K143109 · BTL Industries, Inc. · GEI · Mar 6, 2015 · General, Plastic Surgery
Device Facts
Record ID
K143109
Device Name
XP300
Applicant
BTL Industries, Inc.
Product Code
GEI · General, Plastic Surgery
Decision Date
Mar 6, 2015
Decision
SESE
Submission Type
Special
Regulation
21 CFR 878.4400
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The XP300 device is indicated for the primary treatment of dermatological procedures for non-invasive treatment of periorbital wrinkles and rhytids.
Device Story
XP300 is a microprocessor-driven electrosurgical device for non-invasive periorbital wrinkle and rhytid treatment. It applies high-frequency electromagnetic energy to tissue to induce heating (39°C to 42°C). The system includes a control unit with an 8.4" color touch screen and an applicator with integrated control buttons and a display. The device features an ultrasonic element added to the applicator to improve patient comfort during treatment. Operated by clinicians in a clinical setting, the device provides real-time feedback on therapeutic parameters, power levels, and method type via the user interface. The device monitors patch electrode contact quality to ensure safety. The output is controlled by the clinician to achieve desired tissue heating, which is intended to improve the appearance of wrinkles and rhytids.
Clinical Evidence
No clinical data provided. Substantial equivalence is supported by non-clinical performance testing, including frequency accuracy, carrier waveform, nominal output power, power fluency, and tissue heating verification. Biocompatibility and electrical safety were validated against ISO and IEC standards.
Technological Characteristics
Class II electrosurgical device; 120W nominal power; Monopolar RF energy; 8.4" color touch screen interface; 100-240 VAC power source. Materials conform to ISO 10993-1. Complies with IEC 60601-1 (safety), IEC 60601-1-2 (EMC), IEC 60601-2-2 (HF surgical equipment), and ISO 14971 (risk management). Features ultrasonic element for patient comfort. Not sterile.
Indications for Use
Indicated for non-invasive treatment of periorbital wrinkles and rhytids in patients undergoing dermatological procedures.
Regulatory Classification
Identification
An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized image of three human profiles facing right, with flowing lines suggesting hair or wings. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the image.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
March 10, 2015
BTL Industries Incorporated Mr. Jan Zarsky Director 47 Loring Drive Framingham, Massachusetts 01702
Re: K143109 Trade/Device Name: XP300 Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical, Cutting & Coagulation Device and Accessories Regulatory Class: Class II Product Code: GEI Dated: February 4, 2015 Received: February 9, 2015
Dear Mr. Zarsky:
This letter corrects our substantially equivalent letter of March 6, 2015.
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations. Title 21. Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical
{1}------------------------------------------------
device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 8, please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# Jennifer R. Stevenson -S
For Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known)
Device Name XP300
Indications for Use (Describe)
The XP300 device is indicated for the primary treatment of dermatological procedures for non-invasive treatment of periorbital wrinkles and rhytids.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
_ Over-The-Counter Use (21 CFR 801 Subpart C)
#### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
#### FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## 510(k) Summary
for
### XP300
#### 1. Submission Sponsor
BTL Industries, 47 Loring Drive Framington MA, 01702 USA Phone: 866.285.1656 Fax: 866.499.2502 Contact: Jan Zarsky, Director
2. Submission Correspondent
Mr. Jan Zarsky Director BTL Industries 47 Loring Drive Framington, MA 01702 USA Email: zarsky@btlnet.com
- 3. Date Prepared
January 15, 2015
- 4. Device Identification
| Trade/Proprietary Name: | XP300 |
|----------------------------|----------------------------------------------------------------|
| Common/Usual Name: | Electrosurgical, cutting & coagulation & accessories |
| Classification Name: | Electrosurgical cutting and coagulation device and accessories |
| Classification Regulation: | 21 CFR 878.4400 |
| Product Code: | GEI |
| Device Class: | Class II |
| Classification Panel: | General & Plastic Surgery |
- 5. Legally Marketed Predicate Device(s)
| Trade/Proprietary Name: | XP200 |
|----------------------------|----------------------------------------------------------------|
| Common/Usual Name: | Electrosurgical, cutting & coagulation & accessories |
| Classification Name: | Electrosurgical cutting and coagulation device and accessories |
| Classification Regulation: | 21 CFR 878.4400 |
| Product Code: | GEI |
| Device Class: | Class II |
| Classification Panel: | General & Plastic Surgery |
{4}------------------------------------------------
#### 6. Device Description
The XP300 device is indicated for the primary treatment of dermatologic procedures for noninvasive treatment of periorbital wrinkles and rhytids. The XP300 system is a state-of-theart device to apply therapy by a non-invasive method of high-frequency field.
