The indications for use of the Helio A7 and Move Manual Wheelchairs is to provide mobility to persons limited to a sitting position.
Device Story
Helio A7 and Move are manually operated, user-propelled mechanical wheelchairs; intended to provide mobility to persons limited to a sitting position. Device consists of monocoque aluminum frame, folding cross-brace, rear wheels with hand rims, and front swivel casters. User controls motion, speed, and direction by manual propulsion of hand rims. Used in various settings (home, clinic, community) by the patient. No complex scientific concepts or automated functions; performance relies entirely on user operation. Differences from predicate include vertical axle plate mounting and integrated caster housing design to improve handling and reduce wear. Benefits include improved rigidity, reduced weight, and enhanced propulsion efficiency.
Clinical Evidence
Bench testing only. Compliance with ISO 7176 standards (1, 3, 5, 7, 8, 15) for static stability, brake effectiveness, dimensions/maneuvering, seating/wheel dimensions, and static/impact/fatigue strength. Upholstery meets Cal 117 flammability requirements.
Technological Characteristics
Materials: Aluminum (7000 series for A7, 6000 series for Move), steel fasteners, polyurethane tires, fabric-covered foam upholstery. Design: Monocoque welded frame, folding cross-brace, adjustable axle plates, swivel casters. Manual operation; no energy source. Connectivity: None. Sterilization: N/A.
Indications for Use
Indicated for persons limited to a sitting position requiring a manual wheelchair for mobility.
Regulatory Classification
Identification
A mechanical wheelchair is a manually operated device with wheels that is intended for medical purposes to provide mobility to persons restricted to a sitting position.
Predicate Devices
Sunrise Medical Quickie 2 Manual Wheelchair (K123975)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the logo is an abstract symbol that resembles a stylized caduceus or a series of interconnected human profiles.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
May 1, 2015
Motion Composites c/o Edward A. Kroll Spectre Solutions, Inc. 5905 Fawn Lane Cleveland, OH 44141
Re: K143101
> Trade/Device Name: Helio A7 Manual Wheelchair and Move Manual Wheelchair Regulation Number: 21 CFR 890.3850 Regulation Name: Mechanical Wheelchair Regulatory Class: Class I Product Code: IOR Dated: March 23, 2015 Received: March 24, 2015
Dear Mr. Kroll,
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in
{1}------------------------------------------------
the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# Felipe Aquel -S
Carlos L. Peña, PhD, MS for Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K143101
#### Device Name
Helio A7 and Move Manual Wheelchairs
Indications for Use (Describe)
The indications for use of the Helio A7 and Move Manual Wheelchairs is to provide mobility to persons limited to a sitting position.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
|X | Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for Motion Composites. The logo consists of a blue circle with a stylized blue wing inside of it, followed by the text "MOTION COMPOSITES" in blue. Below the logo is the text "510(k) Premarket Notification".
### Motion Composites Helio A7 and Move Manual Wheelchairs 510(k) Summary (Modified 5/1/2015)
#### l. SUBMITTER
Motion Composites 519 J - Oswald Forest Suite 101 Saint-Roch-del'Achigan Quebec, Canada JOK 3H0 Phone: (450) 588-6555 Fax: (450) 588-0200 Email: info@motioncomposites.com Web Site: www.motioncomposites.com
#### II. DEVICE
Name of Device: Helio A7 and Move Manual Wheelchair Common or Usual Name: Manual Wheelchair Classification Name: Wheelchair, Mechanical (21 CFR 890.3850) Regulatory Class: I Product Code: IOR
#### III. PREDICATE DEVICE
Sunrise Medical Quickie 2 Manual Wheelchair K123975
#### IV. DEVICE DESCRIPTION
The Motion Composites Helio A7 and Move Manual Wheelchairs are manually operated, user propelled, manual, mechanical wheelchairs. Their intended function and use is to provide mobility to persons limited to a sitting position.
