Testing confirmed all clinical user needs were met on Achieva and Ingenia 1.5T/3T systems.
Technological Characteristics
Software-based MR reconstruction enhancement. Runs on host computer (HP Z420, 3.5 GHz, 64GB RAM, Windows 7). Uses 7-peak fat-peak model, B0-field information, water-fat shift correction, and geometry correction. Compatible with gradient echo (FFE) and turbo spin echo (TSE) sequences. Standalone software option; no external network connectivity. No hardware changes required.
Indications for Use
Indicated for use in Magnetic Resonance Imaging of the human body for patients requiring MR imaging, including head imaging and MR Angiography where exogenous contrast media is approved.
Regulatory Classification
Identification
A magnetic resonance diagnostic device is intended for general diagnostic use to present images which reflect the spatial distribution and/or magnetic resonance spectra which reflect frequency and distribution of nuclei exhibiting nuclear magnetic resonance. Other physical parameters derived from the images and/or spectra may also be produced. The device includes hydrogen-1 (proton) imaging, sodium-23 imaging, hydrogen-1 spectroscopy, phosphorus-31 spectroscopy, and chemical shift imaging (preserving simultaneous frequency and spatial information).
Special Controls
*Classification.* Class II (special controls). A magnetic resonance imaging disposable kit intended for use with a magnetic resonance diagnostic device only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized caduceus symbol, which is a staff with two snakes coiled around it, overlaid on three human profiles facing right. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the symbol.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
March 20, 2015
Philips Medical Systems Nederland B.V. Susan Ouick Regulatory Affairs Specialist 595 Miner Rd CLEVELAND OH 44094
Re: K143128
Trade/Device Name: mDIXON XD Regulation Number: 21 CFR 892.1000 Regulation Name: Magnetic resonance diagnostic device Regulatory Class: II Product Code: LNH Dated: February 06, 2015 Received: February 20, 2015
Dear Ms. Quick:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21. Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Robert A Ochs
Robert Ochs. Ph.D. Acting Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
| DEPARTMENT OF HEALTH AND HUMAN SERVICES | |
|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------|
| Food and Drug Administration | |
| | |
| <b>Indications for Use</b> | Form Approved: OMB No. 0910-0120<br>Expiration Date: January 31, 2017<br>See PRA Statement below. |
| 510(k) Number (if known) | K143128 |
| Device Name | mDIXON XD |
| Indications for Use (Describe) | |
| mDIXON XD is a software option intended for use on Achieva, Ingenia 1.5T & 3T MR Systems. It is indicated for use in Magnetic Resonance Imaging of the human body. mDIXON XD is an enhanced 2-point fat-free mDIXON technology, available for gradient echo and spin echo acquisitions. mDIXON XD enables the combination of a fat-free technology with the MultiVane XD motion-free technique for anatomies such as head. mDIXON XD also enables subtractionless MR Angiography in areas of the body where exogenous contrast media has been approved for MR imaging. | |
| Type of Use (Select one or both, as applicable) | |
| <label><input checked="checked" type="checkbox"/>Prescription Use (Part 21 CFR 801 Subpart D)</label> | <label><input type="checkbox"/>Over-The-Counter Use (21 CFR 801 Subpart C)</label> |
| CONTINUE ON A SEPARATE PAGE IF NEEDED. | |
This section applies only to requirements of the Paperwork Reduction Act of 1995.
***DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.***
The burden time for this collection of information is estimated to average 79 hours per response, including the
time to review instructions, search existing data sources, gather and maintain the data needed and complete
and review the collection of information. Send comments regarding this burden estimate or any other aspect
of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
[PRAStaff@fda.hhs.gov](mailto:PRAStaff@fda.hhs.gov)
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of
information unless it displays a currently valid OMB number."
FORM FDA 3881 (8/14) Page 1 of 1 PSC Publishing Services (301) 443-6740 EF
{3}------------------------------------------------
# PHILI
# Philips Medical Systems Nederland B.V.
# 510(k) Summary
# mDIXON XD
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of 21 CFR 807.92.
- I. General Information 21 CFR 807.92 (a)(1), (2)
| Company Name: | Philips Medical Systems Nederland B.V. |
|-----------------------|-------------------------------------------------------------------------------------------------------------------------------------|
| Address: | Veenpluis 4-6<br>5684 PC Best<br>The Netherlands |
| Registration No .: | 3003768277 |
| Contact Person: | Susan Quick<br>595 Miner Rd<br>Cleveland, Oh 44143<br>Tel: (440)-483-2291<br>Fax: (440)-483-4799<br>E-mail: susan.quick@philips.com |
| Prepared (date): | 2014 October 29 |
| Trade Name of Device: | mDIXON XD |
| Classification: | Class II |
| Regulatory Section: | Magnetic Resonance Diagnostic Device.<br>892.1000 |
| Product Code: | 90LNH |
{4}------------------------------------------------
510(k) Summary mDIXON XD
# 21 CFR 807.92(a)(3):Legally marketed predicate device to which substantial equivalence is claimed:
| 1. Primary Predicate Device: mDIXON | |
|--------------------------------------|--------------------------------------------|
| Manufacturer: | Philips Healthcare |
| Predicate Device k#: | K102344 |
| 2. Reference Devices: | MultiVane and e-THRIVE (Ingenia MR System) |
| Manufacturer: | Philips Healthcare |
| Predicate Device k#: | K110151 |
## 21 CFR 807.92(a)(4): Description of the device that is the subject of this premarket notification:
#### Summary of functions of the device and its major components
mDIXON XD provides an improved water/fat reconstruction algorithm (7-peak fat modeling, B0correction and fat-shift correction). mDIXON XD relies on the 2-point mDIXON fat-free technology making use of unrestricted echo times and is capable of generating four image contrasts (Water, Fat, In-Phase, Out-Phase) in one scan. mDIXON XD can be used with the following:
- Multivane XD (motion correction) for the head
- MR Angiography (subtractionless MRA), removes the need for the subtraction of the pre-● and post-contrast enhanced images.
