← Product Code [NIE](/productcode/NIE) · K143102

# Multi-Lumen Acute Hemodialysis Catheter for High Volume Infusions (K143102)

_Arrow International, Inc. (Subsidiary of Teleflex, Inc.) · NIE · Jul 24, 2015 · Gastroenterology, Urology · SEKD_

**Canonical URL:** https://fda-staging.innolitics.com/device/K143102

## Device Facts

- **Applicant:** Arrow International, Inc. (Subsidiary of Teleflex, Inc.)
- **Product Code:** [NIE](/productcode/NIE.md)
- **Decision Date:** Jul 24, 2015
- **Decision:** SEKD
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 876.5540
- **Device Class:** Class 2
- **Review Panel:** Gastroenterology, Urology
- **Attributes:** Therapeutic

## Indications for Use

The device is indicated for use in attaining short-term (less than 30 days) vascular access for hemodialysis, apheresis, rapid fluid administration, intravenous therapy, blood sampling, pressure injection of contrast media, and central venous pressure monitoring. The catheter may also be used in a variety of renal replacement therapies, such as hemofiltration and hemoperfusion. The catheter may be inserted into the jugular, or femoral veins. The maximum pressure injection flow rate is 6 mL / sec.

## Device Story

Multi-lumen acute hemodialysis catheter (3-L AHDC) provides short-term central venous access in healthcare facilities. Device consists of non-tapered, radiopaque polyurethane body with soft, contoured Blue Flex Tip. Used for hemodialysis, apheresis, fluid/drug administration, blood sampling, CVP monitoring, and pressure injection of contrast media (max 6 mL/sec). Inserted by clinicians into jugular, subclavian, or femoral veins. Facilitates renal replacement therapies like hemofiltration and hemoperfusion. Packaged as sterile kit with insertion components. Output is vascular access for clinical intervention; enables rapid fluid/contrast delivery and continuous blood processing. Benefits patient by providing reliable, high-volume access for acute renal and infusion needs.

## Clinical Evidence

No clinical data. Substantial equivalence demonstrated via bench testing only, including performance and material testing in accordance with ISO 10555-1 and ISO 10993-1 standards.

## Technological Characteristics

Radiopaque polyurethane extruded catheter body; Blue Flex Tip. 12 Fr. triple-lumen configuration; 16-25 cm usable lengths. Non-implanted. Sterile, single-use. Tested per ISO 10555-1 (catheter performance) and ISO 10993-1 (biocompatibility).

## Regulatory Identification

A blood access device and accessories is a device intended to provide access to a patient's blood for hemodialysis or other chronic uses. When used in hemodialysis, it is part of an artificial kidney system for the treatment of patients with renal failure or toxemic conditions and provides access to a patient's blood for hemodialysis. The device includes implanted blood access devices, nonimplanted blood access devices, and accessories for both the implanted and nonimplanted blood access devices.(1) The implanted blood access device is a prescription device and consists of various flexible or rigid tubes, such as catheters, or cannulae, which are surgically implanted in appropriate blood vessels, may come through the skin, and are intended to remain in the body for 30 days or more. This generic type of device includes various catheters, shunts, and connectors specifically designed to provide access to blood. Examples include single and double lumen catheters with cuff(s), fully subcutaneous port-catheter systems, and A-V shunt cannulae (with vessel tips). The implanted blood access device may also contain coatings or additives which may provide additional functionality to the device. (2) The nonimplanted blood access device consists of various flexible or rigid tubes, such as catheters, cannulae or hollow needles, which are inserted into appropriate blood vessels or a vascular graft prosthesis (§§ 870.3450 and 870.3460), and are intended to remain in the body for less than 30 days. This generic type of device includes fistula needles, the single needle dialysis set (coaxial flow needle), and the single needle dialysis set (alternating flow needle). (3) Accessories common to either type include the shunt adaptor, cannula clamp, shunt connector, shunt stabilizer, vessel dilator, disconnect forceps, shunt guard, crimp plier, tube plier, crimp ring, joint ring, fistula adaptor, and declotting tray (including contents).

