K143117 · American Orthodontics · FRG · Oct 29, 2015 · General Hospital
Device Facts
Record ID
K143117
Device Name
Wire Sterilization Package with Indicators
Applicant
American Orthodontics
Product Code
FRG · General Hospital
Decision Date
Oct 29, 2015
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.6850
Device Class
Class 2
Indications for Use
Wire Sterilization Package with Indicators is intended to be used to enclose and package American Orthodontics' line of Orthodontic Arch Wires. These packages are shipped unsterile and for single-use only. This packaging offers the health provider (trained dental professional and/or Orthodontist) the option of sterilizing the enclosed arch wire using steam or ethylene oxide gas. The indicators will display a visual color change to steam or ethylene oxide gas and distinguish between processed and unprocessed wire.
Device Story
Device is a single-use pouch manufactured from bleached surgical paper and plastic film; designed to enclose American Orthodontics' arch wires. Pouch features external chemical process indicators that change color upon exposure to steam or ethylene oxide (EO) gas; allowing clinicians to distinguish between processed and unprocessed wires. Used in dental clinics/orthodontic offices by trained professionals. After sterilization, arch wire must be used immediately. Device provides a method for sterilization without affecting wire material or functionality. Benefits include convenient, validated sterilization of arch wires prior to patient use.
Clinical Evidence
Bench testing only. Performance testing validated Sterility Assurance Level (SAL) using biological indicators in worst-case locations. Residual testing for EO followed ISO 10993-7. Indicator validation confirmed color change performance per AAMI/ANSI/ISO 11140-1:2014. Biocompatibility/cytotoxicity testing (ISO 10993-5, USP <87>) performed on arch wires post-sterilization; all samples passed.
Technological Characteristics
Pouch materials: bleached surgical paper (21CFR186.1673), polyester film (21CFR177.1630), polypropylene (21CFR177.1520), and adhesive (21CFR175.105). Chemical indicators meet AAMI/ANSI/ISO 11140-1:2014. Sterilization methods: Steam (132°C, 15 min) or EO (725 mg/L, 55°C, 1 hour).
Indications for Use
Indicated for trained dental professionals and/or Orthodontists to package and sterilize American Orthodontic arch wires using steam or ethylene oxide gas. For single-use only.
Regulatory Classification
Identification
A sterilization wrap (pack, sterilization wrapper, bag, or accessories, is a device intended to be used to enclose another medical device that is to be sterilized by a health care provider. It is intended to allow sterilization of the enclosed medical device and also to maintain sterility of the enclosed device until used.
Predicate Devices
Safe Secure Sterilization Pouch with Steam and Ethylene Oxide Process Indicators (K112591)
{0}------------------------------------------------
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-000
October 29, 2015
American Orthodontics Ms. Trang Adams Regulatory Affairs Specialist 3524 Washington Avenue Sheboygan, WI 53081
Re: K143117
Trade/Device Name: Wire Sterilization Package with Indicators Regulation Number: 21 CFR 880.6850 Regulation Name: Sterilization Wrap Regulatory Class: II Product Code: FRG; JOJ Dated: September 21, 2015 Received: September 29, 2015
Dear Ms. Adams,
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
*Tejashri Purohit-Sheth, M.D.*
Tejashri Purohit-Sheth, M.D. Clinical Deputy Director DAGRID/ODE/CDRH FOR
Erin I. Keith, M.S. Director Division of Anesthesiology, General Hospital. Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K143117
Device Name Wire Sterilization Package with Indicators
#### Indications for Use (Describe)
Wire Sterilization Package with Indicators is intended to be used to enclose and package American Orthodontics' line of Orthodontic Arch Wires. These packages are shipped unsterile and for single-use only. This packaging offers the health provider (trained dental professional and/or Orthodontist) the option of sterilizing the enclosed arch wire using steam or ethylene oxide gas. The indicators will display a visual color change to steam or ethylene oxide gas and distinguish between processed and unprocessed wire.