- Gradient echo (FFE Fast Field Echo) and turbo spin echo (TSE) sequences to achieve ● efficient fat-free scanning compatibility, even at large FOV or sub-millimetric resolutions
mDIXON XD is supported on the following systems:
- 3.0T Ingenia
- 3.0T Achieva ●
- 1.5T Ingenia
- 1.5T Achieva ●
The functionality is supported on all available gradient performance levels. Optimized protocols will be provided for the different performance points. mDIXON XD is supported on the centralized data acquisition systems of the Achieva systems as well as the digitally networked data acquisition system of the Ingenia systems. The data acquisition system is fully transparent to the pulse sequences and mDIXON XD reconstructions.
{5}------------------------------------------------
# 21 CFR 807.92(a)(5): Intended Use
mDIXON XD is a software option intended for use on Achieva. Ingenia I .5T & 3T MR Systems. It is indicated for use in Magnetic Resonance Imaging of the human body. mDIXON XD is an enhanced 2-point fat-free mDIXON technology, available for gradient echo and spin echo acquisitions. mDIXON XD enables the combination of a fat-free technology with the MultiVane XD motion-free technique for anatomies such as head. mDIXON XD also enables subtractionless MR Angiography in areas of the body where exogenous contrast media has been approved for MR imaging.
## 21 CFR 807.92(a)(6): Technological Characteristics:
The main functional units in the software are:
- (software for controlling and diagnosing the hardware) - MRChains
- Methods (acquisition of MR signals by means of MR pulse sequences)
- Reconstruction (transforming the MR signals to images)
- Patient Administration (storing of the images in the database and providing access)
- Viewing (display of images)
The impact of mDIXON XD on these functional units are:
- MRChains: Provide static B0-field information, as measured at system installation time. -
- Methods: Use an existing multi-echo (usually 2 echoes) FFE, TFE or multi-acquisition TSE for mDIXON XD, within the cleared limits provided by the basic MR system. The requested output image types can be specified by the user in the scan protocol.
- -Reconstruction: Add a new calculation function that calculates water, fat, IP and OP images. Apply a fat-peak-model and B0-field information to separate water from fat. Make use of water-fat shift correction and geometry correction.
- -Patient Administration and Viewing: enable storage and display of the output images with appropriate labels (water, fat, IP, OP)
No off-the-shelf software is used for mDIXON XD. The off-the-shelf software used in the basic MR system is cleared. mDIXON XD is not designed to be connected to an external network. mDIXON XD does not require any change of the hardware platform. The major extensions introduced by mDIXON XD are in methods pulse sequence code, and in reconstruction for the enhanced calculation. Those run on the host computer:
Computer characteristics:
- Manufacturer: HP; Model: Z420; Processor clock: 3.5 GHz; RAM: 64 GB RAM; -Processors: six core with hyper threading
- Operating system: Windows 7, 64 bits -
Changes to operator workflow for mDIXON XD are:
- Protocol selection: The operator selects an ExamCard with mDIXON XD protocols -
- Planscan phase: Optionally the operator may want to change the predefined selection of output image-types.
- For mDIXON XD MRA: During scanning no pre-contrast baseline has to be scanned, saving one multi-station table-stroke, compared to conventional MRA.
All other steps are not changed. The generated image types can be viewed, post-processed, printed
{6}------------------------------------------------
and archived as any other image type.
#### 21 CFR 807.92(b)(1): Brief discussion of nonclinical tests submitted, referenced or relied on in this premarket notification:
mDIXON XD has been verified to function with the Achieva and Ingenia 1.5T and 3.0T MR systems.
The verification testing showed the examcards could be loaded, the correct parameters were listed for each scan, there were no issues when turning on SENSE, the four image types could be selected, all scans ran properly and images were provided.
The conclusion from this report is:
All the tests performed for mDIXON XD were successful. Workflow was smooth and no problems occurred.
All defects have been corrected.
#### 21 CFR 807.92(b)(2): Brief discussion of clinical tests submitted, referenced or relied on in this premarket notification:
Clinical user needs are tested as part of validation testing showed that for total body and head/neck examinations, the Image Quality, fat suppression and fat suppression over large FOV was similar or better with mDIXON XD compared to e-Thrive. For cardiac examinations the image quality and fat suppression were better with mDIXON XD. For MultiVane XD in combination with mDIXON TSE, the motion related artifacts were reduced compared to mDIXON TSE alone and the MultiVane XD with mDIXON XD can be used with T2 in head imaging.
The conclusion from testing the device is:
The clinical validation of mDIXON XD has completed successfully. All clinical user needs have passed for mDIXON XD on the Achieva and Ingenia 1.5T and 3T systems. All defects have been analyzed and are verified to be solved and will be closed, no new hazards were identified.
## 21 CFR 807.92(b)(3): The conclusions drawn from the nonclinical and clinical tests that demonstrate that the device is as safe, as effective, and performs as well as or better than the legally marketed device identified in paragraph (a)(3) of this section:
The nonclinical and clinical tests have demonstrated that the device is safe and works according to its intended use.
mDIXON XD software does not introduce new indications for use, nor does the use of the device result in any new potential hazard. Philips Medical Systems considers mDIXON XD Software to be substantially equivalent to the above mentioned predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.