The control unit of the device is fitted with a color touch screen, which significantly facilitates the use of the device. The design of the device enables the operator to see the on-screen information from various angles during the procedure. In addition, the brightness of the screen can be set to match the lighting in the room. The on-screen information will guide you through the entire therapy by means of easy setting of parameters using touch-screen buttons and knobs/keys on the device.
For easier control, the applicator is equipped with buttons, enabling to operate the device during therapy, and a display, which shows the set and indicated parameters. The XP300 device incorporates the ultrasonic feature for added patient comfort.
Any therapeutic parameter can be set easily by simple use of the touch-screen buttons. During the entire therapy time the device informs you about the therapeutic method, the type of the therapy applied, the set power, and other necessary data.
The XP300 consists of the following main components:
- . microprocessor-driven control unit
- . high-frequency electromagnetic generator and ultrasonic element
- . user interface with 8.4" colour touch screen
- applicator with colour screen and control buttons .
The main change between this device and the predicate is the inclusion of the Ultrasonic convenience function which is intended to improve patient's comfort during the procedure
- 7. Indication for Use Statement
The XP300 is indicated for the primary treatment of dermatological procedure for noninvasive treatment of periorbital wrinkles and rhytids.
- 8. Substantial Equivalence Discussion
The following table compares the XP300 to the predicate device, XP200, with respect to intended use, technological characteristics and principles of operation, providing more detailed information regarding the basis for the determination of substantial equivalence.
{5}------------------------------------------------
| Manufacturer | BTL Industries, Inc. | BLT Industries, Inc. | SIGNIFICANT<br>DIFFERENCES |
|-------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Trade Name | XP300 | XP200 | n/a |
| 510(k) Number | Unknown | K122966 | n/a |
| Product Code | GEI | GEI | No difference |
| Regulation<br>Number | 21 CFR 878.4400 | 21 CFR 878.4400 | No difference |
| Regulation Name | Electrosurgical cutting<br>and coagulation device<br>and accessories | Electrosurgical cutting<br>and coagulation device<br>and accessories | No difference |
| Intended Use | Application of heat to<br>the tissue w/RF energy | Application of heat to the<br>tissue w/RF energy | No difference |
| Indications for Use | The XP300 is indicated<br>for the primary<br>treatment of<br>dermatological<br>procedure for non-<br>invasive treatment of<br>periorbital wrinkles and<br>rhytids. | The XP200 is indicated<br>for the primary<br>treatment of<br>dermatological<br>procedure for non-<br>invasive treatment of<br>periorbital wrinkles and<br>rhytids. | No difference |
| Material | The unit is constructed<br>of material that conform<br>with safety standards<br>and requirements. | The unit is constructed of<br>material that conform<br>with safety standards<br>and requirements. | No difference |
| Sterile | No | No | No difference |
| Single-Use | No | No | No difference |
| Interface | Touch-Screen user<br>applied interface to<br>program and set the<br>controls for the patient<br>application; there is a<br>hand-piece utilized to<br>deliver the treatment. | Touch-Screen user<br>applied interface to<br>program and set the<br>controls for the patient<br>application; there is a<br>hand-piece utilized to<br>deliver the treatment. | No difference. |
| The ultrasonic<br>convenience<br>Function | Yes | No | No significant<br>difference.<br><br>The change is being<br>introduced to improve<br>patient's comfort<br>during the application.<br><br>There is no change to<br>the principle of<br>operation. |
| Table 5A – Comparison of Characteristics | |
|---------------------------------------------|--|
| | |
{6}------------------------------------------------
| Manufacturer | BTL Industries, Inc. | BLT Industries, Inc. | SIGNIFICANT<br>DIFFERENCES |
|--------------------------------------------|------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------|----------------------------|
| Trade Name | XP300 | XP200 | n/a |
| Color Touch Screen | 8.4" (2.15 cm)/640x480 pixels | 8.4" (2.15 cm)/640x480 pixels | No difference |
| Energy Type | Radiofrequency | Radiofrequency | No difference |
| Modes of Operation | Monopolar | Monopolar | No difference |
| Nominal Operating power | 120 W | 120 W | No difference |