#### STERILIZATION PARAMETERS
Ethylene Oxide [100% E.O.] Gas
- 1. EO Concentration: 725 mg/L
- 2. Exposure Temperature: 130°F (55°C)
- 3. Exposure Time: 1 hour
- 4. Aeration Time: 8 hours
Upon completion of the process, the EO color indicator will change from Blue to Cocoa Brown.
Steam - Gravity
- 1. Exposure Temperature: 270°F (132°C)
- 2. Exposure Time: 15 minutes
- 3. Dry Time: 30 minutes
Upon completion of the process, the STEAM color indicator will change from Blue to Black.
Please Note: Upon completion of the process, the archwire is to be used immediately.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------------------------------------------------------|
| <span style="text-decoration: underline;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) |
| ☐ Over-The-Counter Use (21 CFR 801 Subpart C) |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for American Orthodontics. The logo consists of the letters "AO" in a stylized font on the left, followed by the words "AMERICAN" and "ORTHODONTICS" stacked on top of each other on the right. The letters and words are all in a dark blue color. The logo is simple and modern, and it is likely used to represent the company's brand.
www.americanortho.com
1 800 558 7687 +1 920 457 5051 office +1 920 457 1485 fax
Page 5-1
# 510(k) Summary 21CFR807.92
### Date Prepared: October 23, 2015
#### Company Information:
American Orthodontics 3524 Washington Avenue Sheboygan, WI 53081 Phone: 920-457-5051 Fax: 920-457-5773
#### Contact Information:
Trang Adams / Regulatory Affairs Specialist
#### Device Information:
Trade Name: Wire Sterilization Package with Indicators Common Name: Wire Sterilization Package with Indicators Classification Name: Sterilization Wrap; Sterilization Process Indicator 510(k) Number: K143117 Classification Code: FRG; JOJ Regulation Number (21CFR): 880.6850; 880.2800
#### Component Device Information:
Common Name: Orthodontic Arch Wires Classification Name: Wire, Orthodontic 510(k) Number: Exempt (Class I) Classification Code: DZC Regulation Number (21CFR): 872.5410 American Orthodontics' Models: Stainless Steel Wires, Nickel Titanium (NiTi) Wires, Beta Titanium Wires, Chromium Cobalt (CrCo) Wires, Copper Nickel Titanium (CuNiTi) Wires
#### Predicate Devices Information:
Primary Predicate:
Product/Trade Name: Safe Secure Sterilization Pouch with Steam and Ethylene Oxide Process Indicators Manufacturer: Safe Secure Packaging Company, Limited 510(k) #: K112591 Classification Name: Sterilization Wrap Classification Code: FRG Regulation Number (21CFR): 880.6850
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for American Orthodontics. The logo consists of the letters "AO" in a stylized font on the left, followed by the words "AMERICAN" and "ORTHODONTICS" stacked on top of each other on the right. The letters and words are all in a dark blue color. The logo is simple and modern, and it is likely used to represent the company's brand.
www.americanortho.com
1 800 558 7687 +1 920 457 5051 office +1 920 457 1485 fax
# Page 5-2
Secondary Predicate: Product/Trade Name: Cross-Checks Indicator Manufacturer: SteriTec Products Mfg. Co., Inc 510(k) #: K951113 Classification Name: Sterilization Process Indicator Classification Code: JOJ Regulation Number (21CFR): 880.2800
American Orthodontics' new device combines a sterilizable pouch (Classification Code: FRG) with sterilization process indicators (Classification Code: JOJ) that contains an Arch Wire (Classification Code: DZC). It was deemed appropriate to choose two predicates containing the classification codes that were suitable to the Wire Sterilization Package with Indicators to demonstrate substantial equivalency.
## Description of the Device:
American Orthodontics' Wire Sterilization Package with indicators is manufactured from bleached surgical paper and plastic film with external chemical process indicators. This packaging is a pouch that is enclosed on all four sides and contains a single arch wire. These packages are shipped unsterile and for single-use only. This packaging offers the health provider (trained dental professional and/or Orthodontist) the option of sterilizing the enclosed arch wire using steam or ethylene oxide gas. Upon completion of the sterilization process, the enclosed arch wire is intended to be used immediately.
