K143070 · Jiangsu Yile Medical Article Co., Ltd. · MEG · Jul 7, 2015 · General Hospital
Device Facts
Record ID
K143070
Device Name
Disposable Safety Self-destructive Syringes, 5 ML
Applicant
Jiangsu Yile Medical Article Co., Ltd.
Product Code
MEG · General Hospital
Decision Date
Jul 7, 2015
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5860
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Disposable Safety Self-destructive Syringe is used to inject fluids into, or withdraw fluids from the body. In addition, it is designed to aid in the prevention of needle stick injuries.
Device Story
Disposable safety syringe for fluid injection or withdrawal; features manual retraction mechanism to prevent needle stick injuries. After use, clinician fully depresses plunger to engage luer assembly; pulling back plunger retracts needle into barrel; lateral pressure on plunger causes controlled fracture, permanently disabling device. Components include polypropylene barrel/plunger/protector, natural rubber piston/O-ring, and 304 stainless steel needle. Used in clinical settings by healthcare professionals. Output is physical fluid delivery or withdrawal; safety feature benefits patient and clinician by reducing sharps injury risk.
Clinical Evidence
Bench testing only. Performance testing included biocompatibility (ISO 10993-5, -10, -11; ASTM F756-13), mechanical performance (ISO 7886-1, ISO 7864, ISO 9626, ISO 23908, ISO 594-1/2), and USP 788 particle count. Simulated clinical use performed per FDA guidance for sharps injury prevention features. Package and shelf-life validation included real-time/accelerated aging, sterility, vacuum leak, dye penetration, and seal strength tests (ASTM F1980, D3078, F1929, F88). All results satisfactory.
Technological Characteristics
Materials: Polypropylene (barrel, plunger, protector), natural rubber (piston, O-ring), 304 stainless steel (needle). Lubricant: Polydimethylsiloxane oil. Dimensions: 5ML capacity, 21G needle. Connectivity: None. Sterilization: EO. Standards: ISO 7886-1, ISO 7864, ISO 9626, ISO 23908, ISO 594-1/2, ISO 10993, ISO 11607, ISO 11737, ASTM F756, F1980, D3078, F1929, F88.
Indications for Use
Indicated for use by healthcare professionals to inject or withdraw fluids from the body while providing needle stick injury prevention.
Regulatory Classification
Identification
A piston syringe is a device intended for medical purposes that consists of a calibrated hollow barrel and a movable plunger. At one end of the barrel there is a male connector (nozzle) for fitting the female connector (hub) of a hypodermic single lumen needle. The device is used to inject fluids into, or withdraw fluids from, the body.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized image of three faces in profile, stacked on top of each other. The faces are oriented to the right and are depicted in a simple, line-art style. The logo is surrounded by the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" in a circular arrangement.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
July 7, 2015
Jiangsu Yile Medical Article Co., Ltd C/O Mr. Charles Mack Principal Engineer International Regulatory Consulting 12226 Washington Lane Parker, AZ 85344
Re: K143070
Trade/Device Name: Disposable Safety Self-destructive Syringes, 5ML Regulation Number: 21 CFR 880.5860 Regulation Name: Piston syringe Regulatory Class: II Product Code: MEG, FMI Dated: May 27, 2015 Received: June 04, 2015
Dear Mr. Mack:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Tejashri Purohit-Sheth, M.D.
Tejashri Purohit-Sheth, M.D. Clinical Deputy Director DAGRID/ODE/CDRH FOR
Erin I. Keith, M.S. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known)
K143070
Device Name
Disposable Safety Self-destructive Syringes, 5ML
Indications for Use (Describe)
The Disposable Safety Self-destructive is used to inject fluids into, or withdraw fluids from the body. In addition, it is designed to aid in the prevention of needle stick injuries.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------|
| <span style="font-family: Arial, sans-serif;"><input checked="true" type="checkbox"/></span> Prescription Use (Part 21 CFR 801 Subpart D) | <span style="font-family: Arial, sans-serif;"><input type="checkbox"/></span> Over-The-Counter Use (21 CFR 801 Subpart C) |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## K143070
# 510(k) Summary
This summary of 510(k) information is being submitted in accordance with requirements with requirements of CFR Part 807.92.