Both are traditional folding cross-brace wheelchairs. They are made of an aluminum frame which utilizes one cross brace system that, when opened, nestles inside of the frame onto 4 hooks to create a box like rigid assembly. Upon the outside of this framework, and to the rear, are assembled two aluminum axle plates. Wheels of varying size and type are connected to the stainless steel axle receivers via stainless steel axles.
On the front end of the frame are assembled two aluminum caster housings. Caster forks are mounted to these housings via steel axles. A variety of caster wheels and tires are then connected to the fork based on user preference.
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for Motion Composites, which includes a blue circle with a stylized feather inside, followed by the company name in blue text. Below the logo, the text "510(k) Premarket Notification" is present. The text is in a smaller font size and is also in blue.
### Device Function
Device function is dependent solely upon the wheelchair user. It does not function on its' own in any manner. The wheelchair user controls motion, speed and direction by propelling themselves using the hand rims located on the rear wheels.
### Scientific Concepts
There are no complex scientific concepts related to the Helio A7 and Move manual wheelchair. They are simple, basic, manually operated mobility devices.
### Significant Physical and Performance Characteristics:
### Design:
- . The Helio A7 and Move utilize a monocoque aluminum frame that is entirely welded and heat threated for greatly improved rigidity over reqular interconnected side frames.
- A completely symmetrical aluminum cross-brace connects the left and right mono side frames of the wheelchair together and serves as the mechanism that allows the frame to fold.
- A set of adjustable axle plates are attached vertically to each side frame and provide the ● wheelchair with ability to move the rear wheel position forward and backward in order to give the user an adjustment as to the center of gravity of the wheelchair. This rear axle vertical position limits torsion and flexion for improved propulsion.
### Materials:
Materials used are:
- . Aluminum frame, support members, wheels and components
- . Steel fasteners and components
- Polyurethane tires
- Fabric covered foam upholstery
### Physical Properties:
The Helio A7 and Move consist primarily of an aluminum frame assembly, a back rest frame, seat and back rest upholstery, large rear wheels with hand rims for self-propelling the chairs and front swivel pivoting casters for turning.
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for Motion Composites, which includes a blue circle with a stylized feather inside, followed by the company name in blue text. Below the logo, there is a horizontal line, and below that, the text "510(k) Premarket Notification" is written. The text is smaller than the logo and is also in blue.
#### INDICATIONS FOR USE V.
The indications for use of the A7 and Move Manual Wheelchairs are to provide mobility to persons limited to a sitting position. This identical to the predicate device.
## VI. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE
The technology and principle of operation for the Helio A7 and Move and the Quickie 2 are identical. They all consist of an aluminum frame with a seat and large rear wheels with hand rims for propelling the device. Smaller, pivoting type casters are mounted on the front of the chairs for steering and turning.
Device function is dependent solely upon the wheelchair user. They do not function on their own in any manner. The wheelchair user controls motion, speed and direction by propelling themselves using the hand rims located on the rear wheels.
If both rear wheels are propelled at the same time, the chair will move forward in a straight direction. If only the left rear wheel is propelled the chair will turn to the right. If only the right rear wheel is propelled the chair will turn to the left.
#### Discussion of Similarities and Differences:
The Helio A7 and Move are substantially equivalent to the predicate Quickie 2 wheelchair (K123975) in technology, function, performance and materials. They have the same indications for use which is to provide mobility for persons restricted in a sitting position.
The frame width, depth and the back cane heights vary slightly between the devices, but since these are only used to better fit the device to the user's needs, these slight differences do not cause any concerns for the safety and effectiveness of the weight limit is the same for all devices and is clearly stated in the proposed device labeling.
With regard to accessories and add-ons, all devices offer the same types of armrests, backrests, hangers and footplates. These accessories are made of the exact same materials for all devices and thus do not raise any questions for the safety and effectiveness of the Helio A7 and Move wheelchairs. The same can be said of back types and extension tubes.