## Special Controls

*Classification.* (1) Class II (special controls) for the implanted blood access device. The special controls for this device are:(i) Components of the device that come into human contact must be demonstrated to be biocompatible. Material names and specific designation numbers must be provided.
(ii) Performance data must demonstrate that the device performs as intended under anticipated conditions of use. The following performance characteristics must be tested:
(A) Pressure versus flow rates for both arterial and venous lumens, from the minimum flow rate to the maximum flow rate in 100 milliliter per minute increments, must be established. The fluid and its viscosity used during testing must be stated.
(B) Recirculation rates for both forward and reverse flow configurations must be established, along with the protocol used to perform the assay, which must be provided.
(C) Priming volumes must be established.
(D) Tensile testing of joints and materials must be conducted. The minimum acceptance criteria must be adequate for its intended use.
(E) Air leakage testing and liquid leakage testing must be conducted.
(F) Testing of the repeated clamping of the extensions of the catheter that simulates use over the life of the device must be conducted, and retested for leakage.
(G) Mechanical hemolysis testing must be conducted for new or altered device designs that affect the blood flow pattern.
(H) Chemical tolerance of the device to repeated exposure to commonly used disinfection agents must be established.
(iii) Performance data must demonstrate the sterility of the device.
(iv) Performance data must support the shelf life of the device for continued sterility, package integrity, and functionality over the requested shelf life that must include tensile, repeated clamping, and leakage testing.
(v) Labeling of implanted blood access devices for hemodialysis must include the following:
(A) Labeling must provide arterial and venous pressure versus flow rates, either in tabular or graphical format. The fluid and its viscosity used during testing must be stated.
(B) Labeling must specify the forward and reverse recirculation rates.
(C) Labeling must provide the arterial and venous priming volumes.
(D) Labeling must specify an expiration date.
(E) Labeling must identify any disinfecting agents that cannot be used to clean any components of the device.
(F) Any contraindicated disinfecting agents due to material incompatibility must be identified by printing a warning on the catheter. Alternatively, contraindicated disinfecting agents must be identified by a label affixed to the patient's medical record and with written instructions provided directly to the patient.
(G) Labeling must include a patient implant card.
(H) The labeling must contain comprehensive instructions for the following:
(
*1* ) Preparation and insertion of the device, including recommended site of insertion, method of insertion, and a reference on the proper location for tip placement;(
*2* ) Proper care and maintenance of the device and device exit site;(
*3* ) Removal of the device;(
*4* ) Anticoagulation;(
*5* ) Management of obstruction and thrombus formation; and(
*6* ) Qualifications for clinical providers performing the insertion, maintenance, and removal of the devices.(vi) In addition to Special Controls in paragraphs (b)(1)(i) through (v) of this section, implanted blood access devices that include subcutaneous ports must include the following:
(A) Labeling must include the recommended type of needle for access as well as detailed instructions for care and maintenance of the port, subcutaneous pocket, and skin overlying the port.
(B) Performance testing must include results on repeated use of the ports that simulates use over the intended life of the device.
(C) Clinical performance testing must demonstrate safe and effective use and capture any adverse events observed during clinical use.
(vii) In addition to Special Controls in paragraphs (b)(1)(i) through (v) of this section, implanted blood access devices with coatings or additives must include the following:
(A) A description and material characterization of the coating or additive material, the purpose of the coating or additive, duration of effectiveness, and how and where the coating is applied.
(B) An identification in the labeling of any coatings or additives and a summary of the results of performance testing for any coating or material with special characteristics, such as decreased thrombus formation or antimicrobial properties.
(C) A Warning Statement in the labeling for potential allergic reactions including anaphylaxis if the coating or additive contains known allergens.
(D) Performance data must demonstrate efficacy of the coating or additive and the duration of effectiveness.
(viii) The following must be included for A-V shunt cannulae (with vessel tips):
(A) The device must comply with Special Controls in paragraphs (b)(1)(i) through (v) of this section with the exception of paragraphs (b)(1)(ii)(B), (b)(1)(ii)(C), (b)(1)(v)(B), and (b)(1)(v)(C), which do not apply.
(B) Labeling must include Warning Statements to address the potential for vascular access steal syndrome, arterial stenosis, arterial thrombosis, and hemorrhage including exsanguination given that the device accesses the arterial circulation.
(C) Clinical performance testing must demonstrate safe and effective use and capture any adverse events observed during clinical use.
(2) Class II (performance standards) for the nonimplanted blood access device.
(3) Class II (performance standards) for accessories for both the implanted and the nonimplanted blood access devices not listed in paragraph (b)(4) of this section.
(4) Class I for the cannula clamp, disconnect forceps, crimp plier, tube plier, crimp ring, and joint ring, accessories for both the implanted and nonimplanted blood access device. The devices subject to this paragraph (b)(4) are exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.