Date: June 24, 2015
- 1. Company and Correspondent making the submission:
Name - Jiangsu Yile Medical Article Co., Ltd Address - Sanhekou Development Zone, Changzhou City Jiangsu Province China 213115 Tel: 86-519-88675299 Fax: 86-519-88679018 Contact - Mr. Feng Xingvuan General Manager Email - charliemack@irc-us.com
Primary contact and correspondent: Name: Charles Mack Address: 12226 Washington Lane, Parker, Arizona 85344 Phone" 931-625-4938 Email: charliemack@irc-us.com
- 2. Device :
Trade/proprietary name: Disposable Safety Self-destructive Syringes, 5ML Common Name : Syringe, Antistick; needle; hypodermic, single lumen Classification Name : Piston syringe, Hypodermic single lumen needle
- 3. Predicate Devices :
InviroSnap Safety Syringe 1ML, 3ML, 5ML, 10ML, 20ML, (K092430)
- 4. Classifications Names & Citations :
21CFR 880.5860, MEG, syringe, antistick; 21CFR880.5570, FMI, needle, hypodermic, single lumen; both classifications are Class 2
{4}------------------------------------------------
#### 5. Description :
The JIANGSU YILE MEDICAL ARTICAL C0., LTD Disposable Safety Self-destructive Syringes, 5MLs are single lumen hypodermic needles with a safety mechanism to prevent needle stick after the usage. The Disposable Safety Self-destructive Syringe is used to inject fluids into, or withdraw fluids from the body. In addition, it is designed to aid in the prevention of needle stick injuries.
The syringes are made of a protector, needle, up connector, 0-ring seal, down connector, barrel, piston and plunger. The syringe and body are constructed of Polypropylene (PP). The piston and 0-ring seal are made of natural rubber. Lumens are constructed of 304 Stainless Steel.
6. Indication for use :
The Disposable Safety Self-destructive Syringe is used to inject fluids into, or withdraw fluids from the body. In addition, it is designed to aid in the prevention of needle stick injuries.
- 7. Technological Characteristics and Comparison with predicate device : Jiangsu Yile Medical Article Co., Ltd believes that the Disposable Safety Selfdestructive Syringes, 5ML are substantially equivalent to the InviroSnap Safety Syringe 1ML, 3ML, 5ML, 10ML, 20ML (K092430).
Please note the comparison table on the following page, which demonstrates the similarities of the Jiangsu Yile Medical Article Co., Ltd Disposable Safety Selfdestructive Syringes, 5MLs and the predicate device.
{5}------------------------------------------------
| Element of comparison | Subject Device | Claimed SE Device |
|--------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Company | JIANGSU YILE MEDICAL ARTICLE CO.,LTD | INVIRO MEDICAL DEVICES, INC. |
| FDA510(K) Number | K143070 | K092430 |
| Device Name | Disposable Safety Self-destructive Syringes, 5ML | InviroSnap Safety Syringe |
| Model Number | 5ML | 1, 3, 5, 10 and 20ML |
| Intended Use | The Disposable Safety Self-destructive Syringe is used to<br>inject fluids into, or withdraw fluids from the body. In<br>addition, it is designed to aid in the prevention of needle<br>stick injuries. | The InviroSnap Safety Syringe is used to inject fluids into, or<br>withdraw fluids from, the body. In addition, the InviroSnap<br>Safety Syringe is designed to aid in the prevention of needle<br>stick injuries. |
| Principle of operation | Identical | After use, the health care professional fully depresses the<br>plunger to engage the luer assembly. Once the luer assembly<br>is engaged, pulling back the plunger causes the Adapter and<br>the attached needle to be withdrawn into the safety of the<br>barrel. In this position against the flange, lateral pressure on<br>the plunger results in a controlled fracture of the plunger. Both<br>the syringe and plunger are discarded in a sharps container.<br>Manually retractable safety syringe with permanent disable |
| Safety feature | Identical | Manually retractable safety syringe with permanent disable |
| Material | | |
| Barrel | PP | PP |
| Plunger | PP | PP |
| Piston | Natural Rubber | Natural Rubber |
| Needle | Stainless Steel | Stainless Steel |
| Needle Protector | PP | PP |
| Lubricant | Polydimethylsiloxane Oil | Silicone Oil |
| Specific drug use | Identical | Conventional drugs |