Axle plates for all devices are made out of the same material, however, the actual mountinq systems differ slightly. Where the Quickie 2 axle plate is mounted horizontally on the frame, the Helio A7 and Move are mounted vertically.
This mechanical change permits better handling and propulsion the inwards flexion of the wheel during wheel pushes. It does not raise any concerns for the safety and effectiveness of the device and actually augments the effectiveness of the Helio A7 and Move when compared to
{6}------------------------------------------------
Image /page/6/Picture/1 description: The image contains the logo for Motion Composites. The logo consists of a blue circle with a stylized blue feather inside of it. To the right of the logo is the text "MOTION COMPOSITES" in blue.
510(k) Premarket Notification
the predicate device. This mounting system is a patented technology exclusive to Motion Composites.
The same can be said for the caster housing that holds the front wheel into place. While the caster housing of the predicate device and those of the Helio A7 and Move are made all made of aluminum, the mounting system of the Helio A7 casters housing is patented and thus slightly different from the one found on the Quickie 2 wheelchair.
This mounting system embeds the caster housing directly into the frame, reducing the effects of wear and tear by minimizing the amount of parts used to fasten the caster to the frame, thus providing a better overall experience to the wheelchair user.
This change as well does not raise any concerns as far as the safety of the overall product against the predicate device, since all casters offer the same effectiveness as per the Helio A7 successfully obtaining its ISO certification. As for the Move caster housing, it uses the same exact design and function as the predicate device.
The seat and back upholstery offering of the Helio is equivalent to the one found on the predicate device. It is made of high-quality durable sailcloth, which consists of 1000 denier fill and 250 denier wrap.
The rear wheels and hand rims of the predicate device and the Helio A7 Move wheelchair are offered in a wide range of different options. They are materials and are widely used in the wheelchair industry. The specific type, material and size of the wheels are chosen depending on the user's needs. These differences do not raise any concerns for the safety and effectiveness of the Helio A7 and Move wheelchairs. We also note the wheel locks are the same on all devices.
Based on the above discussion, Motion Composites believes that the Helio A7 and Move wheelchairs are substantially equivalent to the Quickie 2 wheelchair (K123975). While there are some differences between the Helio A7 and Move and its predicate, these differences are minor and do not alter the intended function and use of the device, nor do they raise any new questions pertaining to safety or effectiveness.
#### VII. PERFORMANCE DATA
The Helio A7 and Move Manual Wheelchairs have been tested to the following standards;
- ISO 7176-1:1999 Determination of Static Stability
- ISO 7176-3:2012 Determination of Effectiveness of Brakes
- ISO 7176-5:2008 Determination of Overall Dimensions, Mass and Maneuvering
- Space
- ISO 7176-7:1998 Determination of Seating and Wheel Dimensions
- ISO-7176-8:1998 Requirements and Test Method for Static Impact and Fatique Strength
{7}------------------------------------------------
510(k) Premarket Notification
- ISO 7176-15 Requirements for Information Disclosure, Documentation and Labeling
The upholstery used meets the requirements of Cal 117 for flammability. These are identical to the materials used in the Motion Composites Helio manual wheelchair. The Helio manual wheelchair was cleared by FDA December 19, 2012 under 510(k) Accession Number K120628.
The performance data for the Quickie 2 is not published. However, on page 6 of the Quickie 2 Owner's Manual, the following statement is made;
"Based on ANSI/RESNA testing, Sunrise Medical recommends the use of a caster wheel with a minimum diameter of 5″ if the wheelchair will be overcoming obstacles of up to ½″ on a reqular basis.
This statement indicates that the Sunrise Quickie 2 has been tested to at least some of the ANS//RESNA standards. These standards are comparable to the ISO 7176 standards which were used to test the Motion Composites A7 and Move wheelchairs. (Note that a copy of the Quickie 2 Owner's Manual is provided in Volume 10 of this submission).