## Predicate Devices

- 12 Fr. and 14 Fr. Two-Lumen Hemodialysis Catheterization Kit with Blue FlexTip® ARROWg+ard Blue® Antimicrobial Catheter for High Volume Infusions ([K993933](/device/K993933.md))
- Bard Power-Trialysis™ Short Term Dialysis Catheter ([K083675](/device/K083675.md))
- Medcomp Tri-Flow® Triple Lumen Short Term Hemodialysis Catheter ([K973561](/device/K973561.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the seal of the Department of Health & Human Services - USA. The seal features a stylized eagle with three heads facing right, representing the department's mission to protect the health of all Americans and provide essential human services. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the eagle.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

July 24, 2015

Arrow International, Inc. (Subsidiary of Teleflex Inc.) % Karl J Nittinger Sr. Regulatory Affairs Specialist, Interventional Access 2400 Bernville Road P.O. Box 12888 Reading, PA 19605

Re: K143102

> Trade/Device Name: Multi-lumen Acute Hemodialysis Catheter for High Volume Infusions Regulation Number: 21 CFR 876.5540 Regulation Name: Blood access device and accessories Regulatory Class: II Product Code: NIE Dated: June 10, 2015 Received: June 15, 2015

Dear Karl J Nittinger,

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. However, you are responsible to determine that the medical devices you use as components in the kit have either been determined as substantially equivalent under the premarket notification process (Section 510(k) of the act), or were legally on the market prior to May 28, 1976, the enactment date of the Medical Device Amendments. Please note: If you purchase your device components in bulk (i.e., unfinished) and further process (e.g., sterilize) you must submit a new 510(k) before including these components in your kit/tray. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be 

{1}------------------------------------------------

found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act): 21 CFR 1000-1050.

In addition, we have determined that your device kit contains antiseptic hand gels, chlorapreporange, lidocaine, and providone iodine swabsticks, which are subject to regulation as drugs.

Our substantially equivalent determination does not apply to the drug components of your device. We recommend you first contact the Center for Drug Evaluation and Research before marketing your device with the drug components. For information on applicable Agency requirements for marketing these drugs, we suggest you contact:

> Director, Division of Drug Labeling Compliance Center for Drug Evaluation and Research Food and Drug Administration 5600 Fishers Lane Rockville, Maryland 20857 (301) 594-0101

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and Part 809), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note

the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 

{2}------------------------------------------------

Page 3 - Karl J Nittinger

796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Herbert P. Lerner -S

for Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

# Indications for Use

510(k) Number (if known) K143102

#### Device Name

Multi-Lumen Acute Hemodialysis Catheter for High Volume Infusions

Indications for Use (Describe)

The device is indicated for use in attaining short-term (less than 30 days) vascular access for hemodialysis, apheresis, rapid fluid administration, intravenous therapy, blood sampling, pressure injection of contrast media, and central venous pressure monitoring. The catheter may also be used in a variety of renal replacement therapies, such as hemofiltration and hemoperfusion. The catheter may be inserted into the jugular, or femoral veins. The maximum pressure injection flow rate is 6 mL / sec.

Type of Use (Select one or both, as applicable)

| <div> <span> <b>       ☑      </b>      Prescription Use (Part 21 CFR 801 Subpart D)     </span> </div> |
|---------------------------------------------------------------------------------------------------------|
| <div> <span>      ☐ Over-The-Counter Use (21 CFR 801 Subpart C)     </span> </div>                      |

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the logo for Arrow International. The word "ARROW" is in large, white, sans-serif font on the top line. Below that, the word "INTERNATIONAL" is in a smaller, white, sans-serif font. The background is a solid blue color.