| Sterilization | Identical | EO Sterilization |
| Needle length | 1 1/2 Inch | 1 1/4 Inch |
| Needle gauge | 21G | 21/22/23G |
| Needle tip configuration | Identical | Tri-Beveled Tip |
| Wall type | Identical | Regular wall |
| Nozzle type | Identical | Needle and hub are separated to the syringe |
| Barrel marking specs | Identical | Conforms to ISO 7886-1 |
| Gradations legibility | Identical | 0.2ml |
| Needle cover color | Identical | Clarity |
| Lubricant composition | Polydimethylsiloxane Oil | Silicone Oil |
| Lubricant amount/cm² | Identical | <0.25mg/cm² |
| Barrel transparency | Identical | Transparency |
| Delivery accuracy | Identical | Conforms to ISO 7886-1 |
| Biocompatibility | Identical | Conform to ISO 10993-1 |
| Syringe type | Identical | Antistick syringe |
| Reuse | Identical | Single use only |
| Sterility | Identical | Sterile |
{6}------------------------------------------------
- 8. Non-clinical and Performance Tests :
| List the actual tests performed on the subject device |
|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| <b>Biocompatibility:</b><br>Cytotoxicity Test (ISO10993-5: 2009)<br>Sensitization Test (ISO10993-10: 2010)<br>Intracutaneous Reactivity Test (ISO10993-10: 2010)<br>Acute Systemic Toxicity Test (ISO 10993-11: 2006)<br>Hemolysis Test (ASTM F756-13) |
| <b>Performance:</b><br>ISO 7886-1:1993/ Corrigendum 1:1995 Sterile hypodermic syringes for single use - Part 1: Syringes for manual use<br>ISO 7864:1993 Sterile hypodermic needles for single use<br>ISO 9626: 2001 Stainless steel needle tubing for the manufacture of medical devices<br>ISO 23908: 2011 Sharps injury protection-requirements and test methods-sharps protection features for single-use hypodermic needles, introducers for catheters and needles used for blood sampling<br>ISO 7886-4:2006 Sterile hypodermic syringes for single use-Part 4: Syringes with re-use<br>ISO 594/1, Conical fittings with a 6% (luer) taper for syringes, needles and certain other medical equipment - part 1: general requirements.<br>ISO 594-2 Second edition 1998-09-01, Conical fittings with a 6% (luer) taper for syringes, needles and certain other medical equipment - part 2: lock fittings.<br>USP 788 Light Obscuration Particle Count Test |
| Simulated clinical use:<br>FDA guidance "Guidance for industry and FDA Stuff, Medical Devices with Sharps Injury Prevention Features, August 9, 2005". |
| Package and Shelf Life:<br>- Real Time Stability Test<br>- Accelerated Aging Test<br>- Sterile test<br>- Vacuum leak test<br>- Dye penetration test<br>- Agar contact-attack test (Microbial barrier properties)<br>- Tensile Seal Strength Test |
| Related standards:<br>ISO11607-1: Packaging for terminally sterilized medical devices - Part1: Requirements for materials, sterile barrier systems and packaging systems.<br>ISO11737-2: Sterilization of medical devices - Microbiological methods - Part 2: Tests of sterility performed in the validation of a sterilization process.<br>ASTM F1980-07: Standard guide for accelerated aging of sterile barrier systems for medical devices.<br>ASTM D3078-02: Standard Test Method for Determination of Leaks in Flexible Packaging by Bubble Emission.<br>ASTM F1929-98: Standard Test Method for Detecting Seal Leaks in Porous Medical Packaging by Dye Penetration.<br>ASTM F88 -09 Standard Test Method for Seal Strength of Flexible Barrier Materials<br>ISO7886-1: Sterile hypodermic syringes for single use - Part1: Syringes for manual use.<br>ISO 23908: Sharps injury protection-requirements and test methods-sharps protection features for single-use hypodermic needles, introducers for catheters and needles used for blood sampling |
| ISO7864: Sterile hypodermic needles for single use.
All test results were satisfactory.
{7}------------------------------------------------
### 9. Conclusions:
In accordance with the Federal Food, Drug and Cosmetic Act, 21 CFR Part 807 and based on the information provided in this premarket notification Jiangsu Yile Medical Article Co., Ltd concludes that the Disposable Safety Self-destructive Syringes, 5ML are substantially equivalent to predicate devices as described herein.
END
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.