This packaging is not offered as a stand-alone product, but is strictly used for the packaging of American Orthodontics' Arch Wires (Classification Code: DZC; 510(k) Exempt).
The bleached surgical paper conforms to the FDA 21CFR186.1673 and can be sterilized using steam or ethylene oxide gas.
The plastic film is sealed to the bleached surgical paper and conforms to 21CFR177.1630 (Polyester), 21CFR175.105 (Adhesive) and 21CFR177.1520 (Polypropylene).
The process indicators meet the performance requirements of AAMI/ANSI/ISO 11140-1:2014.
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for American Orthodontics. The logo consists of the letters "AO" in a stylized font, followed by the words "AMERICAN ORTHODONTICS" in a smaller, sans-serif font. The color of the logo is a dark blue. The logo is simple and modern, and it is likely used to represent the company's brand.
www.americanortho.com
1 800 558 7687 +1 920 457 5051 office +1 920 457 1485 fax
# Page 5-3
## Indications for Use:
Wire Sterilization Package with Indicators is intended to enclose and package American Orthodontics' line of Orthodontic Arch Wires. These packages are shipped unsterile and for single-use only. This packaging offers the health provider (trained dental professional and/or Orthodontist) the option of sterilizing the enclosed arch wire using steam or ethylene oxide gas. The indicators will display a visual color change to signify exposure to steam or ethylene oxide gas and distinguish between processed and unprocessed wire.
### STERILIZATION PARAMETERS
Ethylene Oxide [100% E.O.] Gas
- 1. EO Concentration: 725 mg/L
- Exposure Temperature: 130°F (55°C) 2.
- 3. Exposure Time: 1 hour
- 4. Aeration Time: 8 hours
Upon completion of the process, the EO color indicator will change from Blue to Cocoa Brown.
#### Steam – Gravity
- 1. Exposure Temperature: 270°F (132°C)
- 2. Exposure Time: 15 minutes
- 3. Dry Time: 30 minutes
Upon completion of the process, the STEAM color indicator will change from Blue to Black/Brownish Black.
Please Note: Upon completion of the process, the archwire is to be used immediately.
The Indications for Use statement for American Orthodontics' Wire Sterilization Package with indicators is not identical to the predicates however, the differences do not alter the intended use or safety and effectiveness of the device relative to the predicates. Both devices have the same intended use, to be able to distinguish between processed and unprocessed devices.
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for American Orthodontics. The logo consists of the letters "AO" in a stylized font, followed by the words "AMERICAN" and "ORTHODONTICS" stacked on top of each other. The logo is in a dark blue color. The logo is simple and modern, and it is likely used to represent the company's brand.
www.americanortho.com
1 800 558 7687 +1 920 457 5051 office +1 920 457 1485 fax
# Page 5-4
## Device Comparison:
For detailed comparisons, please reference Section 12: Substantial Equivalence Discussion.
The table below outlines comparisons of the predicates and American Orthodontics' Wire Sterilization Packaging with Indicators to show substantial equivalency.