Further, manual wheelchairs are subject to the 1995 FDA guidance document entitled: "Guidance Document for the Preparation of Premarket Notification [(510(k)] Applications Mechanical and Powered Wheelchairs, and Motorized Three-Wheeled Vehicles." This quidance includes various safety and performance requirements that the Quickie 2 would have to have met or addressed prior to being cleared by FDA. Meeting these requirements also serves as a basis for substantial equivalence of the A7 and Move to its' predicate.
#### VIII. CONCLUSIONS:
Performance testing and compliance with FDA quidance for manual wheelchairs supports the safety of the Helio A7 and Move manual wheelchairs. Since the predicate device was cleared based in part on these data, conformance with the same requirements demonstrate that the Helio A7 and Move will perform as intended in the specified use conditions.
Further, technology, principle of operation and indications for use between Helio A7 and Move and the Quickie 2, are identical. This demonstrates that they perform comparably to the predicate device that is currently marketed for the same intended use. Therefore, the Helio A7 and Move manual wheelchairs are substantially equivalent to the Quickie 2 device.
{8}------------------------------------------------
Image /page/8/Picture/0 description: The image shows the logo for Motion Composites, a company that specializes in the design and manufacturing of high-performance wheelchairs. The logo features a stylized feather design in blue, along with the company name in a sans-serif font. Below the logo, the text "510(k) Premarket Notification" is displayed, indicating that the company has submitted a premarket notification to the FDA for one of its products.
Helio A7 and Move Manual Wheelchairs
# Helio A7 and Move Predicate Device Comparison
| Feature/Specification | Quickie Wheelchair (K123975) | Helio A7 Wheelchair | Move Wheelchair |
|-------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended Use | The Quickie 2 Wheelchair is a<br>manually operated device intended to<br>be used as a means of mobility for<br>persons restrict to a sitting position. | The Helio A7 wheelchair is a manually<br>operated device intended to be used<br>as a means of mobility for persons<br>restrict to a sitting position. | The Move wheelchair is a manually<br>operated device intended to be used<br>as a means of mobility for persons<br>restrict to a sitting position. |
| Primary Materials | Aluminum 6000 series | Aluminum 7000 series | Aluminum 6000 series |
| Folding Method | Collapsible Cross-Brace | Collapsible Cross-brace | Collapsible Cross-brace |
| Frame Width | 11" to 22" | 14" to 22" | 14, 16, 18, 20, 22" |
| Overall width | 20,5" to 28,5" | 21 1/2" to 29 ½" | 21 1/2" to 29 ½" |
| Seat Depth | 10" to 20" | 15" to 21" | 16",18",20 |
| Back Heights | 8,5" to 19" | 8,5" to 19" | 16", 20" |
| Weight Limit : | 250lbs | 250 lbs. | 250 lbs. |
| Seat height : | 16.75" to 22.75". | 13" to 21 1/4" | 13" to 20 3/4" |
| Chair Weight (without<br>footrests) | 27 lbs. | 16.5 lbs. (without footrests) | 19 lbs. (without footrests) |
| Warranty | Lifetime on frame | Frame 5 years, Components 1 year | Frame 5 years, Components 1 year |
| Armrests | Flip back Height adjustable<br>T-Shaped Armrests - Height adjustable<br>Tubular Swing-away | -Flip back armrest, removable, height<br>adjustable (by user, ne tool required)<br>-T-Shaped armrest, removable, height<br>adjustable (by user, no tool required)<br>-Tubular Swing-away | -Flip back armrest, removable, height<br>adjustable (by user, ne tool required)<br>-T-Shaped armrest, removable, height<br>adjustable (by user, no tool required)<br>-Tubular Swing-away |