| Operating temperature | 18°C to 30°C | 18°C to 30°C | No difference |
| Operating Humidity | 60%-75% | 60%-75% | No difference |
| Temperature Treatment Range | 39°C to 42°C | 39°C to 42°C | No difference |
| Power Level Adjustable via Applicator | Yes | Yes | No difference |
| Patch Electrode Contact Quality Monitoring | Yes | Yes | No difference |
| RF Energy Emission Indicator | Yes; Information displayed on the screen of the applicator and on the main screen of the unit. | Yes; Information displayed on the screen of the applicator and on the main screen of the unit. | No difference |
| Applicator Dimensions | 4.25" x 2.6" x 7"<br>(11cm x 7cm x18cm) | 4.25" x 2.6" x 7"<br>(11cm x 7cm x18cm) | No difference |
| Complies with ISO 10993-1 | Yes | Yes | No difference |
| Electrical Safety Testing Passed | Class II, BF | Class II, BF | No difference |
| Energy Source | 100-240 VAC, max 4A,<br>50-60 Hz | 100-240 VAC, max 4A,<br>50-60 Hz | No difference |
| System Dimensions | 16" X 10.6" x 11.9"<br>(41cm x 22cm x 18cm) | 16" X 10.6" x 11.9"<br>(41cm x 22cm x 18cm) | No difference |
| System Weight | 16lb (7.3kg) | 16lb (7.3kg) | No difference |
{7}------------------------------------------------
#### Non-Clinical Performance Data 9.
This Special 510(k) proposes a modification to the applicator by incorporating an ultrasonic element to improve patient's comfort during the application.
This proposed change does not impact the indications for use, technology, principle of operation, patient contact materials, or packaging of the device. Testing has been performed on the modified applicator to demonstrate that treatment heads applied to the patient shall have a convenient patient contact surface temperature.
The change only applies to the applicator therefore XP300 continues to conform to the medical device safety standards applicable to both XP300 and XP200. The device has been evaluated to demonstrate electrical, electromagnetic and mechanical safety. Medical device software life cycle processes have been verified and validated as well as output and biocompatibility.
The system complies with the following standards:
ISO 14971 Medical devices - Application of risk management to medical devices IEC 62304 Medical Device Software - Software Life Cycle Processes
#### MedicalElectricalEquipment:
IEC 60601-1 General requirements for safety
IEC 60601-1-2 Collateral Standard: Electromagnetic compatibility – Requirements and Tests IEC 60601-2-2 Particular requirements for the safety of high frequency surgical equipment IEC 60601-1-6 General requirements for basic safety and essential performance Collateral standard: Usability
EMCRequirements for Medical Equipment: IEC 61000-4-2; IEC 61000-4-3; IEC 61000-4-4; IEC 61000-4-5
BiologicalEvaluationforMedicalDevices: ISO 10993-1; ISO 10993-5; ISO 10993-10
#### PerformanceTesting:
- Frequency Accuracy and Carrier Wave Form .
- . Carrier Wave Nominal Output Power
- Power Fluency
- Tissue Heating:
The performance testing confirmed safety of simultaneous application of radiofrequency and ultrasound energy.
The XP300 meets all the requirements for overall design, biocompatibility, and electrical safety confirms that the output meets the design inputs and specifications. The XP300 complies with the applicable voluntary standards for biocompatibility.
{8}------------------------------------------------
#### 10. Clinical Performance Data
There was no clinical testing required to support the XP 300 medical device as the indications for use are equivalent to the predicate device. These types of devices, including the predicate devices, have been on the market for many years with a proven safety and efficacy for the use of the device. The non-clinical testing detailed in this submission supports the substantial equivalence of the device.
#### 11. Biocompatibility and Sterilization/Shelf-life
The change does not impact the biological safety of the device. The XP200 device biocompatibility was tested according to the ISO 10993-1; ISO 10993-5; ISO 10993-10.
There are no changes in the product life and shelf life compared to that of XP200 device. The device shelf life is 5 years. The product life is 5 years.
The device is not intended to be sterile and is not intended to be sterilized by the user.
#### 12. Statement of Substantial Equivalence
It has been shown in this 510(k) submission that the differences between the XP300 and the predicate device do not raise any questions regarding its safety and effectiveness. The XP300, as designed and manufactured, is determined to be substantially equivalent to the referenced predicate device; XP200.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.