#### 510(k) SUMMARY Multi-Lumen Acute Hemodialysis Catheter for High Volume Infusions

## 1. Submitter Information

| Name:             | Arrow International, Inc (subsidiary of Teleflex Inc.) |
|-------------------|--------------------------------------------------------|
| Address:          | 2400 Bernville Road<br>Reading, PA 19605-9607          |
| Contact Person:   | Karl J. Nittinger<br>Sr. Regulatory Affairs Specialist |
| Telephone Number: | (610) 378-0131, ext. 603384                            |
| Email:            | karl.nittinger@teleflex.com                            |
| Date Prepared:    | July 23, 2015                                          |

### 2. Device Name

| Device Trade Name: | Multi-Lumen Acute Hemodialysis Catheter for High Volume<br>Infusions |
|--------------------|----------------------------------------------------------------------|
| Common Name:       | Acute Hemodialysis Catheter                                          |

Classification Name: Catheter, Hemodialysis, Triple-Lumen, Non-Implanted

### 3. Predicate Devices

Predicate 1: 12 Fr. and 14 Fr. Two-Lumen Hemodialysis Catheterization Kit with Blue FlexTip® ARROWg+ard Blue® Antimicrobial Catheter for High Volume Infusions (K993933)

Predicate 2: Bard Power-Trialysis™ Short Term Dialysis Catheter (K083675)

Predicate 3: Medcomp Tri-Flow® Triple Lumen Short Term Hemodialysis Catheter (K973561)

### 4. Device Description

The Multi-Lumen Acute Hemodialysis Catheter for High Volume (hereafter referred to as the 3-L AHDC) is a short-term, single use catheter designed to provide access to the central venous system in a healthcare facility environment. It consists of a non-tapered, radiopaque polyurethane extruded catheter body with a softer, contoured Blue Flex Tip. The catheter is available in a 12 Fr. triple-lumen configuration with usable lengths of 16-25 cm. The catheters can be used for the injection of contrast media. The maximum recommended pressure injection flow rate is 6 mL/sec.

The catheters are packaged sterile with various components to facilitate insertion.

{5}------------------------------------------------

# 5. Indications for Use

The device is indicated for use in attaining short-term (less than 30 days) vascular access for hemodialysis, apheresis, rapid fluid administration, intravenous therapy, blood sampling, pressure injection of contrast media, and central venous pressure monitoring. The catheter may also be used in a variety of renal replacement therapies, such as hemofiltration and hemoperfusion. The catheter may be inserted into the jugular, subclavian, or femoral veins. The maximum pressure injection flow rate is 6 mL / sec.

## 6. Technological Characteristics and Substantial Equivalence

The 3-L AHDC is substantially equivalent to the Arrow Acute Hemodialysis AGB Catheters (K993933) and the Bard Power-Trialysis Short Term Dialysis Catheter (K083675) in terms of overall design, manufacturing process, functional performance, materials of construction, and indications for use. The proposed catheter is also substantially equivalent to the Medcomp Tri-Flow Triple Lumen Short Term Hemodialysis Catheter (K973561) in terms of overall functional performance.

# 7. Nonclinical Testing

Bench testing was performed to demonstrate substantial equivalence between the predicate and proposed devices. Performance and material testing was completed on the proposed devices in accordance with ISO 10555-1. ISO 10993-1, and FDA Guidance on Premarket Notification [510(k)] Submission for Short-Term and Long-Term Intravascular Catheters.

## 8. Clinical Data

Clinical performance data was not used to determine substantial equivalence.

## 9. Conclusions

The results of the testing performed have demonstrated that the proposed Multi-Lumen Acute Hemodialysis Catheter for High Volume Infusions is safe and performs as intended and therefore is considered substantially equivalent to the cited predicate devices.

---

**Source:** [https://fda-staging.innolitics.com/device/K143102](https://fda-staging.innolitics.com/device/K143102)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