| | Device Name / Manufacturer | | | |
|---------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------|
| Element | Safe Secure Sterilization Pouch<br>with Steam and Ethylene Oxide<br>Process Indicators / Safe Secure<br>Packaging Company, Limited | Cross-Checks / SteriTec<br>Products Mfg. Co., Inc | Wire Sterilization<br>Packaging with<br>Indicators/ American<br>Orthodontics | Substantial<br>Equivalence<br>Analysis |
| 510(k) Number | K112591 | K951113 | K143117 | N/A |
| Classification<br>Code/Regulation<br>Number | FRG<br>880.6850 | JOJ<br>880.2800 | FRG/JOJ<br>880.2800/880.6850 | N/A |
| Intended Use | Safe Secure Sterilization Pouch with<br>Steam and Ethylene Oxide Process<br>Indicators Models: ABHSP, ABSSP, is<br>intended to be used to enclose<br>another medical device that is to be<br>sterilized by a health provider by<br>gravity steam and ethylene oxide<br>(EtO). The recommended steam<br>sterilization cycle parameters are 30<br>minutes at 121°C (250°F). The<br>recommended EtO gas sterilization<br>cycle is 735mg/L of ethylene oxide<br>(EtO) for 1 hour at 55°C (130°F) and<br>50% to 80% RH. The pouch's<br>external chemical ink indicators on<br>the pouches are intended to<br>demonstrate that the device has been<br>exposed to steam or EtO sterilization<br>process and to distinguish between<br>processed and unprocessed devices.<br>The pouch is intended to allow<br>sterilization of the enclosed medical<br>device. | SteriTec CROSS-CHECKS<br>sterilization chemical<br>monitoring strips are<br>designed to be utilized in<br>steam sterilizer operating<br>at 132°C to 135°C (270°F<br>to 276°F). When used as<br>directed, the SteriTec<br>Cross-Checks indicator<br>strips give visible<br>indication that sterilizing<br>conditions were met.<br>During steam<br>sterilization, the check<br>mark at the end of the<br>strip changes from white<br>to black, becoming as<br>dark or darker in<br>intensity than the black<br>reference arrow printed<br>on the strip. | Wire Sterilization Package<br>with Indicators is intended<br>to be used to enclose and<br>package American<br>Orthodontics' line of<br>Orthodontic Arch Wires.<br>These packages are shipped<br>unsterile and for single-use<br>only. This packaging offers<br>the health provider (trained<br>dental professional and/or<br>Orthodontist) the option of<br>sterilizing the enclosed arch<br>wire using steam or<br>ethylene oxide gas. The<br>indicators will display a<br>visual color change to<br>signify exposure to steam or<br>ethylene oxide gas and<br>distinguish between<br>processed and unprocessed<br>wire. | Equivalent |
| Device Design &<br>Components | Pouch: 2-30" Wx2-30"L, Medical-<br>grade paper or Tyvek, Plastic Film<br>Chemical Indicator:<br>not disclosed; printed on surgical<br>paper | Strip: 0.5625"W x 7.66"<br>L, 80 lbs Cover<br>Chemical Indicator:<br>White to Dark; printed on<br>paper strip over laminate<br>of PET | Pouch: 3.5" Wx4.5"L,<br>60g/m² Surgical Paper, Blue<br>Film<br>Chemical Indicator:<br>Steam: Blue to Brownish<br>Black/Black; Ethylene<br>Oxide: Blue to Coco Brown | Equivalent |
| Indicator Agent | Indicating Ink: Proprietary to Safe<br>Secure, yields a color-change | Indicating Ink: Lead<br>based agent to yield<br>color-change | Steam Indicating Ink: Lead<br>based agent to yield color-<br>change<br>Ethylene Oxide Indicating<br>Ink: Water based agent to<br>yield color-change | Equivalent |
| Sterilization<br>Methods | Steam: 121°C for 30 minutes.<br>Ethylene Oxide gas : 55°C for 1 hour | Steam: 132°C - 134°C for<br>3 minutes | Steam: 132°C for 15<br>minutes<br>Ethylene Oxide: 55°C for 1<br>hour | Equivalent |
| Standards Met | ISO 11140-1 (Indicators) | ISO 11140-1 (Indicators) | ISO 11140-1 (Indicators) | Equivalent |
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the logo for American Orthodontics. The logo consists of the letters "AO" in a stylized font, with the word "AMERICAN" written to the right of the letters. Below the word "AMERICAN" is the word "ORTHODONTICS" in a smaller font. The logo is in a dark blue color.
www.americanortho.com
1 800 558 7687 +1 920 457 5051 office +1 920 457 1485 fax
# Page 5-5
## Performance Data:
Please reference Section 15: Biocompatibility and Section 18: Performance Testing - Bench for more details.