| Front end types | Swing-Away<br>Non swing-away | Swing-Away | Swing-Away |
| Back Type | Standard<br>Angle Adjustable<br>Depth Adjustable Standard | -Straight<br>-8 degree bend back<br>-Adjustable angle- straight or 8 degree<br>bend | -8 degree bend back with integrated<br>push handle<br>-Adjustable angle 8 degree bend with<br>integrated push handle |
| Footrest Hangers | 60,70,90 degrees, elevating legrests | Flip in, flip out; 60°, 70° , 90°,<br>elevating legrest | Flip out; 60°, 70° , elevating legrest |
| Footplates | Composites, Foam, Aluminum angle<br>adjustable | -Standard non adjustable<br>-Angle Adjustable Oversized | -Standard non adjustable<br>-Angle Adjustable Oversized |
| Back Upholstery | Low, Medium, Tall, Adjustable, | Standard, Adjustable | Standard, Adjustable |
| Axle Plates | Standard, Curved, Amputee, Offset<br>Standard | Standard, Amputee | Standard, Amputee |
| Wheel sizes | 20,22,25,26 " | 20 ,22,24, 25, 26 " | 20 ,22,24, 25, 26 " |
| Wheel types | Spoke, Composite Mag, Spinergy, One<br>arm drive | -Plastic Mags<br>-Spoke, Newton One<br>-Spoke, Newton Gravity<br>-Spinergy LX<br>-Spinergy Spox | -Plastic Mags<br>-Spoke, Newton One<br>-Spoke, Newton Gravity<br>-Spinergy LX<br>-Spinergy Spox |
| Tire types | Pneumatic, Pneumatic w/ airless<br>insert<br>Polyurethane, Low Profile Full<br>profile<br>Iron Cap | -Pneumatic regular and high pressure<br>-Full and low profile polyurethane<br>-Hard Urethane | -Pneumatic regular and high pressure<br>-Full and low profile polyurethane<br>-Hard Urethane |
| Handrims | Aluminum, Plastic Coated,<br>Projections | -Aluminum Anodized<br>-Plastic Coated (regular or high friction)<br>-Spinergy Flexrim<br>-Natural Fit handrim<br>-Surge Handrim | -Aluminum Anodized<br>-Plastic Coated (regular or high friction)<br>-Spinergy Flexrim<br>-Natural Fit handrim<br>-Surge Handrim |
| Feature/Specification | Quickie Wheelchair (K123975) | Helio A7 Wheelchair | Move Wheelchair |
| Caster types | Polyurethane, semi-pneumatic, Soft | Composites wheels, Pneumatic, Soft | Composites wheels, Pneumatic, Soft |
| | Roll, Pneumatic | Roll | Roll |
| Fork Sizes | 3″,4″,5″,6″,7″ | 3″,4″,5″,7″ | 3″,4″,5″,7″ |
| Fork Stem Sizes | Std,+3/4",+1 ½" | std, +1 , +2 | std, +1 , +2 |
| Caster Options | Multi-position fork caster, pin locks | Multi-position fork caster, pin locks | Multi-position fork caster, pin locks |
| Wheel Locks | Push to lock, pull to lock, scissor lock | Push to lock, pull to lock, scissor lock | Push to lock, pull to lock, scissor lock |
| Anti-tip tubes | Yes | Yes | Yes |
| Target population | Restricted to a sitting position | Restricted to a sitting position | Restricted to a sitting position |
| Standards | Unknown | ISO 7176-(1, 3, 5, 7, 8 & 15) | ISO 7176 - (1, 3, 5, 7, 8 & 15) |
{9}------------------------------------------------
Image /page/9/Picture/0 description: The image shows the logo for Motion Composites, which includes a blue circle with a stylized feather inside. To the right of the logo is the company name, "MOTION COMPOSITES," also in blue. Below the logo and company name, the text "510(k) Premarket Notification" is displayed, indicating the document's purpose.
Helio A7 and Move Manual Wheelchairs
## Helio A7 and Move Predicate Device Comparison
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.