The following performance data were provided:
## MATERIAL COMPATIBILITY TESTING
Material Compatibility Testing for the enclosed arch wires were conducted following ISO 15841:2014 Dentistry – Wires for Use in Orthodontics before and after the sterilization processes. Sterilization of the arch wires using Steam, Formaldehyde and Ethylene Oxide had no effect on the material or functionality of the arch wires.
## PERFORMANCE TESTING
Performance Testing was conducted by placing spores in the most difficult to sterilize locations. Samples were processed at half cycle and biological indicators (BI) removed after exposure. Biological indicators were aseptically transferred to culture media and incubated. Performance testing validated that the cycles are able to achieve the desired Sterility Assurance Level (SAL).
## RESIDUAL TESTING
Residual testing was conducted following ISO 10993-7:2008/(R) 2012 Biological Evaluation of Medical Devices - Part 7: Ethylene Oxide Sterilization Residuals. Test samples met reference guidelines.
## INDICATOR VALIDATION
Indicator Validation testing was conducted following AAMI/ANSI ISO 11140-1 Sterilization of Health Care Products – Chemical Indicators – Part 1: General Requirements. Results verified that the chemical indicators tested changed to an appropriate signal color when tested under the "Pass Cycle" conditions and met the "Pass Cycle" requirements for Class I process indicators.
## BIOCOMPATIBILITY/LEACHABILITY: CYTOTOXICITY TESTING
Cytotoxicity Testing was conducted after sterilization processes following ISO 10993-5: Biological Evaluation of Medical Devices – Part 5: Tests for In-Vitro Cytotoxicity and United States Pharmacopeia/National Foundry, <87> Biological Reactivity Test, In Vitro; Elution Test. American Orthodontics' Wire Sterilization Package with Indicators do not come into contact with patients. The enclosed archwires met the USP and ISO 10993-5 requirements for the test. All test samples PASSED.
{8}------------------------------------------------
Image /page/8/Picture/0 description: The image shows the logo for American Orthodontics. The logo consists of the letters "AO" in a stylized font, followed by the words "AMERICAN ORTHODONTICS" in a sans-serif font. The letters and words are all in a dark blue color. The logo is simple and modern, and it is likely used to represent the company's brand.
www.americanortho.com
1 800 558 7687 +1 920 457 5051 office +1 920 457 1485 fax
# Page 5-6
# BIOCOMPATIBILITY - (ENCLOSED ARCHWIRES)
Archwires were sent to independent labs for biocompatibility testing before sterilization according to ISO 10993-5:2009, Biological Evaluation of Medical Devices – Part 5: Tests for In Vitro Cytotoxicity and ISO 10993-10, Tests for Irritation and Sensitization - Sensitization Test. All test samples PASSED.
Biocompatibility Justification for stainless steel archwires is included in Section 15: Biocompatibility.
## Summary:
The Wire Sterilization Package with Indicators is provided as a convenient and purely optional method for the trained dental professional and/or Orthodontist to sterilize American Orthodontics' arch wires. This does not affect the function of the enclosed arch wires, nor does it affect the original purpose of the package – which is to store and ship individual arch wires.
## Conclusion:
American Orthodontics' Wire Sterilization Package with Indicators has the following similarities to those legally marketed under 510(k) numbers K112591 (Primary Predicate) and K951113 (Secondary Predicate):
- . Same intended use; and
- Same technological characteristics through device design, indicator agent, sterilization . method and incorporation of similar materials.
Generated testing data in accordance to "ANSI/AAMI/ISO 11140-1:2014 Sterilization of health care products – Chemical indicators" demonstrates that the indicators on American Orthodontics' Wire Sterilization Package with Indicators are substantially equivalent to the predicate devices.
Any slight differences do not affect the original function or intended purpose of the device.
Information contained in this 510(k) does not raise new questions or safety and effectiveness, and, demonstrates it is at least as safe and effective as the listed predicates.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.